Abstract
In this third-generation dual-source CT (DSCT) study, we retrospectively investigated radiation dose and image quality of portal-venous high-pitch emergency CT in 60 patients (28 female, mean age 56 years) with a body mass index (BMI) ≥ 30 kg/m2. Patients were dichotomized in groups A (median BMI 31.5 kg/m2; n = 33) and B (36.8 kg/m2; n = 27). Volumetric CT dose index (CTDIvol), size-specific dose estimate (SSDE), dose length product (DLP) and effective dose (ED) were assessed. Contrast-to-noise ratio (CNR) and dose-independent figure-of-merit (FOM) CNR were calculated. Subjective image quality was assessed using a five-point scale. Mean values of CTDIvol, SSDE as well as normalized DLP and ED were 7.6 ± 1.8 mGy, 8.0 ± 1.8 mGy, 304 ± 74 mGy * cm and 5.2 ± 1.3 mSv for group A, and 12.6 ± 3.7 mGy, 11.0 ± 2.6 mGy, 521 ± 157 mGy * cm and 8.9 ± 2.7 mSv for group B (p < 0.001). CNR of the liver and spleen as well as each calculated FOM CNR were significantly higher in group A (p < 0.001). Subjective image quality was good in both groups. In conclusion, third-generation abdominal high-pitch emergency DSCT yields good image quality in obese patients. Radiation dose increases in patients with a BMI > 36.8 kg/m2.
Similar content being viewed by others
Introduction
Obesity, which is defined as a body mass index (BMI) of at least 30 kg/m2, represents a growing issue in the western societies1. Obese patients pose a special diagnostic challenge in the emergency setting, as the image quality of ultrasound and X-ray is limited. In this context, a better image quality is associated with a higher patient radiation dose2. Furthermore, both the gantry diameter and ad-hoc availability of magnetic resonance imaging are limited3,4. Hence, computed tomography (CT) often represents the only non-invasive and easily available examination modality to clarify the medical condition of an obese patient. As an expression of the technical adaptation to an increasingly obese patient population, modern CT scanners have larger gantry diameters of up to 80 cm, patient tables with a maximum weight capacity of 300 kg, a higher tube power simultaneously providing a stable table speed despite a higher table load, as well as a larger field of view5.
In this situation, dual-source CT (DSCT), which was introduced in the mid 2000s, provides simultaneous usage of two tubes3,5,6. However, particularly in young obese patients the optimal balance between radiation dose penalty and potential endangerment of the patient due to reduced image quality caused by dose limitation is essential.
DSCT scanners of the current, third generation have two X-ray tubes with a maximal generator power of 120 kW each, enabling tube currents up to 1300 mA each. In combination with automated dose modulation and iterative reconstruction, these features contribute to image quality and radiation dose optimization in obese patients5,7. Furthermore, third-generation DSCT scanners provide high-pitch protocols with a pitch factor of up to 3.2. This technique allows for the continuous scan of a complete abdominal volume in less than one second5, consequently reducing both motion artifacts and probably (when combined with a reduced rotation time of up to 0.25 s)5 radiation dose according to several studies8,9,10. In obese patients, however, these advantages in turn yield a higher image noise due to a limited tube current: CT tubes generate only a certain amount of X-radiation during the extremely short acquisition times of high-pitch CT protocols. Hence, high-pitch CT could not be applied in a non-weight-selected patient population without taking the risk of a reduced diagnostic confidence in obese patients so far. This was confirmed by high-pitch CT studies which to date mainly analyzed normal-weight individuals10,11,12.
The aim of this study was to evaluate the impact of a high-pitch CT acquisition protocol on radiation dose and image quality in obese patients, who received a portal-venous emergency CT of the abdomen on a third-generation DSCT scanner.
Material and Methods
Patient selection
This retrospective single-center study was approved by the responsible institutional review board (Project Number 811-16) of the Ludwig-Maximilians-University Munich with a waiver for written informed consent. The study was performed in accordance with the Declaration of Helsinki.
All examinations were carried out on a third-generation DSCT scanner (SOMATOM Force, Siemens Healthineers) between February 2015 and December 2016. Within this time span, a total of 60 patients met the following inclusion criteria:
-
Age ≥ 18 years
-
BMI ≥ 30 kg/m2
-
Acute abdominal pain
-
High-pitch abdominal DSCT in portal-venous phase
Non-contrast and low-dose CT, repeated CT scans of the same patient, and examinations of patients with a body weight > 300 kg were excluded from further analysis.
CT acquisition protocol
Table 1 summarizes the CT acquisition parameters. Each patient was examined in dual-source, helical acquisition mode. Concerning the applied high-pitch dual-source mode (Flash mode, Siemens Healthineers)13 of third-generation DSCT, both X-ray tubes start up simultaneously, yielding substantial increase of overall tube current and thus enabling CT examinations of overweight individuals3,5,6,7. The scan range comprised the whole abdomen, extending from the diaphragm to the pubic symphysis. Automated dose modulation was enabled by default, in terms of automated tube current modulation (ATCM; CARE Dose 4D, Siemens Healthineers)12,14,15 and automated tube voltage selection (ATVS; CARE kV, Siemens Healthineers)16,17 with a quality reference tube current time product of 80 mAs and a quality reference tube voltage of 140 kV. The gantry rotation time was 0.25 seconds per rotation, the pitch 1.55, the section thickness 0.75 mm, and the slice collimation 192 × 0.6 mm. A body-weight-adapted non-ionic iodinated contrast agent (Iomeprol 400 mg iodine/ml, 1 ml per kg; Imeron, Bracco Imaging) was administered intravenously at a flow rate of 2.5 mL/s followed by 100 mL of saline. CT images were acquired using the bolus-tracking technique (CARE Bolus, Siemens Healthineers), with the region of interest (ROI) manually placed in the abdominal aorta and the trigger threshold set to 100 Hounsfield units (HU) with a 95-second delay for the portal venous phase. Advanced modeled iterative reconstruction (ADMIRE), strength 3, was enabled in each patient.
Radiation metrics
Individual values of selected tube voltage and tube current time product as well as the volumetric CT dose index (CTDIvol) and dose length product (DLP) were documented from the dose report, which was automatically stored in the picture archiving and communication system (Syngo Imaging 2010, Siemens Healthineers). The DLP was normalized for a typical abdominal scan length of 40 cm14,15. The normalized effective dose (ED) was then calculated by multiplying the normalized DLP with the specific conversion factor k for combined adult abdomen and pelvic CT of 0.017 mSv/mGy * cm18,19.
We furthermore assessed the size-specific dose estimate (SSDE) which according to Christner et al. reflects the patient dose more independent of size20. Analogously to other authors9,20,21, the SSDE was calculated by multiplying the CTDIvol with the size-specific conversion factor fsize according to the AAPM Report 20420,22. The individual fsize was obtained by summing the anteroposterior (AP) and lateral (LAT) diameters from transverse CT images at the mid-liver level (size = AP + LAT)20.
Objective image quality
In each patient, the attenuation was measured by a radiologist with 8 years of experience in abdominal imaging (R.F.). In detail, on portal-venous phase images round or oval ROIs were manually placed within the liver (ROI size, 150–300 mm2), the pancreas (100–200 mm2), the spleen (150–300 mm2), the renal cortex (100–200 mm2) as well as the abdominal aorta (35–150 mm2) and main portal vein (40–80 mm2)15. For each region, three individual measurements were performed and averaged. Measurements of the liver, the abdominal aorta and main portal vein were conducted at the same level. During measurement, the radiologist carefully avoided focal luminal or parenchymal heterogeneities such as calcifications, thrombotic material, focal lesions, ducts and/or artifacts.
To determine image noise and contrast-to-noise ratio (CNR) of the individual anatomic structure, ROIs with a size of 200–400 mm2 were manually placed in the psoas muscles and the adjacent mesenteric fatty tissue. According to Wichmann and colleagues15, image noise was defined as the standard deviation (SD) of the ROI in the mesenteric fat (SDfat), and the organ-specific CNR was calculated as following:
As ATVS was enabled, we furthermore calculated figure-of-merit (FOM) values for each organ in both groups in order to provide objective data of differences in CNR independent from the ED23,24. The FOM CNR can be calculated using the following formula15:
Evaluation of subjective image quality and motion artifacts
All portal-venous abdominal CT images were read in consensus by two radiologists with 8 (R.F.) and 14 (C.G.T.) years of experience in abdominal imaging. Images were randomly analyzed with freely adjustable window settings. The readers were blinded to the individual patient data and CT reports.
Subjective image quality was evaluated by using a five-point scale according to Guimaraes et al.25: 1, excellent image quality; 2, good image quality; 3, fair but comprised image quality; 4, poor image quality; 5, non diagnostic, severe distortion.
Similar to others12, motion artifacts were categorized as ‘none’, ‘minor’ or ‘major’. Minor motion artifacts were defined as ‘moderately reduced image quality, sufficient diagnostic confidence’, while major motion artifacts were defined as ‘severely reduced image quality, no diagnostic confidence’.
Influence of size on radiation dose and image quality
For further analysis of radiation dose and image quality, the study population was dichotomized on the basis of the mean size at the mid-liver level (AP + LATmean)20. In detail, the calculated AP + LATmean was 73.6 ± 6.2 cm (72.0–75.2 cm). Accordingly, 33/60 patients were included in group A (AP + LAT < 73.6 cm) and 27/60 patients in group B (AP + LAT > 73.6 cm).
Statistics
Data analysis was performed using IBM SPSS Statistics for Windows, Version 24.0 (IBM Corp., Armonk, N.Y., USA). A level of significance of α = 0.05 was used throughout the study. Data were initially assessed for normality applying the Kolmogorov-Smirnov test. Continuous variables are provided as mean ± standard deviation (95% confidence interval). Variables that do not follow normal distribution are shown as median (25%; 75% interquartile range).
Variables of the two groups were compared according to the t test if data were normally distributed. The Mann-Whitney U test was used if data were not normally distributed. The Spearman correlation analysis was applied to investigate the impact of BMI on ED.
Results
Of the sixty patients included in this study, 28 were female and 32 male. The mean patient age was 56 ± 17 years.
Patient characteristics
The patient characteristics are summarized in Table 2. The median BMI was 31.5 kg/m2 (30.8 kg/m2; 33.5 kg/m2) in group A and 36.8 kg/m2 (33.2 kg/m2; 40.2 kg/m2) in group B (p < 0.001). Difference in patient’s height between groups did not reach statistical significance (p = 0.051).
Radiation dose
Table 3 and Fig. 1 provide data on radiation dose for both groups. In group A, median selected tube voltage was 100 kV (90 kV; 100 kV) and mean selected tube current time product 213.6 ± 39.4 mAs (199.7–227.6 mAs), while group B was characterized by a median selected tube voltage of 100 kV (100 kV; 120 kV) and a mean selected tube current time product of 259.0 ± 45.4 mAs (241.0–277.0 mAs) (p < 0.001). Minimum and maximum values of selected tube voltage were 80 kV (n = 1, group A) and 130 kV (n = 3, group B).
Mean values of CTDIvol, SSDE and DLP were 7.58 ± 1.84 mGy (6.92–8.23 mGy), 7.99 ± 1.78 mGy (7.36–8.62 mGy) and 363.7 ± 101.5 mGy * cm (327.8–399.7 mGy * cm) for group A, and 12.60 ± 3.74 mGy (11.12–14.07 mGy), 10.99 ± 2.55 mGy (9.99–12.00 mGy) and 630.6 ± 182.4 mGy * cm (558.5–702.8 mGy * cm) for group B (p < 0.001).
Using a 40 cm normalization, mean values for DLP and ED were 303.8 ± 73.9 mGy * cm (277.6–329.0 mGy * cm) and 5.16 ± 1.26 mSv (4.72–5.61 mSv) in group A in comparison to 521.2 ± 157.1 mGy * cm (459.1–583.4 mGy * cm) and 8.86 ± 2.67 mSv (7.80–9.92 mSv) in group B (p < 0.001).
Regarding the total study population (n = 60), we observed a significant positive correlation between BMI and ED (rs = 0.583, p < 0.001).
Objective image quality
Results from objective analysis of image quality are illustrated in Table 4 and Fig. 2. Significant differences of mean values were observed regarding CNR of the liver (group A: 5.58 ± 1.95 (4.89–6.27); group B: 3.93 ± 2.17 (3.07–4.79); p = 0.003) and spleen (group A: 6.43 ± 1.8 (5.79–7.07); group B: 5.44 ± 1.55 (4.83–6.05); p = 0.027). All other pairwise comparisons of organ-specific CNR (pancreas, kidneys, aorta, portal vein) did not reach statistical significance (p ≥ 0.120).
Regarding dose-independent FOM CNR, the values in group A were significantly higher for all measured organs when compared to group B (p < 0.001). The greatest difference was calculated for the liver with a FOM CNR increment factor of 3.4.
Subjective image quality
Table 5 and Fig. 3 provide results of subjective image quality evaluation. According to two readers, the median rating of overall image quality was 2 (good) in both groups; however, difference of ratings between groups reached significance towards a higher image quality in group A (p = 0.035). In detail, examinations in group A were exclusively rated as 1 (n = 13/33) or 2 (n = 20/33), whereas ratings in group B were 1 in 6/27, 2 in 16/27, and 3 in 5/27 examinations. Of the five patients in whom the image quality was rated 3, the mean BMI was 44 kg/m2. For both groups, no minor or major motion artifacts were documented.
Discussion
In the present study we investigated radiation dose and image quality of portal-venous high-pitch emergency CT of the abdomen in obese patients on a third-generation DSCT scanner. We believe, that our findings add relevant information to the field of emergency abdominal CT in overweight individuals, as comparable data are missing. Our results suggest, that the large tube generator capacities of third-generation DSCT in combination with the high-pitch (Flash) mode and iterative reconstruction enable abdominal CT examinations of overweight patients, yielding both good image quality and a reasonable radiation dose.
It has been demonstrated, that third-generation DSCT provides a dose reduction of at least 30% in portal-venous abdominal CT using a pitch factor of 0.6 when compared to second-generation DSCT15. Furthermore, as the high-pitch dual-source mode (in combination with a reduced rotation time) of third-generation DSCT allows for a shorter CT acquisition time10, this technique probably further optimizes radiation dose efficacy in comparison to a standard-pitch protocol8,9,10,26,27. Likewise, in the present high-pitch study the ED of the group with a lower BMI (median BMI 31.5 kg/m2) was yet below the value measured by Wichmann and colleagues who applied similar tube settings but a standard-pitch protocol on the same third-generation DSCT scanner15, even though the BMI was lower in their study (this study: mean ED 5.2 mSv; Wichmann et al.: mean BMI 27.6 kg/m2, mean ED 6.2 mSv). However, as ATCM and ATVS were enabled in our study, the scanner output was automatically adjusted depending on the imaged object and size20. Accordingly, we found a positive correlation between BMI and radiation dose. Furthermore, the mean values of CTDIvol and DLP were significantly increased in the group with a higher BMI (median BMI 36.8 kg/m2) when compared to the group with a lower BMI, resulting in a mean ED of 8.9 mSv in the former group.
For DLP-based calculation of ED, we applied the combined conversion factor for abdominal and pelvic CT according to the ICRP 60 publication (k = 0.017 mSv/mGy * cm)18,19 for reasons of comparability with regard to relevant published studies14,15. Application of the more recent ICRP 103 publication might have yielded a further 7% reduction of ED in both groups according to Christner et al.19; however, this adjustment actually would not have altered the overall conclusion of the present study. Furthermore, the calculated ED in our obesity study (BMI ≥ 30 kg/m2) is possibly overestimated at least in the group with a higher BMI and AP + LAT diameter (> 73.6 cm), respectively, as their organs are more strongly shielded against X-radiation due to the distinct fatty tissue when compared to slimmer individuals28. For a more precise estimation of the ED conversion factor in patients with different BMIs we therefore recommend the application of specific obesity phantoms28. Nevertheless, mean values of other dose descriptors assessed in this study (CTDIvol and DLP), which are more objective than the ED, were still below the diagnostic reference levels provided by the respective federal office for radiation protection (CTDIvol 15 mGy, DLP 700 mGy * cm)29 even in the group with a higher BMI.
The SSDE, representing another dose descriptor which can be derived from the CTDIvol20, considers the individual patient size and thus reflects more precisely the mean absorbed patient dose. To note, the calculated mean of the sum of AP and LAT diameters defining the cut-off for the two groups (lower BMI and higher BMI) was 73.6 cm in the present study. According to Table 1 of AAPM Report 204, this dimension almost exactly corresponds to a SSDE conversion factor of 1 and hence to the CTDIvol obtained for a standard 32 cm diameter phantom22. We assessed a significant increase of SSDE in the group with a higher BMI and diameter, respectively – although less distinct when compared to the increase of CTDIvol. These findings contrast with Christner et al.20 who suggested that the SSDE is more or less independent from patient size. However, Li et al.21, who created whole-body computational phantoms from clinical images of normal- and overweight patients, showed that obesity has a significant effect on dose coefficients, that cannot be predicted using only the body diameter, and that SSDE overestimates organ dose for obese patients.
Regarding dose optimization – besides iterative reconstruction8,9,26 – reduction of tube voltage represents a well-established technique as it increases the intravascular contrast, which is of particular interest in CT angiography. With third-generation DSCT, for example, Meinel et al. demonstrated that lowering the tube voltage from 120 kV to 70–80 kV in coronary CT angiography reduces overall radiation dose by 49–68% while maintaining CNR even in obese patients7. In the present study, attenuation-based ATVS was enabled (in addition to ATCM) throughout the study period as mentioned above. ATVS optimizes – based on the attenuation along the z-axis according to the topogram – dose efficiency dependent on patient-specific mAs curves for all kV levels17. In third-generation DSCT, it has been shown that ATVS mostly selects lower kV values in body CT angiography of normal-weight individuals when compared to second-generation DSCT (90 versus 100 kV), consequently yielding further dose reduction17. In this obesity study, the median tube voltage selected by ATVS was 100 kV in both groups, which also represents a commonly applied value for non-weight selected portal-venous DSCT studies15. However, differences between groups in the present study reached statistical significance towards increased automatically selected values of both tube voltage (maximum 130 kV) and tube current time product in the group with a higher BMI.
Regarding objective image quality, we calculated CNRs and FOM CNRs of various organs and vessels. CNRs of the two parenchymal upper abdominal organs liver and spleen were significantly higher in the group with a lower BMI when compared to the group with a higher BMI. This was possibly caused by the relatively superficial anatomical location of these two organs resulting in a lower image noise in the former group. Furthermore, the dose-independent FOM CNR24,25 actually revealed a distinct difference of objective image quality between groups for each assessed organ, with significantly higher FOM CNRs for the group with a lower BMI. These FOM CNRs also exceeded the calculations by Wichmann et al. who applied a standard-pitch protocol in slimmer patients on the same third-generation DSCT scanner15. Nevertheless, this finding had limited clinical relevance in the present study, as the image quality was high in the majority of examinations even in the group with a higher BMI despite comparably lower FOM CNRs. We indeed noted increased image noise in patients with the highest BMIs (maximum BMI in this study: 54 kg/m2), but overall image quality remained diagnostic in these examinations. To note, in this most obese patient ATCM yielded a maximum combined tube current at the umbilical region of 2488 mA according to the DICOM header. Thus, we state that the tube current limit of both X-ray tubes (2 × 1300 mA) was not reached in this study. Altogether, the applied high-pitch DSCT protocol was robust to confirm (or rule out) therapeutically relevant pathologies of parenchymal organs and the gastrointestinal tract in both groups.
High-pitch DSCT protocols with pitch factors of at least 1.912 allow for a reduction of motion artifacts which is of particular interest in chest CT. In the present abdominal obesity study with a preset pitch of 1.55, no minor or major motion artifacts were documented. The applied pitch, which is still clearly higher than in standard-pitch protocols15, therefore seems to be sufficiently high for abdominal CT. In this context, with previous CT scanner models the pitch had to be decreased in order to accumulate radiation dose and increase image quality in CT examinations of obese patients, if the tube had reached its maximum power limit3. In this study, it was not necessary to further decrease the pitch despite the large patient size, as the tube generator of the utilized third-generation DSCT scanner has larger capacities7.
The results of the present study should be interpreted in the context of the study design and its limitations. First, CT data were only collected from one specific vendor (Siemens Healthineers), and we neither compared high- with standard-pitch CT acquisition protocols nor high- with standard-BMI patients. Thus, further studies are recommended in order to increase generalizability. Second, we did not apply a dedicated low-kV protocol due to concerns of image quality in this obesity study. However, we believe that a further reduction of tube voltage (e.g., 80 kV) may be reasonable at least in individuals with a BMI lower than 37 kg/m2 in order to achieve further dose optimization while maintaining diagnostic confidence. Third, extremely obese patients (weight > 300 kg) were not included, as their positioning is difficult with respect to standard gantry diameters and table weight limits. Future CT development by large manufacturers might cope with the epidemiologic impact of obesity on emergency units. Finally, vascular emergencies were not explicitly evaluated in this portal-venous CT study. However, CNR within large abdominal vessels was high, enabling the diagnosis of therapeutically relevant pathologies.
In conclusion, high-pitch emergency CT of the abdomen in portal-venous phase can be routinely performed in obese patients due to large tube generator capacities of third-generation DSCT. In the present study, radiation dose increased in individuals with a BMI > 36.8 kg/m2 as automated dose modulation (ATCM and ATVS) was enabled. This increase did not exceed diagnostic reference levels. Dose-independent FOM CNR was highest for obese patients with a BMI < 36.8 kg/m2, while image quality remained sufficient in patients with higher BMIs. The applied high-pitch CT protocol should be adapted according to dedicated dose and image quality optimization methods.
Data availability
The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request.
References
Arroyo-Johnson, C. & Mincey, K. D. Obesity Epidemiology Worldwide. Gastroenterol Clin North Am. 45, 571–579 (2016).
Siadecki, S. D., Frasure, S. E., Lewiss, R. E. & Saul, T. High Body Mass Index is Strongly Correlated with Decreased Image Quality in Focused Bedside Echocardiography. J Emerg Med. 50, 295–301 (2016).
Bamberg, F. et al. Challenges for computed tomography of overweight patients [Article in German]. Radiologe. 51, 366–371 (2011).
Ginde, A. A., Foianini, A., Renner, D. M., Valley, M. & Camargo, C. A. Jr. The challenge of CT and MRI imaging of obese individuals who present to the emergency department: a national survey. Obesity (Silver Spring). 16, 2549–2551 (2008).
Siemens Healthineers. SOMATOM Force. Siemens Healthineers, Erlangen, https://www.healthcare.siemens.de/computed-tomography/dual-source-ct/somatom-force (2018).
Petersilka, M., Bruder, H., Krauss, B., Stierstorfer, K. & Flohr, T. G. Technical principles of dual source CT. Eur J Radiol. 68, 362–368 (2008).
Meinel, F. G. et al. Image quality and radiation dose of low tube voltage 3rd generation dual-source coronary CT angiography in obese patients: a phantom study. Eur Radiol. 24, 1643–1650 (2014).
Lim, H. K., Ha, H. I., Hwang, H. J. & Lee, K. High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers. PLoS One. 14, e0211097 (2019).
Gariani, J., Martin, S. P., Botsikas, D., Becker, C. D. & Montet, X. Evaluating the effect of increased pitch, iterative reconstruction and dual source CT on dose reduction and image quality. Br J Radiol. 91, 20170443 (2018).
Amacker, N. A., Mader, C., Alkadhi, H., Leschka, S. & Frauenfelder, T. Routine chest and abdominal high-pitch CT: an alternative low dose protocol with preserved image quality. Eur J Radiol. 81, e392–397 (2012).
Hardie, A. D., Mayes, N. & Boulter, D. J. Use of high-pitch dual-source computed tomography of the abdomen and pelvis to markedly reduce scan time: clinical feasibility study. J Comput Assist Tomogr. 35, 353–355 (2011).
Sabel, B. O. et al. High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population. PLoS One. 11, e0146949 (2016).
Achenbach, S. et al. High-pitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc. Comput Tomogr. 3, 117–121 (2009).
De Cecco, C. N. et al. Second-generation dual-energy computed tomography of the abdomen: radiation dose comparison with 64- and 128-row single-energy acquisition. Comput Assist Tomogr. 37, 543–546 (2013).
Wichmann, J. L. et al. Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT. Eur Radiol. 27, 642–650 (2017).
Spearman, J. V. et al. Effect of Automated Attenuation-based Tube Voltage Selection on Radiation Dose at CT: An Observational Study on a Global Scale. Radiology. 279, 167–174 (2016).
Winklehner, A. et al. Automated attenuation-based tube voltage selection for body CTA: Performance evaluation of 192-slice dual-source CT. Eur Radiol. 25, 2346–2353 (2015).
Menzel, H., Schibilla, H., Teunen, D. eds. European Guidelines on Quality Criteria for Computed Tomography. Publication EUR 16262 EN. Luxembourg: European Commission. 735–738 (2000).
Christner, J. A., Kofler, J. M. & McCollough, C. H. Estimating effective dose for CT using dose-length product compared with using organ doses: consequences of adopting international commission on radiological protection publication 103 or dual-energy scanning. Am J Roentgenol. 194, 881–889 (2010).
Christner, J. A. et al. Size-specific dose estimates for adult patients at CT of the torso. Radiology. 265, 841–847 (2012).
Li, X. et al. Effects of protocol and obesity on dose conversion factors in adult body CT. Med Phys. 39, 6550–6571 (2012).
Boone, J., Strauss, K. & Cody, D. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Report of AAPM Task Group 204 (2011).
Schindera, S. T. et al. Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection–phantom study. Radiology. 246, 125–132 (2008).
Nakaura, T. et al. Abdominal dynamic CT in patients with renal dysfunction: contrast agent dose reduction with low tube voltage and high tube current-time product settings at 256-detector row CT. Radiology. 261, 467–476 (2011).
Guimaraes, L. S. et al. Appropriate patient selection at abdominal dual-energy CT using 80 kV: relationship between patient size, image noise, and image quality. Radiology. 257, 732–742 (2010).
Lu, G. M. et al. High-pitch computed tomography pulmonary angiography with iterative reconstruction at 80 kVp and 20 mL contrast agent volume. Eur Radiol. 24, 3260–3268 (2014).
De Zordo, T. et al. Comparison of image quality and radiation dose of different pulmonary CTA protocols on a 128-slice CT: high-pitch dual source CT, dual energy CT and conventional spiral CT. Eur Radiol. 22, 279–286 (2012).
Ding, A., Mille, M. M., Liu, T., Caracappa, P. F. & Xu, X. G. Extension of RPI-adult male and female computational phantoms to obese patients and a Monte Carlo study of the effect on CT imaging dose. Phys Med Biol. 57, 2441–59 (2012).
Federal Office for Radiation Protection. Publication of updated diagnostic reference levels for diagnostic and interventional X-ray examinations [Article in German]. Federal Office for Radiation Protection, Berlin, http://www.bfs.de/SharedDocs/Downloads/BfS/DE/fachinfo/ion/drw-roentgen.pdf?__blob=publicationFile&v=9 (2016).
Author information
Authors and Affiliations
Contributions
R.F. and C.G.T. conceived and planned the study. R.F., M.I. and R.S. were instrumental in data collection. R.F. and C.G.T. analyzed CT studies. R.F. performed the statistical analysis. R.F. drafted the manuscript and prepared the tables and figures. All authors contributed to the data interpretation, critically reviewed and edited the manuscript for important intellectual content and approved the final version of the manuscript to be published.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Forbrig, R., Ingrisch, M., Stahl, R. et al. Radiation dose and image quality of high-pitch emergency abdominal CT in obese patients using third-generation dual-source CT (DSCT). Sci Rep 9, 15877 (2019). https://doi.org/10.1038/s41598-019-52454-5
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-019-52454-5
- Springer Nature Limited