Abstract
We present a new two-plate linear ion trap mass spectrometer that overcomes both performance-based and miniaturization-related issues with prior designs. Borosilicate glass substrates are patterned with aluminum electrodes on one side and wire-bonded to printed circuit boards. Ions are trapped in the space between two such plates. Tapered ejection slits in each glass plate eliminate issues with charge build-up within the ejection slit and with blocking of ions that are ejected at off-nominal angles. The tapered slit allows miniaturization of the trap features (electrode size, slit width) needed for further reduction of trap size while allowing the use of substrates that are still thick enough to provide ruggedness during handling, assembly, and in-field applications. Plate spacing was optimized during operation using a motorized translation stage. A scan rate of 2300 Th/s with a sample mixture of toluene and deuterated toluene (D8) and xylenes (a mixture of o-, m-, p-) showed narrowest peak widths of 0.33 Th (FWHM).
Similar content being viewed by others
Introduction
Mass spectrometry has become one of the most widely used analytical methods because of its its high sensitivity, selectivity, and fast analysis. Mass spectrometers (MS) have historically been large, laboratory-confined instruments. Though the motivation for developing smaller MS was first driven by space exploration [1, 2], MS that are small enough to be portable now find their way into such diverse fields as environmental monitoring [3,4,5], threat detection [6,7,8], oceanography [9, 10], and clinical diagnosis [11,12,13]. Although portability typically requires a reduction of the physical size, weight, and power consumption of the whole instrument, miniaturizing the mass analyzer often leads to reduced weight, size, and power of other system components. Previous efforts have miniaturized a variety of conventional mass analyzers, including time-of-flight (TOF) [14,15,16], magnetic sector [17, 18], and quadrupole mass analyzers [19,20,21]. Radio frequency (rf) ion traps [22,23,24,25] are of particular promise because of several key characteristics: their simple structures allow less stringent electrode alignment; their tolerances to high pressure that help reduce the vacuum requirements; and their compact size and low weight. Although the resolution in ion traps is typically lower than that in other mass analyzers, they have nonetheless seen a fast pace of development for portable MS [12, 26,27,28,29].
Several issues frequently arise in development of miniaturized mass analyzers. For instance, electric breakdown is more likely as electrode surfaces are closer together and pressures are higher [30]. Electronic stability (rf and amplitude) must meet more stringent absolute requirements when using lower voltages and smaller physical dimensions [31]. The absolute mechanical tolerances must be tighter on the smaller dimensional scale of smaller analyzers [32], often pushing the limits of conventional machining techniques. Space-charge effects are a greater concern in analyzers with smaller dimensions [31, 33], often resulting in decreased sensitivity. With ion traps, this reduced sensitivity has been recovered in two ways: either by using parallel arrays of traps or by using traps with inherently larger storage volumes due to an extended trapping dimension (e.g., linear, rectilinear, or toroidal traps).
Compared with quadrupole ion traps and cylindrical ion traps, toroidal ion traps have a compact structure and a larger trapping capacity for the same characteristic trapping dimension [25], making them appealing for developing a portable MS [26, 34], though mechanical fabrication and assembly may limit miniaturization [35]. The 2D linear ion trap (LIT) also effectively expands the trapping capacity [36], although performance may be more sensitive to geometric deviations, including misalignment of electrodes [23]. The rectilinear ion trap (RIT) substitutes flat surfaces for the hyperbolic electrodes of the LIT, simplifying fabrication for a given level of machining precision [24, 37]. Ion trap arrays based on cylindrical [38,39,40], linear/rectilinear [41, 42], and either in 2-D [42,43,44,45] or 3-D quadrupoles [46, 47] have been extensively explored and developed. A common feature of trap arrays is the involvement of microfabrication techniques to produce very small traps [31].
Microfabrication, including photolithography and micropatterning, provides high precision and accuracy in two dimensions, but limited capabilities in the third (out-of-plane) dimension. To leverage the high precision of in-plane features and reduce or eliminate dependence on out-of-plane fabrication, we previously demonstrated an approach to making ion traps using two plates, with patterning on the facing surfaces of each plate to produce the needed trapping potential [48,49,50]. Several trapping geometries were demonstrated, including a linear ion trap [51]. This two-plate linear ion trap included sets of electrodes lithographically patterned on the facing surfaces of two ceramic (Al2O3) plates. Ions were ejected through laser-drilled slits in the plates. Capacitive voltage dividers created different rf amplitudes for each electrode. The plates were positioned 4–6 mm apart in initial experiments [52], and 1.9 mm apart in later experiments [52], and near-unit mass resolution was demonstrated in both cases. However, subsequent efforts to further miniaturize this two-plate LIT ran into several issues. Foremost among these were difficulties in ion ejection through an increasingly narrow ejection slit.
Although the dimensional scale of lithographic patterning can easily be reduced to make smaller features and smaller ion traps in the two-plate approach, the plate thickness cannot be reduced without making the resulting device more fragile. However, using a plate that is thick compared with the trapping dimensions can lead to problems with ion ejection. In previous efforts (unpublished), we developed and tested a two-plate LIT with plate spacing of 724 μm (y0 = 362 μm). The ejection slit was fabricated to be 100 μm wide and 20 mm long. The plate thickness was still 0.5 mm as with the larger devices, resulting in an ejection slit that was many times deeper than it was wide. Unfortunately, ions were not successfully ejected through this narrow, deep slit. Two factors likely prevented ion ejection: ions traveling at small angles during ejection would not make it through the slit but would hit the walls, and build-up of charge on the inside walls of the slit repelled subsequent ions. In a LIT, ions can pick up energy in the non-ejection direction, which has a secular frequency close to the ejection frequency, resulting in ions being ejected with some angular dispersion. In addition, depending on operating frequency and voltage, ion trajectories in miniaturized ion traps can be larger, relative to the available space between electrodes, than in full-size traps [30], exacerbating the issues with the narrow ejection slit.
One possible solution to the above problem is to eject ions parallel to the plates rather than through a slit. However, simulations show that it is more difficult to obtain desirable combinations of higher-order terms in the trapping field when ions are ejected between and parallel to the plates. Early results from the halo ion trap [49, 53] showed the same difficulties. A slit allows ions to be ejected with a potential in the ejection direction that is carefully controlled all the way to the point of ejection from the trap.
We have opted for another solution, which is to taper the ejection slit profile. The slit walls open out, allowing ions to pass through even with a large angular dispersion. Charge build-up on the slit walls is thereby prevented, and ion ejection can proceed. The tapered slits profile also allows the slit walls to be coated with a conducting or semiconducting material, which is impractical in straight-walled slits. Tapered or open-out slits are used on other ion traps made using metal electrodes, including the original Finnigan quadrupole ion trap, the linear ion trap [23], and the rectilinear ion trap [24]. In these cases, the slit is not to reduce charge build-up but simply to improve the efficiency of ejection for ions emerging with some angular dispersion.
The previous plate design also had several issues in making electrical connections between the front and back side of the plates, and in making electrical connection with the printed circuit board behind the ceramic plates. These issues were solvable at the dimension scale of the full-size device, but were difficult to miniaturize further. Both of these issues are solved in the present design by patterning the connections on the trapping side of the plates and wire-bonding the connections from ceramic plate to printed circuit board.
The above issues presented significant challenges in ion trap miniaturization using the two-plate approach. This article demonstrates that a tapered ejection slit combined with single side patterning and connection solves these issues, enabling further miniaturization of this mass analyzer. In addition, performance is significantly improved even for the full-scale device, in terms of mass resolution and also ruggedness and ability to operate for long periods of time without signal loss due to charge build-up. Further, in characterizing this new design, one of the two plates was mounted on a motorized translation stage so that plate spacing could be optimized in vacuum during operation.
Experimental
Design and Fabrication
Whereas our prior efforts all used ceramic substrates, in the present study we used borosilicate glass. Glass has similar electrical and structural properties as alumina ceramic, but is more amenable to the tapering cuts used to make the ejection slit. The fabrication process is shown in Figure 1. Plates were milled and diced to 57 × 37 mm, with a thickness of 0.50 mm. A 2.5-mm long, 0.50 mm-wide ejection slit was machined in the glass with a back side taper angle of 45°. Holes were milled into the plates for plate alignment during assembly. Aluminum electrodes and traces were patterned onto the plates using photoresist and photomasks. After patterning, a 100 nm Ge layer was evaporatively deposited on both sides of the plates, covering both the trapping side and the back side of each plate.
The electrode pattern is shown in Figure 1b. Five pairs of rf electrodes are used on each plate, with the positions and dimensions identical to those of the previous ceramic plate design, with a few exceptions. This was done to allow comparison between the designs to be as close as possible. Axial confinement of ions was accomplished using patterned end bars on each plate. The PCB and glass plate were assembled as shown in Figure 2a, b.
The specifications and the optimized rf amplitudes for the five pairs of rf electrodes are shown in Table 1. Electrodes #0 surrounded the ejection slits, and were grounded on one plate but floated (nominally at ground) with an applied AC resonance waveform on the other, electrodes #1 were always grounded, #4 and #5 were always applied with the full rf driving voltage, and the voltages applied on #2 and #3 electrodes were scaled as 0.14 times of the full driving amplitude to adjust the shape of the trapping field [54].
After patterning, each glass plate was attached to a printed circuit board (PCB) using vacuum-grade epoxy (Torr-Seal; Agilent Technologies, Lexington, MA, USA) as seen in Figure 2a, b. The connection pads of the plate were then wire-bonded to corresponding pads on the PCB. In addition, a second PCB with capacitors was attached to the back side of the mounting PCB to establish rf amplitudes and conduct the applied waveforms to the trapping electrodes. Although fairly fragile by themselves, glass plates bonded to the PCBs and mounted in an ion trap assembly can withstand routine handling in the laboratory with no problems.
Mounting and Assembly
Previous designs relied on mounting screws and sapphire balls (for plate separation and alignment), resulting in a rigid assembly. To provide flexibility with the present design, one of the two plates was mounted onto a translation stage and the other plate was fixed. Plate alignment was adjusted prior to putting the instrument in vacuum. Plate spacing was controlled using a motorized stage and was adjustable in vacuum. The assembly is shown in Figure 2c. To mount the moveable plate on the motorized stage (MAX 343, Thorlab Inc., Newton, NJ, USA), a plate holder was designed that can be attached to the glass plate and its support PCB. The stage was mounted on a base, and the top plate was fixed onto a ceiling piece. The bottom plate was immobilized on the stage together with and right above an electron multiplier (Detector Technology, Inc., Palmer, MA, USA). An electron gun (Torion Inc., American Fork, UT, USA) was mounted on the side of the trap with the electron beam directed along the axis of the trap. The two plates were pre-aligned using four sapphire balls (Swiss Jewel Company) before the bottom plate was completely immobilized on the stage. The bottom plate was first loosely placed above the detector, and the two plates were separated apart so that the four balls could be placed in the alignment holes on the bottom plate. The bottom plate was then moved closer to the top plate by the actuator to the position where the four balls were seated in the alignment holes on the top plate. This position was stored in the motorized stage controller. Next, the bottom plate was immobilized and the plates were separated apart again to remove the alignment balls. Finally, the bottom plate was moved back to the stored position. The repeatability of the piezo positioners is 50 nm in each direction. However, larger variances in position are expected when plates are replaced due to imprecision of mounting the plates in the positioning system.
Experiments
The electronic timing control of resonance ejecting ions was similar to previous experiment [51] but with a shorter period (60 ms for the entire experiment). The ionization time was reduced to 0.5 ms; ions were trapped with the rf at 200 V0-p and cooled for 1.825 ms. Ions were then ramped out by sweeping the rf voltage from 200 to 560 V0-p. An AC resonance waveform was applied during the rf ramp with an optimized frequency of 620 kHz and amplitude of 1 V0-p. A +5 V DC voltage applied to the two end bars contained the ions axially inside in the trap.
The sample used in this experiment was a mixture of toluene and deuterated toluene (D8-toluene) at a volume ratio of 1:1, and a mixture of xylenes (o-, m-, p-). Sample pressure for toluene and D8-toluene was 1–2 × 10−6 Torr, whereas the sample pressure for xylene mixtures was 4–5 × 10−6 Torr, nearly three times higher than that of toluene mixtures, so that the fragment ions m/z 91 as well as their isotopic peak m/z 105 and m/z 107 could be seen within a wider range of plate spacing distances. The electron gun (e-gun) acceleration voltage was −62 V for toluenes and was decreased to −52 V for xylenes to reduce the fragmentation. A −100 V/+23 V pair of voltages were used on the gate to block or focus the electron beam. The electron multiplier was used with −1600 V applied. As plates and coaxial cables added about 100 pF to the circuit load, the drive frequency was reduced to 1.35 MHz instead of the rf power supply’s built-in frequency of 1.6 MHz (Ardara Technologies, L.P., Lewiston, NY, USA). The background pressure was in the 10−7 Torr level. Sample was preloaded in a glass tube that was connected to a leak valve (203 variable leak; Granville Phillips, Boulder, CO, USA). Sample vapor was leaked into the vacuum. Helium at 3–4 × 10−3 Torr was introduced for collisional cooling of the trapped ions. Variations of plate spacing and scan rate were explored to observe the effect on signal intensity and mass resolution. Computer simulations utilizing SIMION 8.1 were used to plan experiments and anticipate optimal conditions (voltages, plate spacing, etc.).
Results and Discussion
Simulations showed an optimized plate spacing of 4.4 mm with driving frequency at 1.6 MHz, while best resolution was achieved in practice when the spacing was 5.00 mm with the same applied rf. Although end cap stretching can improve performance in many ion traps because of improvement in higher-order terms of the trapping field [55], these higher-order terms were taken into account in the simulations. It may be that the fields were not exactly as intended, or that some other factor is present that was not adequately accounted for in the SIMION simulations. Such factors may include the effect of the deposited germanium layer, or fabrication tolerances of the slit. We observed small plate-to-plate differences, including a different optimal AC frequency when different plates were used, which may be the result of fabrication variance or inconsistent alignment.
The spectra for the toluene mixture and xylenes are shown in Figure 3. The scan rate was 2300 Th/s. In Figure 3a the toluene molecular ion peak at m/z 92 and the H-loss peak at m/z 91 were completely resolved, as were the corresponding peaks at m/z 98 and 100 from D8-toluene. Peaks at m/z 105, 106, 107, 112, 113, and 114 are expected products of ion-neutral reactions (methyl abstraction) within the trap, including the various combinations of deuterated and non-deuterated species [48]. In Figure 3b, the molecular ion peak at m/z 106 from xylene is well resolved from the H-loss peak at m/z 105 and the 13C peak at m/z 107. The S/N ratio in the xylene spectrum was higher than that in toluene mixture spectrum as the xylene pressure was much higher [56].
Spectra at different plate spacings for the toluene mixture and xylenes are shown in Figures 4 and 5, respectively. The AC frequency, AC amplitude, end-bar voltage, rf trapping voltage, and rf were same as above and constant for all these experiments. Plate spacing in this experiment can be changed without having to break vacuum, so a relatively constant pressure was maintained, greatly reducing changes in signal intensity attributable to sample or helium pressure. Thus, the variation of signal intensity should be due primarily to differences in plate spacing. In the process of collecting toluene mixture spectra, the smallest spacing was 4.8 mm as the peak intensity was too weak to observe if the plate spacing was closer than 4.8 mm. When the spacing was increased from 4.8 mm to 5.0 mm, a clear increase of signal intensity was observed, similar to the xylene mixture. On the other hand, the resolution in toluene mixture spectra seemed to vary little at small spacing from 4.8 mm to 5.0 mm, compared with the resolution variation in xylene mixture spectra when the spacing was increased from 4.2 mm to 5.0 mm. When the plate spacing was further increased from 5.0 mm to 5.5–5.6 mm, peak intensity from toluene (m/z 91 and 92) started to decrease more obviously than ion peaks from D8-toluene. A clear decrease in resolution was also seen in spectra of both samples, indicating there was a maximum point for both toluene mixture and xylene mixture within the variation range. That optimal point occurs at 5.0 mm under the current conditions, at which the minimum peak width (full width of a peak at its half maximum height, FWHM) was 0.35 Th in toluene spectra (m/z 98) and 0.32 Th in xylene spectra (m/z 106), corresponding to a resolution of 119 and 140, respectively. The variation in resolution with plate spacing was probably due to the variation in higher order components of the trapping field at different plate spacings. The variation in peak intensity might be because the e-gun was fixed relative to the stationary plate, and was not always directed precisely along the trapping center. However, a more likely explanation is that the ion ejection efficiency changed with the magnitude of the higher-order field components. With a greater contribution of octopole and dodecapole, ions more quickly fall out of resonance during AC excitation, reducing the fraction that gain sufficient energy to be ejected at that point in the scan. Another phenomenon observed is that when the plate spacing is increased, all the peaks gradually shift to higher ejection rf voltages (as seen in Figures 4 and 5), consistent with the increased effective trap size (y0).
Figure 6 shows the fraction of octopole and dodecapole as a function of plate spacing, calculated using SIMION and using a previously published procedure for calculating the potential on the germanium surface [52]. This procedure cannot be used for ion trajectory calculations and differs from the procedure used above to predict performance. Interestingly, the optimal experimental performance occurs when the octopole is still negative relative to the quadrupole, and even the sum of octopole and dodecapole is slightly negative.
It is well known that resolution in ion traps is dependent on the scan rate [57].
Figure 7 shows the spectra for toluene and D8-toluene at different scan rates ranging from 1700 to 12,000 Th/s, with each spectrum averaged 50 times. Plate spacing was always at the optimal value of 5.0 mm. It can be seen that the resolution (FWHM) was better with slower scan rate. At 3700 Th/s, the m/z 91 and 92 peaks started to combine. Peaks at m/z 98 and 100 remain resolved even with reduced resolution.
An issue with previous LITs made using patterned plates is that charge build-up in the slits prevented long-term operation. Signal intensities would decrease over a timescale of minutes, and would return to normal only after the system was turned off for several tens of minutes. Although the cause of this behavior was previously unclear, the absence of the problem with the present design points to charge build-up as the culprit. With the present design we saw no reduction in signal even after many hours of continuous operation. This likely is the result of the tapered slit because other aspects of the design have not significantly changed (e-gun operation, trap design). From this we also conclude that the signal reduction shown in Figures 4 and 5 is likely the result of differences in ejection efficiency due to different electric fields with each value of plate spacing, and not due to charge build-up.
The above results demonstrate that the tapered slit design is effective at solving one of the main limitations of miniaturizing the two-plate LIT. Specifically, the taper allows ion ejection to be unimpeded by either collision with the slit wall or by build-up of charge on the slit wall, without compromising plate ruggedness by requiring that the plate thickness scale with the trapping dimensions. In addition, the improved electrical connections with the new design proved to be more reliable and easier to miniaturize. The tapered slit can be machined much narrower than what was used in the present study; smaller versions have already been fabricated, patterned, and mounted, and are being prepared for testing.
Conclusions
Key improvements for a two-plate planar LIT were demonstrated to work and to provide solutions encountered in efforts to miniaturize the previous design. Better than unit resolution was achieved by optimizing the plate spacing under a given voltage distribution at a scan rate of 2300 Th/s. This new design consistently showed stable spectra with good resolution. The fabrication process for this one-side ion trap is easier and faster, and the connections are more durable than those of the old one, again an advantage for further miniaturization efforts.
References
Rushneck, D.R., Diaz, A.V., Howarth, D.W., Rampacek, J., Olson, K.W., Dencker, W.D., Smith, P., McDavid, L., Tomassian, A., Harris, M., Bulota, K., Biemann, K., Lafleur, A.L., Biller, J.E., Owen, T.: Viking gas chromatrograph mass spectrometer. Rev. Sci. Instrum. 49, 817–834 (1978)
Hoffman, J.H., Hodges, R.R., Duerksen, K.D.: Pioneer venus large probe neutral mass spectrometer. J. Vac. Sci. Technol. 16, 692–694 (1979)
Bouvier-Brown, N.C., Carrasco, E., Karz, J., Chang, K., Nguyen, T., Ruiz, D., Okonta, V., Gilman, J.B., Kuster, W.C., de Gouw, J.A.: A portable and inexpensive method for quantifying ambient intermediate volatility organic compounds. Atmos. Environ. 94, 126–133 (2014)
Barreira, L.M.F., Parshintsev, J., Karkkainen, N., Hartonen, K., Jussila, M., Kajos, M., Kulmala, M., Riekkola, M.L.: Field measurements of biogenic volatile organic compounds in the atmosphere by dynamic solid-phase microextraction and portable gas chromatography-mass spectrometry. Atmos. Environ. 115, 214–222 (2015)
Eckenrode, B.A.: Environmental and forensic applications of field-portable GC-MS: an overview. J. Am. Soc. Mass Spectrom. 12, 683–693 (2001)
Graichen, A.M., Vachet, R.W.: Using metal complex ion-molecule reactions in a miniature rectilinear ion trap mass spectrometer to detect chemical warfare agents. J. Am. Soc. Mass Spectrom. 24, 917–925 (2013)
Nagashima, H., Kondo, T., Nagoya, T., Ikeda, T., Kurimata, N., Unoke, S., Seto, Y.: Identification of chemical warfare agents from vapor samples using a field-portable capillary gas chromatography/membrane-interfaced electron ionization quadrupole mass spectrometry instrument with Tri-Bed concentrator. J. Chromatogr. A 1406, 279–290 (2015)
Smith, P.A., Lepage, C.J., Lukacs, M., Martin, N., Shufutinsky, A., Savage, P.B.: Field-portable gas chromatography with transmission quadrupole and cylindrical ion trap mass spectrometric detection: Chromatographic retention index data and ion/molecule interactions for chemical warfare agent identification. Int. J. Mass Spectrom. 295, 113–118 (2010)
Mach, P.M., Winfield, J.L., Aguilar, R.A., Wright, K.C., Verbeck, G.F.: A portable mass spectrometer study targeting anthropogenic contaminants in Sub-Antarctic Puerto Williams, Chile. Int. J. Mass Spectrom. In Press, Corrected Proof, Available online 10 December (2016)
Short, R.T., Toler, S.K., Kibelka, G.P.G., Rueda Roa, D.T., Bell, R.J., Byrne, R.H.: Detection and quantification of chemical plumes using a portable underwater membrane introduction mass spectrometer. TrAC Trends Anal. Chem. 25, 637–646 (2006)
Ye, H., Gemperline, E., Li, L.: A vision for better health: Mass spectrometry imaging for clinical diagnostics. Clin. Chim. Acta 420, 11–22 (2013)
Li, L., Chen, T., Ren, Y., Hendricks, P.I., Cooks, R.G., Ouyang, Z.: Mini 12, miniature mass spectrometer for clinical and other applications - introduction and characterization. Anal. Chem. 86, 2909–2916 (2014)
Molina, M.A., Zhao, W., Sankaran, S., Schivo, M., Kenyon, N.J., Davis, C.E.: Design-of-experiment optimization of exhaled breath condensate analysis using a miniature differential mobility spectrometer (DMS). Anal. Chim. Acta 628, 155–161 (2008)
Huang, Z., Tan, G., Zhou, Z., Chen, L., Cheng, L., Jin, D., Tan, X., Xie, C., Li, L., Dong, J., Fu, Z., Cheng, P., Gao, W.: Development of a miniature time-of-flight mass/charge spectrometer for ion beam source analyzing. Int. J. Mass Spectrom. 379, 60–64 (2015)
Gao, W., Tan, G., Hong, Y., Li, M., Nian, H., Guo, C., Huang, Z., Fu, Z., Dong, J., Xu, X., Cheng, P., Zhou, Z.: Development of portable single photon ionization time-of-flight mass spectrometer combined with membrane inlet. Int. J. Mass Spectrom. 334, 8–12 (2013)
Getty, S.A., Brinckerhoff, W.B., Cornish, T., Ecelberger, S., Floyd, M.: Compact two-step laser time-of-flight mass spectrometer for in situ analyses of aromatic organics on planetary missions. Rapid Commun. Mass Spectrom. 26, 2786–2790 (2012)
Chen, E.X., Russell, Z.E., Amsden, J.J., Wolter, S.D., Danell, R.M., Parker, C.B., Stoner, B.R., Gehm, M.E., Glass, J.T., Brady, D.J.: Order of magnitude signal gain in magnetic sector mass spectrometry via aperture coding. J. Am. Soc. Mass Spectrom. 26, 1633–1640 (2015)
Li, D., Guo, M., Xiao, Y., Zhao, Y., Wang, L.: Development of a miniature magnetic sector mass spectrometer. Vacuum 85, 1170–1173 (2011)
Wright, S., Malcolm, A., Wright, C., O'Prey, S., Crichton, E., Dash, N., Moseley, R.W., Zaczek, W., Edwards, P., Fussell, R.J., Syms, R.R.A.: A microelectromechanical systems-enabled, miniature triple quadrupole mass spectrometer. Anal. Chem. 87, 3115–3122 (2015)
Orient, O.J., Chutjian., A., Farkanian, V.: Miniature, High-resolution, Quadrupole Mass-Spectrometer Array. Rev. Sci. Instrum. 68, 1393–1397 (1997)
Cheung, K., Velasquez-Garcia, L.F., Akinwande, A.I.: Chip-scale quadrupole mass filters for portable mass spectrometry. J. Microelectromech. Syst. 19, 469–483 (2010)
Badman, E.R., Johnson, R.C., Plass, W.R., Cooks, R.G.: A Miniature Cylindrical Quadrupole Ion Trap: Simulation and Experiment. Anal. Chem. 70, 4896–4901 (1998)
Schwartz, J.C., Senko, M.W., Syka, J.E.P.: A two-dimensional quadrupole ion trap mass spectrometer. J. Am. Soc. Mass Spectrom. 13, 659–669 (2002)
Ouyang, Z., Wu, G., Song, Y., Li, H., Plass, W.R., Cooks, R.G.: Rectilinear ion trap: concepts, calculations, and analytical performance of a new mass analyzer. Anal. Chem. 76, 4595–4605 (2004)
Lammert, S.A., Plass, W.R., Thompson, C.V., Wise, M.B.: Design, optimization and initial performance of a toroidal rf ion trap mass spectrometer. Int. J. Mass Spectrom. 212, 25–40 (2001)
Contreras, J.A., Murray, J.A., Tolley, S.E., Oliphant, J.L., Tolley, H.D., Lammert, S.A., Lee, E.D., Later, D.W., Lee, M.L.: Hand-Portable Gas Chromatograph-Toroidal Ion Trap Mass Spectrometer (GC-TMS) for Detection of Hazardous Compounds. J. Am. Soc. Mass Spectrom. 19, 1425–1434 (2008)
Yang, M., Kim, T., Hwang, H., Yi, S., Kim, D.: Development of a palm portable mass spectrometer. J. Am. Soc. Mass Spectrom. 19, 1442–1448 (2008)
He, M., Xue, Z., Zhang, Y., Huang, Z., Fang, X., Qu, F., Ouyang, Z., Xu, W.: Development and characterizations of a miniature capillary electrophoresis mass spectrometry system. Anal. Chem. 379, 2236–2241 (2015)
Gao, L., Song, Q., Patterson, G.E., Cooks, R.G., Ouyang, Z.: HandHeld recilinear ion trap mass spectrometer. Anal. Chem. 78, 5994–6002 (2006)
Tian, Y., Higgs, J.M., Li, A., Barney, B.B., Austin, D.E.: How far can ion trap miniaturization go? Parameter scaling and space-charge limits for very small cylindrical ion traps. J. Mass Spectrom. 49, 233–240 (2014)
Smith, S.A., Mulligan, C.C., Song, Q., Noll, R.J., Cooks, R.G., Ouyang, Z.: In: March, R.E., Todd, J.F.J. (eds.) Practical aspects of trapped ion trap mass spectrometry: ion traps for miniature, multiplexed, and soft-landing technologies. CRC Press, Boca Raton (2010)
Xu, W., Chappell, W.J., Cooks, R.G., Ouyang, Z.: Characterization of electrode surface roughness and its impact on ion trap mass analysis. J. Mass Spectrom. 44, 353–360 (2009)
Avinash, K., Agarwal, A.K., Jana, M.R., Sen, A., Kaw, P.K.: Space-charge effect in the paul trap. Phys. Plasmas. 2, 3569–3572 (1995)
Smith, P.A., Lepage, C.R.J., Savage, P.B., Bowerbank, C.R., Lee, E.D., Lukacs, M.J.: Use of a hand-portable gas chromatograph-toroidal ion trap mass spectrometer for self-chemical ionization identification of degradation products related to O-ethyl S-(2-diisopropylaminoethyl) methyl phosphonothiolate (VX). Anal. Chim. Acta 690, 215–220 (2011)
Lammert, S.A., Rockwood, A.A., Wang, M., Lee, M.L., Lee, E.D., Tolley, S.E., Oliphant, J.R., Jones, J.L., Waite, R.W.: Miniature toroidal radio frequency ion trap mass analyzer. J. Am. Soc. Mass Spectrom. 17, 916–922 (2006)
Hager, J.W.: A new linear ion trap mass spectrometer. Rapid Commun. Mass Spectrom. 16, 512–526 (2002)
Wang L., Xu, F., Ding, C.: Performance and geometry optimization of the ceramic-based rectilinear ion traps. Rapid Commun. Mass Spectrom. 26, 2068–2074 (2012)
Chaudhary, A., van Amerom, F.H.W., Short, R.T.: Development of microfabricated cylindrical ion trap mass spectrometer arrays. J. Microelectromech. Syst. 18, 442–448 (2009)
Cruz, D., Chang, J.P., Fico, M., Guymon, A.J., Austin, D.E., Blain, M.G.: Design, microfabrication, and analysis of micrometer-sized cylindrical ion trap arrays. Rev. Sci. Instrum. 78, 015017–015109 (2007)
Ouyang, Z., Badman, E.R., Cooks, R.G.: Characterization of a serial array of miniature cylindrical ion trap mass analyzers. Rapid Commun. Mass Spectrom. 13, 2444–2449 (1999)
Fico, M., Maas, J.D., Smith, S.A., Costa, A.B., Ouyang, Z., Chappell, W.J., Cooks, R.G.: Circular arrays of polymer-based miniature rectilinear ion traps. Analyst 134, 1338–1347 (2009)
Wang, Y., Zhang, X., Zhai, Y., Jiang, Y., Fang, X., Zhou, M., Deng, Y., Xu, W.: Mass selective ion transfer and accumulation in ion trap arrays. Anal. Chem. 86, 10164–10170 (2014)
Li, X., Jiang, G., Luo, C., Xu, F., Wang, Y., Ding, L., Ding, C.: Ion trap array mass analyzer: structure and performance. Anal. Chem. 81, 4840–4846 (2009)
Badman, E.R., Cooks, R.G.: A Parallel Miniature Cylindrical Ion Trap Array. Anal. Chem. 72, 3291–3297 (2000)
Pau, S., Pai, C.S., Low, Y.L., Moxom, J., Reilly, P.T.A., Whitten, W.B., Ramsey, J.M.: Microfabricated Quadrupole Ion Trap for Mass Spectrometer Applications. Phys. Rev. Lett. 96, 120801 (2006)
Xu, W., Li, L., Zhou, X., Ouyang, Z.: Ion sponge: a 3-dimentional array of quadrupole ion traps for trapping and mass-selectively processing ions in gas phase. Anal. Chem. 86, 4102–4109 (2014)
Wilpers, G., See, P., Gill, P., Sinclair, A.G.: A monolithic array of three-dimensional ion traps fabricated with conventional semiconductor technology. Nat. Nanotechnol. 7, 572–576 (2012)
Zhang, Z.P., Peng, Y., Hansen, B.J., Miller, I.W., Wang, M., Lee, M.L., Hawkins, A.R., Austin, D.E.: Paul trap mass analyzer consisting of opposing microfabricated electrode plates. Anal. Chem. 81, 5241–5248 (2009)
Austin, D.E., Wang, M., Tolley, S.E., Maas, J.D., Hawkins, A.R., Rockwood, A.L., Tolley, H.D., Lee, E.D., Lee, M.L.: Halo ion trap mass spectrometer. Anal. Chem. 79, 2927–2932 (2007)
Peng, Y., Hansen, B.J., Quist, H., Zhang, Z.P., Wang, M., Hawkins, A.R., Austin, D.E.: Coaxial ion trap mass spectrometer: concentric toroidal and quadrupolar trapping regions. Anal. Chem. 83, 5578–5584 (2011)
Hansen, B.J., Niemi, R.J., Hawkins, A.R., Lammert, S.A., Austin, D.E.: A lithographically patterned discrete planar electrode linear ion trap mass spectrometer. J. Microelectromech. Syst. 22, 876–883 (2013)
Li, A., Hansen, B.J., Powell, A.T., Hawkins, A.R., Austin, D.E.: Miniaturization of a planar-electrode linear ion trap mass spectrometer. Rapid Commun. Mass Spectrom. 28, 1338–1344 (2014)
Wang, M., Quist, H.E., Hansen, B.J., Peng, Y., Zhang, Z., Hawkins, A.R., Rockwood, A.L., Austin, D.E., Lee, M.L.: Performance of a halo ion trap mass analyzer with exit slits for axial ejection. J. Am. Soc. Mass Spectrom. 22, 369–378 (2011)
Zhang, Z., Quist, H., Peng, Y., Hansen, B.J., Wang, J., Hawkins, A.R., Austin, D.E.: Effects of higher-order multipoles on the performance of a two-plate quadrupole ion trap mass analyzer. Int. J. Mass Spectrom. 299, 151–157 (2011)
Gill, L.A., Amy, J.W., Vaughn, W.E., Cooks, R.G.: In situ optimization of the electrode geometry of the quadrupole ion trap1. Int. J. Mass Spectrom. 188, 87–93 (1999)
Moxom, J., Reilly, P.T.A., Whitten, W.B., Ramsey, J.M.: Sample pressure effects in a micro ion trap mass spectrometer. Rapid Commun. Mass Spectrom. 18, 721–723 (2004)
Kaiser, R.E., Cooks, R.G., Stafford, G.C., Syka, J.E.P., Hemberger, P.H.: Operation of a quadrupole ion trap mass spectrometer to achieve high mass/charge ratios. Int. J. Mass Spectrom. Ion Process. 106, 79–115 (1991)
Acknowledgements
The authors are grateful for financial support from the National Science Foundation (USA), Chemical Measurement and Imaging Program, award #1404886.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Tian, Y., Decker, T.K., McClellan, J.S. et al. Improved Miniaturized Linear Ion Trap Mass Spectrometer Using Lithographically Patterned Plates and Tapered Ejection Slit. J. Am. Soc. Mass Spectrom. 29, 213–222 (2018). https://doi.org/10.1007/s13361-017-1759-z
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13361-017-1759-z