Introduction

It is unknown how to titrate PEEP in patients with COVID-19-related ARDS. Previous randomized controlled trials in patients with ARDS found that a higher PEEP versus a lower PEEP strategy did not reduce mortality rate [1,2,3]. However, higher PEEP did reduce mortality rate in patients with severe ARDS and tended to increase mortality rate in the subgroup with mild ARDS [4]. Apparently, there are patient subgroups that benefit from higher PEEP and subgroups that do not benefit from higher PEEP.

Initially, COVID-19-related ARDS was thought to be typical ARDS according to the Berlin definition of ARDS [5]. Later, a phenotype consisting of preserved respiratory system compliance and low lung recruitability was described. The application of lower PEEP was advised [6, 7]. Subsequent studies found that respiratory mechanics of patients with COVID-19-related ARDS and typical ARDS, including respiratory system compliance, were similar between groups. Significant heterogeneity was observed in COVID-19-related ARDS similar to non-COVID ARDS [8, 9]. The high PEEP-FiO2 table, currently recommended for the treatment of COVID-19-related ARDS, does not take into account patient heterogeneity [10].

From a pulmonary perspective, PEEP titration is finding a compromise between minimal alveolar overdistention and collapse. The recommended PEEP for lung protective ventilation strategies ranges between 5 cmH2O and 24 cmH2O [11]. A strategy consisting of recruitment maneuvers and titrated PEEP resulted in increased long-term mortality in patients with ARDS [12]. Several studies failed to find a benefit of a one-size-fits-all PEEP strategy, and more research into tailoring of PEEP to the individual patient is recommended [13]. Therefore, we think it is crucial to quantify the amount of alveolar overdistention and collapse at the bedside among parameters reflecting respiratory mechanics. EIT can be used for detecting and quantify regional alveolar overdistention and collapse and allows for personalized PEEP titration [14, 15] In a recent randomized controlled trial in ARDS patients, He et al. [16] showed titration using EIT resulted in a decoupling between PEEP and FiO2, but no difference in long-term outcome compared to a PEEP/FiO2 table. Hsu et al. [17] compared PEEP titrated using EIT to PEEP set based on pressure–volume loops. They found the EIT leads to lower PEEP and a higher survival rate.

We retrospectively describe a cohort of patients with COVID-19-related ARDS in whom an EIT-guided PEEP trial was used to personalize PEEP. The aim of this study was to compare PEEP set by EIT and baseline PEEP according to the high PEEP-FiO2 table [1]

Methods

Study design and inclusion criteria

This is a retrospective analysis of a cohort study conducted between March 1 and June 1 2020 in the general intensive care unit (ICU) of the Erasmus MC, Rotterdam, The Netherlands. The first 15 patients in this study have been described previously [18] All patients that met the following criteria were included: 1. age ≥ 16 years; 2. established COVID-19 following a SARS-CoV-2 positive polymerase chain reaction; 3. moderate-to-severe ARDS according to the Berlin definition of ARDS [5]; and 4. intubated and on controlled mechanical ventilation. The Erasmus MC is a tertiary referral hospital, and some patients were intubated and mechanically ventilated elsewhere before transfer to our ICU. A PEEP trial guided by EIT was performed following admission to the ICU according to our local COVID-19 mechanical ventilation protocol. The PEEP trial was performed once in every patient and was not routinely repeated. We did not perform a PEEP trial if patients had a contra-indication for EIT belt placement (e.g., pacemaker, spinal cord injury), inadequate EIT signal (e.g., thoracic bandages, undrained pneumothorax), or hemodynamic instability (MAP < 60 mmHg despite optimization of fluid status and/or use of vasopressors). The Medical Ethical Committee of the Erasmus MC approved this study. According to Dutch legislation, no informed consent was required for the retrospective use of anonymized patient data.

Study protocol

All patients were ventilated in pressure control mode. Baseline PEEP was set by the attending clinician. The protocol was prescribed using the high PEEP-FiO2 table, but the clinician had the freedom to choose the PEEP and FiO2 combination within the limits of the table [1]. Patients were fully sedated (Richmond Agitation-Sedation Scale -5) with continuous intravenous infusion of propofol, midazolam and/or opiates. Persisting spontaneous inspiratory efforts were prevented with increased sedation or neuromuscular blockade (rocuronium). Mean arterial blood pressure (MAP) was measured continuously, and noradrenalin was administered to maintain MAP above 65 mmHg prior to the PEEP trial. The fraction of inspired oxygen (FiO2) was titrated to obtain a peripheral oxygen saturation (SpO2) between 92 and 95%.

The PEEP trial was guided by one of two EIT devices, based on availability: Pulmovista 500, Dräger, Germany or Enlight 1800, Timpel, Brazil. An EIT belt containing surface electrodes was placed in the transversal plane at the 4th–5th intercostal parasternal space according to manufacturer’s instructions. Regional ventilation data were visualized on screen during the entire PEEP trial without repositioning the EIT belt. The PEEP titration tool of the EIT devices was used to guide PEEP titration.

A decremental PEEP trial was performed starting at the baseline PEEP according to the high PEEP-FiO2 table (PEEPbase). We increased the airway pressure until PEEP was 10 cmH2O above PEEPbase with a minimum of 24 cmH2O, corresponding to the maximum PEEP advised by the high PEEP-FiO2 table [1]. The imposed driving pressure (i.e., the difference between PEEP and peak pressure) remained unchanged during the trial. In case of hypotension (MAP < 60 mmHg) or desaturation (SpO2 < 88%), PEEP was limited to the highest airway pressure without hypotension or desaturation. We aimed to maintain PEEP for at least one minute in order to establish a constant electrical impedance signal. PEEP was reduced in steps of 2 cmH2O every 30 s until continuous EIT monitoring showed evident collapse as compared to maximum PEEP. To confirm a further increase in collapse, PEEP was lowered an additional 2 cmH2O. Subsequently, we performed a small recruitment maneuver at the highest PEEP used during the PEEP trial for 30 s. The PEEP titration tools of both EIT devices provided a percentage of relative alveolar overdistention and collapse at every PEEP step. PEEP was set (PEEPset) at the lowest PEEP step above the intersection of the curves representing relative alveolar overdistention and collapse, as described previously (see Fig. 2C) [18, 19].

Mechanical ventilation, SpO2 and hemodynamic parameters were recorded at PEEPbase and after 30 min of PEEPset. Plateau airway pressure (Pplat) and total PEEP were measured during an inspiratory and expiratory hold procedure, respectively. We used the last arterial blood gas before and the first arterial blood gas after the decremental PEEP trial for calculation of the PaO2/FiO2 ratio at PEEPbase and PEEPset, respectively. Patient characteristics were extracted from the patient information system.

The primary goal of this study was to compare PEEPbase with PEEPset. Secondary goals were to compare respiratory mechanics and oxygenation before and after the PEEP trial. Subsequently, we identified the patients in whom PEEPset according to the EIT-guided PEEP trial was decreased by ≥ 2 cmH2O (PEEPlower) or was increased by ≥ 2 cmH2O (PEEPhigher) as compared to PEEPbase. Patients with a change in PEEPset of less than 2 cmH2O as compared to PEEPbase were assigned to a third group: PEEPequal. The change in percentage of relative alveolar overdistention and collapse were reported between PEEP 24 cmH2O and PEEP 12 cmH2O, because both PEEP levels were reached during the PEEP trial in 93% (n = 70) of patients.

Statistical analysis

Data were presented as mean (standard deviation), median [25th-75th percentile] or count (percentage). Data were tested for normality using the Shapiro–Wilk test. The Student independent T test or Mann–Whitney U test was used for the comparison between two groups. A one-way ANOVA or the Kruskal–Wallis test was used for the comparison between three groups. The Student dependent T test or Wilcoxon signed-rank test was used to compare changes from baseline within patients. A repeated measures ANOVA or Friedman test was used to compare changes over more than two levels. The Chi-square test was used to compare frequencies. Bonferroni correction was applied to correct for multiple testing. Spearman’s rank correlation coefficient (ρ) was used for calculation of correlations between variables. A p value of ≤ 0.05 was considered to be statistically significant.

Results

Seventy-five mechanically ventilated patients with COVID-19-related ARDS were included in this retrospective cohort study. Patients had a median age of 64 years [54–71] and a body mass index (BMI) of 30.4 kg/m2 (5.8). Median APACHE IV score at ICU admission was 70 (27) and median time since intubation was 3 days [1–8].

In the entire cohort, we did not observe a difference between the median PEEP level before and after the PEEP trial (Table 1). After the PEEP trial, there was a small increase in static compliance and tidal volume. In 31% of patients (n = 23), PEEP was decreased by ≥ 2 cmH2O, and in 32% of patients (n = 24), PEEP was increased by ≥ 2 cmH2O (Fig. 1). The remaining 28 patients (37%) were assigned to PEEPequal group. EIT images of a representative patient from the PEEPlower group and the PEEPhigher group are shown in Fig. 2. In five patients (7%), a PEEP of 24 cmH2O could not be applied due to desaturation (n = 4) or hypotension (n = 1). Desaturation occurred only in the PEEPlower group. One (1%) pneumothorax was observed following central catheter placement.

Table 1 PEEPbase versus PEEPset
Fig. 1
figure 1

Change in PEEP following the EIT-guided PEEP trial. All 75 patients in this cohort are shown on the x-axis. On the y-axis, the change in PEEP (PEEPset–PEEPbase) is presented. We identified the patients in which PEEPset according to the EIT-guided PEEP trial was decreased by ≥ 2 cmH2O (PEEPlower in blue) or was increased by ≥ 2 cmH2O (PEEPhigher in green) as compared to PEEPbase. Patients with a change in PEEPset of less than 2 cmH2O as compared to PEEPbase were assigned to a third group: PEEPequal (in orange)

Fig. 2
figure 2

Overdistention and collapse for two typical patients. a — A patient assigned to the PEEPhigher group. The top row displays ventilation distribution at four levels of PEEP (left to right: 28, 20, 16, and 10 cmH2O). Black indicates no ventilation, various shades of blue indicate ventilation, and white indicates the region where most ventilation is detected. At high PEEP levels, ventilation occurs in the dorsal lung regions, whereas the center of ventilation shifts ventrally at lower PEEP levels. The distribution of alveolar overdistention (orange) and alveolar collapse (white) is shown in the bottom row. At high PEEP levels, only a small amount of ventilation is detected in the ventral region as a result of overdistention. At low PEEP levels, ventilation shifts ventrally as a result of alveolar collapse. High PEEP resulted in a relatively small increase in alveolar overdistention and a large decrease in alveolar collapse. This patient was considered to have high recruitability and total PEEP was set at the lowest PEEP step above the intersection of both curves: 20 cmH2O., b — A patient assigned to the PEEPlower group. The top row displays ventilation distribution at four levels of PEEP (left to right: 24, 20, 12, and 6 cmH2O). A decrease in PEEP resulted in an increase in ventilation (light blue to white), and even at a low PEEP level of 6 cmH2O, ventilation in the dorsal lung regions is more or less preserved. At high PEEP levels, the relative amount of alveolar overdistention is 29%, which indicates severe alveolar overdistention. In contrast, at low PEEP levels, the amount of alveolar overdistention is significantly reduced, and only a small amount of alveolar collapse is identified (8%). This patient was considered to have low recruitability, and PEEP was set at 10 cmH2O., c — This plot represents the relative amount of alveolar overdistention and collapse as measured by EIT during a decremental PEEP trial. PEEP was set above at the lowest PEEP step above the intersection of the curves representing relative amount of alveolar overdistention and collapse (vertical lines). The patient in PEEPhigher group (orange triangles) had lower amounts of relative overdistention and higher amounts of alveolar collapse at the same PEEP level as compared to the patient assigned to a patient in PEEPlower group (blue dots)

Patients in the PEEPlower group had a lower BMI, a higher APACHE IV score, a longer time between onset of symptoms and intubation, a higher d-dimer concentration at ICU admission, and a higher incidence of pulmonary embolism as compared to the other groups (Table 2). In the entire cohort, we found a statistically significant correlation between PEEPset and BMI (ρ = 0.59, p value < 0.001), and time between onset of symptoms and intubation (ρ = − 0.42, p value < 0.001). We did not observe a correlation between PEEPset and APACHE IV score (ρ = − 0.10, p value 0.41) or d-dimer concentration (ρ = − 0.06, p value 0.66). PEEPset resulted in an increase in tidal volume at the same driving pressure (static compliance was not significantly changed after Bonferroni correction) compared to PEEPbase in patients in the PEEPlower group, but not in the other two groups (Table 3). In addition, we observed a significant reduction in plateau pressure in the PEEPlower group. In the patients assigned to PEEPhigher group, we observed a significant increase in plateau pressure. There was a trend for a higher PaO2/FiO2 ratio, but this was not significant after Bonferroni correction.

Table 2 Comparison of baseline characteristics between PEEP groups
Table 3 Comparison of respiratory mechanics between PEEP groups

Additional file 1: Tables S1–4 in de Supplementary Materials show the respiratory mechanics at the baseline PEEP, the highest and lowest PEEP during the decremental PEEP trial and the set PEEP immediately after it was set for the entire cohort (Additional file 1: Table S1) and the PEEPlower (Additional file 1: Table S2), PEEPequal (Additional file 1: Table S3) and PEEPhigher (Additional file 1: Table S4) groups.

The relative percentages of alveolar collapse and overdistention at PEEPset did not differ between groups (Table 4). In the PEEPlowergroup, relative alveolar collapse increased by 8.3% (3.6–15.4%) in the dependent lung region and relative alveolar overdistention decreased by 28.3% (22.2–43.2%) in the non-dependent lung region as a result of a PEEP decrease from 24 to 12 cmH2O. In contrast, in PEEPhigher group, this was 24.2% (20.1–29.2%) and 15.1% (4.4–26.2%), respectively.

Table 4 Comparison of alveolar collapse and overdistention between PEEP groups

Discussion

Based on the EIT-guided PEEP trial, PEEP was decreased in 31% of patients and increased in 32% of patients. We found a significant positive correlation between PEEPset and BMI. Patients in PEEPlower group had improved respiratory mechanics after the PEEP trial, a lower BMI, longer time between onset of symptoms and intubation, and a higher incidence of pulmonary embolism during ICU admission. In patients in the PEEPlower group, an increase in PEEP resulted in major alveolar overdistention and a small amount of recruitment on EIT. In PEEPhigher group, we observed a significant increase in plateau pressure and improved oxygenation after the PEEP trial. In addition, an increase in PEEP resulted in significant alveolar recruitment and small amounts of alveolar overdistention on EIT. Hence, the latter group should be considered as recruitable. The PEEP trial was relatively safe, as 5% of patients had a desaturation and 1% of patients was hypotensive during the PEEP trial.

PEEPset resulted in a trend toward improved respiratory mechanics in the PEEPlower group and improved oxygenation in the PEEPhigher group. Both an improved driving pressure and improved oxygenation after a change in PEEP are associated with reduced mortality rate in patients with ARDS [20]. Therefore, we should aim to identify the patients that are likely to respond to PEEP, i.e., recruitability.

Recruitability is the amount of collapsed lung tissue that has the potential for reaeration at higher airway pressures [21]. An increase in PEEP in the patients in PEEPlower group resulted in major alveolar overdistention and a small amount of alveolar recruitment, whereas the patients in PEEPhigher group had significant alveolar recruitment and less alveolar overdistention. In patients with COVID-19-related ARDS, alveolar recruitment does not necessarily result in an increase in static compliance [22]. Thus, patients in PEEPlower group were considered to have low recruitability, patients in PEEPequal group had intermediate recruitability and patients in PEEPhigher group had high recruitability.

Until now, we tended to focus on the identification of patients that had high recruitability [23]. However, it might also be beneficial to identify the patients that have low recruitability and are prone to alveolar overdistention. Patients with low recruitability had a lower BMI, a higher incidence of pulmonary embolism, and a longer time between onset of symptoms and intubation. Patients with obesity have lower transpulmonary pressures and lower end-expiratory long volumes as a result of higher pressure from the chest wall [24]. BMI has a positive correlation with recruitability and the use of higher PEEP, as higher PEEP increases transpulmonary pressure and reduces alveolar collapse [18]. In addition, patients in PEEPlower group had a higher incidence of pulmonary embolism during ICU admission. These findings suggest that disturbed pulmonary perfusion, resulting in a ventilation-perfusion mismatch, caused hypoxemia in these patients. Nevertheless, all patients had a reduced static compliance, possibly leading to disturbed minute ventilation or increased dead space fraction as well [25]. Patients in PEEPlower group had a longer time between onset of symptoms and intubation. This could indicate that these patients may have had some form of patient self-inflicted lung injury or pulmonary fibrosis [26]. Unfortunately, we had too few CT scans at the day of PEEP titration to test this hypothesis. The PEEP trial did not reach a maximum PEEP of 24 cmH2O in four (5.3%) patients because of desaturation. These four patients were assigned to the PEEPlower group and had large amounts of alveolar overdistention. Desaturation at high PEEP could be a clear indication of ventilation-perfusion mismatch, likely due to alveolar overdistention.

An observational cohort performed in the Netherlands found a median PEEP titrated by the clinician of 14.0 cmH2O (11.0–15.0) [27]. Two small observational cohorts that used EIT to titrate PEEP found a median PEEP of 12.0 cmH2O [28, 29]. In our EIT-guided population, we found a higher median PEEP of 18.0 cmH2O (14.0–20.0) as compared to the other studies. Explanations are the relatively high BMI in our cohort and long duration of mechanical ventilation in the cohort of Sella et al. [29]. In addition, there is no consensus on how to interpret EIT data obtained during a PEEP trial [14, 30].

In our study, total PEEP was arbitrarily set at the PEEP level above the intersection of the curves representing relative alveolar overdistention and collapse [18, 19]. We chose this method as it is an intuitive and simple approach that can be performed at the bedside, but arguably assumes that both alveolar overdistention and collapse are equally harmful [31]. Both Perrier et al. [28] and Sella et al. [29] chose to set PEEP at the intersection of both curves itself, whereas Franchineau et al. [32] chose to limit alveolar collapse at 15%, independent of alveolar overdistention. The last approach favors alveolar collapse over alveolar overdistention and likely resulted in a lower set PEEP as compared to the method used in this study. Future research should focus on the best approach to titrate PEEP based on EIT data and its association with clinical outcomes.

Previous randomized controlled trials in patients with ARDS compared PEEP titrated using EIT to conventional methods. In patients with mild-to-severe ARDS, He et al. [16] showed EIT resulted in a similar PEEP compared to the PEEP/FiO2 table, but was decoupled from FiO2. In patients with moderate-to-severe ARDS, Hsu et al. [17] showed PEEP and mortality rate were lower using EIT compared to pressure–volume loops, but mortality rate was high overall (31% in de EIT group and 56% in the control group, versus 21–27% in the study by He et al. [16] and 29% in the current study). In our study, PEEP was not changed on average for the entire cohort after titration using EIT, but was changed with ≥ 2 cmH2O in the majority of patients.

This study has several limitations. First, this retrospective analysis was not prespecified in the study protocol and results should be considered hypothesis-generating. The main purpose of this EIT-guided PEEP trial protocol was to improve clinical practice. As a consequence, mechanical ventilation parameters were only recorded at PEEPbase and PEEPset, limiting a more accurate retrospective analysis of the PEEP trials and EIT data at every PEEP step. A major limitation of this study is the lack of randomization and of the sequence of interventions. All patients received PEEP set by the clinician using the PEEP-FiO2 table first and then the EIT-guided PEEP trial. A part of the improvements in oxygenation and respiratory mechanics may be due to the PEEP trial itself, instead of the titration of PEEPset. This is noticeable in the changes in respiratory mechanics for the PEEPequal group. Second, only patients with COVID-19-related ARDS were included in this study. Although respiratory mechanics in non-COVID-19-related ARDS and typical ARDS seem to be similar, it is uncertain whether results can be generalized to the non-COVID-19-related ARDS population [8, 9]. Third, maximum and minimum PEEP reached in all trials varied. The estimation of the amount of collapse and overdistention is based on the maximum compliance for each EIT pixel. It is probable or even likely maximum compliance is not reached for all pixels, e.g., due to residual collapse in the dependent lung at the highest PEEP level. Therefore, approximately 0% alveolar collapse at PEEP 24 cmH2O does not necessarily mean that application of higher or prolonged airway pressures cannot result in additional alveolar recruitment. Fourth, we performed a PEEP trial with small steps of 2 cmH2O and a short step duration of 30 s. Some other studies report larger steps and longer duration for similar PEEP trials [15, 28, 32]. There is a tradeoff between step size, step duration and the time it takes to complete the protocol. After a change in PEEP, respiratory mechanics can change in multiple ways with different time frames. By rapidly changing PEEP, we did not allow for slow effects like slow derecruitment, morphological changes to the abdomen and diaphragm, changing hemodynamics and changes in pO2 and pCO2. In addition, as a result of the large numbers of patients with COVID-19, we chose a time-efficient study protocol. Fifth, hemodynamic monitoring was limited to continuous measurement of blood pressure and heart rate. PEEP titration is more than balancing alveolar overdistention and collapse, as PEEP influences cardiac output as well [33]. Although the PEEP trials had limited effects on blood pressure and heart rate, we cannot exclude a decrease in cardiac output. In addition, we did not assess pulmonary perfusion with EIT. Hence, EIT-guided PEEP titration might have resulted in optimal ventilation, but not necessarily in an optimal ventilation-perfusion match. Sixth, ventilation distribution assessed by EIT is measured in only a small cross-sectional slice of the lung. Ventilation distribution changes when the EIT belt is placed more cranially or caudally, further complicating EIT-guided PEEP titration [34]. Seventh, we used devices from two manufacturers to perform the EIT measurements. Although the devices apply the same algorithm by Costa et al. [15] to derive the relative collapse and overdistention, results could vary due to differences in belts, reconstruction models and algorithms. Additional file 1: Tables S5–12 in the supplementary materials show the results presented in Tables 1, 2, 3, 4 split by EIT device. Considering the limited data, it seems possible the Timpel Enlight 1800 gives higher values overdistention at high PEEP compared to the Dräger Pulmovista 500. Due to the small amount of measurements with the Timpel device (n = 7), we were not able to properly compare the devices. Overall, considering only the measurements with the Dräger device (n = 68) does not change our interpretation or conclusions.

In conclusion, a PEEP trial guided by EIT as compared to PEEP titration based on the PEEP-FiO2 table resulted in a clinically relevant change in PEEP in 63% of patients with COVID-19-related ARDS. We found a significant positive correlation between set PEEP and BMI. Patients in whom PEEP was decreased had a lower BMI, a longer time between onset of symptoms and intubation, and a higher incidence of pulmonary embolism. Our results support the hypothesis that PEEP should be personalized in patients with COVID-19-related ARDS in order to reduce the total amount of alveolar overdistention and collapse, i.e., too low or too high PEEP.