Summary
The use of ion-selective electrodes for direct potentiometric water hardness determinations is restricted to relatively soft waters if they are to be competitive with other techniques in terms of uncertainty on the water hardness scale. The selectivity factorK PotMgCa must be within 1.1% of unity if a hardness of 10 degrees on the German scale is to be measured to within 0.1 degrees in the composition rangec Mg:c Ca=0.25 to 25. The expressionK PotMgNa ·a 2Na / (a Mg+a Ca) must be less than 0.001 if freedom from sodium interference is to be achieved. Under these conditions a 1% relative standard deviation can be obtained for an uncertainty of the EMF⩽100μV.
Zusammenfassung
Die direkte potentiometrische Bestimmung der Wasserhärte mit ionenselektiven Elektroden kann nur für relativ weiches Wasser verwendet werden, sofern sie mit anderen Methoden konkurrenzfähig sein soll. Der SelektivitätsfaktorK PotMgCa sollte nicht mehr als 1,1% von 1 abweichen, um eine Messung von 10 deutschen Härtegraden auf 0,1 Grad genau zu ermöglichen, wenn gleichzeitig das Mischungsverhältnisc Mg:c Ca im Bereich 0,25 bis 25 schwankt. Der AusdruckK PotMgNa ·a 2Na /(a Mg +a Ca) sollte 0,001 nicht überschreiten, wenn eine Störung durch Natrium vermieden werden soll. Unter den obigen Bedingungen läßt sich eine Wasserhärte auf 1% genau bestimmen, wenn die Unsicherheit der gemessenen EMK⩽100μV.
Similar content being viewed by others
References
J. W. Ross, Jr., in: Ion-Selective Electrodes (R. A. Durst, Ed.), NBS Special Publication,314, 1969.
Divalent Cation Electrode, Model 93-32, Instruction Manual, Orion Research, Inc., Cambridge, Mass. (1975).
R. G. Bates and M. Alfenaar, in: Ion-Selective Electrodes (R. A. Durst, Ed.), NBS Special Publication,314, 1969.
M. S. Mohan and R. G. Bates, Clin. Chemistry21, 864 (1975).
P. C. Meier, D. Ammann, W. E. Morf, and W. Simon, in: Medical and Biological Applications of Electrochemical Devices (J. Korytá, Ed.), Chichester, New York, Brisbane, Toronto: Wiley. 1980.
W. J. Hamer and Y.-Ch. Wu, J. Phys. Chem. Ref. Data1, 1047 (1972).
R. N. Goldberg and R. L. Nuttall, J. Phys. Chem. Ref. Data7, 263 (1978).
W. Stumm and E. Stumm-Zollinger, Chimia22, 325 (1968).
S. K. A. G. Hassan, G. J. Moody, and J. D. R. Thomas, Analyst105, 147 (1980).
D. Erne, W. E. Morf, S. Arvanitis, Z. Cimerman, D. Ammann, and W. Simon, Helv. Chim. Acta62, 994 (1979).
H. F. Osswald, R. E. Dohner, T. Meier, P. C. Meier, and W. Simon, Chimia31, 50 (1977).
Metrohm Information10, 3 (1980).
D. G. Rands and E. Stoica, J. Am. Water68, 309 (1976).
I. Sekerka and J. F. Lechner, Talanta22, 459 (1975).
D. Midgley and K. Torrance, Potentiometric Water Analysis, Chichester, New York, Brisbane, Toronto: Wiley. 1978.
M. Whitfield and J. V. Leyendekkers, Analyt. Chim. Acta45, 383 (1969).
M. Whitfield, J. V. Leyendekkers, and J. D. Kerr, Analyt. Chim. Acta45, 399 (1969).
T. P. Hadjiioannou and D. S. Papastathopoulos, Talanta17, 399 (1970).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Meier, P.C., Erne, D., Cimerman, Z. et al. Direct potentiometric water hardness determination using ion-selective electrodes. Mikrochim Acta 73, 317–327 (1980). https://doi.org/10.1007/BF01196318
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF01196318