Abstract
Absorption spectroscopy is widely used to detect samples with spectral specificity. Here, we propose and demonstrate a method for enhancing the sensitivity of absorption spectroscopy. Exploiting multiple light scattering generated by a boron nitride (h-BN) scattering cavity, the optical path lengths of light inside a diffusive reflective cavity are significantly increased, resulting in more than ten times enhancement of sensitivity in absorption spectroscopy. We demonstrate highly sensitive spectral measurements of low concentrations of malachite green and crystal violet aqueous solutions. Because this method only requires the addition of a scattering cavity to existing absorption spectroscopy, it is expected to enable immediate and widespread applications in various fields, from analytical chemistry to environmental sciences.
Similar content being viewed by others
Introduction
Absorption spectroscopy detects a trace of a specific substance by measuring the absorbance of light at various wavelengths1,2. Due to its instrumentational simplicity and high spectral specificity, absorption spectroscopy has been exploited in various fields, such as analytical chemistry3,4, atmospheric science5,6, geology7, and atomic physics8. The measured absorbance is proportional to the optical path length (OPL) l according to the Beer-Lambert law:
where I0 and I are the measured intensities in the absence of a sample and in the presence of a sample, respectively, A is the dimensionless absorbance of the sample, ε is the molar attenuation coefficient of the sample, and c is the molar concentration of the solution9.
The sensitivity of absorption spectroscopy can be enhanced by increasing l. For instance, multi-path absorption cells with mirrors have been employed10,11,12,13. Such systems can significantly amplify the OPL but generally require a bulky system and sophisticated alignment. Attenuated total reflection spectroscopy exploits multiple total internal reflections in a crystal, which creates an evanescent wave14,15. However, this method requires a complicated configuration, such as the exact optical contact between a sample and a crystal16. The need for a method of using simpler and more compact equipment to increase the OPL has been developed17.
Multiple light scattering has been exploited to increase the OPL and overcome the limitations of conventional optical elements18, including superlens19, temperature20, pressure sensing21, wavelength meters22, fiber-based spectrometers23, and non-resonant lasers24. The existing approaches employing the multiple scattering regime are represented by an integrating sphere25,26,27,28. Absorption spectroscopy using an integrating sphere has the advantage of increasing the OPL without the demand for precise alignment of a beam and optical components25.
Recently, Martin et al.29 suggested multiple scattering for enhancing the sensitivity of absorption spectroscopy. Dielectric microspheres were added to a sample solution to induce multiple scattering, which effectively increased the OPL. However, this method has various limitations: it requires the addition of microspheres for each measurement. Once mixed with microspheres for measurement, the remaining beads prevented the reuse of the sample. In addition, the degree of increase in the OPL varies depending on the quantity and size of the beads, as well as the refractive index (RI) difference between the microspheres and the sample. Thus, the enhancement factor varies according to the experimental environment.
Here, we present a simple but powerful method for increasing the OPL in absorption spectroscopy. By introducing a scattering cavity to existing absorption spectroscopy to enclose a sample, the sensitivity of the measurements is significantly improved, because the light is trapped inside a scattering cavity and interacts with a sample numerous times before exiting. This method does not perturb a sample or require complex instruments. In addition, the method does not require calibration for each measurement and has not interfered with the sample condition. Although there are methods that showed the increase of the OPL using an integrating sphere, our method can be an alternative in the practical perspective because the present scattering cavity is easily combined with an existing commercial cuvette and a spectrometer in the minimal modification. With this regard, we demonstrate that the limit of detection (LOD) can be lowered to what extent with a commercial cuvette.
Materials and methods
In the conventional method, light passes through a cuvette only once. In the proposed method, the diffusive surface of a scattering cavity scatters light, causing the diffusively reflected light to pass through a longer path, and the reflected light is collected several times (Fig. 1a). The experimental setup was composed of a halogen lamp (OSL1-EC, Thorlabs, Inc.), a custom-made scattering cavity, and a spectrometer (HR4000, Ocean Optics, Inc.) (Fig. 1b). Two linear polarizers (LPVISE100-A, Thorlabs, Inc.) were employed as the beam power attenuators. A short-pass filter (under 750 nm, FES0750, Thorlabs, Inc.) was utilized to filter the wavelengths that exceeded the spectral range of the spectrometer. Note that linear polarizers and short-pass filters are not essential components.
The scattering cavity was made of > 99.5% purity hexagonal boron nitride (h-BN). h-BN exhibits minimal absorbance30,31, and high diffuse reflectance in the visible region32 makes h-BN suitable for absorption spectroscopy. Due to its graphite-like atomic structure, h-BN exhibits excellent mechanical properties and exhibits satisfactory machinability33,34,35,36. The diffuse reflectance of h-BN was calibrated using a spectrophotometer (Lambda 1050, Perkin Elmer, Inc.) (Fig. 1c). h-BN has a high diffuse reflectance of more than 80% at wavelengths longer than 500 nm. The inverse adding-doubling method37,38 was used to estimate the absorption and scattering coefficients of h-BN. The estimated absorption coefficient \(\mu_{a}\) and reduced scattering coefficient \(\mu_{s}^{\prime }\) were 0.023 mm−1 and 129 mm−1, respectively, at 532 nm. The front and top views of the scattering cavity are shown in Fig. 1d. To prevent the direct passing of normal incident light, the exit hole was offset from the entrance with a height difference of 10 mm. The height difference makes the light hit the wall of the scattering cavity and multiply-reflected. Generally, the roughness of the surface will change the specular reflection portion of the reflected light. In our cavity geometry, however, the offset between the entrance and exit holes prevents the domination of the direct or few specular reflections, similar to the function of baffles in a conventional integrating sphere.
Results
Measurement of the normalized intensity and absorbance with the scattering cavity
To verify this method, the absorption spectra of malachite green and crystal violet were measured. The differentiation of substances from the absorption spectrum of their mixtures has been a difficult task in absorption spectroscopy, especially when two or more substances have overlapped absorption peaks39. We demonstrated the proposed method for pure solutions. Malachite green and crystal violet have high water solubility and exhibit maximum absorption at wavelengths of 617 nm and 590 nm, respectively40,41. We measured the malachite green aqueous solutions at various concentrations (Fig. 2). It was difficult to discern the malachite green solution with the naked eye when the concentration was below 1 µM (Fig. 2a). The absorption spectra were measured from the sample solution and deionized (DI) water, which correspond to I and I0, respectively, in Eq. (1). The normalized spectrum I/I0 using the conventional method was measured in the same manner. Figure 2b presents the normalized spectra using the proposed and conventional methods for 0.2 µM malachite green aqueous solution. The absorption dip at 617 nm can be clearly observed in the proposed method, whereas the conventional method provides a low signal-to-noise ratio (SNR).
The normalized spectra of malachite green and crystal violet aqueous solutions at various concentrations are shown in Fig. 3. For validation, five measurements were acquired from the same sample solution (Fig. 3a,b) and five individual solutions of the same concentration (Fig. 3c,d). We discovered that the absorption dips were consistently observed at 617 nm and 590 nm for malachite green and crystal violet, respectively, which shows an agreement with the known values. In addition, the linear relationship between concentration and absorbance in Eq. (1) is shown with high R-square values in the graphs of the absorbance (Fig. 3a–d). The linearity of the Beer-Lambert law with the effective OPL is valid for the solutions of low concentrations25. These results show the robustness of the presented method using multiple scattering.
Enhancement factor for the sensitivity of detection
According to the results using the scattering cavity in Fig. 3, it is possible to see a distinct dip of the normalized intensity. To quantify the enhancement in absorbance (−log(I/I0)) of the proposed method, we calculated the ratio of the absorbances between the proposed method and the conventional method (control) using the same sample solution (malachite green and crystal violet, 1 µM) (Fig. 4a,b). The normalized intensity depending on the wavelength contains 25 data for each case. Figure 4c,d show the enhancement factor as a function of the wavelength for both aqueous solutions. For wavelengths near the dips, the enhancement factors were almost constant regardless of the wavelength. The highly variable enhancement in the low-absorbance regimes is due to the near-zero absorbance of the control experiments. The averaged enhancement factors at the wavelengths near the dip were 10.22 times and 10.41 times for malachite green and crystal violet, respectively (Fig. 4e). Because the increment in the OPL is not dependent on the sample, the result of a similar enhancement for the two samples is reasonable.
Limit of detection
The LOD is the lowest concentration of a sample that discernable with the absence of the sample42 and can be lower by enhancing the measured absorbance. To determine the LOD of the proposed system, we used highly diluted malachite green solutions (Fig. 5). The LOD was obtained by interpolating the concentration that yields the absorbance criterion μ0 + 3σ0, where μ0 and σ0 are the mean and standard deviation, respectively, of the results using pure water (0 µM) 42,43. The evaluation of the LOD via pure water (blank solution) fundamentally reflects the systematic error of the detection system.
Figure 5a shows the absorbance measurements for the LOD calibration. We employed malachite green solutions at concentrations of 0.040 µM and 0.004 µM for the conventional system (control) and proposed system, respectively. Figure 5b,c show the shaded regions in Fig. 5a. The absorbance was calculated from the mean and standard deviation of the 0 µM (pure water) absorbance results. We found that the absorbance criteria were μ0 + 3σ0 = 0.01 in both cases. Note that the absorbance criteria should be the same in both cases because the zero fluctuations are solely derived from the sensitivity and precision of the detection system. In the conventional method (control), the linear correlation was maintained up to 0.120 µM, but the deviation increased for concentrations lower than 0.100 µM (Fig. 5b). The calibrated LOD was 0.059 µM, which was consistent with the data observations. In the proposed method, it is noteworthy that the linearity was maintained at substantially lower concentrations (minimum of 0.010 µM) than in the conventional method (Fig. 5c). The calibrated LOD was 0.004 µM, which is lower than 1/10 of the LOD of the control.
Conclusion
In summary, we present a scattering cavity to enhance the sensitivity of absorption spectroscopy via a significant increase in the OPL. The scattering cavity was composed of h-BN and led to diffusive reflection on its surface, achieving an improvement in the absorbance by more than 10 times. Moreover, the LOD was drastically lowered in the scattering cavity, and malachite green could be detected even at 0.004 µM using a commercial cuvette. We expect that a higher diffuse reflectance will further improve the enhancement factor. For example, a commercially available highly reflective surface shows > 99% diffuse reflectance in the visible and near-infrared range44; many common ceramics (such as alumina and zirconia) or synthetic polymers (such as PTFE) with high reflectance can be utilized instead. Furthermore, the proposed method may be improved by optimizing the size of the entrance and the scattering cavity for further enhancement of the OPL24.
It should be emphasized that the proposed method using multiple scattering is more robust against misalignment than the methods using specular reflection of a mirror. In addition, the proposed method enables quick and simple measurement of absorbance using a plastic cuvette, which is commonly employed in analytical chemistry and can be readily applicable to existing spectrometers. Due to the effectiveness of the proposed method in detecting trace amounts of a sample mixed in a liquid solvent, the method is expected to widen the detection range for a substance and the wavelength of interest, expanding the research area. We believe that our system can be beneficial for practical applications that require low LOD in water45, such as those in the food industry46, pathological diagnosis47, and biochemical sensing48,49.
References
McManus, J. B., Kebabian, P. L. & Zahniser, M. Astigmatic mirror multipass absorption cells for long-path-length spectroscopy. Appl. Opt. 34, 3336–3348 (1995).
Platt, U. & Stutz, J. Differential absorption spectroscopy. in Differential optical absorption spectroscopy 135–174 (Springer, 2008).
BV, D.C. & Hydraulics, D. Absorption spectroscopy. (1962).
Fuller, M. et al. Chemical characterization of tribochemical and thermal films generated from neutral and basic ZDDPs using X-ray absorption spectroscopy. Tribol. Int. 30, 305–315 (1997).
Brown, S. S. Absorption spectroscopy in high-finesse cavities for atmospheric studies. Chem. Rev. 103, 5219–5238 (2003).
Werle, P., Mücke, R. & Slemr, F. The limits of signal averaging in atmospheric trace-gas monitoring by tunable diode-laser absorption spectroscopy (TDLAS). Appl. Phys. B 57, 131–139 (1993).
Chen, Y. et al. Applications of micro-fourier transform infrared spectroscopy (FTIR) in the geological sciences: A review. Int. J. Mol. Sci. 16, 30223–30250 (2015).
Van Loon, A. T. Analytical atomic absorption spectroscopy: Selected methods (Elsevier, London, 2012).
Dean, J. A. & Dean, J. Analytical chemistry handbook (McGraw-Hill, New York, 1995).
White, J. U. Long optical paths of large aperture. JOSA 32, 285–288 (1942).
Herriott, D. R. & Schulte, H. J. Folded optical delay lines. Appl. Opt. 4, 883–889 (1965).
Romanini, D., Kachanov, A., Sadeghi, N. & Stoeckel, F. CW cavity ring down spectroscopy. Chem. Phys. Lett. 264, 316–322 (1997).
Robert, C. Simple, stable, and compact multiple-reflection optical cell for very long optical paths. Appl. Opt. 46, 5408–5418 (2007).
Averett, L. A., Griffiths, P. R. & Nishikida, K. Effective path length in attenuated total reflection spectroscopy. Anal. Chem. 80, 3045–3049 (2008).
Fahrenfort, J. Attenuated total reflection: A new principle for the production of useful infra-red reflection spectra of organic compounds. Spectrochim. Acta 17, 698–709 (1961).
Grdadolnik, J. ATR-FTIR spectroscopy: Its advantage and limitations. Acta Chim. Slov. 49, 631–642 (2002).
Tuzson, B., Mangold, M., Looser, H., Manninen, A. & Emmenegger, L. Compact multipass optical cell for laser spectroscopy. Opt. Lett. 38, 257–259 (2013).
Park, J. H., Park, J., Lee, K. & Park, Y. Disordered optics: Exploiting multiple light scattering and wavefront shaping for nonconventional optical elements. Adv. Mater. 32, 1903457 (2020).
Park, J.-H. et al. Subwavelength light focusing using random nanoparticles. Nat. Photon. 7, 454 (2013).
Trivedi, V. et al. Optical temperature sensor using speckle field. Sens. Actuat. A 216, 312–317 (2014).
Kim, K. et al. Remote sensing of pressure inside deformable microchannels using light scattering in Scotch tape. Opt. Lett. 41, 1837–1840 (2016).
Mazilu, M., Vettenburg, T., Di Falco, A. & Dholakia, K. Random super-prism wavelength meter. Opt. Lett. 39, 96–99 (2014).
Redding, B. & Cao, H. Using a multimode fiber as a high-resolution, low-loss spectrometer. Opt. Lett. 37, 3384–3386 (2012).
Lee, K., Ma, H. J., Rotermund, F., Kim, D. K. & Park, Y. Non-resonant power-efficient directional Nd: YAG ceramic laser using a scattering cavity. Nat. Commun. 12, 1–10 (2021).
Hodgkinson, J., Masiyano, D. & Tatam, R. P. Using integrating spheres as absorption cells: Path-length distribution and application of Beer’s law. Appl. Opt. 48, 5748–5758 (2009).
Pope, R. M. & Fry, E. S. Absorption spectrum (380–700 nm) of pure water II. Integrating cavity measurements. Appl. Opt. 36, 8710–8723 (1997).
Villanueva, Y., Veenstra, C. & Steenbergen, W. Measuring absorption coefficient of scattering liquids using a tube inside an integrating sphere. Appl. Opt. 55, 3030–3038 (2016).
Elterman, P. Integrating cavity spectroscopy. Appl. Opt. 9, 2140–2142 (1970).
Koman, V. B., Santschi, C. & Martin, O. J. Multiscattering-enhanced absorption spectroscopy. Anal. Chem. 87, 1536–1543 (2015).
Watanabe, K., Taniguchi, T. & Kanda, H. Direct-bandgap properties and evidence for ultraviolet lasing of hexagonal boron nitride single crystal. Nat. Mater. 3, 404–409 (2004).
Wang, J., Ma, F. & Sun, M. Graphene, hexagonal boron nitride, and their heterostructures: properties and applications. RSC Adv. 7, 16801–16822 (2017).
Spiridonov, D., Henaish, A., Vokhmintsev, A. & Weinstein, I. Diffuse reflectance spectral features of hexagonal boron nitride nanopowder. in AIP conference proceedings, Vol. 1886 020021 (AIP Publishing LLC, 2017).
Ooi, N., Rajan, V., Gottlieb, J., Catherine, Y. & Adams, J. Structural properties of hexagonal boron nitride. Modell. Simul. Mater. Sci. Eng. 14, 515 (2006).
Eichler, J. & Lesniak, C. Boron nitride (BN) and BN composites for high-temperature applications. J. Eur. Ceram. Soc. 28, 1105–1109 (2008).
Zunger, A., Katzir, A. & Halperin, A. Optical properties of hexagonal boron nitride. Phys. Rev. B 13, 5560 (1976).
Song, L. et al. Large scale growth and characterization of atomic hexagonal boron nitride layers. Nano Lett. 10, 3209–3215 (2010).
Prahl, S. A., van Gemert, M. J. & Welch, A. J. Determining the optical properties of turbid media by using the adding–doubling method. Appl. Opt. 32, 559–568 (1993).
Prahl, S. Everything I think you should know about Inverse Adding-Doubling 1–74 (Oregon Medical Laser Center, 2011).
Belal, T. S., Daabees, H. G., Abdel-Khalek, M. M., Mahrous, M. S. & Khamis, M. M. New simple spectrophotometric method for determination of the binary mixtures (atorvastatin calcium and ezetimibe; candesartan cilexetil and hydrochlorothiazide) in tablets. J. Pharm. Anal. 3, 118–126 (2013).
Modirshahla, N. & Behnajady, M. A. Photooxidative degradation of Malachite Green (MG) by UV/H2O2: Influence of operational parameters and kinetic modeling. Dyes Pigm. 70, 54–59 (2006).
Gupta, A., Pal, A. & Sahoo, C. Photocatalytic degradation of a mixture of Crystal Violet (Basic Violet 3) and Methyl Red dye in aqueous suspensions using Ag+ doped TiO2. Dyes Pigm. 69, 224–232 (2006).
Long, G. L. & Winefordner, J. D. Limit of detection. A closer look at the IUPAC definition. Anal. Chem. 55, A712–A724 (1983).
Shrivastava, A. & Gupta, V. B. Methods for the determination of limit of detection and limit of quantitation of the analytical methods. Chronic Young Sci. 2, 21 (2011).
Kokaly, R. F. & Skidmore, A. K. Plant phenolics and absorption features in vegetation reflectance spectra near 1.66 μm. Int. J. Appl. Earth Observ. Geoinform. 43, 55–83 (2015).
Heberer, T. Tracking persistent pharmaceutical residues from municipal sewage to drinking water. J. Hydrol. 266, 175–189 (2002).
Terry, L. A., White, S. F. & Tigwell, L. J. The application of biosensors to fresh produce and the wider food industry. J. Agric. Food Chem. 53, 1309–1316 (2005).
Bieschke, J. et al. Ultrasensitive detection of pathological prion protein aggregates by dual-color scanning for intensely fluorescent targets. Proc. Natl. Acad. Sci. 97, 5468–5473 (2000).
Slack, J. A. et al. Biochemical markers of cardiac injury in normal, surviving septic, or nonsurviving septic neonatal foals. J. Vet. Intern. Med. 19, 577–580 (2005).
Caucheteur, C., Guo, T. & Albert, J. Review of plasmonic fiber optic biochemical sensors: Improving the limit of detection. Anal. Bioanal. Chem. 407, 3883–3897 (2015).
Acknowledgements
This work was supported by KAIST UP program, BK21+ program, KAIST Advanced Institute for Science-X, The.Wave.Talk, National Research Foundation of Korea (NRF) (2017M3C1A3013923, 2015R1A3A2066550, 2018K000396, 2018R1A6A3A01011043, 2021R1C1C2009220).
Author information
Authors and Affiliations
Contributions
Y.P. conceived the project. J.O. and K.L. developed the instrument and the analytical tools. J.O. conducted the experiment. All authors wrote the manuscript. Y.P. provided supervision.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Oh, J., Lee, K. & Park, Y. Enhancing sensitivity in absorption spectroscopy using a scattering cavity. Sci Rep 11, 14916 (2021). https://doi.org/10.1038/s41598-021-94028-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-94028-4
- Springer Nature Limited