Abstract
Pulmonary artery blood temperature fluctuations are a principle source of variability of thermodilution cardiac output measurements. Two signal processing strategies were investigated to decrease this variability; a subtractive noise cancellation algorithm, developed from a heat balance model of heat transport through the right heart, and a band pass integration technique. Blood temperature fluctuations were recorded from the inferior and superior vena cavae and the pulmonary artery in four dogs. The recorded temperature signals were processed off line to investigate the performance of the signal processing strategies. The findings indicate that the accuracy of the heat transport model, while dependent on the accuracy of the sensing of caval blood temperature, is sufficient for cancellation of a significant fraction of the pulmonary artery thermal noise power if the amplitude of the fluctuations is large. The use of the subtractive cancellation algorithm improved the standard deviation of thermodilution curve area determinations by a factor of five, relative to area estimates made without noise cancellation. Band pass integration produced a more modest improvement in area estimate reproducibility. These results demonstrate that it is possible to reduce, but not eliminate, the variability of thermodilution cardiac output measurements with the described processing strategies.
Similar content being viewed by others
Abbreviations
- UUCOM:
-
University of Utah Cardiac Output Monitor
- V a :
-
volume of atrial mixing chamber
- T a :
-
temperature of the blood in the atrial mixing chamber
- t :
-
time
- F ivc :
-
volumetric flow of blood in the inferior vena cava
- T ivc :
-
temperature of the blood in the inferior vena cava
- F suc :
-
volumetric flow of blood in the superior vena cava
- T suc :
-
temperature of the blood in the superior vena cava
- F a :
-
volumetric flow of blood through the atrium
- V ed :
-
end diastolic volume of the ventricle
- T p :
-
temperature of the blood in the pulmonary artery
- i :
-
heart beat index
- F p :
-
volumetric flow of blood through the pulmonary artery
- F r :
-
flows ratio,F ivc/F svc
- \(\bar \in \) :
-
mean modeling error
- VAR(res) :
-
variance of the modeling error
- C.D. :
-
coefficient of determination of model
- N :
-
number of data segments
References
Afonso, S.; Rowe, G.G.; Castillo, C.A., Crumpton, C.W. Intravascular and intracardiac blood temperatures in man. J. Appl. Physiol. 17:706–708; 1962.
Afonso, S.; Herrick, J.F., Rowe, G.G.; Crumpton, C.W. Temperature variations in the venous system of dogs. Am. J. Physiol. 203:278–282; 1962.
Antman, S. Foundations of indicator dilution theory. In: D.A. Bloomfield, ed. Dye Curves: The Theory and Practice of Indicator Dilution, Baltimore: University Park Press; 1974: pp. 24–29.
Ehlers, K.C.; Mylrea, K.C.; Calkins, J.M. Investigation of a continuous heating/cooling technique for cardiac output measurement. Ann. Biomed. Eng. 15:551–565; 1987.
Elkayam, U.; Berkley, R.; Azen, S.; Weber, L.; Geva, B.; Henry, W.L. Cardiac output by thermodilution technique: effect of injectate volume and temperature on the accuracy and reproducibility in the critically ill patient. Chest 84:418–422; 1983.
Hamilton, W.F.; Riley, R.L.; Attyah, A.M.; Cournand, A., et al. Comparison of the Fick and dye injection methods of measuring the cardiac output in man. Am. J. Physl. 153:309–321; 1948.
Johnson, R.W.; Normann, R.A. Mathematical and mechanical modeling of heat transport through the heart. Ann. Biomed. Eng. 15:603–617; 1987.
Johnson, R.W.; Normann, R.A. Signal processing strategies for the enhancement of the signal to noise ratio of thermodilution measurements. Ann. Biomed. Eng. 16:265–278; 1987.
Normann, R.A.; Johnson, R.W.; Messinger, J.E.; Sohrab, B. A continuous cardiac output computer based on thermodilution principles. Ann. Biomed. Eng. 17:61–73; 1989.
Moreno, A.H.; Burchell, A.R. Respiratory regulation of splanchnic and systemic venous return in normal subjects and in patients with hepatic cirrhosis. Surg., Gyn., Obst. 154:257–267; 1982.
Pantalos, G.M. The effect of heart rate on left ventricular energy balance in chronically instrumented, anaesthetized dogs. Ohio State University; 1983. Ph.D Dissertation.
Philip, J.H.; Long, M.C.; Quinn, M.D.; Newbower, R.S. Continuous thermal measurement of cardiac output. IEEE Trans. on Biomed. Eng. 31:393–399; 1984.
Wessel, H.U.; James, G.W., Paul, M.H. Effects of respiration and circulation on central blood temperature of the dog. Am. J. Physiol. 211:1403–1412; 1966.
Widrow, B.; Stearns, S.D. Adaptive signal processing. Englewood Cliffs, NJ: Prentice-Hall; 1985.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Johnson, R.W., Normann, R.A. Central venous blood temperature fluctuations and thermodilution signal processing in dogs. Ann Biomed Eng 17, 657–669 (1989). https://doi.org/10.1007/BF02367469
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF02367469