Abstract
A sealed tube dry combustion method that is simple, accurate, and precise for ultramicro determination of halogens in organic compounds has been developed. Samples below 1 mg are heated at 580 °C for 1 h in an electric furnace. After the combustion and absorption of the gaseous halogen, the interior of the tube is rinsed. Subsequent titration is carried out potentiometrically with silver nitrate and yielded standard deviations of 0.21% for chlorine, 0.17% for bromine, and 0.25% for iodine.
Similar content being viewed by others
References
D. G. Newman, C. Tomlinson,Mikrochim. Acta [Wien] 1961, 73.
W. I. Awad, Y. A. Gawargious, S. S. M. Hassan, N. E. Milad,Anal. Chim. Acta 1966,36, 339.
W. Kirsten,Fresenius' Z. Anal. Chem. 1961,181, 1.
W. Kirsten, K. Hozumi,Mikrochim. Acta [Wien] 1962, 777.
K. Hozumi,Anal. Chem. 1963,35, 666.
K. Hozumi, W. Kirsten,Anal. Chem. 1963,35, 1522.
C. E. Childs, E. E. Meyers, J. Cheng, E. Laframboise, R. B. Balodis,Microchem. J. 1963,7, 266.
W. J. Kirsten,Mikrochim. Acta [Wien] 1960, 272.
M. Bigois, C. Carriere, P. Linet,Microchem. J. 1978,23, 110.
J. K. Kreling, F. Block, G. T. Louthan, J. Dezwaan,Microchem. J. 1986,34, 158.
A. M. Quinn, K. W. M. Siu, G. J. Gardner, S. Berman,J. Chromatogr. 1986,370, 203.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Tanaka, Y., Okazaki, A. & Hozumi, K. Ultramicro determination of organic halogens using sealed tube dry combustion method. Mikrochim Acta 105, 169–173 (1991). https://doi.org/10.1007/BF01243544
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF01243544