Abstract
A new analytical technique using ICP-MS was applied to the determination of237Np in some environmental samples. The accuracy and precision of the new method were assessed by comparison with those of conventional NAA and α-spectrometric method as well as by analysis of the intercomparison sample of Ravenglass NRPB silt. The results obtained by ICP-MS were in good agreement with those by NAA or α-spectrometry at a relative deviation of 2–9%. The detection limit is 0.02 mBq/ml (26 mBq=1 ng).
Similar content being viewed by others
References
A. R. BYRNE, J. Environ. Radioactivity, 4 (1986) 133.
C. K. KIM, R. SEKI, N. IKEDA, Radioisotopes, 37 (1988) 229.
R. K. SCHULZ, G. T. WINK, L. M. FUJII, Soil. Sci., 132 (1981) 71.
M. SAKANOUE, M. YAMAMOTO, K. KOMURA, J. Radioanal. Nucl. Chem., 115 (1987) 71.
E. HOLM, A. AARKROG, S. BALLESTRA, J. Radioanal. Nucl. Chem., 115 (1987) 5.
A. L. GRAY, Analyst, 100 (1975) 289.
R. S. HOUK, V. A. FASSEL, G. D. FLESCH, H. S. SVEC, A. L. GRAY, C. E. TAYLOR, Anal. Chem., 52 (1980) 2283.
H. P. LONGERICH, B. J. FRYER, D. F. STRONG, Spectrochim. Acta, 42B (1987) 101.
D. W. BOOMER, M. J. POWELL, Anal. Chem., 59 (1987) 2810.
J. KYME, Chem. Geol., 68 (1988) 31.
J. R. DEAN, L. EBDON, R. MASSEY, J. Anal. At. Spectrom., 2 (1987) 369.
R. E. SERFASS, J. J. THOMPSON, R. S. HOUK, Anal. Chim. Acta, 188 (1986) 73.
T. G. TING, M. JANGHORBANI, Anal. Chem., 58 (1986) 1334.
H. KAWAGUCHI, Anal. Sci., 4 (1988) 339.
M. SELBY, G. M. HIEFTJE, Int. Lab., Jan/Feb (1988) 14.
D. S. POPPLEWELL, G. J. HAM, J. Radioanal. Nucl. Chem., 115 (1987) 191.
S. H. TAN, G. HORLICK, J. Anal. At. Spectrom., 2 (1987) 745.
A. R. DATE, Y. Y. CHEUNG, M. E. STUART, Spectrochim. Acta, 42B (1987) 3.
A. L. GRAY, J. G. WILLIAMS, J. Anal. At. Spectrom., 2 (1987) 599.
S. H. TAN, G. HORLICK, Appl. Spectrosc., 40 (1986) 445.
R. S. HOUK, Anal. Chem., 58 (1986) 97A.
A. L. GRAY, Spectrochim. Acta, 40B (1985) 1525.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Kim, CK., Takaku, A., Yamamoto, M. et al. Determination of237Np in environmental samples using inductively coupled plasma mass spectrometry. Journal of Radioanalytical and Nuclear Chemistry, Articles 132, 131–137 (1989). https://doi.org/10.1007/BF02060985
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02060985