Abstract
X-ray grating interferometry is a promising imaging technique sensitive to attenuation, refraction and scattering of the radiation. Applications of this technique in the energy range between 80 and 150 keV pose severe technical challenges and are still mostly unexplored. Phase-contrast X-ray imaging at such high energies is of relevant scientific and industrial interest, in particular for the investigation of strongly absorbing or thick materials as well as for medical imaging. Here we show the successful implementation of a Talbot-Lau interferometer operated at 100 keV using a conventional X-ray tube and a compact geometry, with a total length of 54 cm. We present the edge-on illumination of the gratings in order to overcome the current fabrication limits. Finally, the curved structures match the beam divergence and allow a large field of view on a short and efficient setup.
Similar content being viewed by others
Introduction
X-ray radiography and computed tomography (CT) are standard imaging techniques in materials and life sciences for the nondestructive examination of samples or diagnostic tasks in medicine. The underlying contrast mechanism relies on the different X-ray attenuation properties of different materials or tissue types. The dominant physical effects contributing to attenuation are the photoelectric effect and incoherent (Compton) scattering. Besides attenuation, the wave nature of X-rays reveals another contrast mechanism, which is the phase shift. The interaction contributing to phase shifts is coherent (Rayleigh) scattering1.
The attenuation and phase shift properties are described by the complex refractive index n = 1 − δ + iβ. The imaginary part β is related to the attenuation coefficient by µ = 4πβ(λ)/λ, while the real part δ(λ) determines the phase shift ϕ = 2πδ(λ)/λ.
While the attenuation can be measured with an X-ray detector as the reduction of the beam intensity, the phase is not directly observable. Therefore, an optical system is needed to convert the phase shift into intensity modulations.
Phase-sensitive imaging is a desirable modality, since it delivers a complementary source of contrast with respect to absorption by providing direct access to the electron density1. Moreover, the combination with attenuation enables the determination of the effective atomic number2. An enhanced contrast-to-noise ratio (CNR) compared to attenuation for certain materials or tissues3,4 has also been demonstrated.
The vast majority of phase-sensitive techniques, including crystal analyzer based5,6 or interferometric7,8 methods rely on X-ray beams of high spatial and temporal coherence, which are available only at synchrotron sources. Inline phase contrast9,10,11 and Talbot interferometry12,13,14 need high spatial coherence but are available on polychromatic microfocus sources. Phase-contrast imaging using X-ray beams of low temporal and spatial coherence such as conventional low-brilliance X-ray tubes have been demonstrated with coded apertures15 and Talbot-Lau interferometry16. Analyzer-based systems have been recently extended to tube sources17,18 but only at energies up to the tungsten Kα line at 60 keV. In addition to phase sensitivity, crystal analyzers and Talbot interferometers also provide — with different retrieval mechanisms — information about the integrated local small angle scattering power from microscopic density fluctuations in a specimen19. This signal is known under the name of dark-field, scatter or visibility reduction contrast.
High-energy Talbot interferometry has been reported so far using a synchrotron source at nominal energies of 82 keV20. Using a low-brilliance X-ray tube, Talbot-Lau interferometry was applied so far at 60 keV design energy21. Medical imaging applications may benefit from phase contrast at higher energies: chest or abdominal radiography or CT require an acceleration voltage between 100 and 150 kVp. Other potential applications are homeland security or chip failure analysis, which require high energies for the visualization of materials of high density and atomic number.
Results
We introduce a method for phase contrast imaging which works on conventional X-ray sources, covers the entire diagnostic X-ray energy range and is compatible with compact imaging arrangements. The approach is based on Talbot-Lau interferometry16 and employs an edge-on illumination approach for the grating design and arrangement. Our solution removes one of the major hurdles which prevented grating interferometry from being applied at high energies so far, namely the fabrication of gratings with high aspect ratios. The aspect ratio, given by
where p is the grating period and h the structure height, is limited by the fabrication process, usually photolithography13 or X-ray lithography22 since grating structures tend to collapse or deform (e.g. due to capillary forces) if the aspect ratio is too high. Moreover, when using a broad spectrum, photons above the design energy should also be efficiently blocked by the gratings, thus requiring even higher aspect ratios. The largest aspect ratios achieved by current fabrication techniques23,24 are around 60. For a given setup length these parameters depend on the target energy E according to and h ∝ E3 and therefore R ∝ E7/214. If at E = 25 keV an aspect ratio for the absorption grating around R = 30 is sufficient, it would have to be at least 128 for E = 100 keV. With our design, we reach an aspect ratio of 143, satisfying the above condition with a transmission of less than 1% at 100 keV. As a comparison, a recent interferometer implemented at a third generation synchrotron facility20 reported an aspect ratio of 21 with a transmission as large as 20% at 82 keV.
Our design introduces edge-on illuminated, circularly aligned structures. Edge-on illumination (Fig. 1), as opposed to face-on illumination, exploits the dimension along the grating lines to form a high aspect ratio of the structures in the direction of the beam34. The effective structure height of the grating is then determined by the grating dimension along the grating lines, which essentially allows arbitrarily high aspect ratios.
Increasing the aspect ratio of the gratings typically leads to a reduction of the field of view due to the change of the grating transmission function at high incident angles. In order to overcome this problem, the grating lines are circularly aligned with a radius equal to the distance to the source. This allows to achieve an arbitrarily large field of view in a fan-beam geometry, a significant improvement compared to face-on based and glancing angle25 approaches.
The combination of edge-on illumination and circularly aligned structures enables phase-contrast imaging at arbitrary design energies and with a maximum field of view in the horizontal direction (x direction). These advantages come at the expense of a limited field of view in the vertical direction (y direction), which is, depending on the X-ray detector, typically a few pixels. However, radiographic 2D imaging can be obtained by scanning the sample or a thin fan beam. The scanning technique has been demonstrated to deliver less dose than the conventional approach based on the illumination of a large area. In digital mammography, for instance, where dose is a critical issue, Philips' MicroDose system combines a scanning approach with a highly collimated fan beam26. Thanks to the high collimation, the dose deposited on patients has been reported to be significantly lower than with other instruments based on the illumination of a large area detector27. Similarly, for tomographic images, the approach allows single slice CT or full 3D imaging in scanning mode.
Grating design and fabrication is nonstandard and involves a complex mask design, as shown in Fig. 2. Multiple gratings can reside on a silicon chip with their specific structure length and curvature. For the current experiments, a symmetric interferometer with a grating period of p = 2.8 µm for all gratings has been used. The design energy is 100 keV and the beam splitter grating periodically shifts the phase by zero and π at this energy13. Using gold as the phase shifting material, a structure length of h1 = 19.8 µm is required. The analyzer grating is an absorption mask for sensing slight changes of the interference pattern generated by the beam splitter14. With a structure length of h2 = 800 µm it can absorb more than 90% of the incoming X-rays up to energies of 160 keV. The beam splitter and analyzer grating are separated at the first fractional Talbot order28, resulting in an intergrating distance of 158 mm. However, the precise position of the analyzer grating along the beam axis is not critical and a change in visibility of less than 1% is observed by displacing it by as much as 5 mm. The source grating splits the relatively large focal spot (~ 1 mm) into an array of individually coherent, but mutually incoherent sources16. It is also made of gold structures with a length of h0 = h2 = 800 µm.
Due to the large spectral acceptance28,29 of the interferometer (50 keV to more than 160 keV) and the high attenuation efficiencies of the source and analyzer gratings (more than 90% up to 160 keV), the voltage of the X-ray source was set to the maximum of 160 kV. With a structure height of approximately 100 µm, the field of view in the vertical direction is limited to one detector pixel row. In the horizontal direction, the field of view is only limited by the grating size to 30 mm, but wider gratings can be fabricated with the same method and the available technology on larger wafers. In addition to the standard components (source, camera, interferometer), two optical slits, one in front of the source grating, the other in front of the camera, were required for the collimation of the beam in the vertical direction.
Fig. 3 shows the first X-ray images acquired with our new setup, based on edge-on illuminated grating interferometry powered by a conventional X-ray source operated at 160 kVp and with a design energy of 100 keV. Several resistors and integrated circuits located on different layers on an electronic chip can be identified. The radiographs were acquired in scanning mode, using a step size of 100 µm along the y axis, covering a total imaging area of 2 × 2 cm2. For a better comparison of the magnified phase and attenuation images, the latter has been replaced with the differential attenuation image, which was obtained by calculating the derivative along the horizontal axis. In the attenuation image, the contrast of the soldering points of the integrated circuit is reduced underneath the resistors, while in the phase image, they can clearly be identified. The reduced contrast of the soldering points in the absorption image is due to beam hardening. The spectrum impinging on these soldering point is hardened by the resistors in the upper layer, resulting in lower absorption contrast. Due to the weaker energy dependence of phase shifts (1/E compared to 1/E3), phase-contrast images are less sensitive to beam hardening30, which explains the lower contrast reduction of the soldering points under the resistors in the phase image of the chip. This example shows one of the potential benefits of phase contrast X-ray imaging at high energies, which may be useful to identify flaws in multilayered structures such as electronic chips.
Discussion
Possible usages of edge-on grating interferometry encompass applications ranging from medical imaging to non-destructive testing and homeland security. All these applications seek for high (greater than 100 keV) nominal energies and compact designs. Particularly for medical application, the operation of a phase contrast imaging device at high energies is imperative. When imaging patients, a crucial issue is the dose, which depends mainly on the absorption of the X-rays and decreases with the third power of the photon energy31. On the other hand, the phase interaction is only inversely proportional to the energy and is therefore relatively much more relevant than absorption as the energy increases, possibly leading to a decreased exposure for an equivalent image quality. Clinical operation of such a device will imply measuring patients who normally breath, move and shake. Fast acquisitions are therefore required, in order to keep the images sharp and artifact free. As a consequence, compact, efficient imaging systems need to be developed to provide imaging capabilities at sufficient speed. Our circular grating alignment matches the structures of the diffracting and absorbing elements to the divergent beam, especially for compact geometries. In conclusion, short, high-energy phase-contrast systems will enable the efficient investigation of high-density materials or thick samples, adding information on electron density and integrated small-angle scattering power to the conventional absorption-based signal. This will improve material discrimination and density sensitivity in future X-ray or even neutron32 investigations.
Methods
Edge-on illuminated gratings were manufactured by Microworks GmbH, Germany, using a LIGA process24. Each grating resides on a 5 × 60 mm2 silicon chip and several grating chips are fabricated on a single 4 inch silicon wafer. The experimental arrangement for a design energy at 100 keV is a symmetric Talbot-Lau interferometer with a grating period of p = 2.8 µm for all gratings. The distance from the source grating to the analyzer grating is 32 cm and the source grating is positioned 23 cm away from the source.
The X-ray source is a tungsten target COMET MXR-160HP/11 X-ray tube with a maximum output voltage of 160 kV and a current of 10 mA. In the experiment, it was set to the maximum voltage. The focal spot size is approximately 1 mm. The detector is a CCD camera from Finger Lakes Instruments. A cesium iodide (CsI:Tl) scintillator of 600 µm thickness converts the X-rays to visible light and is coupled to an optical lens projecting the image onto the CCD. The effective pixel size is 80 µm. The collimating slit right after the source is 25 µm wide, while the second slit before the detector is 100 µm.
In Fig. 3, image acquisition involved 24 phase steps per line and an exposure time of 15 seconds per step. The long exposure times are mostly constrained by the low average visibility of the gratings (5%). The exposure time was chosen in order to get a low noise in the differential phase image. The signal-to-noise ratio (SNR) is proportional to the visibility and the square root of the exposure time33. This implies that the exposure times can easily drop by an order of magnitude as these gratings become comparable in quality to those developed in the last ten years. Smaller regions of these gratings already exhibit a visibility up to 14%, indicating that this goal is reachable as the fabrication becomes more reliable and uniform.
References
Als-Nielsen, J. & McMorrow, D. Elements of modern X-ray physics (Wiley, London, 2011), 2nd edn.
Qi, Z., Zambelli, J., Bevins, N. & Chen, G.-H. Quantitative imaging of electron density and effective atomic number using phase contrast CT. Phys. Med. Biol. 55, 2669–77 (2010).
Pfeiffer, F. et al. High-resolution brain tumor visualization using three-dimensional x-ray phase contrast tomography. Phys. Med. Biol. 52, 6923–6930 (2007).
McDonald, S. A. et al. Advanced phase-contrast imaging using a grating interferometer. J. Synch. Radiat. 16, 562–572 (2009).
Davis, T., Gao, D., Gureyev, T., Stevenson, A. & Wilkins, S. Phase-contrast imaging of weakly absorbing materials using hard X-rays. Nat. 373, 595–598 (1995).
Chapman, D. et al. Diffraction enhanced x-ray imaging. Phys. Med. Biol. 42, 2015–2025 (1997).
Bonse, U. & Hart, M. An x-ray interferometer. Appl. Phys. Lett. 6, 155–156 (1965).
Momose, A., Takeda, T., Itai, Y. & Hirano, K. Phase-contrast X-ray computed tomography for observing biological soft tissues. Nat. Med. 2, 473–475 (1996).
Snigirev, A., Snigireva, I., Kohn, V., Kuznetsov, S. & Schelokov, I. On the possibilities of x-ray phase contrast microimaging by coherent high-energy synchrotron radiation. Rev. Sci. Instr. 66, 5486–5492 (1995).
Wilkins, S., Gureyev, T., Gao, D., Pogany, A. & Stevenson, A. Phase-contrast imaging using polychromatic hard X-rays. Nat. 384, 335–338 (1996).
Cloetens, P., Barrett, R., Baruchel, J., Guigay, J. & Schlenker, M. Phase objects in synchrotron radiation hard x-ray imaging. J. Phys. D: Appl. Phys. 29, 133–146 (1996).
Cloetens, P., Guigay, J., De Martino, C., Baruchel, J. & Schlenker, M. Fractional Talbot imaging of phase gratings with hard x rays. Opt. Lett. 22, 1059–61 (1997).
David, C., Nöhammer, B., Solak, H. & Ziegler, E. Differential x-ray phase contrast imaging using a shearing interferometer. Appl. Phys. Lett. 81, 3287–3289 (2002).
Momose, A. et al. Demonstration of X-Ray Talbot Interferometry. Jpn. J. Appl. Phys. 42, L866–L868 (2003).
Munro, P., Ignatyev, K., Speller, R. & Olivo, A. Phase and absorption retrieval using incoherent X-ray sources. Proc. Natl. Acad. Sci. 2012, 2–7 (2012).
Pfeiffer, F., Weitkamp, T., Bunk, O. & David, C. Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources. Nat. Phys. 2, 258–261 (2006).
Nesch, I. et al. The design and application of an in-laboratory diffraction-enhanced x-ray imaging instrument. The Rev. Sci. Instr. 80, 093702 (2009).
Parham, C., Zhong, Z., Connor, D. M., Chapman, L. D. & Pisano, E. D. Design and implementation of a compact low-dose diffraction enhanced medical imaging system. Acad. Radiol. 16, 911–917 (2009).
Pfeiffer, F. et al. Hard-X-ray dark-field imaging using a grating interferometer. Nat. Mater. 7, 134–137 (2008).
Willner, M. et al. Quantitative X-ray phase-contrast computed tomography at 82 keV. Opt. Expr. 21, 4155–4166 (2013).
Donath, T. et al. Phase-contrast imaging and tomography at 60 keV using a conventional x-ray tube source. The Rev. Sci. Instr. 80, 053701 (2009).
Mohr, J. et al. High aspect ratio gratings for x-ray phase contrast imaging. AIP Conf. Proc. 1466, 41–50 (2012).
David, C. et al. Fabrication of diffraction gratings for hard X-ray phase contrast imaging. Microelectron. Eng. 84, 1172–1177 (2007).
Kenntner, J. et al. Front- and backside structuring of gratings for phase contrast imaging with x-ray tubes. In: Proceedings of SPIE, vol. 7804, 780408 (2010).
Stutman, D. & Finkenthal, M. Glancing angle Talbot-Lau grating interferometers for phase contrast imaging at high x-ray energy. Appl. Phys. Lett. 101, 91108 (2012).
Åslund, M. Digital Mammography with a Photon Counting Detector in a Scanned Multislit Geometry. Ph.D. thesis. (2007).
Oduko, J. M., Young, K. C. & Burch, A. A survey of patient doses from digital mammography systems in the uk in 2007 to 2009. In: Martí, J., Oliver, A., Freixenet, J. & Martí, R. (eds.) Digital Mammography, vol. 6136 of Lecture Notes in Computer Science, 365–370 (Springer, Berlin Heidelberg, 2010).
Weitkamp, T. et al. X-ray phase imaging with a grating interferometer. Opt. Expr. 13, 6296–6304 (2005).
Thuering, T. et al. Energy resolved X-ray grating interferometry. Appl. Phys. Lett. 102, 191113 (2013).
Chabior, M. et al. Beam hardening effects in grating-based x-ray phase-contrast imaging. Med. Phys. 38, 1189 (2011).
Momose, A. Recent Advances in X-ray Phase Imaging. Jpn. J. Appl. Phys. 44, 6355–6367 (2005).
Grünzweig, C. et al. Design, fabrication and characterization of diffraction gratings for neutron phase contrast imaging. The Rev. Sci. Instr. 79, 053703 (2008).
Raupach, R. & Flohr, T. G. Analytical evaluation of the signal and noise propagation in x-ray differential phase-contrast computed tomography. Phys. Med. Biol. 56, 2219–2244 (2011).
David, C. & Stampanoni, M. “A method for x-ray phase contrast and dark-field imaging using an arrangement of gratings in planar geometry”., EP10167569.
Acknowledgements
We thank Gordan Mikuljan, Peter Modregger and István Mohácsi from the Paul Scherrer Institute (PSI), Switzerland, for the work on the mechanical design, the scientific advice and the SEM images respectively, Joachim Schulz and Marco Walter from Microworks GmbH, Germany, for the competent support on grating design issues, Christian Kottler and Vincent Revol from Centre Suisse d'Électronique et de Microtechnique (CSEM), Switzerland for the fruitful discussions on the design of the system. This work has been partially supported by the Competence Centre for Materials Science and Technology (CCMX) of the ETH Board, Project Nr. 61 and by the ERC Grant ERC-2012-StG 310005-PhaseX.
Author information
Authors and Affiliations
Contributions
C.D. and M.S. conceived the edge-on illumination scheme for grating interferometry. T.T., Z.W. and M.A. designed the first experiment. T.T. and M.A. performed the experiment and analyzed the data. T.T. and M.A. wrote the manuscript with contributions from all authors.
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article's Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder in order to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Thüring, T., Abis, M., Wang, Z. et al. X-ray phase-contrast imaging at 100 keV on a conventional source. Sci Rep 4, 5198 (2014). https://doi.org/10.1038/srep05198
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/srep05198
- Springer Nature Limited
This article is cited by
-
Quantitative X-ray phase contrast computed tomography with grating interferometry
European Journal of Nuclear Medicine and Molecular Imaging (2021)
-
Simulation study on X-ray phase contrast imaging with dual-phase gratings
International Journal of Computer Assisted Radiology and Surgery (2019)
-
X-ray Dark-field Radiography - In-Vivo Diagnosis of Lung Cancer in Mice
Scientific Reports (2017)
-
Large field of view, fast and low dose multimodal phase-contrast imaging at high x-ray energy
Scientific Reports (2017)
-
Realistic wave-optics simulation of X-ray phase-contrast imaging at a human scale
Scientific Reports (2015)