Background

Cough is a physiological response whose purpose is to eliminate secretions from the airway tract. Thus, the inability to cough forcefully enough to be effective would increase the risk of pulmonary complications such as atelectasis or pneumonia. This inability is observed in neuromuscular or respiratory diseases and is particularly likely after abdominal or thoracic surgery, when pain, surgical injury and/or the residual effect of anesthetics come into play [1,2,3].

Because a cough maneuver can be made voluntarily and easily by a patient without training, cough pressure measurements at various sites can be obtained readily in clinical situations, in contrast to the often used maximum mouth expiratory pressure recorded during a static artificial maneuver [4, 5], which must be learned. Cough pressure is usually measured with balloon catheters that record maximum gastric or esophageal pressures (Pga, Pes) [4,5,6]. It is not always practical to use these catheters outside research scenarios, however, and they can cause discomfort [4, 7]. Furthermore, they cannot be used in certain situations, such as during postoperative recovery from gastrointestinal surgery. Other points of measurement that might potentially be used to reflect cough pressure include central venous pressure (Pcv), which has been used as an alternative to Pes [8]; bladder pressure (Pbl), which has been compared to Pga [8, 9]; and rectal pressure (Prec), which has been used to measure intra-abdominal pressure as an alternative to Pbl [10]. However, none of these pressures have yet been used to evaluate expiratory muscle strength, even though many hospitalized patients have a catheter already placed in the superior vena cava or in the bladder.

We hypothesized that these catheters could be suitable for measuring cough pressure without compromising patient comfort, as occurs with the placement of a gastric or esophageal balloon. We also reasoned that placement of a rectal balloon catheter, which does not cause the nausea associated with esophageal or gastric balloons, could also measure cough pressure comfortably. If these hypotheses are correct, wider clinical use of cough pressure to reflect respiratory muscle strength and possible risk of respiratory compromise might be facilitated. The aim of this study was to evaluate the use of Pcv, Pbl, and Prec as alternatives to Pes or Pga for the measurement of cough pressure.

Methods

Patients

Adult patients scheduled for open-midline laparotomy for colon cancer surgery, which required placement of central venous and bladder catheters, were enrolled prospectively. Exclusion criteria included rectal surgery, chronic obstructive pulmonary disease [11], neuromuscular disorders [1], chronic pain, and factors that could impede an adequate recording of the research protocol variables.

The study was approved by the clinical research ethics committee of Parc de Salut Mar (CEIC-Parc de Salut Mar) and by the Spanish Agency for Medicines and Health Products (AEMPS). All patients signed an informed consent form before entering the study, and we provided each with an insurance policy to cover care in the event of adverse events related to the procedures.

Interventions and measurements

All patients underwent forced spirometry measurement (Datospir 500, SIBEL, Barcelona, Spain) the day before surgery. Reference values were those for a Mediterranean population [11].

Pressures were measured with catheters placed in the esophagus, stomach, superior vena cava, bladder and rectum as follows:

  • Pes and Pga were measured with compliance balloon catheters (esophageal catheter Jaeger 720,199, Viasys Healthcare, Hoechberg, Germany) as previously described [5]. The catheters were introduced nasally under local anesthesia and the balloons were filled with 1–2 mL of air.

  • Pcv was measured from the distal port of a double-lumen catheter (CV-26702-E, Arrow, Erding, Germany) placed through the subclavian or internal jugular veins [12]. Correct positioning was checked with the Pcv waveform [13].

  • Pbl was measured with a transurethral (Foley) catheter inserted after the bladder was drained and 50 mL of a 0.9% saline solution was instilled [8, 14, 15].

  • Prec was measured with a compliance balloon inserted 10 cm inside the rectum and filled with 5 mL of air [10].

All the pressure curves were displayed on a screen and recorded with a data acquisition system (Acknowledge and MP100, Biopac, Santa Barbara, CA, USA) for off-line analysis. Patients lay in supine position and all the pressure transducers were calibrated and aligned with the axillary midline.

The correct placement of all catheters was assessed by asking the patient to perform a sharp sniff and a cough maneuver while the researcher monitored the signal on the computer screen [5]. Once all catheters were inserted and after a 3-min resting period, baseline respiratory pattern and pressures were recorded.

Cough pressure was then measured at all points. Patients were asked, always by the same researcher (L.G.A.), to cough as forcefully as possible [5] from total lung capacity (TLC). Two or three trials were performed to allow the patient to practice the maneuver before data recording started. Next, the pressures generated by three valid maneuvers, separated by pauses of 5–10 s, were recorded and the difference between the baseline pressure at relaxed end-expiratory lung volume and the peak pressure attained during each cough from TLC was registered, as previously described [4,5,6]. Maneuvers were considered valid if the patient followed the instruction to cough and the expected cough pressure waveform was observed on the screen [5].

Soon after the end of the protocol, the patients were asked to indicate which catheters caused the least and the most discomfort.

Statistical analysis

The sample size was calculated to allow us to detect an intraclass correlation coefficient (ICC) value defined as excellent [16] with a statistical power of 80% based on the range of cough pressure values obtained at the different sites during an earlier study [6]. The ICC was used as the measure of the reliability within each patient of the cough pressure measurements at each site and to reflect agreement between alternative sites and Pes. Bland–Altman plots of mean differences were used to compare Pga, Pcv, Pbl and Prec with the gold-standard (Pes) [17]. Pes rather than Pga was chosen as the gold standard because the former is measured in the chest, where a cough effort becomes effective. Maneuvers that generated the maximum Pes were chosen for comparison between sites. That maximum value was then compared to the pressures generated by the patient during the same maneuver at each of the other sites.

A p value of < 0.05 was considered statistically significant in the ICC analysis. Statistical analysis was performed using IBM SPSS (IBM, Armonk, NY, USA) and STATA (STATA Corp., College Station, TX, USA) software.

Results

Participants

Eleven patients initially consented to participate in the study, but one withdrew consent as catheters were about to be inserted. Data from one patient were lost because of technical problems. Thus, we analyzed data for nine patients. Their characteristics are summarized in Table 1. No adverse events were observed during the insertion of the catheters.

Table 1 Demographic, anthropometric and functional data

Test results

The cough pressure curves for all sites were congruent. Figure 1 shows the pressure curves for four patients.

Fig. 1
figure 1

Waveforms at all five measurement sites. Congruent waveforms recorded for four patients at all five measurement sites. Pcv = central venous pressure; PES = esophageal pressure; Pga = gastric pressure; PBL = bladder pressure; Prec = rectal pressure

The median (first quartile, third quartile) for the maximum Pes values for the nine patients was 112 (89,148) cmH2O. The other median cough pressures recorded at the same time as the maximum Pes values were as follows: Pga 105 (92,156), Pcv 102 (91,149), Pbl 118 (93,157), and Prec 103 (88,150) cmH2O.

Individual measurements recorded for patients at each site are shown in Table 2. Bold-face values show the maneuvers that generated maximum Pes.

Table 2 Measurements of cough pressure at five sites

The ICCs showed excellent repeatability between the three pressures recorded for each patient at each site (within-site repeatability) (p < 0.001) and excellent agreement between alternative sites and Pes (p < 0.004) (Table 3).

Table 3 Analysis of cough pressure measurement methods

The median (first quartile, third quartile) differences between maximum PES and pressures at other sites were as follows: Pga 0.7 (− 1.8,13.5), Pcv − 2.1 (− 4.2,5.2), Pbl 5.8 (4,12.9), and Prec − 2.5 (− 9.5,12) cmH2O. Bland–Altman plots of the differences between Pes and each of the other pressures are shown in Fig. 2. Pcv and Prec were the alternative-site pressures that showed the best agreement (smallest average differences from Pes). The average differences between Pbl and Pes were slightly larger, Pbl usually registering higher values than Pes.

Fig. 2
figure 2

Bland–Altman plots of differences between the maximum (optimal) Pes and the pressure measurements at the alternative sites. Circles indicate the difference between PES and the alternative measured pressures. The central line indicates the average difference between the compared measurements. The upper and lower (dotted) lines indicate the 95% confidence interval. Values are expressed in cmH2O. Pcv = central venous pressure; PES = esophageal pressure; Pga = gastric pressure; PBL = bladder pressure; Prec = rectal pressure

The rectal catheter was the least unpleasant for all the patients, whereas the ones introduced nasally into the esophagus and stomach were the most uncomfortable for seven patients (4 men and 3 women). The bladder catheter was the most uncomfortable for 2 men.

Discussion

This study demonstrates that cough pressure can be measured with central venous or rectal catheters as alternatives to conventional esophageal balloon catheters. Bladder catheters could also be used although recorded Pbl values were systematically slightly higher than the gold-standard Pes values.

The ICC analysis indicated excellent repeatability between measurements at the same site, evidence of the precision of cough pressure measurements. In a few patients, cough pressures were slightly lower and differences among repeated maneuvers higher than for the other subjects, but in a series of continuous non-selected patients this minimal variability can be expected. In fact, 5–10% of adult outpatients cannot achieve reproducible pulmonary function results when repeated maneuvers are performed with coaching [18]. The largest source of within-subject variability of spirometry is improper performance of the test [19]. For these reasons, when voluntary maneuvers are recorded several measurements must always be performed and the best one will be chosen.

The ICC analysis also demonstrated excellent agreement between alternative sites and Pes, and the Bland–Altman plots showed small differences lacking clinical significance between Pes and Pga, Pcv and Prec. These data therefore suggest that the alternative measurements are reliably accurate. Thus, Pcv or Prec would be valid candidates to choose as surrogates for Pes. The greater difference between Pbl and Pes, on the other hand, shows that pressure behaves somewhat differently at the bladder, something the clinician would need to bear in mind. Furthermore, given that Pbl was usually higher than Pes, we can conclude that Pbl was precise but less accurate than pressure measurements at the other alternative sites. This slight but systematic difference between Pes and Pbl would mean that the bladder catheter would be the last-choice alternative to the esophageal catheter. Pbl could nevertheless be useful in hospitalized patients who already have a bladder catheter in place, so as to avoid disturbing the patient by placing an additional one. The clinician must always take into account this fixed bias in relation to Pes when interpreting the pressures.

Although Pga has been used widely to reflect cough pressure in studies of respiratory muscle strength [4,5,6, 20] and was measured at the same time as Pes in this study, we designated Pes as the gold standard in the Bland–Altman analysis because it is recorded in the chest, where cough effort takes place. Cough pressures have fluctuated in previous studies of expiratory muscle strength measured with Pes or Pga [4, 6, 20,21,22] because the study populations varied. Higher cough pressures are observed in young, male, and tall subjects as well as in chronic coughers. We measured cough pressure in a specific surgical population, accounting for differences between our results and previously reported values.

Our study was performed under conditions relevant to clinical situations. Patients were in supine position, in which the pressure transducers were all at the same approximate level, favoring reliable comparison between measurement sites. Patients carrying central venous or Foley (bladder) catheters are usually confined to a bed. The cough maneuver was performed from TLC in order to achieve a standardized test measurement [4] and because it is usual to take a deep breath before a cough [5]. Maximum levels of respiratory muscle strength and hence pressure are expected from TLC [5, 23].

The practical implication of our study is that cough pressure can be measured using the technique that best fits the clinical condition of an individual patient. A central venous catheter would be the first choice if one has been inserted. A bladder catheter could also be used provided the clinician bears in mind the larger difference between Pbl and Pes discussed above. If no catheter has been inserted, a good choice would be a rectal catheter, which our patients found to be the least uncomfortable. Pes remains the gold standard, but its main disadvantage, that the insertion of a balloon catheter through the nose causes discomfort [4], was confirmed by our patients. In addition, discomfort can cause esophageal contractions that can impede correct measurement in a considerable percentage of patients [7]. For these reasons, esophageal and gastric balloons should probably be reserved for selected patients or volunteers under experimental conditions.

This study has limitations. The results probably cannot be extrapolated to patients with chronic cough, in whom voluntary cough pressure can be higher than in healthy individuals [20]. In addition, the results possibly cannot be extrapolated to scenarios in which cough is triggered by nerve stimulation [24] or to patient types we excluded. Similar results might well be obtained in these scenarios, but further studies would be needed to confirm that hypothesis.

Our results facilitate further investigation and patient management in many settings. An important scenario is the postoperative period after abdominal or thoracic surgery, where cough effort is reduced [1,2,3] and where patients are at risk of respiratory complications [25].

Conclusions

Pressure generated with a cough maneuver from TLC in supine position can be measured in the esophagus, stomach, superior vena cava, or rectum indistinctly. Bladder catheters could also be used, although the recorded pressures would usually be slightly higher than Pes. These results support assessing expiratory muscle strength for clinical or research purposes without using an additional invasive catheter in most hospitalized patients. If no invasive catheter has already been placed for clinical purposes, a minimally invasive catheter can be chosen for assessing cough pressure.