Reduction of blood trauma in roller pumps for long-term perfusion

  • Published:
World Journal of Surgery Aims and scope Submit manuscript

Abstract

As part of an attempt to develop a roller pump for long-term (2 weeks) cardiac assist, factors generating blood trauma in roller pumps were re-examined. Fresh human blood was pumped for 4 hr in a simple closed circulatory loop. Samples drawn at regular intervals were examined for plasma hemoglobin, lactate dehydrogenase, platelet function, and activation of coagulation factors. Experiments were performed in which tubing diameters were varied (polyvinyl chloride, 1/2- and 3/8-inch internal diameters), roller diameters were varied (3/4, 1 1/8, and 1 3/4 inches), and roller occlusion was varied from total occlusion to 0.020-inch underocclusion. The results showed more hemolysis when the roller diameters and the tubing diameters were smaller and when the rollers were more occlusive. Changes in platelet function and activation of coagulation factors were observed, but were not correlated with occlusion. These results suggest that when roller pumps are used for long-term cardiac support, the diameter of the rollers should be as large as possible, the tubing diameter should be as large as practical, and the occlusion setting should be as open as possible while still maintaining required flow rates. New techniques for setting occlusion are suggested.

Résumé

Les facteurs engendrant des traumatismes sanguins des pompes de circulation extra-corporelle ont été réexaminés pour la mise au point d'une pompe pour des assistances cardiaques de longue durée (deux semaines). Du sang frais humain a été mis en circulation fermée pour des durées de 4 heures. Des échantillons prélevés à intervalles réguliers, furent analysés, en particulier pour ce qui est de l'hémoglobine plasmatique, de la lactate déshydrogénase, des fonctions plaquettaires, et de l'activation des facteurs de coagulation. Les expériences furent menées avec des tubes de diamètres variables (en chlorure de polyvinile dont le diamètre interne était de 1/2 et 3/8 d'inch), le diamètre des galets (3/4, 1 1/8 et 1 3/4 d'inch) et le caractère occlusif des galets variant depuis une occlusion totale jusqu'à une sub-occlusion de 0.02 inch. Les résultats ont montré une hémolyse plus importante lorsque le diamètre des galets et des tubes étaient plus petits et lorsque les galets étaient plus occlusifs. Des modifications dans les fonctions plaquettaires et dans l'activation des facteurs de coagulation ont été observées, mais n'étaient pas corellées avec le degré occlusif de la pompe. Les résultats suggèrent que lorsqu'on utilise une assistance cardiaque au long cours et qu'on emploie des pompes à galet le diamètre des galets doit être aussi grand que possible, le diamètre des tubes aussi important qu'il est réalisable et le réglage de l'occlusion le plus lâche possible, tout en maintenant des débits de pompe suffisants. De nouvelles techniques pour régler l'occlusion sont proposées.

Resumen

Como parte de un esfuerzo destinado a desarrollar una bomba de rodillo para asistencia cardíaca a largo término (dos semanas), se procedió a revisar los factores generadores de trauma sanguíneo en las bombas de rodillo. Sangre humana fresca fué bombeada por períodos de cuatro horas en un simple circuito circulatorio cerrado. Las muestras tomadas a intervalos regulares fueron examinadas para hemoglobina plasmática, deshidrogenasa láctica, función plaquetaria y activación de los factores de coagulatión. Se realizaron experimentos en los cuales los diámetros de los tubos fueron variables (cloruro de polivinil, diámetro interior de 1/2 y 3/8 pulgadas), los diámetros de los rodillos fueron diferentes (3/4, 1 1/8 y 1 3/4 pulgadas) y la oclusión del rodillo variable entre oclusión total y oclusión de 0.020 pulgadas. Los resultados demostraron mayor hemólisis con rodillos y tubos de diámetro pequeño y con rodillos de mayor oclusión. Se observaron alteraciones en la función plaquetaria y en la activación de factores de coagulación, pero éstas no se correlacionan con el grado de oclusión. Estos resultados sugieren que cuando se utilicen bombas de rodillo por largos períodos de soporte, el diámetro de los rodillos debe ser lo mayor posible, el diámetro de los tubos debe ser tan amplio como sea práctico, y el nivel de oclusión debe ser fijado tan abierto como sea posible para todavía mantener las tasas de flujo requeridas. Se sugieren nuevas técnicas para fijar el nivel de oclusión.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Crane, K.A., Brown, D., Anderson, R., Begelman, K.M.: Further decrease in subclinical hemolysis utilizing 12.7 mm tubing in the arterial roller head. Ann. Thorac. Surg.35:463, 1983

    Google Scholar 

  2. Bernstein, E.F., Gleason, L.R.: Factors influencing hemolysis with roller pumps. Surgery61:432, 1967

    Google Scholar 

  3. McCaughan, J.S., Jr., McMichals, H., Schuder, J.C., Kirby, C.K.: Use of a totally occlusive pump as a flowmeter with observations of hemolysis caused by occlusive and nonocclusive pump-oxygenator. Surgery44:210, 1957

    Google Scholar 

  4. Head, L.R., Coenen, J.P., Angola, E., Nogueira, C., Mendelsohn, D., Kay, E.B.: Operation of roller pump for extracorporeal circulation. J. Thorac. Cardiovasc. Surg.39:210, 1960

    Google Scholar 

  5. Durusz, M., Christman, E., Williams, E., Tyers, G.: Roller pump induced tubing wear: Another argument in favor of arterial line filtration. J. Extra-Corp. Tech.12:2, 1980

    Google Scholar 

  6. Bietz, G., Schermeher, T., Donnelly, T.: Long-term tubing fatigue characteristics. J. Extra-Corp. Tech.11:4, 1979

    Google Scholar 

  7. Esmond, W.G., Attar, S., Baker, B., Chyba, J., Demetriades, A.D., Cowley, R.A.: An improved 360° single roller spring loaded blood pump. Tr. Am. Soc. Artif. Intern. Organs7:167, 1961

    Google Scholar 

  8. Kusserow, B.K., Kendall, L.W.: In vitro changes in the corpuscular elements of blood flowing in tubular conduits. Tr. Am. Soc. Artif. Intern. Organs9:262, 1963

    Google Scholar 

  9. Saleem, A., Blifeld, C., Saleh, S.A., Yawn, D.H., Mace, M.L., Schwartz, M., Crawford, E.S.: Viscoelastic measurement of clot formation: A new test of platelet function. Ann. Clin. Lab. Science13:115, 1983

    Google Scholar 

  10. Von Kaulla, K.N., Ostendorf, P., von Kaulla, E.: The impedance machine: A new bedside coagulation recording device. J. Med.6:73, 1975

    Google Scholar 

  11. Kusserow, B.K., Machanic, B., Collins, F.M., Jr., Clapp, J.F., III: Changes observed in blood corpuscles after prolonged perfusions with two types of blood pumps. Tr. Am. Soc. Artif. Intern. Organs11:122, 1965

    Google Scholar 

  12. Galletti, P.: Laboratory experience with 24-hour partial heart-lung bypass. J. Surg. Res.5:97, 1965

    Google Scholar 

  13. Bernstein, E.F., Castaneda, A.R., Varco, R.L.: Some biologic limitations to prolonged blood pumping. Tr. Am. Soc. Artif. Intern. Organs11:118, 1965

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Supported in part by a grant from the Norton Company.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Noon, G.P., Kane, L.E., Feldman, L. et al. Reduction of blood trauma in roller pumps for long-term perfusion. World J. Surg. 9, 65–71 (1985). https://doi.org/10.1007/BF01656257

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01656257

Keywords

Navigation