Skip to main content
Log in

Die Durchstichkappentechnik

Quantitative Untersuchungen über den Einfluß der Durchstiche und anderer Faktoren auf die Dichtigkeit von Durchstichkappen

  • Originalabhandlungen
  • Published:
Fresenius' Zeitschrift für analytische Chemie Aims and scope Submit manuscript

Summary

The tightness of self-sealing elastomer septums during puncture operations was investigated with a high-vacuum apparatus. During the punctures the pressure in the apparatus was recorded in the range smaller than 10−5 up to 5 · 10−3 Torr. The penetrated air was collected by a Toepler pump and also measured. Influences of different factors during punctures regarding tightness and lifetime of the septums were ascertained.

Under normal experimental conditions the septums were still high-vacuum tight after 100 punctures and the quantity of the penetrated air was under 0.5 ml.

Zusammenfassung

Mit einer Hochvakuumapparatur wurden Untersuchungen über die Dichtigkeit von Durchstichkappen beim Einstechen und Herausziehen einer Kanüle gemacht. Während des Durchstechens wurde der Druck in der Apparatur im Bereich von <10−5 bis 5 · 10−3 Torr registriert und die eingedrungene Luft mit einer Toepler-Pumpe gesammelt und gemessen. Der Einfluß verschiedener Faktoren auf die Dichtigkeit und Lebensdauer der Durchstichkappen beim Durchstechen wurde ermittelt. Unter normalen Versuchsbedingungen waren die Durchstichkappen nach 100 Durchstichen noch hochvakuumdicht, und die Menge der bei diesen 100 Durchstichen eingedrungenen Luft lag unter 0,5 ml.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Adler, N.: Anal. Chem. 36, 2291–2295 (1964).

    Google Scholar 

  2. Fehér, F., G. Kuhlbörsch u. H. Luhleich: Z. Anorg. Allg. Chem. 303, 294 bis 302 (1960).

    Google Scholar 

  3. — — —: Z. Naturforsch. 14b, 466–467 (1959).

    Google Scholar 

  4. Luhleich, H.: Dissertation, Universität Köln 1958.

  5. —: Kerntechnik 7, 326–329 (1965).

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

Herrn Prof. Dr. E. Bagge und Herrn Prof. Dr. E. Fischer danke ich für das Interesse an der vorliegenden Arbeit. Herrn E. Neumann, der mit eigenen Ideen zum Gelingen der vorliegenden Arbeit beigetragen hat, gilt mein besonderer Dank. Den Herren P. Pflaum und N. Pillmann, die mir bei der Durchführung der Versuche geholfen haben, bin ich zu Dank verpflichtet.

Siehe auch [2–5].

Rights and permissions

Reprints and permissions

About this article

Cite this article

Luhleich, H. Die Durchstichkappentechnik. Z. Anal. Chem. 232, 347–354 (1967). https://doi.org/10.1007/BF00533840

Download citation

  • Received:

  • Revised:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00533840

Navigation