Abstract
To design and investigate a comparison card to evaluate the glenoid bone defect compared with Sugaya method. 33 patients with bony Bankart lesions were included. The comparison card and Sugaya method were performed on two occasions by three participants. The intra-group correlation coefficient (ICC) analysis and the inter-group correlation coefficient analysis of two measurements was performed. The concordance of the two methods was assessed using Bland–Altman analysis. Firstly, the percentage of defect measured by Sugaya method was 10.32 ± 8.38, and the comparison card method was 10.26 ± 8.41, 10.15 ± 8.23, and 10.62 ± 8.48, separately. There was no statistically significant difference (P > 0.05). The second measurement showed it was 10.37 ± 8.39 for Sugaya method, and 10.23 ± 8.37, 10.15 ± 8.35, 10.54 ± 8.49 for the comparison card, without a statistically significant difference (P > 0.05). For the comparison card, the intra- and inter-observer ICC values were all > 0.75. In the first measurement, Bland–Altman analysis demonstrated agreement between the two methods (bias, −0.03; SD, 0.48; − 0.97– 0.91; 95% CI, − 0.1999– 0.1413). Agreement was also found between them (bias, 0.07; SD, 0.61; − 1.13– 1.26; 95% CI, − 0.1509– 0.2812) in the second measurement. The comparison card method has similar accuracy with Sugaya method, which is of great reliability and convenience.
Similar content being viewed by others
Introduction
Fractures of the anteroinferior aspect of the glenoid rim, known as bony Bankart lesions, can occur frequently in the setting of traumatic anterior shoulder dislocation, causing anterior instability of shoulder1,2. An osseous Bankart lesion is usually secondary to direct violence or recurrent erosion of shoulder dislocation, manifested as compression or bone defect, mainly located at the anterior rim of glenoid. Glenoid bone loss can be seen in about 40% of patients with first discovered dislocation and 85% of patients with recurrent dislocation3. Previous studies have demonstrated that the bone loss was associated with great risk of re-dislocation after surgical repair4. Therefore, it is crucial to evaluate the glenoid bone loss threshold value for therapeutic decision-making, and accurate preoperative quantification is paramount in preventing surgical failure 5,6.
The evaluation of glenoid defect is of controversy. Several methods have been described to measure the amount of glenoid bone defect, including radiogram, Computed Tomography (CT) scan, Magnetic Resonance Imaging (MRI), and arthroscopy7. But in determining surgical procedure, preoperative 3D CT could provide the most accurate, reliable, and reproducible estimation of glenoid bone loss6. Sugaya method is widely used and highly recognized based on the circle formed by inferior glenoid8. However, the usage is limited in some circumstances due to the complicated software calculation. Also it’s hard to draw an exact curve if the defect shape was irregular. Nevertheless, it is increasingly necessary to develop a new method to detect and measure the glenoid defect with convenience and accuracy. The purpose of our study is to design a comparison card to evaluate the glenoid bone defect quickly and accurately, and validate its intra-observer and inter-observer reliability at the same time.
Results
A total of 33 patients with bony Bankart lesions were included in the study, including 26 males and seven females from August 2016 to May 2020, with an average of 37.5 years of age.
All patients were measured by the two methods. The first results from different observers had no statistically significant difference (P > 0.05). The second results after three months were not statistically different among observers as well (P > 0.05) (Table 1). The intra- and inter-observer ICC values were all > 0.75, and the reliability was good (Table 2).
In the first measurement, concordance was found between Sugaya method and the comparison card (bias, − 0.03; SD, 0.48; lower limit, − 0.97; upper limit, 0.91; 95% CI, − 0.1999−0.1413) using Bland–Altman analysis. In the second measurement, concordance was also found between the two methods (bias, 0.07; SD, 0.61; lower limit, − 1.13; and upper limit, 1.26; 95% CI, − 0.1509−0.2812). It was showed that most plot were within the average area of ± 1.96, showing good equivalence. Figure 1a, b showed the difference in glenoid defect between Sugaya and comparison card method was clinically nonsignificant.
Discussion
Glenoid rim fractures can be classified into three types: Type I, a displaced avulsion fracture with attached capsule; Type II, a medially displaced fragment malunited to the glenoid rim; and Type III, erosion of the glenoid rim with less than 25% (Type IIIA) or greater than 25% (Type IIIB) deficiency9. Any fracture of glenoid rim can decrease the glenohumeral contact area and disrupt the inferior glenohumeral ligament, altering the static stabilizers and predisposing patients to recurrence. Likewise, glenoid bone loss, either from an acute low-energy compression fracture or chronic attenuation secondary to recurrent shoulder dislocations, may also contribute to anterior instability by decreasing the articular arc length and reducing the surface area that resists shear and axial forces10. Kim11 reported that if the defect was less than 15%, surgical repairment of labrum could restore the stability. But if the defect was greater than 25%, osseous reconstruction was required to obtain better postoperative shoulder function. Therefore, it was critical to precisely quantify the bone loss for making surgical decisions.
The assessment of bone loss, ranging from classic radiographs, through CT with 2D and 3D reconstructions and MRI has been widely described7. Radiogram is easy to obtain and the loss of sclerotic glenoid line (LSGL) on anteroposterior radiographs is a moderately sensitive but highly specific finding for anterior glenoid rim defects12. However, not being able to quantify the glenoid bone defect has limited the clinical usage. MRI is irreplaceable in evaluating soft tissue and reducing fluorescence exposure, some studies have demonstrated its usage in measuring glenoid bone defects, but not as accurate as three-dimensional CT13. Arthroscope is an invasive method, and the position of bare spot is still controversial, which may overestimate the defect size.
Compared with other methods, 3D CT is the golden standard for measuring glenoid defects. The best-circle method is widely used and well recognized, including Barchilon method, Sugaya method, and Pico method7. In Barchilon method, the anterior area is calculated as the ratio between the depth and radius of inferior glenoid circle14. Because of the irregular defect edges, different measurement points have different distances from the center of circle. Therefore, the measurement error cannot be ignored and the applicationis limited.
The principle and calculation of Sugaya method and Pico method are similar. In Sugaya method8 we draw a circle that adequately fit the inferior aspect of glenoid. If there were free bone fragments, their area could be calculated by computer software directly. If there were no bone fragments, the missing part of circle was equivalent to the bone defect, and the proportion of defects equaled the ratio between the missing area and best-fit circle. Similarly, Pico method15 is to perform a CT scan of bilateral shoulders to reconstruct the enface view of both glenoids. By drawing a best-fit circle in the contralateral glenoid, we moved this circle to the affected side. The area of missing part is calculated by software to obtain the amount of bone defect and defect ratio.
Magarelli16 have demonstrated that Pico method was reliable in measuring glenoid bone defects. However, both Sugaya method and Pico method require the aid of professional software and a certain calculation process, which is difficult for physicians to master and popularize. Due to the limited popularity of computer software and difficulty in software learning, some surgeons made an approximate estimation of defect through visual inspection. But the accuracy is hard to guarantee. Therefore, it is of great essential to develop an method that is easy to use and can measure the glenoid defect accurately at the same time.In this study, we combined digital graphic analysis and Riemann Integral algorithm to design a comparison card (vector diagram) and used the statistical analysis to validate the inter-observer and intra-observer reliability of the comparison card. In addition, we compared the performance of the comparison card method and Sugaya method, which had similar accuracy. Although the design process was a little complicated, we can print the shape and lines representing various amounts of bone loss on transparent film and use it as a template. This will aid surgeons in using zoom feature to correctly size the glenoid to match and read the defect ratio directly in future clinical practice. This design isn’t limited by structure variation and is applicable in different centers. Besides, it can avoid the complicated calculation process and software learning, and be mastered without learning process. Nevertheless, the comparison card makes it possible for physicians in different levels to measure tbone defect precisely and quickly, relieving them from complicated software manipulations.
There are some limitations of this study. First the limited sample size was one of the shortcomings of this study. Also Moroder has found if there were some degrees between observer and en face view of the glenoid, there would be some bias in the measurement of bone loss17. In this method, we still need a 3D image of the glenoid to measure, which might bring in some bias.
Conclusion
The comparison card method has similar accuracy with Sugaya method, which is of great reliability and convenience.
Methods
Participants
We included patients with unilateral anterior glenohumeral instability with osseous Bankart lesion, who were admitted to the Third Department of Orthopedics, Zhongshan Hospital of Guangdong Traditional Chinese Medicine University from August 2016 to May 2020. Approval was granted by the Ethics Committee of Traditional Chinese Medicine Hospital of Zhongshan (2021ZXZY-LLK-316). Written informed consent was obtained from patients. All methods were performed in accordance with the relevant guidelines and regulations.
Exclusion criteria: shoulder joint deformity, rheumatism, rheumatoid arthritis, severe osteoarthritis and other diseases that may cause abnormal glenoid morphology.
CT data
The CT images were taken with a continuous axial 0.625-mm slice thickness. The Digital Imaging and Communications in Medicine (DICOM) data of CT images were exported and analyzed using Mimics software 17.0 (Belgium Materialize Company), to make a multiplanar reconstruction.
Design the comparison card
Taking a circle (radius = 4) as an example Fig. 2), the area of quarter circle was πr2/4 = 4π, then we calculated the coordinate of k value by calculus method to get two equal green area, as shown in Fig. 2,
k = 1.615.
Therefore, a circle (radius = r) could be divided into n segments with equal area.
Find k, k1…kn as shown in Fig. 3.
Drew a standard circle with a diameter of 50 mm in Geomagic Design Direct (USA Raindrop Company). Drew two vertical lines across the center as x-axis and y-axis, and the coordinate of center was (0,0) (Fig. 4).
Based on the requirement of single unit area as 2.5%, ten equal grids of a quarter circle should be obtained. Based on the calculus formula, n = 5, we got:
k = 3.94, k1 = 7.99, k2 = 12.29, k3 = 17.17.
The area of ten equal units was approximately 49 mm2, according to k*y = 49 mm2, k1*y1 = 49 mm2…
It could be calculated: y = 12.45, y1 = 12.13 y2 = 11.42 y3 = 10.06 y4 = 6.51 as shown in Fig. 5.
Re-mirror the 1/4 circle to obtain a complete circle (Fig. 6a). We used Adobe Illustrator (Adobe company) to vectorize the picture and remove the background to obtain the comparison card with a high pixel resolution (Fig. 6b).
Using of comparison card
The comparison card was transparent. Superimposed the card with the en face image of glenoid obtained from 3D CT scan (Fig. 7). We ensured that the center of comparison card and the best-fit circle of inferior glenoid coincided with each other. Then we zoomed the feature to correctly size the glenoid to match the card. As the area of each grid accounted for 2.5%, the defect percentage (p) equaled the number of grids (n) in the defect area multiple 2.5%: p = n*2.5%.
If the bone defect could not fill the comparison card grid as shown in Fig. 8, it needed to be estimated by direct reading. To minimize the estimation error, the measurement accuracy could be improved by setting tolerance. Specifically defined as λ = 1.25% ± α, λ was the actual estimated value, α was the limit of tolerance. Through direct comparison and observation, if more than half of the grid was filled, it was calculated as 1.25% + α; if less than half of the grid, it was calculated as 1.25%-α; it was calculated as 1.25% when half of the grid was filled exactly. For statistical calculation, α was 0.5%. The limit error of a single grid was 0.75%, and the limit error of four grids was 3%, which was less than the current clinical requirement of 5%. Nevertheless, the accuracy requirement could be met.
Sugaya method
Consideringthe inferior aspect of glenoid resembled a circle, we drew a circle that fit the outline of rim (Fig. 9), and calculated the circle area as S by Mimics 17.0. We draw along the defect contour using a free curve to completely divide the circle to obtain an irregular surface, and the area was A. The percentage of bone defect (D) was the ratio between area A and the area of assumed fitting circle (S), D = A/S. Bony defects could be divided into large (> 20%), medium (5–20%), and small (< 5%).
Statistical analysis
The percentage of glenoid bone loss was represented as mean ± standard deviation. The comparison card and Sugaya method were used to measure the glenoid bone defect. Each measurement was performed on two separate occasions by a radiologist and two orthopedic surgeons, and there was 12 weeks’ interval between two measurements. The intra-group correlation coefficient (ICC) analysis was performed by three participants in 12 weeks, and the average value of inter-group correlation coefficient analysis was performed onthe two measurements by every participant. The reliability was considered good (ICC > 0.75), fair (0.4 < ICC < 0.75) or poor (ICC < 0.4)18. Statistical analysis was conducted using SPSS software version 25 (IBM, the U.S.). All tests were two-side, and P value less than 0.05 were considered statistically significant.
Agreement between the two methods was performed using Bland–Altman analysis. The average value of 3 observers was used as the final data of each glenoid. Statistical analysis was performed using MedCalc, version 15 (MedCalc Software Ltd).
Data availability
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
References
Dekker, T. J. et al. Risk factors for recurrence after arthroscopic instability repair-the importance of glenoid bone loss >15%, patient age, and duration of symptoms: A matched cohort analysis. Am. J. Sports Med. 48, 3036–3041. https://doi.org/10.1177/0363546520949840 (2020).
Nolte, P. C., Elrick, B. P., Bernholt, D. L., Lacheta, L. & Millett, P. J. The bony bankart: Clinical and technical considerations. Sports Med. Arthrosc. Rev. 28, 146–152. https://doi.org/10.1097/jsa.0000000000000286 (2020).
Griffith, J. F. et al. Prevalence, pattern, and spectrum of glenoid bone loss in anterior shoulder dislocation: CT analysis of 218 patients. AJR Am. J. Roentgenol. 190, 1247 (2008).
Shaha, J. S. et al. Redefining “critical” bone loss in shoulder instability: Functional outcomes worsen with “subcritical” bone loss. Am. J. Sports Med. 43, 1719–1725. https://doi.org/10.1177/0363546515578250 (2015).
Huijsmans, P. E. et al. Quantification of a glenoid defect with three-dimensional computed tomography and magnetic resonance imaging: A cadaveric study. J. Shoulder Elbow Surg. 16, 803–809. https://doi.org/10.1016/j.jse.2007.02.115 (2007).
Rerko, M. A., Pan, X., Donaldson, C., Jones, G. L. & Bishop, J. Y. Comparison of various imaging techniques to quantify glenoid bone loss in shoulder instability. J. Shoulder Elbow Surg. 22, 528–534. https://doi.org/10.1016/j.jse.2012.05.034 (2013).
Skupinski, J., Piechota, M. Z., Wawrzynek, W., Maczuch, J. & Babinska, A. The bony bankart lesion: How to measure the glenoid bone loss. Pol. J. Radiol. 82, 58–63 (2017).
Sugaya, H., Moriishi, J., Dohi, M., Kon, Y. & Tsuchiya, A. Glenoid rim morphology in recurrent anterior glenohumeral instability. J. Bone Joint Surg. Am. 85, 878–884. https://doi.org/10.2106/00004623-200305000-00016 (2003).
Bigliani, L. U., Newton, P. M., Steinmann, S. P., Connor, P. M. & McLlveen, S. J. Glenoid rim lesions associated with recurrent anterior dislocation of the shoulder. Am. J. Sports Med. 26, 41–45. https://doi.org/10.1177/03635465980260012301 (1998).
Piasecki, D. et al. Glenoid bone deficiency in recurrent anterior shoulder instability: Diagnosis and management. J. Am. Acad. Orthop. Surg. 17, 482–493. https://doi.org/10.5435/00124635-200908000-00002 (2009).
Kim, Y. K., Cho, S. H., Son, W. S. & Moon, S. H. Arthroscopic repair of small and medium-sized bony Bankart lesions. Am. J. Sports Med. 42, 86–94. https://doi.org/10.1177/0363546513509062 (2014).
Jankauskas, L., Rudiger, H. A., Pfirrmann, C. W., Jost, B. & Gerber, C. Loss of the sclerotic line of the glenoid on anteroposterior radiographs of the shoulder: A diagnostic sign for an osseous defect of the anterior glenoid rim. J. Shoulder Elbow Surg. 19, 151–156. https://doi.org/10.1016/j.jse.2009.04.013 (2010).
e Souza, P. M. et al. Recurrent anterior glenohumeral instability: The quantification of glenoid bone loss using magnetic resonance imaging. Skeletal. Radiol. 43, 1085–1092 (2014).
Barchilon, V. S., Kotz, E., Barchilon Ben-Av, M., Glazer, E. & Nyska, M. A simple method for quantitative evaluation of the missing area of the anterior glenoid in anterior instability of the glenohumeral joint. Skeletal. Radiol. 37, 731–736. https://doi.org/10.1007/s00256-008-0506-8 (2008).
Baudi, P. et al. How to identify and calculate glenoid bone deficit. Chir. Organi. Mov. 90, 145–152 (2005).
Magarelli, N. et al. Intra-observer and interobserver reliability of the “Pico” computed tomography method for quantification of glenoid bone defect in anterior shoulder instability. Skeletal Radiol. 38, 1071–1075. https://doi.org/10.1007/s00256-009-0719-5 (2009).
Moroder, P. et al. The effect of scapula tilt and best-fit circle placement when measuring glenoid bone loss in shoulder instability patients. Arthroscopy 34, 398–404. https://doi.org/10.1016/j.arthro.2017.08.234 (2018).
Ulijaszek, S. J. & Kerr, D. A. Anthropometric measurement error and the assessment of nutritional status. Br. J. Nutr. 82, 165–177. https://doi.org/10.1017/s0007114599001348 (1999).
Acknowledgements
This work was supported by Medical Research Fund of Guangdong Province (B2022100); Department of Science and Technology of Guangdong Province, Joint Fund of Provincial Basic and Applied Basic Research with enterprise (Public Health and Medical Health) (2021A1515220133); Zhongshan Social Welfare and Basic Research Project (2020B1058); Peking University People’s Hospital Scientific Research and Development Funds (RDG2021-01, RDL2021-08).
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Material preparation was performed by J.Y.C., J.B.J. and Z.X.H., data collection was performed by Y.F.W. and Y.C.Z., analysis was performed by D.W.G. and J.H.C. The first draft of the manuscript was written by L.C. and Y.C.Z., all authors commented on previous versions of the manuscript. All authors reviewed and approved the final manuscript.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Chen, L., Zhang, Y., Wu, Y. et al. Designing and validating a comparison card method for quantification of glenoid bone defect. Sci Rep 12, 16717 (2022). https://doi.org/10.1038/s41598-022-20908-y
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-022-20908-y
- Springer Nature Limited