Key points

  • An isolated beating pig heart MRI-compatible model was prepared and submitted to human physiological conditions of cardiac output.

  • In seven pig hearts, four-dimensional flow sequences were successfully acquired and intracardiac and coronary flow patterns could be visualised.

  • In two hearts, valve function and regurgitation after TAVR were also studied.

Background

Large animal models, in particular sheep and pigs, have provided indispensable and valuable insights in cardiac anatomy and physiology during health and disease, as they are easy to manipulate and to reproduce [1]. Additionally, working animal models have demonstrated to be realistic training models for cardiologic intervention [2]. However, laboratory animal husbandry of sheep or pigs is expensive, and medical experiments with large animals are subject to increased ethical objections in many countries. The use of isolated pig hearts harvested from pigs slaughtered for human consumption might be an acceptable alternative, as it leads to a reduction in the use of live animals. Additionally, isolated hearts have the advantage of controlled physiological conditions (e.g., controlled blood pressure) and experimental settings (e.g., surgical intervention or drug administration) [3].

A promising non-invasive imaging modality to measure three-dimensional (3D) blood flow is the time-resolved three-dimensional phase-contrast magnetic resonance imaging (MRI), i.e., four-dimensional (4D) flow MRI [4, 5]. This technique facilitates accurate 3D visualisation and quantification of blood flow, in conjunction with the quantification of cardiac parameters such as cardiac output (CO), stroke volume (SV), and regurgitation fraction [6]. Studies showed that 4D flow MRI is valuable for intracardiac blood flow visualisation [7], vortex detection [8], and flow assessment through all four heart valves [9, 10], as well as for detecting flow alterations after TAVR and other surgical procedures in the aorta [11,12,13,14]. Traditionally, transthoracic or transesophageal echocardiography is used to follow-up after surgical interventions like transcatheter aortic valve replacement (TAVR) [15]. However, it remains challenging to measure the extent and type of regurgitation or deviations in flow patterns with echocardiography [15, 16].

Using 4D flow MRI in animal models, in vivo experiments of juvenile pigs have demonstrated that intraventricular flow patterns change under drug-induced stress [17]. This technique has also revealed disturbed left ventricle (LV) flow patterns after mitral annuloplasty in sheep, correlated with the size of the annuloplasty ring [18].

In this study, we introduce an MRI-compatible platform for 4D flow measurements in isolated beating pig hearts, with real blood actively pumped by the LV at physiological pressures. This provides in vivo blood flow behaviour, resulting in physiological coronary flow and myocardial perfusion with good-to-excellent MRI contrast [3, 19]. In a previous study, Schuster et al. [3] presented an MRI-compatible isolated beating pig heart model to investigate cardiac perfusion. Its design was based on the Langendorff model [20], which allows for coronary perfusion and LV contraction, but not for LV filling and output. The working heart model presented in this study is similar to that described by Vaillant et al. [21]. It actively pumps blood in both ventricles, which, in contrast to Langendorff perfusion, causes natural coronary filling, mimicking physiological heart function and providing real control on the relevant physiological parameters.

We investigate the feasibility and reproducibility of 4D flow MRI in an ex vivo beating pig heart platform by measuring flow parameters and performing flow visualisation in five repeated pig heart experiments with native valves. In two additional experiments, we test the feasibility of the working platform for evaluating transcatheter aortic valve replacement (TAVR) procedures. We hypothesised that this platform may help investigate the influence of different surgical procedures on intracardiac flow and the performance of MRI sequences in different pathophysiological settings, without needing a live animal experiment.

Methods

Preparation

In this study, seven hearts were retrieved from pigs slaughtered for human consumption (Dutch Landrace hybrids, approximately 110 kg live weight). The protocols at the slaughterhouse and during the experiment were in agreement with European Community regulations 1069/2009 and 142/2011 regarding the use of slaughterhouse byproducts for research and were approved by the associated legal authorities of animal welfare. 4D flow MRI was applied in five pig hearts with native valves. In two additional experiments described in Additional file 3, the native values of two other hearts were replaced in the preparation phase of the experiment with TAVR valves by a specialist (A.d.W.) with 12 years of experience. All experimental and post-processing steps were the same as for hearts with native values. The prosthetic valves available were the CoreValve (29 mm, Medtronic Inc., Minneapolis, MN, USA) and Edwards SAPIENS XT (26 mm, Edwards Lifesciences LLC, Irvine, CA, USA). The prosthetic valve sizes were not selected according to the pigs’ annulus size, which is why the Edwards SAPIENS XT valve was additionally stabilised by a purse string suture.

Each slaughterhouse pig heart was harvested and arrested with crystalloid cardioplegic solution after a very short warm ischaemic time. Approximately 10 L of blood was collected from the same animal and heparinised. The heart was transported in an iced cardioplegic solution and connected to the MRI-compatible isolated beating pig heart platform [19] (PhysioHeartTM platform, LifeTec Group BV, Eindhoven, The Netherlands) on average 4 h after death. During preparation, the pericardial sac was discarded and the right pulmonary veins, vena cava inferior, vena cava superior, and the left azygos vein were tied off. A 27-mm cannula was inserted into the opened left pulmonary vein and secured with a purse string suture. The connecting tubes were compliant for blood flow and connected to preload simulating pulmonary resistances. A cannula with an internal diameter of 24 mm was inserted into the ascending aorta and fixed downstream to the aortic valve.

The aortic tube drained in an afterload simulating body resistance (systolic/diastolic pressure approximately 120/80 mmHg). A CO of approximately 4.5 L/min was maintained during the experiments to match human conditions. The CO was real time controlled via a flow sensor and regulated by adapting the preload and afterload resistances. A 17-mm cannula was inserted in the pulmonary artery, which was connected directly to a reservoir. The balanced blood was warmed and oxygenated by a heart lung machine outside the MRI room. The blood returned as a nutrient-rich, warm, oxygenated solution of 5% CO2 and 25% O2 and with a temperature of 38 °C. A schematic overview of the connections of pre- and afterload is shown in Fig. 1a.

Fig. 1
figure 1

Experimental setup of the MRI-compatible beating pig heart platform. a Experimental setup of the MRI-compatible PhysioHeartTM platform (LifeTec, Eindhoven, The Netherlands). The heart was connected to a preload and afterload system. The venous blood (purple) was heated and oxygenated outside the MRI room and pumped back to the experiment afterwards (red). The dotted lines indicate (1) coronary perfusion during the Langendorff mode and (2) the cross clamping of the aorta and the pulmonary vein. Ao Aorta, LA Left atrium, LV Left ventricle, PA Pulmonary artery, PV Pulmonary vein, RV Right ventricle. b The heart during Langendorff perfusion, attached to the aortic cannula and the pulmonary vein cannula. c The resuscitated, beating pig heart installed in the platform between the two coils on the patient bed before scanning. d The heating device and oxygenator outside the MRI room. e Setup during the scan, with two tubes leading to the MRI operator room

Reperfusion

To resuscitate the heart, a Langendorff perfusion [20] mode was created by cross clamping preload and afterload systems (Fig. 1b). Via a side port in the aortic cannula, blood was pumped retrograde in the aorta at approximately 80 mmHg closing off the aortic valve forcing flow into the coronary system. After perfusion of the myocardium, the deoxygenated blood drained in the coronary sinus, the right atrium, and the right ventricle. Through the pulmonary artery, blood was pumped back to the reservoir. The heart was left in Langendorff perfusion to recover for about 30 min until a steady state (in terms of a physiological colour and temperature, a stable sinus rhythm, a constant coronary flow, and constant pressures) was obtained (Additional file 1: Video S1).

Hereafter, the platform was switched to the working mode (Fig. 1c): Langendorff perfusion was stopped and the aorta and left pulmonary vein (preload) were opened, resulting in blood actively pumped by the LV with sufficient preload leading to a physiological CO. Coronary filling was created by LV function and aortic pressure only. To ensure a stable heart rate (HR) during MRI scanning, a pacemaker was attached to the heart. The paced HR was approximately 5–10 beats per min (bpm) over the irregular HR of the resuscitated heart. As soon as the heart was beating with a stable rhythm (approximately 20 min after switching from the Langendorff mode to the working mode), the setup was inserted into the MRI scanner (Fig. 1e) and the acquisitions were started.

4D flow MRI

All acquisitions were performed with a 3-T scanner (Ingenia, Philips Healthcare, Best, The Netherlands) using retrospectively triggered 4D flow MRI. Two medium flex coils (diameter 10 cm) were connected to the isolated beating pig heart platform below and above the heart (Fig. 1c). Electrocardiography sensors were attached via copper wires on the hearts’ surface and were used as a cardiac trigger signal. Twenty-four cardiac frames at a temporal resolution of 21 ms were acquired covering the cardiac cycle.

The 4D flow MRI scan had a field of view of 150 × 150 × 150 mm3 and a non-interpolated spatial resolution of 2.3 × 2.3 × 2.3 mm3. Echo time, repetition time, and flip angle were 2.2 ms, 5.2 ms, and 8°, respectively. To reduce the typically long acquisition times of 4D flow MRI, the scan was accelerated three times using k-t principal component analysis (Gyrotools, Zürich, Switzerland) resulting in a scan time of 12 min [22]. k-t principal component analysis acquisitions undersample k-space regularly over time, together with an interleaved training scan of the k-t space centre. The images are recovered during reconstruction by exploiting the relevant signal correlations available in the training data, which are represented as temporal basis functions and can be derived using a principal component analysis. The data was reconstructed using CRecon (Gyrotools, Zürich, Switzerland) with a k-t regularisation parameter of λ = 1. The velocity encoding was 100 cm/s.

Flow measurements

Visualisation of 4D flow MRI data and quantification of velocity and flow were done with GTFlow (Gyrotools, Zurich, Switzerland). For flow calculation, a region of interest (ROI) was chosen in the ascending aorta close to the sinotubular junction and downstream to the aortic valve. Net flow, forward flow, and backward flow were determined. The net flow was defined as the spatially averaged flow through the defined ROI:

$$ Q(t)={\int}_{\mathrm{ROI}}v\left(\mathbf{r},t\right)\ {d}^2\mathbf{r} $$

with v(r, t) being the velocity (pointing either in forward or backward direction) at position r and cardiac phase t.

Likewise, forward was defined as:

$$ Q{(t)}_{\mathrm{forward}}={\int}_{\mathrm{ROI}}v{\left(\mathbf{r},t\right)}_{v>0}\ {d}^2\mathbf{r} $$

and backward flow was defined as:

$$ Q{(t)}_{\mathrm{backward}}={\int}_{\mathrm{ROI}}v{\left(\mathbf{r},t\right)}_{v<0}\ {d}^2\mathbf{r}. $$

For the per cent quantification of the regurgitation fraction for one cardiac cycle T, forward flow volume was calculated as

$$ {V}_{\mathrm{forward}}={\int}_0^TQ{(t)}_{\mathrm{forward}}\ dt $$

and backward flow volume was calculated as:

$$ {V}_{\mathrm{backward}}={\int}_0^TQ{(t)}_{\mathrm{backward}}\ dt. $$

The per cent regurgitation fraction was based on the ratio RF = Vbackward/|Vforward|100. The stroke volume in millilitres was defined as SVflow = Vforward − |Vbackward|. The COflow in litres per minute was calculated by the product of the mean HR during the acquisition and SVflow.

In one heart, the 4D flow MRI scan was repeated three times within the same scan session at time points 0 min, 12 min, and 1 h 5 min.

Volume measurements

The LV volume was quantified for each heart using a 4D segmentation tool of velocity data from Medis (Medis medical imaging systems, Leiden, The Netherlands). Stroke volume was calculated as the difference between end-diastolic volume (EDV) and end-systolic volume (ESV). The per cent ejection fraction (EF) was defined as EF = SVvolume × 100/EDV. The size (long axis, short axis, and area) of the aortic annulus was measured during systole at the height of the aortic valve. Cardiac output was similarly calculated as the product of mean HR and SVvolume.

Statistical analysis

A statistical comparison between the flow and volumetric measurements was done using a Bland-Altman analysis and linear regression. Normal distribution of the data was tested using a Shapiro-Wilk test. The significance level was set to p < 0.05.

Results

For all seven hearts, the installation in the ex vivo beating pig heart platform and resuscitation was feasible. 4D flow MRI acquisitions were successful, and flow visualisation was feasible in all hearts. In Fig. 2, snapshot examples from one heart are shown. Figure 2a is a photograph of the heart before the MRI scan for anatomical reference. Figure 2b depicts reconstructed 4D flow MRI velocity vectors at peak systole. Velocity vectors of blood flow in the LV and the left atrium as well as coronary flow in the left anterior descending artery, the left circumflex artery, and the filling and ejection of the right ventricle were visualised. In the aortic root, late systolic vortex formation and filling of the coronary arteries were observed in all hearts as exemplified in Fig. 2c. Figure 2 d and e display velocity vectors of the LV at diastole and end-diastole in detail. In the left atrium and LV, filling and vortex formation after closing of the mitral valve were observed. Quantification of net flow Q(t) could be performed in ROIs downstream to the aortic valve (indicated in Fig. 2e), resulting in similar flow curves for all hearts as shown in Fig. 2f.

Fig. 2
figure 2

Velocity vector plots of one representative heart and aortic flow curves for all five hearts. a Anatomical reference of the heart before the experiment. b Vector plots of intracardiac flow in the ejection phase. c Vortex formation and coronary filling in the aortic root of the heart during diastole. For better visualisation of intracardiac flow, d shows the vector plots of the segmented LV at end-systole and e at end-diastole, where vortex formation can be observed. f Net flow downstream to the aortic valve for all five experiments, measured in an ROI indicated in e. Ao Aorta, LA Left atrium, LAD Left anterior descending artery, LCX Left circumflex artery, LV Left ventricle, PA Pulmonary artery, PV Pulmonary vein, ROI Region of interst, RV Right ventricle

Velocity vectors in the LV of all five hearts are presented in Fig. 3, showing filling and ejection of blood in the LV. In all hearts, late diastolic vortex formation was found as shown in the vector plots in Fig. 3c. A video of the vector plots over the cardiac cycle can be found in Additional file 2: Video S2. In Table 1, HR, SV, CO, and regurgitation fraction for these hearts are given. The cardiac functional parameters, averaged for all five valves, were HR = 105 ± 12 bpm, SVflow = 31 ± 6 mL, COflow = 3.3 ± 0.9 L/min, and regurgitation fraction = 16 ± 9%. Aortic annulus sizes were on average 31 ± 1 mm (mean ± standard deviation) for the long axis, 24 ± 2 mm for the short axis, and 600 ± 31 mm2 for the area.

Fig. 3
figure 3

Velocity vector plots of all five experiments at three time points in the cardiac cycle. Vector plots for the segmented left ventricle (LV) of all five experiments during (a) systole and (b) diastole. During end-diastole (c), vortex formation could be observed in all five hearts

Table 1 Cardiac parameters for the five hearts

Figure 4 shows Bland-Altman plots for CO and SV calculated from flow data and LV volume segmentations. The SVvolume calculated from LV volume was on average 11 mL higher than the SVflow calculated from flow data. COvolume was thus on average 1.2 L/min higher than COflow.

Fig. 4
figure 4

Linear regression and Bland-Altman plots for stroke volume and cardiac output calculated from flow and from volumetric measurements. Comparison of the two techniques used for cardiac output (CO) and stroke volume (SV) estimation using Bland-Altman analysis. a Linear regression and Bland-Altman plots for SV calculated from aortic flow data (SVflow) and from left ventricle (LV) volume (SVvolume). b Linear regression and Bland-Altman plots for CO calculated from aortic flow data (COflow) and from LV volume (COvolume)

The results of the flow curves, HR, COflow , and SVflow for the one heart where the 4D flow MRI scan was repeated three times within the same scan session at 0 min, 12 min, and 1 h 5 min are summarised in Additional file 3: Figure S1. These parameters were similar for all three scans.

The results of the two hearts in which a TAVR procedure was performed are summarised in Additional files 4, 5, 6, 7 and 8.

Discussion

In this study, 4D flow MRI was successfully applied in five pig hearts independently beating after installation and resuscitation in an MRI-compatible ex vivo beating pig heart platform. The platform provided detailed, comprehensive visualisations of time-resolved 3D, i.e., 4D, intracardiac blood flow and quantification of cardiac functional parameters. In literature, only two other setups have demonstrated ex vivo beating pig hearts in an MRI-compatible setup [3, 21], none of them using 4D flow MRI, as these scans can be very time consuming. In fact, advanced acceleration and reconstruction techniques are necessary to perform these scans in a feasible acquisition time. Additionally, the use of slaughterhouse byproducts as shown in this study is beneficial to reduce the number of in vivo animal experiments in view of ethical considerations as well as animal housing labour and costs.

The feasibility of the ex vivo beating pig heart platform for cardiac intervention was demonstrated in two additional experiments in which a TAVR procedure was conducted. In these TAVR experiments, 4D flow MRI was able to quantify and visualise the type and extent of aortic regurgitations, a typical complication after valve replacement procedure that is difficult to measure with echocardiography [15, 16].

Literature values for the SV of a healthy 60-kg pig’s LV are typically 65–81 mL [23] (in humans, 95 mL [24]). Literature values for the porcine HR are typically 85–114 bpm [23] (in humans, 60–100 bpm). This results in a porcine CO of 5–10 L/min [25] in comparison to a 4–8 L/min in humans. However, in our ex vivo beating pig heart platform, CO was artificially regulated via the preload, independently of the actual heart size. Therefore, at an experimental HR of 100–130 bpm, the porcine SV of the experiments shown in this study was expected to be lower and around 38–50 mL. Additionally, the SV calculated from flow measurements was slightly lower than that from volumetric measurements, since coronary flow could not be considered. In spite of the feasibility of visualising coronary flow, quantification was hampered by heart motion and spatial resolution.

In in vivo 4D flow MRI experiments of 12 juvenile pigs, Cesarovic et al. [17] found a stress HR of 124 ± 3 bpm which matched our experimental HR, similar also to the experimental HR reported by Vaillant et al. [21]. In their experiments, they found EDV = 55 ± 8 mL, ESV = 19 ± 6 mL, SV = 35 ± 4 mL, and an EF = 65% ± 7%. However, the difference between the in vivo experiment and the ex vivo beating pig heart experiment from this study is that pigs slaughtered for human consumption were used for the ex vivo model. The pigs reached a final live weight of approximately 110 kg, which is much more than a juvenile pig (30 kg). That weight difference was expressed in the large ESV and EDV measured in this experiment.

In general, complex flow patterns (such as vortices) are more difficult to capture with echocardiography as with 4D flow MRI. In our study, two late diastolic vortices were observed in the LV of all hearts (Figs. 2e and 3c). These vortices were also described by Elbaz et al. [8], who reported on vortex flow analysis during LV filling using 4D flow MRI in normal humans, and by Witschey et al. [18], who studied LV flow dynamics using 4D flow MRI in sheep.

This study has some limitations. The main idea was to use slaughterhouse waste material to allow an easy and cost-efficient setup. However, the use of pig hearts from large pigs raised for human consumptions limits the similarity between pig and human hearts. Additionally, an artificial human-like CO could limit the viability of the heart, which is why it might be an option to regulate the CO to another value. The hearts stayed vital for approximately 5 h, however would not survive several resuscitations, which limits their use for pre-procedure versus post-procedure measurements in the case of TAVR or any other intervention. As described by Driessen et al. [26], other cardiac MRI techniques such as black-blood, bright blood-balanced steady-state-free precession, perfusion, angiography, and T1-mapping sequences, are important to describe the full extent of cardiac function. As real blood is used, the ex vivo beating pig heart platform can be used for many of these MRI techniques, as well as for contrast-enhanced imaging, e.g., late gadolinium enhancement, as contrast agent could simply be added to the preload. A limitation for MRI techniques in general, however, could be the missing surrounding tissue, improving static phase offset corrections or shimming and which could also cause susceptibility artefacts. Also, the attachment of pacemaker lead cables on the heart’s surface can create some distortions. In general, the setup is also compatible with standard echocardiography instruments as well as pressure sensors and optical flow sensors. Another limitation is the use of five pig hearts in this feasibility study. The limited number of samples exaggerates the variation shown in the Bland-Altman plots. Including more hearts could reduce this variability.

An advantage of the isolated beating pig heart setup for MRI is that receiver radiofrequency coils can be placed close to the hearts’ surface, which results in high signal to noise and a high-flow contrast, even in very small anatomical structures such as the coronary arteries. The relatively small differences in velocity fields between the five hearts indicate good reproducibility of the experiments.

As shown in Additional files 4, 5, 6, 7 and 8 this platform has a potential for experiments on the performance of prosthetic valves or surgical strategies. Furthermore, with the isolated beating pig heart platform, aortic, pulmonary, and mitral regurgitation can be simulated. Complex settings of multiple regurgitations can be tested and investigated using 4D flow MRI. This is important since for example mitral regurgitation is common in patients with severe aortic regurgitation and aortic stenosis [27].

In conclusion, this study demonstrated the feasibility of 4D flow MRI data in a physiological working pig heart model pumping real blood and simulating physiological conditions. The isolated beating pig heart platform can allow for investigating new MRI sequences, pathophysiologic haemodynamics in relation with implanted artificial heart valves or also drug administration.