Summary
A study of comparative intraocular pressure measurements was made using the hand-held applanation tonometers of Perkins and Draeger in the sitting and lying patient. A series of 100 patients was investigated by two observers taking the pressures with both tonometers under blind conditions. The statistical evaluation of the data revealed an average pressure increment when lying down of 2,5 to 3 mm Hg for both observers and both tonometers. This postural response is fully reversible when getting up and an overshoot-phenomenon statistically significant is noted.
The pressure increment when lying down is not related to initial pressure level and there is no clear relationship to the age of the patients tested.
Interobserver variation as well as variations of both tonometers are evaluated. The effect of repeated tonometry in this study is considered.
Some aspects of position influence on the intraocular pressure are discussed.
Zusammenfassung
Es wird über vergleichende Augeninnendruckmessungen am sitzenden und liegenden Patienten mit den Handapplanationstonometern von Perkins und Draeger berichtet. Es wurden 100 Patienten von zwei Untersuchern unter “Blind-Bedingungen” mit beiden Tonometern der Augeninnendruck gemessen. Die statistische Auswertung der Ergebnisse zeigte einen durchschnittlichen Anstieg des Augendruckes beim Wechsel vom Sitzen zum Liegen um 2,5–3 mm Hg für beide Untersucher und beide Tonometer. Dieser Positionseffekt zeigte sich reversibel beim Wiederaufrichten, wobei der Augendruck statistisch signifikant unter das Niveau der initialen Druckmessung abfiel. Es fand sich keine Beziehung des Positionseffektes zur Höhe des intraokularen Druckes und zum Alter der untersuchten Patienten. Die Varianz der Meßergebnisse von Untersucher zu Untersucher sowie von Tonometer zu Tonometer wurde statistisch untersucht. Der Einfluß der wiederholten Applanation auf die Meßergebnisse wurde berücksichtigt. Einige Aspekte des Einflusses der Körperlage auf den intraokularen Druck werden diskutiert.
Similar content being viewed by others
Literatur
Armaly, M.F., Salamoun, G.S.: Schiötz and applanation tonometry. Arch. Ophthal.70, 603–609 (1963)
Bynke, H., Krakau, C.E.T., Wilke, K.: Repeated applanation tonometry in optic atrophy. Acta ophthal. (Kbh.)50, 240–246 (1972)
Draeger, J.: Über ein lageunabhängiges Applanationstonometer. Klin. Mbl. Augenheilk.149, 905–911 (1966)
Draeger, J.: Principle and clinical application of a portable applanation tonometer. Invest. Ophthal.6, 132–134 (1967)
Dunn, J.S., Brubaker, R.F.: Perkins applanation tonometer. Clinical and laboratory evaluation. Arch. Ophthal.89, 149–151 (1973)
Goldmann, H., Schmidt, Th.: Über Applanationstonometrie. Ophthalmologica (Basel)134, 221–242 (1957)
Hetland-Eriksen, J.: On tonometry. 5. The pressure of glaucomatous eyes in the sitting and the lying positions by means of the Goldmann applanation tonometer. Acta ophthal. (Basel)44, 515–521 (1966)
Kindler-Loosli, C., Schmidt, Th.: Der intraokulare Druck nach Änderung der Körperlage. Albrecht v. Graefes Arch. klin. exp. Ophthal.194, 17–21 (1975)
Krakau, C.E.T., Wilke, K.: On repeated tonometry. Acta ophthal. (Kbh.)49, 611–614 (1971)
Krieglstein, G.K.: The postural responses in normal and glaucomatous eyes. Glaukom-Symposium, Univ. Augenklinik Würzburg, 11.–13. Okt. 1974 Würzburg
Krieglstein, G.K., Langham, M.E.: Glaukom ohne Hochdruck. Ein Beitrag zur Ätiologie. Klin. Mbl. Augenheilk.166, 18–23 (1975)
Krieglstein, G.K., Langham, M.E.: Influence of body position on the intraocular pressure of normal and glaucomatous eyes. Ophthalmologica (Basel)171, 132–145 (1975)
Krieglstein, G.K., Waller, W.K.: Goldmann applanation versus handapplanation and Schiötz indentation tonometry. Albrecht v. Graefes Arch. klin. exp. Ophthal.194, 11–16 (1975)
Langham, M.E., Leydhecker, W., Krieglstein, G.K., Waller, W.K.: Pneumatonographic studies in normal and glaucomatous eyes. Adv. Ophthal.32 (1976) (in press)
Linner, E., Rickenbach, C., Werner, H.: Comparative measurements of the pressure in the aqueous veins and conjunctival veins using different methods. Acta ophthal. (Kbh.)28, 469–478
Leydhecker, W.: The intraocular pressure. Clinical aspects. 15. Meeting of the Association of Eye Research, Univ. Augenklinik Würzburg 13.–16. Okt. 1974, Würzburg
Makabe, R.: Orthostatische zephale Hypotension und Auge. Klin. Mbl. Augenheilk.166, 56–58 (1975)
Moses, R.A.: Adler's physiology of the eye, 5. Aufl. St. Louis: Mosby 1970
Perkins, E.S.: Hand-held applanation tonometer. Brit. J. Ophthal.49, 591–593 (1965)
Schmidt, Th.: Tonometrie und Differentialtonometrie, Tension und Rigiditätsbestimmung. Ophthalmologica (Basel)19, 123–252 (1965)
Weigelin, E., Niesel, P.: Der Netzhautarteriendruck bei Änderung der Körperlage. Albrecht v. Graefes Arch. Ophthal.150, 374–384 (1950)
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Krieglstein, G.K., Brethfeld, V. & v. Collani, E. Vergleichende Tonometrie mit lageunabhängigen Handapplanationstonometern. Albrecht von Graefes Arch. Klin. Ophthalmol. 199, 101–113 (1976). https://doi.org/10.1007/BF02385206
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02385206