Specific gas chromatography detectors

Part II: Electrolytic conductivity detector

Spezifische detektoren in der gas-chromatographie

Teil II: Detektor für elektrolytische Leitfähigkeit

Détecteurs spécifiques en chromatographie en phase gazeuse

2ème partie: Détecteur à conductivité électrolytique

  • Serial
  • Published:
Chromatographia Aims and scope Submit manuscript

Summary

The operation of conductometric detectors is based on the measurement of the electrolytic conductivity of water. The standard commercial detector can be made specific for halogen, nitrogen or sulfur. Also other uses of conductometry as gas chromatography detection principle have been described.

In the standard heteroelement-sensitive mode of operation, the column effluent is combusted under oxidative or reductive conditions, interfering combustion products are selectively and quantitatively removed, and the conductivity of a water solution of the component of interest is measured. The conductivity measurement as GC detection principle can doubtlessly be extended to other applications. Linearity of response and, in most cases also response predictability, together with a comparatively high sensitivity of measurements make it an interesting detection tool.

Zusammenfassung

Die Arbeitsweise von Leitfähigkeits-Detektoren beruht auf der Messung der elektrolytischen Leitfähigkeit von Wasser. Der kommerzielle Standard-Detektor kann für Halogen, Stickstoff oder Schwefel spezifisch gemacht werden. Es sind auch noch andere Anwendungen der Leitfähigkeitsmessung als gas-chromatographisches Nachweisprinzip beschrieben worden. Bei der üblichen, auf Heteroelemente ansprechenden Arbeitsweise wird der Ablauf aus der Säule unter Oxydation oder Reduktion verbrannt, störende Verbrennungsprodukte werden selektiv und quantitativ entfemt, und die Leitfähigkeit einer wäßrigen Lösung der interessierenden Komponente wird gemessen. Die Leitfähigkeitsmessung als GC-Nachweisprinzip kann zweifellos auch auf andere Anwendungen ausgedehnt werden. Die Linearität des Ansprechverhaltens und — in den meisten Fällen — die Voraussehbarkeit des Ansprechverhaltens, im Verein mit verhältnismäßig hoher Meßempfindlichkeit, lassen dieses Verfahren als ein interessantes Nachweissystem erscheinen.

Résumé

Le fonctionnement des détecteurs conductimétriques est basé sur la mesure de la conductivité électrolytique de l'eau. Le détecteur standard commercialisé peut être rendu spécifique pour les halogènes, l'azote ou le soufre. D'autres applications mettant en jeu la conductimétrie comme principe de détection en chromatographie en phase gazeuse ont également été décrites. Dans le mode opératoir standard, lorsque le détecteur est sensible aux hétéroéléments, l'effluent de la colonne est «brûlé» en atmosphère oxydante ou réductrice; les produits de combustion qui donnent la même réponse sont éliminés de façon sélective et quantitative, et l'on mesure ainsi la conductivité d'une solution aqueuse correspondant au composant auquel on s'intéresse. La mesure de la conductivité utilisée comme principe de détection en CG peut sans doute être étendue à d'autres applications. La linéarité de la réponse et, dans la plupart des cas, la prévisibilité de la réponse, jointe à une sensibilité relativement élevée, font de ce mode de détection un outil de travail très intéressant.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Boer H., Proc. 4th World Petrol Congr., Rome 1955, Sect. V., S1

  2. Piringer O., Tataru E., Pascalau M., J. Gas Chromatogr.2, 104 (1964)

    CAS  Google Scholar 

  3. Piringer O., Pascalau, M., J. Chromatog.8, 410 (1962)

    Article  CAS  Google Scholar 

  4. Coulson D. M., J. Gas. Chromatog.3, 134 (1965)

    CAS  Google Scholar 

  5. Coulson D. M., J. Gas Chromatog.4, 285 (1966)

    CAS  Google Scholar 

  6. Sternberg J. C., Jones D. T. L., Pittsburgh Conf. Anal. Chemistry and Appl. Spectroscopy, Cleveland, Ohio, March 5–9, 1970

  7. Rhoades J. W., Johnson D. E., J. Chromatog. Sci.,8, 616 (1970)

    CAS  Google Scholar 

  8. Mohnke M., Piringer O., Tataru E., J. Chromatogr. Sci.6, 117 (1968)

    CAS  Google Scholar 

  9. Coulson D. M., in Advances in Gas Chromatography, Vol. 3J. C. Giddings, R. A. Keller, Eds., M. Dekker, Inc., New York, 1966; pp. 200–214.

    Google Scholar 

  10. US patent 3, 309. 845

  11. Dijkstra A., Fabrie C. C. M., Kaleman G., Lauboo C. J., Thissen J. A. I., J. Gas Chromatog.2, 180 (1964)

    CAS  Google Scholar 

  12. Product of Tracor Inc., Austin, Texas

  13. Patchett G. G., J. Chromatog. Sci.8, 155 (1970)

    CAS  Google Scholar 

  14. Johnson D. E., Rhoades J. W., Wheeler R. J., presented at the 150th National ACS Meeting Sept. 1967

  15. Frink R. C., David D. J., Pittsburgh Conf. Anal. Chemistry and Appl. Spectroscopy, Cleveland, Ohio, March 5–9, 1970

  16. Hartman C. H., J. Chromatog. Sci.7, 163 (1969)

    Google Scholar 

  17. Unpublished results from our laboratory

  18. Cramer M. F., Carroll J. J.

  19. Johns T. Braithwaite Jr. C. H., Residue Revs.5, 45 (1964)

    CAS  Google Scholar 

  20. T. M. Todd, Tracor Inc, private communication

  21. Stevenson, G. W., Anal. Chem.38, 1949 (1966)

    Article  Google Scholar 

  22. Krejci M., Dressler M., in Chromatog. Revs., M. Lederer, Ed., Vol. 13, pp. 1–59. Elsevier, Amsterdam, 1970

    Google Scholar 

  23. Jentzsch D., Otte E., Detektoren in der Gas-Chromatographie, Akad Verlagsgesellschaft, Frankfurt/Main 1970, pp. 195–197

    Google Scholar 

  24. Gudzinowiez B. J., in The Practice of Gas Chromatography,L. S. Ettre, A. Zlatkis Eds., Interscience Publishers N. Y., 1967, pp. 303–306

    Google Scholar 

  25. Coulson, D. M., 1968 Eastern Analytical Symposium New York, N. Y., No. 13-15, 1968

Download references

Author information

Authors and Affiliations

Authors

Additional information

Part I: Chromatographia4, 425 (1971)

Present address: Packard Instrument, D-6000 Frankfurt/Main, Germany

Rights and permissions

Reprints and permissions

About this article

Cite this article

Selucky, M.L. Specific gas chromatography detectors. Chromatographia 5, 359–366 (1972). https://doi.org/10.1007/BF02315258

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02315258

Keywords

Navigation