Abstract
We compared the 95% response time (95% RT) of two tissue oxygen tonometers under two sets of circumstances. We first evaluated the devices during normoxia, hyperoxia, and anoxia in vitro, using a transcutaneous PO2 electrode (PtcO2) as the reference. The responses to normoxia and to different grades of hyperoxia were examined in vivo in 8 healthy volunteers to assess the relationship between changes in subcutancous PO2 and PtcO2, an estimate of arterial PO2 (PaO2). One subcutaneous method (ScA) used a technique based on a polarographic needle electrode in situ connected to an ammeter; the second method (ScB) was based on a blood gas analyzer system first described by Hunt (Lancet 164;2:1370). ScA and PtcO2 both responded to stepwise changes in ambient oxygen concentration (21–100%) in vitro within 10 seconds; the 95% RT of ScA was 1.39±0.5 to 2.39±0.8 minutes and that of PtcO2 was 0.32±0.1 to 0.49±0.1 minutes. ScB had a lag of 3 minutes, and the 95% RT was 6.75±0.5 to 8.2±0.8 minutes. In contrast to the results in vitro, the response of ScA to changes in FiO2 in vivo was delayed compared with the rapid response of PtcO2, reflecting the physiologic delay of tissue PO2 in response to increased PaO2. The time lag and the long 95% RT of ScB were even more evident in vivo. ScA reacted three to four times faster than ScB, both in vitro and in vivo, to changes in the oxygen environment. The in vitro 95% RT of ScA to changes in ambient oxygen varied from 2 to 3.5 minutes. In contrast, the 95% RT of ScB was 8 to 9 minutes. PtcO2 had the fastest 95% RT, from 0.4 to 0.5 minutes. The results suggest that the subcutaneous needle electrode method (ScA) provides close to real-time assessment of tissue PO2.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Hunt TK. A new method of determining tissue oxygen tension. Lancet 1964;2:1370–1371
Niinikoski J. Heughan C, Hunt TK. Oxygen tensions in human wounds. J Surg Res 1972;12:77–82
Chang N, Goodson WH, Gottrup F, Hunt TK. Direct measurement of wound and tissue oxygen tension in post-operative patients. Ann Surg 1983;197:470–478
Wimberley PD, Gronlund Pedersen K, Thode J, et al. Transcutaneous and capillary PCO2 and PO2 measurements in healthy adults. Clin Chem 1983;29(8):1471–1473
Reed RL, Maier RV, Del Landicho MT, et al. Correlation of hemodynamic variables with transcutaneous PO2 measurements in critically ill adult patients. J Trauma 1985;25:1045–1053
Niinikoski J, Hunt TK. Measurement of wound oxygen with implanted silastic tube. Surgery 1972;71:22–26
Niinikoski J, Laaksonen V, Meretoja O, et al. Oxygen transport to tissue under normovolemic moderate and extreme hemodilution during coronary bypass operation. Ann Thorac Surg 1981;31:134–143
Kuttila K, Niinikoski J. Peripheral tissue perfusion during coronary artery bypass grafting. Perfusion 1989;4:25–31
Kuttila K, Niinikoski J. Peripheral perfusion after cardiac surgery. Crit Care Med 1989;17:217–220
Kuttila K, Niinikoski J. Effect of indomethacin on peripheral tissue perfusion after coronary artery bypass surgery. Scand J Thorac Cardiovasc Surg 1989;23:247–251
Nelimarkka O, Halkola L, Niinikoski J. Effect of graded hemorrhage on renal cortical perfusion in dogs. Am J Surg 1981;141:235–239
Nelimarkka O, Halkola L, Niinikoski J. Distribution of renal cortical and medullary tissue oxygenation in hemorrhagic shock. Acta Chir Scand 1982;148:213–219
Nelimarkka O, Niinikowski J. Renal venous oxygen tension as an indicator of tissue hypoxia in hemorrhagic shock. Crit Care Med 1986;14:128–131
Nelimarkka O, Halkola L, Niinikoski J. Renal hypoxia and lactate metabolism in hemorrhagic shock in dogs. Crit Care Med 1984;12:656–661
Nelimarkka O, Niinikoski J. Oxygen and carbon dioxide tensions in the canine kidney during arterial occlusion and hemorrhagic hypotension. Surg Gynecol Obstet 1984;158:27–32
Korsback C, Nyman N, Scheinin T. Small bowel and liver tissue PO2 and PCO2 during hypovolemic shock and intravenous vasopressin infusion. Ann Chir Gynecol 1984;73:236–240
Makisalo H, Soini H, Lalla M, Hockerstedt K. Subcutaneous and liver tissue oxygen tension in hemorrhagic shock: an experimental study with whole blood and two colloids. Crit Care Med 1988;16:857–861
Makisalo H, Soini H, Nordin A, Hockerstedt K. Effects of bicarbonate therapy on tissue oxygenation during resuscitation of hemorrhagic shock. Crit Care Med 1989;17:1170–1174
Makisalo H, Soini H, Nordin A, Hockerstedt K. Tissue oxygenation in predicting “irreversible” hemorrhagic shock. Surg Res Commun 1989;5:175–184
Dixon WJ, ed. BMDP Manual, vol 2. BMDP Statistical Software. Los Angeles: University of California Press, 1989
Hunt TK, Rabkin J, Jensen JA, et al. Tissue oximetry: an interim report. World J Surg 1987;11:126–132
Lubbers DW. Theoretical basis of the transcutaneous blood gas measurements. Crit Care Med 1981;9:721–733
Author information
Authors and Affiliations
Additional information
Supported by the Paulo Foundation and the Orion Foundation of Research.
We thank the local distributor of Radiometer, Oriola Oy, for the loan of TINA.
Rights and permissions
About this article
Cite this article
Soini, H.O., Takala, J. Measurement of tissue oxygen tension: Comparison between two subcutaneous oxygen tonometers. J Clin Monitor Comput 7, 227–231 (1991). https://doi.org/10.1007/BF01619264
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01619264