Skip to main content
Log in

Influence of erythrocyte velocity on impedance plethysmographic measurements

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

Impedance plethysmography has been widely used clinically. However, uncertainty exists concerning the source of the impedance signal. The present study was designed to indicate the relative contributions made by blood volume changes and blood flow changes. Anin vitro flow system containing a pulsatile pump, a bovine carotid artery, a rigid tube, and a terminal resis tance was assembled. Impedance recordings were made on both the expansile artery and the rigid tube. The impedance signal from the rigid tube was taken as the flow-related component. The signal from the expansile artery represented the combined effects of volume changes and flow changes. The recordings were made over the full physiological range of pulse pressures and peak flow values, using both normal saline and blood (haematocrit =26±4%). No impedance change was detected in the rigid vessel using saline as the perfusate. When blood was used, the impedance signal from the rigid tube averaged about 10% of that from the expansile vessel and its morphology was different. Changes in erthrocyte velocity produced a small but definite change in the impedance signal recorded from the artery.

Sommaire

L'usage de la pléthysmographie d'impédance est largement répandu en clinique. Il reste toutefois des doutes au sujet de la provenance du signal d'impédance. Cette étude vise à établir les contributions apportées par les changements de volume et de circulation du sang. Après construction d'un système de circulationin vitro, composé d'une pompe pulsatile, d'une carotide de boeuf, d'un tube rigide et d'une résistance terminale, on enregistra l'impédance sur l'artère et le tube. Le signal émis par l'artère souple se révéla être une représentation des effets combinés des changements de volume et de circulation. Les enregistrements furent effectués sur toute la gamme physiologique des tensions différentielles et des valeurs de circulation maximale, à l'aide de saline normale et de sang (hématocrite =26±4%). Aucun changement d'impédance ne fut détecté dans le vaisseau rigide, lors de l'emploi de la saline somme perfusat. Avec le sang, le signal d'impédance émis par le tube rigide se situait autour de 10% de celui du vaisseau souple, et sa morphologie était différente. Des changements se produisirent également dans la vélocité érythrocyte, qui entraînèrent un changement léger, mais très net, dans le signal d'impédance enregistré dans l'artère.

Zusammenfassung

Widerstandsplethysmographie wird im klinischen Bereich weitverbreitet angewendet. Es besteht jedoch eine gewisse Untersicherheit hinsichtlich der Quelle des Impedanzsignals. Die vorliegende Arbeit zielt darauf ab, darzulegen, welche Auswirkungen Veränderungen in Blutvolumen und Blutdurchfluß hierauf haben. Einin vitro Durchflußsytem mit einer Pulsierpumpe, der Halsschlagader eines Rinds, einem starren Rohr und einem Endwiderstand wurde zusammengebaut. Der Widerstand wurde sowohl an der dehnbaren Arterie wie an dem starren Rohr gemessen. Das Widerstandssignal an dem starren Rohr wurde als durchflußbezogene Komponente angenommen. Das Signal von der dehnbaren Arterie stellte die kombinierte Wirkung von Volumen- und Durchflußveränderungen dar. Es wurde Messungen für den gesamten physiologischen Pulsdruckbereich und Spitzendurchflußwerte angestellt, wobei sowohl normale Salzlösung wie Blut (Hämatokrit= 26±4%) verwendet wurde. Im starren Gefäß wurde bei der Verwendung von Salzlösung als Perfusat keinerlei Veränderung des Widerstands beobachtet. Wurde Blut verwendet, ergab sich für das starre Rohr ein Widerstandssignal, das durchschnittlich bei etwa 10% des Signals des dehnbaren Gefäßes lag und dessen Morphologie unterschiedlich war. Veränderungen in der Erythrocytengeschwindigkeit ergaben eine kleine, aber definitive Veränderung des Widerstandssignals bei der Arterie.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Anderson, F. A., Jr.,Peura, R. A., Penney, B. C. andWheeler, H. B. (1977) Comparison of electrical impedance and mechanical plethysmographic techniques in the human calf. Proceedings: Association for the Advancement of Medical Instrumentation 12th Annual Meeting 281.

  • Edgerton, R. H. (1974) Conductivity of sheared suspensions of ellipsoid particles with applications to blood flow,IEEE Trans.,BME-21, 33–43.

    Google Scholar 

  • Frewer, R. A. (1972) Effect of frequency changes on electrical conductance.Med. & Biol. Eng.,10, 734–744.

    Google Scholar 

  • Fricke, H. andMorse, S. (1926) The electric resistance and capacity of blood for frequencies between 87 kHz-4·5 MHz.J. Gen. Physiol.,9, 153–167.

    Article  Google Scholar 

  • Geddes, L. A. andDa Costa, C. P. (1973) The specific resistance of canine blood at body temperature.IEEE Trans. BME-20, 51–53.

    Google Scholar 

  • Gollan, F. andNamon, R. (1970) Electrical impedance of pulsatile blood flow in rigid tubes and in isolated organs.Ann. NY Acad. Sci.,170, 2, 568–576.

    Google Scholar 

  • Hirsch, E. M. andLiebman, F. M. (1974) A rheographic comparison of electrical conductance in volume-restricted and a volume-unrestricted vascular segment. Federation Proceedings33, 334.

    Google Scholar 

  • Kanai, H., Nakayama, K. andOno, N. (1973) Impedance of blood: the effects of the orientation of red blood cells. Digest of the 10th International Conference on Medical and Biological Engineering 328.

  • Kanai, H., Sakamoto, K. andMiki, M. (1976) Impedance of blood; the effects of red cell orientation. Digest of the 11th International Conference on Medical and Biological Engineering 238–239.

  • Nyboer, J. Electrical impedance plethysmography. (1959) Charles, C. Thomas, Springfield, Ill.

    Google Scholar 

  • Powers, S., Schaffer, C., Boba, A. andNakamura, Y. (1958) Physical and biological factors in impedance plethysmography.Surgery 44, 53–62.

    Google Scholar 

  • Sigman, E., Kolin, A. andKatz, L. N. (1937) Effects of motion on electrical conductivity of the blood.Am. J. Physiol.,118, 708–719.

    Google Scholar 

  • Swanson, D. K. andWebster, J. G. (1976) Origin of the electrical impedance pulse in the limbs. Proceedings 29th Annual Conference on Engineering in Medicine & Biology 18, 324.

  • Tanaka, K., Kanai, H., Nakayama, K. andOno, N. (1970) The impedance of blood: the effects of red cell orientation and its application.Japan J. Med. Electron. & Biol. Rng.,8, 14–21.

    Google Scholar 

  • Velick, S. andGorin, N. (1940) The electrical conductance of suspensions of ellipsoids and its relation to the study of avian erythrocytes.J. Gen. Physiol. 23, 753–771.

    Article  Google Scholar 

  • Wheeler, H. B., O'Donnell, J. A., Anderson, F. A., Jr.,Penney, B. C., Peura, R. A., Benedict, K., Jr. (1975) Bedside screening for venous thrombosis using impedance phlebography.Angiology,26, 2, 199–209.

    Google Scholar 

  • Wheeler, H. B., Penney, B. C., Anderson, F. A., Jr., andPeura, R. A. (1978) Impedance plethysmography: theoretical and experimental basis in non-invasive diagnostic techniques in vascular disease.in Bernstein, E. F. (Ed.), St Louis, Missouri, C. V. Mosby.

    Google Scholar 

  • Young, D. G., Cox, R. H., Stover, E. K. andErdman, W. J. II, (1967) Evaluation of quantitative impedance ment. III Blood flow determinationin vivo.Am. J. Phys. Med.,46, 1450–1457.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Peura, R.A., Penney, B.C., Arcuri, J. et al. Influence of erythrocyte velocity on impedance plethysmographic measurements. Med. Biol. Eng. Comput. 16, 147–154 (1978). https://doi.org/10.1007/BF02451914

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02451914

Keywords

Navigation