Summary
A peroxyoxalate chemiluminescence detection system for liquid chromatography is described. The excitation of fluorophores is generated by the reaction of bis-(2,4,6-trichlorophenyl)oxalate or bis(2,4-dinitrophenyl)oxalate and hydrogen peroxide, which are added to the column effluent. The influence of the solvents and the concentrations of the reagents have been investigated. The influence of the flow cell volume on sensitivity and on band broadening have also been studied and a “chemical band narrowing effect” has been observed. Different types of apparatus have been compared for detection of the emitted light. The system has been used for the detection of the dansyl derivative of a drug with a secondary amine functional group in serum samples. The detection limits are in the 1–10pg range.
Similar content being viewed by others
References
S. Kobayashi, K. Imai, Anal. Chem.52, 424 (1980).
S. Kobayashi, J. Sekino, K. Honda, K. Imai, Anal. Biochem.112, 99 (1981).
K. Honda, J. Sekino, K. Imai, Anal. Chem.55, 940 (1983).
K. W. Sigvardson, J. W. Birks, Anal. Chem.55, 432 (1983).
M. S. Gandelman, J. W. Birks, J. Chromatogr.242, 21 (1982).
A. G. Mohan, N. J. Turro, J. Chem. Educ.51, 528 (1974).
W. R. Seitz in “Methods in Enzymology”, Vol. 57M. A. DeLuca, Ed., Academic Press, London, 1978, p. 456.
D. C. Williams, G. F. Huff, W. R. Seitz, Anal. Chem.48, 1003 (1976).
P. Lechtken, N. J. Turro, Mol. Photochem.6, 95 (1974).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
de Jong, G.J., Lammers, N., Spruit, F.J. et al. Optimization of a peroxyoxalate chemiluminescence detection system for the liquid chromatographic determination of fluorescent compounds. Chromatographia 18, 129–133 (1984). https://doi.org/10.1007/BF02258768
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02258768