Introduction

High tibial osteotomy (HTO) with medial opening wedge is a surgical technique to reconstruct the bony structure of knees and correct their mechanical axis [1, 2]. This technique may be used to manage the early stage of knee osteoarthritis (KOA), particularly in relatively young patients (40–60 years) [3, 4]. Currently, HTO with locking plate and screws is popular because it efficiently slows the progressive degeneration of the knee joint and holds the requirement of terminable joint replacement [5, 6].

To maintain the corrected alignment after HTO, the use of locking plate and screws helps to establish a strong support for the osteotomied tibia till bone union [7,8,9]. In current practice of HTO, traditional titanium locking plate and screw fixation yields acceptable outcomes, and approximately 90% of the patients who returned to work or sport did so within 1 year [6]. However, some complications following HTO such as screw breakage or loss of reduction are not uncommon [10]. To obtain better results, further studies focused either on the structural stability established between the bone and plate or on the creation of beneficial biomechanical stimuli from interfragmentary movement via flexible/dynamic fixation were made in the past [11,12,13]. In these biomechanical studies, some issues including the geometry and configuration of plate, the use of wedge spacer, and even the interfragmentary movement promoting callus formation during bony healing [7, 11, 14,15,16] were investigated. However, to our note, a more fundamental factor, the screw, was less mentioned and rarely investigated. In addition, considering the major clinical issue about screw breakage, it is worth further analyzing and demonstrating the screw stress in HTO-related studies.

Nevertheless, it is difficult to explore the internal stress of screws, plates, and bones, because no sensor is sufficiently small to insert into the body without disturbing the responses. By contrast, the finite element (FE) method, a numerical simulation based on the loading condition and measured material properties of the bone and metallic implants, helps to calculate the internal stresses and has been used in many biomechanical studies [17,18,19,20]. Therefore, this study aimed to explore the mechanical stability and screw stress of HTO fixed with titanium locking plate and screws comparing various lengths and types (flexible or traditional) of locking screws using the FE method.

Methods

Solid model

An intact solid knee joint model consisting of the tibia, distal femur, and meniscus was directly created using open-source computed tomography images of the Visible Human Project (from National Institutes of Health, https://www.nih.gov/) in this study. The contours of the bones in each section were retrieved by examining relatively higher gray values than the surrounding tissues using the Avizo Version 6 (VSG SAS, Bordeaux, France) software. Furthermore, the cortical and cancellous bones were demarcated according to their gray values. The three-dimensional bone models were created based on the retrieved contours of each section. The bone models were then imported into the CAD software SolidWorks 2014 (Dassault Systèmes SolidWorks Corp., Waltham, MA, USA) for creating the meniscus cartilage. The space between the distal femur condyles and the proximal tibia was modeled as the meniscus cartilage, and the shape of the meniscus was modified on the basis of a previous study [21]. Then, the HTO model with medial open wedge and metallic locking plate and screw fixation was created in SolidWorks, too. Virtual planes were defined to divide the tibia into two parts, namely the medial wedge and rest of the tibia. The angle of the medial open wedge was set to 12°, and the length of the lateral hinge (apex of the wedge to tibial border) was set to 10 mm [22]. The volume of the medial wedge was then removed in order to mimic the open wedge of HTO without bone graft.

The TomoFix plate (TomoFix, DePuy Synthes, Oberdorf, Switzerland) and locking screws were used to fix the tibia with HTO (Fig. 1a). The length and width of the TomoFix plate were 115 and 16 mm, respectively. The plate thickness was 3 mm. Two types of locking screws were used in this study, namely traditional locking screw and far-cortical locking (flexible fixation) screws. The tibia shaft was fixed using far-cortical locking and traditional locking screws, and the proximal tibia was fixed using a traditional locking screw only. Six types of screw configuration for HTO fixation were analyzed (Fig. 1b–g). Conventional 30- and 55-mm locking screws were used at the proximal tibia (above the wedge), and three types of screw, namely one-cortical (30 mm), two-cortical (35–50 mm), and far-cortical locking screws (35–50 mm), were used for the tibia shaft (below the wedge). The diameter of the traditional locking screws was 5 mm. For the far-cortical locking screw, the outer diameter of the thread part (distal) was 5 mm, and the diameter of the proximal shaft was 3.5 mm [23]. The bone around the proximal screw shaft of the far-cortical locking screw within 0.75 mm was also removed. Then, a 0.75-mm gap between the proximal screw shaft (of the far-cortical locking screw) and the tibia was created, and the screw shaft did not touch the bone during the initial unloading state. The screw insertion into the tibia was mimicked using Boolean operation to ensure that no space existed between the bone and the thread of screws. The thread of the screw head and the screw hole of the plate were simplified as a cylinder.

Fig. 1
figure 1

a Solid model. Short proximal screw with b one-cortical, c two-cortical, and d far-cortical locking screws. Long proximal screw with e one-cortical, f two-cortical, and g far-cortical locking screws. h Boundary condition used in this study

FE model

The solid model was then imported into ANSYS Workbench V 17 (Swanson Analysis Systems, Inc., Houston, PA, USA) for generating the mesh and simulation. The ligaments around the knee joint, including medial collateral, lateral collateral, anterior cruciate, and posterior cruciate ligaments, were modeled using tension-only springs (Table 1). These springs were created by linking the origin and inserting sites of the corresponding ligaments. The longitudinal stiffness of these springs was set according to the cross-sectional area, the Young’s modulus obtained from a previous study [24], and the length in the present model. The elastic moduli of the cortex and trabecular bone was set to 12 GPa and 430 MPa, respectively [25, 26], and their Poisson’s ratios were set to 0.3. The Young’s modulus and Poisson’s ratio of the meniscus were set to 100 MPa and 0.1, respectively [25]. The metallic implants, including the plate and screws, were made from titanium, and the plasticity of titanium was utilized. Hence, the Young’s modulus and Poisson’s ratio of titanium in the linear elastic phase were defined to 110 GPa and 0.3, respectively. The yield strength and tangent modulus in the plastic phase were set to 800 and 1250 MPa, respectively. The material properties of titanium were defined according to the engineering database in ANSYS Workbench.

Table 1 Material properties of the springs used in this study to simulate ligaments

The thread part of the screws and the surrounding bone were assigned bonding contact behavior, as were the screw head and screw hole of the plate. The contact between the proximal screw shaft of the far-cortical screw and the bone nearby (gap distance of 0.75 mm) was a frictional surface-to-surface contact (if the contact occurred). Furthermore, the contact behaviors between the bones at the osteotomy site were frictional. The contact behavior between the meniscus and the distal femur condyles was frictionless. The friction coefficients of metal-to-bone and bone-to-bone were 0.3 and 0.45, respectively [27]. A 1500 N (about two times the body weight) compressive force was applied on the superior surface of the distal femur to simulate the physiological load of the knee joint in normal gait [22, 28,29,30]. The degree of freedom of the distal femur was set to zero along the X and Y axes, and movement was allowed only in the vertical direction (Z axis). The distal end of the tibia was fixed (Fig. 1h).

Validation

For validation of this FE model, the maximum interfragmentary movements of the FE model with one-cortical and far-cortical locking screws at the tibia shaft (the screw length at the proximal tibia was 55 mm) were compared with those of Roderer et al.’s study [11]. The interfragmentary movement in the FE model and Roderer et al.’s study was similar (Fig. 2). The differences of interfragmentary movement between the present FE model and Roderer et al.’s study were just 0.17 mm (under 1000 N) and 0.03 mm (under 500 N) with the flexible fixation screw and traditional locking screw, respectively.

Fig. 2
figure 2

Comparison of interfragmentary movements (mm) with flexible fixation screw (left) and traditional locking screw (right) in the FE model and Roderer et al.’s study

Incidence

Equivalent stress (also formulated in terms of the von Mises stress) was used as an index for evaluating stress on the metallic screw and plate, whereas maximum principle stress was used for evaluating stress on the bones. The maximum displacement of the tibia and the deformation of the osteotomy gap were used for analyzing HTO stability.

Results

With identical proximal screw length, the different fixation methods using conventional locking screws on the tibial shaft, one-cortical and two-cortical, shared similar gap deformation as well as the tibia displacement (Table 2). Compared with the conventional locking methods, the use of far-cortical locking screws resulted in a larger tibia displacement as well as gap deformation at the osteotomy site (Table 2 and Fig. 3). The maximum gap deformation with the far-cortical locking screw was 2.03 (from 0.219 to 0.445) and 2.05 (from 0.216 to 0.442) times that of the two-cortical screws when the proximal screw lengths were 30 and 55 mm, respectively.

Table 2 Maximum displacement of the tibia, gap deformation, and stress of the implants and bone
Fig. 3
figure 3

Total displacement of the tibia with short (left) and long (right) screws at the proximal tibia

The equivalent stress of the far-cortical locking screws was higher than that of the conventional locking screws (Fig. 4), with both one- and two-cortical locking, whereas the equivalent stress of the plate was slightly lower with the use of far-cortical locking screw than with the use of traditional locking screws (Table 2 and Fig. 5). The maximum equivalent stress of the far-cortical locking screw was 530.88 and 541 MPa with short and long screws, respectively, at the proximal tibia. At the proximal tibia, the far-cortical locking screw also resulted in higher maximum principle stress of the lateral hinge of the osteotomized tibia than one- and two-cortical locking screws (Table 2 and Fig. 6). The maximum equivalent stress of the lateral hinge with far- zcortical, two-cortical, and one-cortical screw was 73.2, 58.78, and 58.82 MPa, respectively.

Fig. 4
figure 4

Equivalent stress of the screws with short (top row) and long (bottom row) screws at the proximal tibia

Fig. 5
figure 5

Equivalent stress of the plate with short (left) and long (right) screws at the proximal tibia

Fig. 6
figure 6

Maximum principle stress of the lateral hinge with short (top row) and long (bottom row) screws at the proximal tibia

Discussion

This study investigated the mechanical responses of metallic implants (namely, the plate and screws) of various lengths, types, and configurations in HTO. The stress differences in the plate and screws depending on screw configurations were demonstrated. Furthermore, this study indicates the risk factor for screw failure in HTO with a titanium locking plate and various locking screws. The results can serve as a reference for surgeons performing HTO. A balance between stability and flexible fixation is essential.

The far-cortical locking screws and plate can decrease the stiffness of locking plates and screws and increase interfragmentary movements at the fracture or osteotomy site [31]. Furthermore, they promote callus formation during the bone healing process because of force stimulation at the fracture site. The decreasing diameter of the proximal shaft of the screw creates a gap between the screw and the cortex near the plate just after implantation without bone ingrowth (i.e., the screw hole is larger than the screw), which provides increased flexibility [13]. The FE simulations in this study and Roderer et al.’s study [11] have indicated that gap deformation is larger at the site when using HTO with flexible fixation than the conventional locking plate. Therefore, this technique is regarded as a potential solution for patients with delayed bone healing or nonunion after HTO.

The flexible fixation plate and screw increase interfragmentary movement, but the high screw stress demonstrated in the present FE model must not be ignored. The maximum equivalent stress of the screw in the far-cortical locking screw and plate system is higher than that in the conventional locking screw system. Because the proximal screw shaft of the far-cortical locking screw is thinner than that of the conventional locking screw and the proximal screw shaft does not contact with the surrounding bone because of a gap 0.75 mm, the thin screw must sustain the complete load. By contrast, the conventional locking screw is in contact with the tibia at the anterior region, and hence, the load is shared with the bones. Therefore, the equivalent stress of the far-cortical locking screw is higher than that of the conventional locking screw. The peak equivalent stress of the far-cortical locking screw (541 MPa) is less than the yield stress of titanium (800 MPa), but not substantially so. Owing to the high stress of the far-cortical locking screw, the risk of a far-cortical locking screw breakage is higher than that of a conventional locking screw after cyclic loadings. Moreover, the equivalent stress of the traditional locking screw is much lower than that of the far-cortical locking screw. Therefore, the risk of screw breakage is lower for conventional locking screws.

In clinical practice, the breakage of the lateral hinge during HTO is a matter of concern, and it may be multifactorial [32, 33]. In the present study, the maximum principle stress at the lateral hinge demonstrated no obvious difference between the different screw lengths and configurations. Further, based on the slightly increased maximum principle stress at the lateral hinge (approximately 24%, from 58.78 to 73.2 MPa), the authors believe the use of far-cortical locking screw does not play a major role in the breakage of the lateral hinge.

In Roderer et al.’s study, a dynamic locking screw system (DePuy Synthes, DePuy Synthes, Oberdorf, Switzerland) was used to increase the movement possible with the conventional locking screw at the osteotomy site. However, far-cortical locking screws and plates (MotionLoc Screw, Zimmer, Warsaw, USA) were used in this study because the dynamic locking screw system is not available in our country, and far-cortical locking screws are widespread. Furthermore, the intended use of dynamic locking screws and far-cortical locking screws is the same. Hence, far-cortical locking screw was used to replace the dynamic locking screw in this study.

We acknowledge some limitations in this study. First, the stress of the bone developed from the distracting process of the open wedge is not considered. Second, the thread at the proximal shaft of the far-cortical screw is neglected. Because the thread is for screw removal and does contact with the surrounding bone after fixation, smoothening the screw shaft does not affect the stability of the tibia in HTO. Third, only the maximum load in full knee extension was simulated, but the other knee positions and the muscle force were not considered. Fourth, all materials were simplified as isotropic and homogeneous.

Conclusions

The study results suggest that far-cortical locking screws and plates can increase interfragmentary movement in HTO; however, the screw stress is relatively high. Increasing the protection time (partial weight duration) is suggested to decrease the risk of screw breakage in HTO through fixation with titanium far-cortical locking screws and plates.