Abstract
A method for the calibration of OH temperature measurements using laser-induced fluorescence is demonstrated. The technique depends on the thermal dissociation of water vapor in a furnace. The calibration was demonstrated for the excitation of theA 2 Σ +(υ′ = 3) −Χ 2 ∏(υ″ = 0) band with a tunable KrF laser. Consistent discrepancies of 3% between calculated and measured temperatures with a standard deviation of 8 % were observed.
Similar content being viewed by others
References
J.H. Bechtel: Appl. Opt.18, 2100 (1979)
R. Cattolica: Appl. Opt.20, 1156 (1981)
K.J. Rensberger, J.B. Jeffries, R.A. Copeland, K. Kohse-Höinghaus, M.L. Wise, D.R. Crosley: Appl. Opt.28, 3556 (1989)
A. Lawitzki, I. Plath, W. Stricker, J. Bittner, U. Meier, K. Kohse-Höinghaus: Appl. Phys. B50, 513 (1990)
A. Arnold, B. Lang, T. Bouché, T. Heitzmann, G. Schiff, W. Ketterle, P. Monkhause, J. Wolfrum: Ber. Bunsenges. Phys. Chem.92, 1388 (1992)
G.H. Dieke, H.M. Crosswhite: J. Quant. Spectrosc. Radiat. Transfer2, 97 (1962)
P.M. Doherty, D.R. Crosley: Appl. Opt.23, 713 (1984)
J.H. Grinstead, G. Laufer, R.H. Krauss, J.C. McDaniel, Jr.: Appl. Opt.33, 1115(1993)
D.E. Heard, D.R. Crosley, J.B. Jeffries, G.P. Smith, A Hirano: J. Chem. Phys.96, 4366 (1992)
I.L. Chisdey, D.R. Crosley: J. Quant. Spectrosc. Radiat. Transfer23, 187(1980)
R.A. Copeland, M.L. Wise, D.R. Crosley: J. Phys. Chem.92, 5710(1988)
T.M. Quagliaroli, G. Laufer, R.H. Krauss, J.C. McDaniel, Jr.: AIAA J.31, 520 (1993)
D.G. Fletcher: AIAA 93-0049 (1993)
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Quagliaroli, T.M., Laufer, G. & McDaniel, J.C. Calibration of OH laser-induced fluorescence temperature measurements using thermally dissociated H2O. Appl. Phys. B 59, 635–638 (1994). https://doi.org/10.1007/BF01081185
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01081185