Abstract—
A Cherenkov proton-beam monitor has been investigated. Its electronic system consisting of a preamplifier and an amplifier of detector signals is described. The measured waveform is compared to the simulated monitor signals during proton detection. The relative variance of the photomultiplier-tube gain for single photoelectrons has been measured. Good agreement is observed between the results of measurements with the monitors and the induction sensor of the proton-beam current. The monitor is incapable of detecting photons and electrons with energies below 170 keV, which is important for suppressing low-energy background.
Similar content being viewed by others
REFERENCES
Boag, J.W., Ionization Chamber in Radiation Dosimetry, New York: Academic, 1968.
McDonald, J., Velissaris, C., Viren, B., Diwan, M., Erwin, A., Naples, D., and Ping, H., Nucl. Instrum. Methods Phys. Res., Sect. A, 2003, vol. 496, p. 293. https://doi.org/10.1016/S0168-9002(02)01768-0
Abramov, A.I., Kazanskii, Yu.A., and Matusevich, E.S., Osnovy eksperimental’nykh metodov yadernoi fiziki (Fundamentals of Experimental Methods for Nuclear Physics), Moscow: Atomizdat, 1970.
Reinhardt-Nikoulin, P., Gaidash, V., Menshov, A., Mirzojan, A., and Feschenko, A., Probl. At. Sci. Technol., Ser.: Nucl. Phys. Invest., 2006, no. 2 (46), p. 113.
D’yachenko, A.F., Vestn. Khar’k. Nats. Univ. im. V. N. Karazina. Ser. Fiz., 2010, no. 887, p. 118.
Geant4 Collab., Agostinelli, S., et al., Nucl. Instrum. Methods Phys. Res., Sect. A, 2003, vol. 506, p. 250. https://doi.org/10.1016/S0168-9002(03)01368-8
Sauli, F., Report CERN 77-09, Geneva: CERN, 1977.
Particle Data Group, Tanabashi, M., et al., Phys. Rev. D, 2018, vol. 98, p. 030001. https://doi.org/10.1103/PhysRevD.98.030001
Birks, J.B., The Theory and Practice of Scintillation Counting, New York: Pergamon, 1967.
Dorenbos, P., Hass, J.T.M., and Eijk, C.W.E., IEEE Trans. Nucl. Sci., 1995, vol. 42, no. 6, p. 2190. https://doi.org/10.1109/TNS.2011.2141683
Hrisoho, A. and Truong, K., Preprint LAL-RT/79-4, Orsay, 1979.
AD8138. Differential ADC Driver. http://www.analog.com.
CAEN DT5742. 16 Channel 12 Bit Waveform Digitizer. http://www.caen.it.
Bellamy, E.H., Bellettiny, G., Budagov, J., Cervelli, F., Chirikov-Zorin, I., Incagli, M., Lucchesi, D., Pagliarone, C., Tokar, S., and Zetti, F., Nucl. Instrum. Methods Phys. Res., Sect. A, 1994, vol. 339, p. 468. https://doi.org/10.1016/0168-9002(94)90183-X
CAEN DT5720. 4 Channel 12 Bit Waveform Digitizer. http://www.caen.it.
Active Voltage Divider VD124K/T. http://www.photonis.com.
ACKNOWLEDGMENTS
In conclusion, I consider it my pleasant duty to express my gratitude to Yu.K. Gavrilov, S.V. Akulinichev, A.B. Neganov, and V.E. Postoev for their support and help with my work.
Funding
This work was carried out using the equipment of the Accelerator Center for Neutron Research of the Structure of Matter and Nuclear Medicine Center for Collective Use of the INR RAS with the support of the Ministry of Education and Science of the Russian Federation under the Subsidy Agreement (no. 14.621.21.0014 dated August 28, 2017), unique identifier RFMEFI62117X0014.
Author information
Authors and Affiliations
Corresponding author
Additional information
Translated by N. Goryacheva
Rights and permissions
About this article
Cite this article
Djilkibaev, R.M. A Cherenkov Proton-Beam Monitor. Instrum Exp Tech 64, 508–515 (2021). https://doi.org/10.1134/S002044122104014X
Received:
Revised:
Accepted:
Published:
Issue Date: