Introduction

Pelvic fractures constitute up to 2–8% of all skeletal fractures. Pelvic fractures are usually unstable and are caused by high-energy trauma with a disability rate of 60% and a mortality rate of more than 13% [1, 2]. Unstable pelvic ring fractures often require simultaneous fixation of both the anterior and posterior pelvic rings to avoid serious complications, including chronic pain and sexual, bowel and bladder dysfunction [3,4,5].

Traditional fixation methods include open reduction internal fixation with plates from the anterior and posterior pathways, external fixation, and minimally invasive percutaneous screw fixation [5, 6]. With the advancement of medical technology and the popularization of minimally invasive concepts, the clinical application of open reduction and plate fixation is less frequent at present [6]. The percutaneous IS screw fixation is becoming increasingly popular worldwide with the advantages of less surgical trauma, less bleeding, fewer infection rates, fewer complications, earlier mobilization, and benefits for second-stage anatomical reconstruction surgery [6]. However, due to the unique anatomical structure of the posterior pelvic ring and the numerous influencing factors during C-arm fluoroscopies, such as morbidly obese patients or bowel gas, it is challenging to ensure the best position of each screw through manual operation under X‐ray monitoring [7]. Only a 4°deviation during the iliosacral screw insertion can lead to penetration of the anterior cortex of the sacrum or S1 foramina [8]. Several studies showed that the malposition rates of IS screw under traditional fluoroscopic guide were 2–15%, and the neurologic injury rates were 1–7% [8, 9]. Another limitation is that static two-dimensional (2D) intraoperative fluoroscopic imaging is inadequate for three-dimensional (3D) fragment alignments, necessitating repeat intraoperative images, which leads to prolonged radiation exposure to the patient and medical staff [10].

Furthermore, up to 44% of patients have Sacral 1 dysmorphism, which may influence the position of the screw [11, 12]. Percutaneous insertion is a highly demanding and challenging operative technique due to the complex pelvic anatomy, requiring skillful correlation of fluoroscopic images and bony landmarks [13]. Therefore, we need to find a safe and effective method to solve these problems.

Robotics and artificial intelligence applied to orthopedics have become an exciting topic. Various robot-assisted orthopedic surgery (RAOS) is currently available on the market. Each addresses specific surgeries and is characterized by special features that may involve different requirements or modus operandi. RAOS has been utilized in a variety of orthopedic operations, including fracture fixation for traumatic surgery, total hip and total knee arthroplasty [14, 15], spine surgery [16, 17], bone tumor surgery [18], arthroscopy [19], and fracture fixation for traumatic surgery [20]. However, few RAOS focus on screw implantation in pelvic fractures. The third RAOS, “TiRobot” (TINAVI Medical Technologies Co. Ltd., Beijing, China) from China has been certified by the China Food and Drug Administration (CFDA). It is the only robot that has been reported in the literature for posterior pelvic injuries. To our knowledge, there has been only one meta-analysis comparing TiRobot‑assisted versus fluoroscopy-assisted percutaneous sacroiliac screw fixation for pelvic ring injuries. However, the previous meta-analysis only included four observational studies, and all included studies are English articles. Considering that the TiRobot is only used in China, the restriction of the previous meta-analysis to English-language publications potentially limits the power that could be obtained with the inclusion of patient enrollment from Chinese-language studies. Finally, the previous meta-analysis did not do the pooled analysis of screw accuracy, reduction accuracy, and overall excellent and good rates of Majeed score. Thus, based on the current studies comparing TiRobot‑assisted versus fluoroscopy-assisted percutaneous sacroiliac screw fixation. We performed a meta-analysis, which includes English studies and Chinese studies. Moreover, our results included pooled analysis of screw accuracy, reduction accuracy, and overall excellent and good rates of Majeed score, which would provide a more exact conclusion and could supplement the previous meta-analysis.

Methods

The current meta-analysis was registered on PROSPERO (International prospective register of systematic reviews), and the registration number was CRD42022318898. We strictly followed the PRISMA (preferred reporting items for systematic review and meta-analysis) guidelines to conduct this analysis according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses statement [21].

Search strategy

We identified studies involving TiRobot‑assisted and fluoroscopy-assisted percutaneous sacroiliac screw fixation for pelvic ring injuries in electronic databases up to April 2022. The databases included Web of Science, Embase, PubMed, Cochrane, Controlled Trials Register, Cochrane Library, Highwire, CBM, CNKI, VIP, and WanFang database up to April 2022. The keywords used were “TiRobot,” “robot,” “robotic,” “pelvic fracture,” “screw fixation,” “percutaneous,” and “pelvic ring injury” in conjunction with Boolean operators “AND” or “OR.” There was no limitation on publication dates and language. We also reviewed the references of included studies and previous systematic reviews to identify other published or unpublished studies.

Inclusion criteria

All randomized controlled trials (RCTs) and non-randomized controlled trials (non-RCTs) comparing TiRobot‑assisted and fluoroscopy-assisted percutaneous sacroiliac screw fixation for pelvic ring injuries were identified and included in the search strategy. These studies should meet the following inclusion criteria: (1) All papers with data on pelvic fracture were included. (2) TiRobot‑assisted percutaneous sacroiliac screw fixation for pelvic ring injuries was involved. (3) The comparator was conventional fluoroscopy-assisted percutaneous sacroiliac screw fixation in the original comparative study. (4) At least one of the following indexes was reported: radiation exposure time of screw implantation, frequency of intraoperative fluoroscopy, guide wire attempts, intraoperative blood loss, operative time, fracture healing time, screw accuracy, Majeed score, overall excellent and good rates of Majeed score and reduction accuracy. The operative time was defined from the start after sterile draping to the end of skin closure, which means that for the robot group, the time for robot preparation is also included in the intraoperative time. During each X-ray exposure, the fluoroscopy system calculated and displayed the real-time exposure seconds. The total fluoroscopy time of each screw implantation was recorded, which included the fluoroscopy for verifying the correct position of the screw placement after screw-setting. Intraoperative blood loss is the sum of the amount of blood through the suction apparatus and the bleeding volume at the gauze. We also excluded studies that had unclear or incomplete sample data.

Data extraction process

Two independent researchers scanned the titles and abstracts of all literature searched, and they independently extracted the available data from each study. After excluding the trials which did not meet the inclusion criteria, we read the full text of the literature that might meet the inclusion criteria to determine whether this literature ultimately met the inclusion criteria. We extracted data based on the following: (1) research features (i.e., authors, year of publication, type of study, and patient population), (2) population information (i.e., number of patients, gender, age, BMI, and type of injury), (3) pelvic fracture information (i.e., fracture type, comorbidity, the average time from injury to surgery), and (4) outcome information. We addressed the disagreements by discussion or by involving a third author. We consulted the corresponding author to request missing data. If the necessary results are omitted, the first author will contact the authors by email or other means to obtain more data if necessary.

Assessment of studies

Two authors independently evaluated the methodological quality. We assessed the non-randomized studies using the nine-star Newcastle–Ottawa Scale (NOS), a validated tool suitable for evaluating the quality of non-randomized studies [22]. According to the Cochrane Handbook for Systematic Reviews of Interventions [23], the methodological quality and basis of the RCTs were assessed as follows: randomization, allocation concealment, blind method, selective reporting, group similarity at baseline, incomplete outcome data, compliance, timing of outcome assessments, and intention-to-treat analysis. Two researchers independently assessed the studies, and disagreements between them were resolved through discussions with a third author or consensus.

Ethical consideration

Ethical approval is not required because this study is based on the existing literature. The findings of this systematic review will be disseminated through a peer-reviewed journal.

Statistical analysis

Heterogeneity was evaluated using the chi-square statistic and was quantified using I2. P values ≤ 0.1 or I2 value > 50% suggested high heterogeneity. Thus we used the randomized-effects model. Otherwise, we used the fixed-effects model [21]. In each study, we used the relative risk (RR) and relevant 95% confidence interval (CI) to measure dichotomous variables such as overall excellent and good rates of Majeed score, reduction accuracy, and screw accuracy. The mean difference (MD) or standard MD was used to assess continuous outcomes such as radiation exposure time of screw implantation, frequency of intraoperative fluoroscopy, guide wire attempts, intraoperative blood loss, operative time, fracture healing time, and Majeed score with a 95% confidence interval (CI). We used statistical algorithms to estimate the standard deviation for those studies that provided only continuous variables for means and range [24]. If P values were less than 0.05, we considered the results as a statistically significant difference. Sensitivity analysis was used to assess the stability of the results (if necessary). Pooled effects of this meta-analysis were calculated using STATA SE version 15.0 (The Nordic Cochrane Center, The Cochrane Collaboration, Copenhagen, Denmark).

Results

Search results

The literature search and selection process are shown in Fig. 1. Finally, eleven publications from 2017 to 2022 were included in our meta-analysis. The detailed literature screening process is shown in the PRISMA flow diagram in Fig. 1. We identified 237 relevant citations from the databases according to the literature search strategy described earlier. After deleting 188 duplicates, we obtained 49 articles. After reviewing the titles and abstracts of the 49 remaining articles, 28 irrelevant clinical studies were excluded. By reading the 21 full-text articles, we excluded another ten for the following reasons: (systematic) reviews, non-compare groups, cadaver research, animal research, and no useful outcome data. The remaining eleven articles were deemed appropriate. Finally, we identified 628 patients assessed in 2 RCTs [25, 26] and 9 non-RCTs [27,28,29,30,31,32,33,34,35]

Fig. 1
figure 1

We identified 237 relevant citations from the databases. After deleting 188 duplicates, we obtained 49 articles. After reviewing the titles and abstracts of the 49 remaining articles, 28 irrelevant clinical studies were excluded. By reading the 21 full-text articles, we excluded another 10. The remaining 11 articles were deemed appropriate

In Wang’s study [34], two subgroups compared TiRobot‑assisted versus conventional fluoroscopy-assisted percutaneous sacroiliac screw fixation. One is the S1 group (only S1 screws were implanted), and the other is the S2 group (only S2 screws were implanted) according to the location of the implanted screw. So we divided the study into two groups, Wang 2020(S1) and Wang 2020(S2).

Study characteristics and quality

The detailed baseline characteristics and general intervention information are presented in Tables 1, 2, 3, and 4. The articles were published in English and Chinese between 2017 and April 2022.

Table 1 Summary of studies’ characteristics
Table 2 Summary of patient demographic details for each study
Table 3 Detailed characteristics of pelvic fracture information
Table 4 Detailed characteristics of outcome information

Risk-of-bias assessment

The methodological quality of the involved studies ranged from seven to nine (Table 5). The risk-of-bias summary and risk-of-bias graph for RCTs are presented in Table 6. As a result, the overall quality of the included studies was considered adequate.

Table 5 Risk-of-bias assessment for the studies included in the meta-analysis (NOS)
Table 6 Methodological assessment according to six domains of potential biases (Cochrane Risk-of-Bias Tool)

Outcomes of meta-analysis

The radiation exposure time of screw implantation

Eight studies reported radiation exposure time of screw implantation; we found statistical heterogeneity between the two groups (x2 = 83.9, df = 7, P = 0.000; I2 = 91.7%; see Fig. 2); thus, a random-effects model was used. The pooled results showed that compared with conventional percutaneous sacroiliac screw fixation, TiRobot would decrease radiation exposure time of screw implantation [MD = − 3.93, 95% CI (− 5.06, − 2.8), P = 0.000; Fig. 2].

Fig. 2
figure 2

Eight studies reported radiation exposure time of screw implantation; we found statistical heterogeneity between the two groups (x2 = 83.9, df = 7, P = 0.000; I2 = 91.7%); TiRobot would decrease radiation exposure time of screw implantation [MD = − 3.93, 95% CI (− 5.06, − 2.8), P = 0.000]. Three studies reported the frequency of intraoperative fluoroscopy. We found some statistical heterogeneity between the two groups (x2 = 4.58, df = 2, P = 0.101; I2 = 56.4%). TiRobot would decrease the frequency of intraoperative fluoroscopy [MD = − 2.62, 95% CI (− 3.28, − 1.96), P = 0.000]. Five studies reported guide wire attempts. We found some statistical heterogeneity between the two groups (x2 = 71.62, df = 4, P = 0.000; I2 = 94.4%). TiRobot would decrease guide wire attempts [MD = − 4.52, 95% CI (− 6.27, − 2.77), P = 0.000]

Frequency of intraoperative fluoroscopy

Three studies reported the frequency of intraoperative fluoroscopy. We found some statistical heterogeneity between the two groups (x2 = 4.58, df = 2, P = 0.101; I2 = 56.4%; see Fig. 2), and thus a random-effects model was used. The pooled results showed that compared with conventional percutaneous sacroiliac screw fixation, TiRobot would decrease the frequency of intraoperative fluoroscopy [MD = − 2.62, 95% CI (− 3.28, − 1.96), P = 0.000; see Fig. 2].

Guide wire attempts

Five studies reported guide wire attempts. We found some statistical heterogeneity between the two groups (x2 = 71.62, df = 4, P = 0.000; I2 = 94.4%; see Fig. 2), and thus a random-effects model was used. The pooled results showed that compared with conventional percutaneous sacroiliac screw fixation, TiRobot would decrease guide wire attempts [MD = − 4.52, 95% CI (− 6.27, − 2.77), P = 0.000; see Fig. 2].

Intraoperative blood loss

Five studies reported intraoperative blood loss. We found statistical heterogeneity between the two groups (x2 = 26.76, df = 4, P = 0.000; I2 = 85.1%; see Fig. 3), and thus a random-effects model was used. The pooled results showed that compared with conventional percutaneous sacroiliac screw fixation, TiRobot would decrease intraoperative blood loss [MD = − 1.10, 95% CI (− 1.85, − 0.34), P = 0.005; see Fig. 3].

Fig. 3
figure 3

Five studies reported intraoperative blood loss. We found statistical heterogeneity between the two groups (x2 = 26.76, df = 4, P = 0.000; I2 = 85.1%). TiRobot would decrease intraoperative blood loss [MD = − 1.10, 95% CI (− 1.85, − 0.34), P = 0.005]. Eight studies reported operative time. We found statistical heterogeneity between the two groups (x2 = 184.87, df = 7, P = 0.000; I2 = 96.2%). Both groups experienced a similar operative time (MD = − 1.24, 95% CI (− 2.51, − 0.02), P = 0.055]

Operative time

Eight studies reported operative time. We found statistical heterogeneity between the two groups (x2 = 184.87, df = 7, P = 0.000; I2 = 96.2%; see Fig. 3), and thus a random-effects model was used. The pooled results showed that patients in both groups experienced a similar operative time [MD = − 1.24, 95% CI (− 2.51, − 0.02), P = 0.055; see Fig. 3].

Fracture healing time

Four studies reported fracture healing time. We did not find statistical heterogeneity between the two groups (x2 = 1.80, df = 3, P = 0.616; I2 = 0%; see Fig. 4), and thus a fixed-effects model was used. The pooled results showed that patients in both groups experienced similar fracture healing time [MD = − 0.13, 95% CI (− 0.42,0.15), P = 0.365; see Fig. 4].

Fig. 4
figure 4

Four studies reported fracture healing time. We did not find statistical heterogeneity between the two groups (x2 = 1.80, df = 3, P = 0.616; I2 = 0%). Both groups experienced similar fracture healing time [MD = − 0.13, 95% CI (− 0.42, 0.15), P = 0.365]. Four studies reported a Majeed score. We did not find statistical heterogeneity between the two groups (x2 = 0.53, df = 3, P = 0.912; I2 = 0%). TiRobot would improve the Majeed score [MD = 0.3, 95% CI (0.03, 0.57), P = 0.031]

Majeed score

Four studies reported a Majeed score. We did not find statistical heterogeneity between the two groups (x2 = 0.53, df = 3, P = 0.912; I2 = 0%; see Fig. 4), and thus a fixed-effects model was used. The pooled results showed that compared with conventional percutaneous sacroiliac screw fixation, TiRobot would improve the Majeed score [MD = 0.3, 95% CI (0.03, 0.57), P = 0.031; see Fig. 4].

Overall excellent and good rates of Majeed score

Six studies reported overall excellent and good rates of Majeed score. We did not find high statistical heterogeneity between the two groups (x2 = 7.19, df = 5, P = 0.207; I2 = 30.5%; see Fig. 5), and thus a fixed-effects model was used. The pooled results showed that compared with conventional percutaneous sacroiliac screw fixation, TiRobot has higher overall excellent and good rates of Majeed score [RR = 1.11, 95% CI (1.02, 1.20), P = 0.018; see Fig. 5].

Fig. 5
figure 5

Six studies reported overall excellent and good rates of Majeed score. We did not find high statistical heterogeneity between the two groups (x2 = 7.19, df = 5, P = 0.207; I2 = 30.5%). TiRobot would have higher overall excellent and good rates of Majeed score [RR = 1.11, 95% CI (1.02, 1.20), P = 0.018]

Overall excellent and good rate of screw accuracy

Eight studies reported an overall excellent and good rate of screw accuracy. We found some statistical heterogeneity between the two groups (x2 = 12.49, df = 7, P = 0.086; I2 = 43.9%; see Fig. 6), and thus a random-effects model was used. The pooled results showed that compared with conventional percutaneous sacroiliac screw fixation, TiRobot would improve screw accuracy [RR = 1.09, 95% CI (1.02, 1.16), P = 0.011; see Fig. 6].

Fig. 6
figure 6

Eight studies reported an overall excellent and good rate of screw accuracy. We found some statistical heterogeneity between the two groups (x2 = 12.49, df = 7, P = 0.086; I2 = 43.9%). TiRobot would improve screw accuracy [RR = 1.09, 95% CI (1.02, 1.16), P = 0.011]. Five studies reported an overall excellent and good rate of reduction. We found some statistical heterogeneity between the two groups (x2 = 7.97, df = 4, P = 0.093; I2 = 49.8%). Both groups experienced a similar excellent and good rate of reduction accuracy [RR = 1.04, 95% CI (0.95, 1.13), P = 0.426]

Overall excellent and good rate of reduction

Five studies reported an overall excellent and good rate of reduction. We found some statistical heterogeneity between the two groups (x2 = 7.97, df = 4, P = 0.093; I2 = 49.8%; see Fig. 6), and thus a random-effects model was used. The pooled results showed that patients in both groups experienced a similar excellent and good rate of reduction accuracy [RR = 1.04, 95% CI (0.95, 1.13), P = 0.426; see Fig. 6].

Discussion

The TiRobot system comprises surgical planning and robot control software, a robot, spatial calibration components, an optical tracking system, a central control station, and matching tools. The operation process of the robot consists of four steps: ① Device connection and skin disinfection: The equipment was placed and connected; then, the skin on the pelvis was sterilized. The surgeon put a tracer on the opposite side of the anterior superior iliac spine. The surgeon placed a sterile C-type arm sleeve on the mechanical arm and connected a positioning device. ② Fluoroscopy: The C-arm X-ray machine fluoroscopy is used for pelvis inlet and outlet view, ensuring all ten markers on the positioning device are visible. ③ Surgical planning and mechanical arm movement: Surgical planning was carried out based on the inlet and outlet view to design the percutaneous sacroiliac screw point, angle, and length. After the surgical planning, the manipulator's arm moves the pilot sleeve according to the planned path, which is the percutaneous sacroiliac screw entry point and angle. ④ Screw-setting: The surgeon inserted the Kirschner wire along the pilot sleeve and then inserted the sacroiliac screw along the Kirschner wire. The screw position was confirmed by the C-arm X-ray machine.

There is only one meta-analysis comparing TiRobot‑assisted versus conventional fluoroscopy-assisted percutaneous sacroiliac screw fixation for pelvic ring injuries. However, all included studies in the previous meta-analysis were observational studies, the number of the included studies was too small (only four articles), and all included studies were English articles. Furthermore, the previous meta-analysis mixed radiation exposure time (seconds) and frequency of intraoperative fluoroscopy (No.) together that may have high heterogeneity. The strengths of our meta-analysis included: First, we included not only English studies but also Chinese studies. Considering that the TiRobot is only used in China, Chinese-language studies were also very important for the meta-analysis. Second, our meta-analysis included more studies (628 patients in 11 studies). Third, our results added pooled analysis of screw accuracy, reduction accuracy, and overall excellent and good rates of Majeed score, which could be a supplement for the previous meta-analysis. Fourth, we analyzed radiation exposure time (seconds) and frequency of intraoperative fluoroscopy (No.) separately, which would provide a more exact conclusion.

Like the previous meta-analyses, our meta-analysis also found that the TiRobot will result in fewer guide wire attempts, less intraoperative blood loss, and better screw accuracy. There were no significant differences in terms of fracture healing time. However, our study came to a different conclusion in terms of Majeed score and operative time. We found that the TiRobot will improve the Majeed score, and we did not see a significant difference in operative time. For some newly added outcomes, we found TiRobot will result in less radiation exposure time of screw implantation, less frequency of intraoperative fluoroscopy, and higher overall excellent and good rates of Majeed score and screw accuracy. We did not find a significant difference in terms of reduction accuracy.

Our meta-analysis showed that TiRobot resulted in less radiation exposure time of screw implantation, less frequency of intraoperative fluoroscopy, and guide wire attempts. The real‐time optical tracking technology of TiRobot makes the repetition of X-rays unnecessary during the operation, decreasing the fluoroscopy frequency and the fluoroscopy time and improving the guide wire insertions. However, we must admit that total radiation exposure time includes not only the evaluated “radiation exposure time of screw implantation(s)” but also the total radiation exposure time, which is needed to prepare the robot procedure before the screw insertion. Before the screw implantation, the surgeon needs to use the C-arm X-ray machine fluoroscopy for pelvis inlet and outlet view, ensuring all ten markers on the positioning device are visible. This step is an extra radiation exposure step compared to conventional groups. We think although the Robot-assisted system needs fluoroscopy for the screw position planning, the overall radiation exposure of both the patients and the surgical team was still lower, providing a more significant safety level. The robotic arm can significantly reduce the deviation caused by the instability of human manual operation, improve the accuracy and safety of the process, and reduce the secondary damage caused by repeated needle insertion in traditional surgery.

The evidence based on this current study confirms that there may be no difference in operative time between the two groups, which is inconsistent with the previous meta-analysis. Theoretically, less radiation exposure time of screw implantation, less frequency of intraoperative fluoroscopy, and guide wire attempts will shorten the operation time and improve the operation efficiency. However, the operation time of the robot system includes the noninvasive time of the robot system and the invasive time of the actual operation. Steps such as collecting fluoroscopic images, positioning the cursor, and planning are extra steps compared to a conventional group and therefore take time. The real invasive operation time is only about 20 min. We also cannot ignore the factors of proficiency, the surgeon's experience, and the operation team's tacit understanding. It is believed that by improving the proficiency in the operation process of the surgical robot-assisted navigation system, surgeons can further shorten noninvasive operation time and the overall operation time. This discrepancy may be because we included some earlier studies in which surgeons lacked expertise in utilizing the robot-assisted system in the early period. With the experience, proficiency, and degree of tacit understanding of cooperation gradually increasing, the surgeons can significantly shorten the duration of operations.

Intraoperative blood loss was notably lower in the TiRobot group, potentially due to the reduced incision length [31, 33], reduced guide wire attempts, and better precision of screw placement. Although there was a statistically significant difference in intraoperative blood loss with the robot group relative to the conventional group, it is unknown whether this statistical difference is clinically meaningful. As well as the length of incision, this may suggest less damage to patients, and the use of TiRobot may be more conducive to postoperative recovery.

There is no significant difference in fracture healing time between the two groups. Despite TiRobot offering better screw accuracy and less radiation exposure; it does not change the degree of comminution, reduction accuracy, and the pelvis’s blood supply, which plays a crucial role in healing duration [36]

Accurate screw path planning and position are crucial to achieving enhanced outcomes in robot-assisted fixation [31]. Screw positioning, evaluated by Florian Gras classification, was notably improved in the TiRobot group in our meta-analysis. Reduction is a crucial step in the surgical treatment of bone fractures to achieve anatomical alignment and facilitate healing [37, 38]. The Matta score evaluates the reduction quality of pelvic fractures [38]. Our meta-analysis found no significant difference between overall excellent and good rate of reduction accuracy according to Matta criteria. Although TiRobot provides real‐time optical tracking technology, it does not change the reduction manner.

Postoperative outcomes measured using Majeed scores have shown more favorable results in the TiRobot group, inconsistent with the previous meta-analysis. The previous meta-analysis didn't find a significant difference in Majeed scores. It seems better accuracy of screw placement positively affects postoperative outcomes [39]. We think the reason for this discrepancy may be the increased number of the included studies in our meta-analysis that can give a significant difference conclusion. However, this result may have some limitations, because the Majeed scores in our included studies were assessed at different follow-up times. In Liu’s study [33], the mean follow‐up duration was 5.4 months (4–12 months). In Wang’s study [34], the mean follow-up time was 11.23 months (5–20 months). In Wang’s study [35], the mean follow-up time was 13.2 months (12–15 months).

We must admit that orthopedic robot systems are expensive and require complete supporting facilities. It is difficult for grassroots hospitals to implement them. Current initiatives should focus not only on improving operative outcomes but also on lowering costs. Potential economic savings gained by using robots in pelvic fracture surgery can be estimated by considering the decreased operative time, patient length of stay, fluoroscopic exposure, and surgical revision rates that robots may increasingly provide in the future. Reduced revision pelvic fracture surgery rates may ultimately be the most effective means of cost savings in this patient population due to the increased direct costs of additional surgery and the indirect cost of prolonged patient disability.

We should consider several limitations before interpreting these findings. First, some studies with a small sample size may lead to unreliable results during meta-analysis. Second, we only included two randomized controlled trials; the other nine studies were observational studies, which may have reduced the quality of the evidence for this meta-analysis. Although we have included all related studies thus far and tried to collect more data to make this meta-analysis, more prospective randomized trials are needed to confirm the results and conclusions. Third, none of the studies evaluated the cost-effectiveness of TiRobot. Cost is the main barrier to robot-assisted surgery, which includes the devices and the associated learning curve and surgeon training. They may be more feasible in large hospitals with a large number of patients [27]. We hope that future studies focus on the overall cost of adopting TiRobot and other systems. Fourth, we didn't compare complications between the two groups because too many factors can affect the occurrence of complications, making the complication comparison high heterogeneity.

Conclusion

TiRobot-assisted fixation has less intraoperative fluoroscopy and intraoperative blood loss, superior screw accuracy, and postoperative outcomes compared with conventional fixation. TiRobot also has no significant effect on operative time, fracture healing time, and reduction accuracy. Given the relevant possible biases in our meta-analysis, we required more adequately powered and better-designed RCT studies with long-term follow-up to reach a firmer conclusion.