Vergleichende Tonometrie mit lageunabhängigen Handapplanationstonometern

Comparative intraocular pressure measurements with position independent hand-applanation tonometers

  • Published:
Albrecht von Graefes Archiv für klinische und experimentelle Ophthalmologie Aims and scope Submit manuscript

Summary

A study of comparative intraocular pressure measurements was made using the hand-held applanation tonometers of Perkins and Draeger in the sitting and lying patient. A series of 100 patients was investigated by two observers taking the pressures with both tonometers under blind conditions. The statistical evaluation of the data revealed an average pressure increment when lying down of 2,5 to 3 mm Hg for both observers and both tonometers. This postural response is fully reversible when getting up and an overshoot-phenomenon statistically significant is noted.

The pressure increment when lying down is not related to initial pressure level and there is no clear relationship to the age of the patients tested.

Interobserver variation as well as variations of both tonometers are evaluated. The effect of repeated tonometry in this study is considered.

Some aspects of position influence on the intraocular pressure are discussed.

Zusammenfassung

Es wird über vergleichende Augeninnendruckmessungen am sitzenden und liegenden Patienten mit den Handapplanationstonometern von Perkins und Draeger berichtet. Es wurden 100 Patienten von zwei Untersuchern unter “Blind-Bedingungen” mit beiden Tonometern der Augeninnendruck gemessen. Die statistische Auswertung der Ergebnisse zeigte einen durchschnittlichen Anstieg des Augendruckes beim Wechsel vom Sitzen zum Liegen um 2,5–3 mm Hg für beide Untersucher und beide Tonometer. Dieser Positionseffekt zeigte sich reversibel beim Wiederaufrichten, wobei der Augendruck statistisch signifikant unter das Niveau der initialen Druckmessung abfiel. Es fand sich keine Beziehung des Positionseffektes zur Höhe des intraokularen Druckes und zum Alter der untersuchten Patienten. Die Varianz der Meßergebnisse von Untersucher zu Untersucher sowie von Tonometer zu Tonometer wurde statistisch untersucht. Der Einfluß der wiederholten Applanation auf die Meßergebnisse wurde berücksichtigt. Einige Aspekte des Einflusses der Körperlage auf den intraokularen Druck werden diskutiert.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  • Armaly, M.F., Salamoun, G.S.: Schiötz and applanation tonometry. Arch. Ophthal.70, 603–609 (1963)

    CAS  PubMed  Google Scholar 

  • Bynke, H., Krakau, C.E.T., Wilke, K.: Repeated applanation tonometry in optic atrophy. Acta ophthal. (Kbh.)50, 240–246 (1972)

    CAS  Google Scholar 

  • Draeger, J.: Über ein lageunabhängiges Applanationstonometer. Klin. Mbl. Augenheilk.149, 905–911 (1966)

    CAS  PubMed  Google Scholar 

  • Draeger, J.: Principle and clinical application of a portable applanation tonometer. Invest. Ophthal.6, 132–134 (1967)

    CAS  PubMed  Google Scholar 

  • Dunn, J.S., Brubaker, R.F.: Perkins applanation tonometer. Clinical and laboratory evaluation. Arch. Ophthal.89, 149–151 (1973)

    CAS  PubMed  Google Scholar 

  • Goldmann, H., Schmidt, Th.: Über Applanationstonometrie. Ophthalmologica (Basel)134, 221–242 (1957)

    CAS  Google Scholar 

  • Hetland-Eriksen, J.: On tonometry. 5. The pressure of glaucomatous eyes in the sitting and the lying positions by means of the Goldmann applanation tonometer. Acta ophthal. (Basel)44, 515–521 (1966)

    CAS  Google Scholar 

  • Kindler-Loosli, C., Schmidt, Th.: Der intraokulare Druck nach Änderung der Körperlage. Albrecht v. Graefes Arch. klin. exp. Ophthal.194, 17–21 (1975)

    Article  CAS  Google Scholar 

  • Krakau, C.E.T., Wilke, K.: On repeated tonometry. Acta ophthal. (Kbh.)49, 611–614 (1971)

    CAS  Google Scholar 

  • Krieglstein, G.K.: The postural responses in normal and glaucomatous eyes. Glaukom-Symposium, Univ. Augenklinik Würzburg, 11.–13. Okt. 1974 Würzburg

  • Krieglstein, G.K., Langham, M.E.: Glaukom ohne Hochdruck. Ein Beitrag zur Ätiologie. Klin. Mbl. Augenheilk.166, 18–23 (1975)

    CAS  PubMed  Google Scholar 

  • Krieglstein, G.K., Langham, M.E.: Influence of body position on the intraocular pressure of normal and glaucomatous eyes. Ophthalmologica (Basel)171, 132–145 (1975)

    CAS  Google Scholar 

  • Krieglstein, G.K., Waller, W.K.: Goldmann applanation versus handapplanation and Schiötz indentation tonometry. Albrecht v. Graefes Arch. klin. exp. Ophthal.194, 11–16 (1975)

    CAS  Google Scholar 

  • Langham, M.E., Leydhecker, W., Krieglstein, G.K., Waller, W.K.: Pneumatonographic studies in normal and glaucomatous eyes. Adv. Ophthal.32 (1976) (in press)

  • Linner, E., Rickenbach, C., Werner, H.: Comparative measurements of the pressure in the aqueous veins and conjunctival veins using different methods. Acta ophthal. (Kbh.)28, 469–478

  • Leydhecker, W.: The intraocular pressure. Clinical aspects. 15. Meeting of the Association of Eye Research, Univ. Augenklinik Würzburg 13.–16. Okt. 1974, Würzburg

  • Makabe, R.: Orthostatische zephale Hypotension und Auge. Klin. Mbl. Augenheilk.166, 56–58 (1975)

    CAS  PubMed  Google Scholar 

  • Moses, R.A.: Adler's physiology of the eye, 5. Aufl. St. Louis: Mosby 1970

    Google Scholar 

  • Perkins, E.S.: Hand-held applanation tonometer. Brit. J. Ophthal.49, 591–593 (1965)

    CAS  PubMed  Google Scholar 

  • Schmidt, Th.: Tonometrie und Differentialtonometrie, Tension und Rigiditätsbestimmung. Ophthalmologica (Basel)19, 123–252 (1965)

    Google Scholar 

  • Weigelin, E., Niesel, P.: Der Netzhautarteriendruck bei Änderung der Körperlage. Albrecht v. Graefes Arch. Ophthal.150, 374–384 (1950)

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Krieglstein, G.K., Brethfeld, V. & v. Collani, E. Vergleichende Tonometrie mit lageunabhängigen Handapplanationstonometern. Albrecht von Graefes Arch. Klin. Ophthalmol. 199, 101–113 (1976). https://doi.org/10.1007/BF02385206

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02385206

Navigation