Abstract
Developing clinical tools that assess bone matrix quality could improve the assessment of a person’s fracture risk. To determine whether Raman spectroscopy (RS) has such potential, we acquired Raman spectra from human cortical bone using microscope- and fiber optic probe-based Raman systems and tested whether correlations between RS and fracture toughness properties were statistically significant. Calculated directly from intensities at wavenumbers identified by second derivative analysis, Amide I sub-peak ratio I1670/I1640, not I1670/I1690, was negatively correlated with Kinit (N = 58; R2 = 32.4%) and J-integral (R2 = 47.4%) when assessed by Raman micro-spectroscopy. Area ratios (A1670/A1690) determined from sub-band fitting did not correlate with fracture toughness. There were fewer correlations between RS and fracture toughness when spectra were acquired by probe RS. Nonetheless, the I1670/I1640 sub-peak ratio again negatively correlated with Kinit (N = 56; R2 = 25.6%) and J-integral (R2 = 39.0%). In best-fit general linear models, I1670/I1640, age, and volumetric bone mineral density explained 50.2% (microscope) and 49.4% (probe) of the variance in Kinit. I1670/I1640 and v1PO4/Amide I (microscope) or just I1670/I1640 (probe) were negative predictors of J-integral (adjusted-R2 = 54.9% or 37.9%, respectively). While Raman-derived matrix properties appear useful to the assessment of fracture resistance of bone, the acquisition strategy to resolve the Amide I band needs to be identified.
Similar content being viewed by others
Introduction
The increase in fracture risk with aging is disproportionate to the decrease in bone mass that occurs after ~50 years of age in both women and men. A comprehensive, robust prediction of fracture risk, based on multiple aspects of bone, could improve the identification of those requiring an intervention, thereby reducing the economic burden and poor quality of life that fractures impose. Numerous changes in bone at multiple length-scales contribute to age- and disease-related decreases in the overall resistance of bone to fracture1. For example, lower mechanical competence of bone at the apparent level (independent of macro-structure but not micro- or ultrastructure) have been associated with an increase in cortical porosity (micro-scale)2, an increase in glycation-mediated, non-enzymatic collagen crosslinks (molecular-scale)3, an increase in degree of mineralization (micro-scale)4, an increase in mineral crystal size (nano-scale)5, an increase in collagen denaturation (nano-scale)6, and a decrease in matrix-bound water (molecular-scale)7; whereas, bending strength of the femoral neck and compressive strength of lumbar vertebra at the whole-bone level have been associated with decreases in bone mass8, thickness of the cortices9, volumetric bone mineral density (BMD)10, and trabecular thickness11.
Dual-energy X-ray absorptiometry (DXA) measures a subject’s areal BMD and bone mineral content at multiple sites, whereas current high-resolution peripheral quantitative computed tomography (HR-pQCT) and micro-magnetic resonance imaging (μMRI) provide clinical measurements of cortical micro-structure and trabecular architecture at peripheral sites. Such measurements including volumetric BMD from HR-pQCT are important contributors to bone fracture risk12. None of these clinical measurements can accurately distinguish individuals with osteoporosis from those without osteoporosis13, which is one reason why guidelines for treating a patient for osteoporosis are based on risk factors, not solely on imaging measurements of bone14. One barrier to establishing a robust, patient-specific predictor of fracture risk is the lack of a diagnostic tool that is sensitive to the contribution of the bone matrix to the fracture resistance of bone1. To date, there are no FDA-approved techniques for in vivo assessment of bone matrix quality.
The current available methods to assess characteristics of the bone matrix composition ex vivo include: (i) vibrational spectroscopy techniques such as Fourier Transform Infrared (FTIR) and Raman spectroscopy (RS), (ii) microscopy techniques such as scanning acoustic microscopy (SAM), atomic force microscopy (AFM), and quantitative backscatter and transmission electron microscopy imaging (qBEI and TEM), and (iii) wide-angle and small-angle X-ray diffraction/scattering techniques (WAXS and SAXS). While all these techniques provide useful information about the bone matrix15, RS is the only method that is both sensitive to composition and organization of all three primary components of bone (mineral, organic matrix, and water)16,17,18 and has clinical feasibility in the near term19.
Of the emerging techniques to assess the quality of a patient’s bone matrix, the OsteoProbe measures the resistance of periosteal bone to impact micro-indentation (bone material strength index or BMSi)20 whereas ultra-short echo-time (UTE) MRI measures the concentrations of matrix-bound water (Cbw) and pore water (an indicator of porosity)21. The key determinants of BMSi and Cbw have yet to be identified. Therefore, RS is particularly well suited to help fulfill the unmet need of clinically assessing the quality of the bone matrix and to help identify mechanisms that increases fracture risk (beyond mineral density and bone structure).
RS measurements of the bone matrix calculated from either integrated area or peak intensity include: mineral-to-matrix ratio (ν1PO4/Amide I, ν1PO4/Proline and ν2PO4/Amide III), Type-B carbonate substitution (CO3/ν1PO4), crystallinity (the inverse of the line-width of the ν1PO4 peak at half the height from baseline or half-maximum; 1/FWHM)22, collagen crosslink or matrix maturity ratio (Amide I sub-band ratio at ~1670 cm−1 and ~1690 cm−1)23, and helical status of collagen (Amide I sub-peak ratio at ~1670 cm−1 and ~1640 cm−1)24. Using standard laboratory RS instruments (integrated into a compound microscope), RS measures were previously observed to significantly change with disease22,25 and drug therapy16 as well as to correlate with selected mechanical properties of bovine cortical bone24,26,27, rat cortical bone (aging study)22, and mouse cortical bone (genetic knock-out study)23. To the best of our knowledge, there are only two studies showing Raman measures can predict mechanical properties of human bone at the apparent-level5,28. In the study published by Yerramshetty and Akkus5, crystallinity explained 15.8% of the variance in elastic modulus and 7.1% of the variance in yield stress when RS analysis of tensile specimens were pooled from different quadrants of the femur mid-shaft. In the fracture toughness study involving specimens from one quadrant of the femur mid-shaft24, ν1PO4/Amide I ratio explained 10.9% the variance in J-integral and 11.6% of the variance in crack initiation toughness, whereas Type-B carbonate substitution explained 7.7% of the variance in crack growth toughness. Acquisition and data processing methods could potentially affect these correlations.
Using the emerging method of spatially offset RS (SORS)19,29,30,31,32,33 and Raman tomography34, several groups have acquired Raman spectra from bone through soft tissue in vivo, showing the feasibility of using RS to assess the bone matrix of patients19,35. However, a significant challenge to the translation of RS into the clinic is a meaningful analysis of the spectra that provides measurements differentiating strong, tough bone from fragile, brittle bone. There is currently no consensus on which RS measures are reliable predictors of the fracture resistance of human bone. Moreover, there are few guidelines on which methods for calculating RS measures improve sensitivity to differences in mechanical properties of bone. Lastly, to date, there is no evidence showing that Raman analysis predicts mechanical properties of human bone when acquiring the spectra with a fiber-optic Raman probe that essentially involves a different optical set-up compared to research-grade Raman micro-spectroscopy instrument (Fig. 1).
To advance RS as a laboratory tool and to motivate the development of clinical Raman probes for native bone, we hypothesized that (i) both fiber-optic RS and commercial Raman micro-spectroscopy can both predict fracture toughness of human cortical bone through the direct analysis of Amide I sub-peak ratios (i.e., without band fitting), (ii) fiber optic RS properties of the Amide I band are comparable with Raman micro-spectroscopy properties, and (iii) RS does not simply provide a surrogate of BMD and provides unique predictors of fracture toughness properties.
Results
Correlations between Raman properties and fracture toughness properties
Both microscope- and fiber optic-based RS produced well-resolved Raman spectrum of human cortical bone in terms of peak locations and shapes of the bands, but with some subtle differences in the spectra between the instruments (Fig. 2). When analyzing the spectra collected by Raman micro-spectroscopy, all RS properties were directly related to age (Table 1). The correlation between ν1PO4/Amide I and age was only significant when female donors were removed, and CO3/ν1PO4 on average was higher for male than female donors at a given age (Supp. Mater. Fig. 1). The mineral-to-matrix ratio (MMR) by ν1PO4/Proline and the matrix maturity ratio by direct calculation of the ~I1670/I1690 ratio did not correlate with the fracture toughness properties, while other traditional RS properties (e.g., ν1PO4/Amide I and CO3/ν1PO4) and another Amide I sub-peak ratio ~I1670/I1610 negatively correlated with all three fracture toughness properties, explaining 5.0% to 36.2% of the variance (Fig. 3A–D and Table 1). The Amide I sub-peak ratio ~I1670/I1640 alone explained 47.4% of the variance in J-int (Fig. 3B and Tables 1), 32.4% of the variance in Kinit (Fig. 3D and Tables 1), and 17.6% of the variance in Kgrow (Table 1). Interestingly, there were no significant correlations between sub-band area ratios and the fracture toughness properties (Table 1) when these ratios were determined using the Amide I sub-band fitting technique.
Further addressing the possibility that the method of calculating RS properties influences the ability to predict fracture toughness, we also determined RS properties from integrated area ratios with and without the secondary linear baselines. In general, peak area ratios correlated with peak intensity ratios (Supp. Mater. Fig. 2), and the correlations between peak intensity ratios and fracture toughness properties (Table 1) remained significant when the secondary linear baseline correction was not applied prior to calculation (Supp. Mater. Table 1). However, the correlation between area ratio for ν1PO4/Amide I and Kint (as well as J-int) was no longer significant without the secondary linear baseline. Several other area ratios, CO3/ν1PO4 and ν1PO4/CH2, did not correlate with J-int and Kinit, respectively, regardless of whether secondary linear baseline correction was used (Supp. Mater. Table 1).
When 9 of 16 randomly selected spectra from the polished longitudinal surface were analyzed, all R2 values were reduced as compared to when RS properties from 32 averaged spectra were correlated with fracture toughness properties (Supp. Mater. Table 2). These values did not appear to depend on whether the 9 or 32 spectra per donor were averaged before peak ratio calculations or whether 9 or 32 peak ratios from individual spectrum were averaged per donor (Supp. Mater. Table 2).
Similar to the results using the micro-spectroscopy system, Raman properties calculated from the spectra obtained by the fiber optic probe-based system had similar correlations between RS properties and age with CO3/ν1PO4 having the highest coefficient of determination (Table 2). There were however less significant correlations between probe-based RS properties and the fracture toughness properties (Table 2). The J-integral still correlated with ν1PO4/Amide I (Fig. 4A), but Kinit no longer correlated with ν1PO4/Amide I (Fig. 4C). The Amide I sub-peak ratios ~I1670/I1610 and ~I1670/I1640, not the matrix maturity ratio, still negatively correlated with all three fracture toughness properties (Fig. 4B,D, and Table 2). The Amide I sub-peak ratios correlated across the two RS systems (Fig. 5A–C).
Multivariate explanation of fracture toughness properties
To determine whether I1670/I1640 added value, it was included as covariate along with age and vBMD, two known determinants of fracture toughness, in general linear models (GLMs). Furthermore, other Raman properties with either the first or second highest R2 value (Table 1), were also included as covariates. The linear combination of vBMD and age together explained 40.7% of the variance in Kint, whereas the linear combination of I1670/I1640 and age together only explained 35.2% of the variance (Table 3). Interestingly, when combining this Amide I sub-peak ratio with vBMD and age, all 3 variables were significant predictors (Table 3) explaining 50.2% of the variance in Kinit. The relative contributions to this variance were similar among 3 variables with age having the lowest standardized β coefficient (Table 3). vBMD did not become a significant explanatory variable of Kgrow when it was included as a covariate of age; and although age correlated with Kgrow, it was no longer a significant explanatory variable when I1670/I1640 was included as covariate (Table 3). Combining age with crystallinity did not help explain the variance in Kgrow. With respect to the variance in J-int, age and vBMD together explained only 12.7% in which age had the higher β coefficient or stronger contribution (Table 3). When including I1670/I1640 as a covariate, age was no longer a significant predictor of J-int. Age and V1PO4/Amide I however were both significant explanatory variables in which the β coefficient of this mineral-to-matrix ratio was 2 times greater than that of age (Table 3). As such, the linear combination of ~I1670/I1640 and V1PO4/Amide I provided the best-fit model to predict J-int (Table 3).
Similar to the Amide I sub-peak ratio from Raman micro-spectroscopy, the fiber optic probe-based I1670/I1640 was a significant covariate of age and vBMD in which the 3 variables explained 49.4% of the variance in Kinit (Table 4). In this GLM, the contribution of age and I1670/I1640 to the variance was nearly equivalent (Table 3). The probe-based I1670/I1640 did not add value when included as covariate to age in predicting Kgrow (Table 4). Although the probe-based ν1PO4/Amide I had a weak linear correlation with J-int (Table 2), it significantly explained the variance in J-int (adjusted-R2 = 33.1%) when combined with age and vBMD, which were also significant covariates. Also, ν1PO4/Amide I had the strongest contribution to the variance in J-int (i.e., highest β coefficient) compared to age and vBMD (Table 4) Nonetheless, probe-based ~I1670/I1640 without age, vBMD, or ν1PO4/Amide I provided the best explanation of the variance in J-int (Table 4).
Discussion
Matrix quality is an essential contributor to the overall fracture resistance of bone1,36. Unlike strength, which assesses the internal resistance of a material to irreversible deformation, fracture toughness is a measure of the ability of a material to resist crack growth. Given that a fracture is the culmination of cracks growing to catastrophic size, fracture toughness is a useful material property of bone to predict with non-destructive surrogates. Herein, we found that RS measurements obtained by a commercially available Raman micro-spectroscopy instrument correlated with several fracture toughness properties of human cortical bone. More importantly, several RS properties remained significantly correlated with the fracture toughness properties when the RS data were acquired using a fiber optic probe-based RS instrument. Also supporting the potential clinical utility of RS, a property related to collagen quality (~I1670/I1640) significantly explained the variance in both J-int and Kinit. It further improved the explanation of the variance in both J-int and Kinit when known determinants of fracture toughness (age and vBMD) were included as covariates.
The resistance of bone to fracture involves numerous toughening mechanisms at different length scales across the hierarchical organization of the tissue37. These toughening mechanisms at the nanoscale include uncoiling of the triple helical collagen molecules and sliding of individual collagen molecules and mineralized collagen fibrils37,38,39,40. As with all materials, bone has pre-existing flaws from which cracks can initiate and grow (e.g., lacunae, fatigue-generated microcracks), and upon loading the ‘worst-case flaw’ (i.e., sharpened micro-notch in a single-edge notched beam or SENB specimen), the crack grows when the stress near notch tip reaches a critical value (Kinit). The resistance to further crack propagation, as reflected by Kgrow, depends on the ability of bone tissue to deflect the crack, thereby increasing the energy required to create fracture surfaces37. As such, additional toughening mechanisms at the submicron-to-micron level govern crack propagation including accumulation of diffuse damage in front of the crack tip41, mineralized collagen fibrils bridging cracks42, and crack deflection at cement lines43 or shifting lamellae with alternating collagen fibril direction44. We found that the ~I1670/I1640 ratio was better at explaining the variance in Kinit than the variance in Kgrow, suggesting that crack initiation is more highly dependent on collagen integrity than crack growth. This is not a surprising outcome given that Kgrow is also largely influenced by larger scale toughening mechanisms such as cement line density, cortical porosity, and tissue heterogeneity. Thus, finding a strong correlation between RS properties and Kgrow may be rather difficult because RS properties are mostly associated with nanoscale toughening mechanisms. As such, a multifactorial assessment approach is likely necessary to accurately predict a patient’s fracture risk.
In previous studies, ~I1670/I1640 ratio increased upon thermally induced collagen denaturation and mechanically induced diffuse damage in bovine cortical bone24 as well as being higher in human cortical bone subjected to fatigue by rotating beam tests45. Unal et al.24 found that the ~I1670/I1640 ratio was negatively correlated with toughness and post-yield toughness of the bovine cortical bone obtained by three-point bending tests. ~I1670/I1640 ratio was thus proposed as a spectroscopic biomarker of the helical structure of collagen I, specifically indicative of a transition from a triple helical structure to less-ordered structure with perturbations in the molecular arrangement of α1 and α2 chains of collagen I24. Importantly, this Amide I sub-peak ratio in present study was also negatively correlated with fracture toughness properties, while the more widely reported matrix maturity ratio (~I1670/I1690) was not correlated with fracture toughness even when determined as a sub-band area ratio (~A1670/A1690). Raman sub-peaks of the Amide I band of bone are also sensitive to the disruption of collagen enzymatic crosslinking46, tissue aging47, degree of mineralization48, ionizing radiation49,50, and hydration45. More recently, the ~I1670/I1640 ratio was found to be higher in diabetic mouse bone with lower toughness compared to non-diabetic mice bone with higher toughness51, implying it could be sensitive to advanced glycated end-products (AGEs), although fluorescence AGEs and pentosidine, an AGE crosslink, were not different between the groups. The ~I1670/I1640 ratio was also recently reported to be sensitive to in vitro glycation of human cortical bone as well52. Cumulatively, all these findings indicate that several of the sub-peaks of Amide I are measures of the perturbation on the helical structure of collagen molecules, and perhaps reflect the capacity of collagen fibrils to dissipate energy, namely ~I1670/I1640 (Fig. 2B).
In our previous Raman micro-spectroscopy study using nearly the same donor set (N = 62, instead of N = 58) and acquisition of Raman spectra prior to fracture toughness testing28 in the region of crack propagation, we found either very weak positive correlations (ν1PO4/Amide I and CO3/ν1PO4) or no correlations between RS properties and fracture toughness properties; whereas in the present study, there were significant negative correlations for most of RS properties (Table 1). There are several differences in data collection process between the studies that potentially explain the discrepancy. In the previous study, we only collected 9 spectra at a 1 μm spot size within 0.25 mm2 area over the intended crack propagation region and calculated peak ratios for each Raman spectrum before averaging the RS properties per donor28; whereas in this study, we collected 32 spectra randomly distributed throughout the entire two longitudinal surfaces of neighboring bone specimens (~69 mm2 ± ~23 mm2) and determined the RS properties from averaged spectrum per donor (i.e., averaging the spectra minimizes noise). The number of spectra analyzed rather than averaging spectra before property calculation affected the coefficient of determination (Supp. Mater. Table 2). This is perhaps not surprising since bone composition is inherently heterogeneous due to osteonal remodeling. Acquiring spectra from many sites helps capture the overall bulk composition of each human sample. Furthermore, with Raman micro-spectroscopy, inherent laser polarization bias (i.e., sensitivity to collagen fibril orientation) is more prominent with higher numerical aperture (NA) of the objective53. In the previous study, spectra were acquired with 50 × (NA = 0.75) objective compared to 20 × (NA = 0.40) objective in the current study. While the orientation of the bone samples relative to the polarization axis of the laser was similar between studies (different research-grade Raman instruments though), differences in the sensitivity to fibril orientation may have caused differences in the ν1PO4/Amide I measurements between the studies. The negative correlations, albeit weak, between fracture toughness and MMR (Table 1) were unexpected, but perhaps a relatively low organic matrix (higher ν1PO4/Amide I) confers more brittle-like behavior.
Custom fiber optic probe-based RS have been developed for in vivo diagnostic applications, namely the detection of cancer54. Spatially offset Raman spectroscopy (SORS) is a variant of fiber optic RS that allows data collection from a target layer through a turbid sample, and first developed for collecting Raman spectra of bone through the skin29. SORS analysis of bone is an emerging technique in osteoporosis research19 and in detection of other bone diseases (i.e., osteogenesis imperfecta or OI)35. Thus far, two pilot studies showed the feasibility of in vivo, non-invasive clinical RS measurements to detect chemical compositional differences between osteoporotic and healthy subject with principal component analysis (PCA)19 and between OI and healthy subject with the analysis of v1PO4/Amide III ratio35. Neither of the studies analyzed sub-peak ratios of the Amide I band. As such, to date, a clinically viable tool that assesses collagen integrity has not been established, even though type I collagen has long been thought to be a primary determinant of toughness55 and fracture toughness56.
In the present study, a fiber optic probe-based RS that does not preserve the primary polarization of the laser (Fig. 1) was in direct contact with prepared bone samples and provided measurements that partially explain the variance in fracture toughness properties (Table 2), even when collecting five Raman spectra from one longitudinal surface (Kinit and J-int vs. ~I1670/I1640 provided in Supp. Mater. Fig. 3). It remains to be seen whether further optimization of Raman probe can improve the ability of RS to predict the fracture toughness of cortical bone (e.g., collection of more Raman spectra per bone, different acquisition parameters and/or additional fiber optics to improve the signal-to-noise, use of polarization preserving fiber optics to capture orientation of the mineralized collagen fibrils, inclusion of a filter to select for water peaks that exist at higher wavenumbers, and using different spectrograph to improve spectral resolution).
This present work had several limitations. Because bone segments nearer to the site of crack propagation were used in other studies, we collected the RS data from distant segments of variable size (~25–35 mm distal from the micro-notch in the axial direction), instead of collecting Raman spectra prior to mechanical testing. These remaining segments from the original bone strip were subjected to more freeze-thaw cycles than the SENB specimens, and this may affect the strength of the correlations. Also, since the spatial and spectral resolution as well as the acquisition time including the number of acquired spectra was different between the two 2 RS instruments (research-grade RS had ~21% higher signal-to-noise ratio than did probe-based RS), there is a possibility that the number of significant correlations (Tables 1 and 2) could be equivalent between the two configurations (Fig. 1) once probe-based RS is optimized. The fiber optic probe also directly contacted the longitudinal surface of samples extracted the femur mid-shaft, and so, we do not know yet: (1) whether reported correlations will persist in spectra collected through skin and periosteum at a clinically accessible site such as the tibial mid-shaft and (2) whether bone matrix quality recorded at the mid-shaft will be predictive of fracture risk at relevant sites such as the femoral neck. Thus, additional work is necessary to determine whether a SORS technique or a minimally invasive technique (under local anesthesia) can adequately resolve the Amide I band using radiant exposures that do not damage tissues before the clinical assessment of collagen quality in vivo can be realized.
As discussed in our previous publication28, reported correlation strengths between fracture toughness properties and various bone properties are typically weak-to-moderate (r = ~0.5 or R2 = ~25%). Again, this is likely due to the numerous toughening mechanisms that bone possesses and part of the rationale behind explaining fracture toughness with several independent explanatory variables (Tables 3 and 4). While RS-derived properties, namely ~I1670/I1640 and ν1PO4/Amide I, improved upon age and vBMD in predicting fracture toughness (i.e., its inclusion either increased the adjusted coefficient of determination and/or superseded these other known determinants), nearly half the variance in Kinit and J-int was not explained. This is perhaps not surprising for two reasons: (i) R-curve testing is inherently stochastic because the location of the micro-notch (which may or may not be near a pore or a cement line) and the features that propagating crack encounters (causing random deflections) cannot be controlled and (ii) other independent factors such as bound water7, porosity and osteonal area57, collagen network connectivity58, and small-scale heterogeneity in the matrix59 likely influence fracture toughness and are not necessarily related to RS-derived mineral-to-matrix ratio or marker of collagen I helical order.
In conclusion, Raman spectroscopy-derived matrix properties, namely the direct calculations of ~I1670/I1640 and v1PO4/Amide I, significantly correlated with the fracture toughness of human cortical bone, though correlations were not particularly strong being highest for the overall energy dissipated (R2 = 47% and 36%, respectively). Moreover, the significant correlations persisted when the spectra were acquired with fiber optic probe-based RS, and ~I1670/I1640 helped age and vBMD explain crack initiation toughness of human cortical bone. Development of probe-based RS instrument using either percutaneous or transcutaneous strategy could advance the clinical assessment of bone matrix quality.
Materials and Methods
Bone specimen preparation and study design
The Institutional Review Board at Vanderbilt University deemed that the use of cadaveric bone did not qualify as human subject research and so informed consent was not necessary. The sources of the femurs (Musculoskeletal Tissue Foundation, National Disease Research Interchange, and Vanderbilt Donor Program) did not provide personally identifiable information (i.e., tissue was de-identified). All methods for processing and analyzing the bone followed relevant guidelines and regulations. We described the preparation of the mechanical specimens and measurement methods in our previous study7, and so they are briefly summarized herein. The proximal end of ~70 mm strips of cortical bone from cadaveric femur mid-shafts (lateral quadrant) were machined into single-edge notched beam (SENB) specimens (length x thickness x width of ~20–30 mm × ~2.5 mm × ~5 mm). The crack path region was scanned by micro-computed tomography (µCT) at an isotropic voxel size of 5 µm, and the corresponding volumetric BMD (vBMD) was determined. The SENB specimens were subjected to three-point bending using a progressive, cyclic loading protocol (loaded to + 0.07 mm at 0.01 mm/s and then-unload to −0.04 mm at 0.015 mm/s) with a short dwell period prior to the next load cycle7. The span was ~20 mm (4 × width), and the crack propagate perpendicular to the osteonal direction7. A non-linear fracture mechanics approach (R-curve testing) was used to determine transverse fracture toughness properties, namely crack initiation (Kinit), crack growth toughness (Kgrow) and J-integral (J-int) or the overall energy dissipated to propagate the crack to failure7. Kgrow could not be calculated for specimens that fractured in a brittle manner. Additional specimens distal to the SENB were also cut (Fig. 1) from the remaining strip of cortical bone. For the present study, the most distal segment was available for 58 (28 males and 30 females spanning 21 to 101 years of age) of the original 62 donors (mean ± SD): length of 18.6 ± 6.2, width of 5.1 ± 0.7, and 2.6 ± 0.3). Two distal samples from male donors (21 years and 91 years of age) were not available for the fiber optic probe-based RS measurements. Samples were soaked in phosphate buffered saline and stored at −20 °C when not being analyzed and thawed to room temperature prior to Raman analysis.
Raman spectroscopy
In this study, we collected Raman spectra from bone specimens using two RS systems: (i) confocal Horiba RS (Xplora, Horiba Jobin Yvon, Edison, NJ) with a 785 nm diode laser and with a 1200 lines/mm grating providing ~1.25 cm−1 spectral resolution and (ii) portable fiber optic probe-based RS. The probe-based RS involved: (i) an imaging spectrograph (Holospec f/1.8i, Kaiser Optical Systems, Ann Arbor, MI) coupled to a thermoelectrically cooled CCD camera (PIXIS: 256BR, Princeton. Instruments, Princeton, NJ), providing ~3.50 cm−1 spectral resolution, (ii) a 785 nm diode laser (Innovative Photonic Solutions, Monmouth Junction, NJ), and (iii) a custom-made fiber optic probe (EmVision, Loxahatchee, FL) consisting of one excitation and six collection fibers (each 300 µm in diameter) configured as a ring shape (Fig. 1). Wavelength calibration of the portable probe-based RS system was done using a neon-argon lamp. Naphthalene and acetaminophen standards were also used to determine the exact excitation wavelength for subsequent Raman shift calculations. The spectral response of the system was further corrected using a tungsten lamp calibrated by the National Institute of Standards and Technology.
For Raman micro-spectroscopy, the long axis of each specimen was aligned parallel to the axis of the primary laser polarization, and thirty-two Raman spectra per specimen were each obtained as the average of 12 consecutive spectra per spot with a 5-second acquisition using a 20x objective (NA = 0.40). Laser power was ~35 mW. For fiber-optic RS with a larger laser spot size than a 20x objective (~300 µm vs. ~2.5 µm), ten spectra per sample were each obtained as the average of 10 consecutive spectra per spot with 3-second acquisition, and laser power was set up at ~80 mW. The long axis was not specifically aligned with the polarization axis of the laser because fiber optics scramble the orientation of the light (Supp. Mater. Fig. 4). Raman data collection were randomly distributed throughout the entire two longitudinal surfaces of bone specimens (sixteen Raman spectra and five Raman spectra per surface for research-grade RS and fiber optic RS, respectively). Since the bone specimens were not immersed in PBS during the acquisition of the multiple spectra, some dehydration occurred. To verify that this does not affect the spectra, we collected spectra from 6 bone specimens before and after 20 min in air which is the maximum time for total spectra collection in this study. We found that there were no apparent differences in the RS measures between these two time points (Supp. Mater. Fig. 5) indicating partial air-drying for 20 minutes did not affect significantly the RS properties.
Raman data analysis
Raman spectra were processed using LabSpec 5 software (Horiba Jobin Yvon, Edison, NJ) and a custom MATLAB script. First, Raman raw spectra collected at thirty-two or ten sites per bone specimen were averaged. Then, background fluorescence was removed from all averaged spectra by subtracting a 5th-order polynomial function from the base of the raw spectra (Supp. Mater. Fig. 6A). Then, the averaged spectra without normalizing to a mean or selected peak intensity were further smoothed to minimize noise using a proprietary de-noising (D-n) algorithm provided by the LabSpec 5 software. Before calculating RS properties, a secondary 5-point linear baseline was applied to minimize residual fluorescence (Supp. Mater. Fig. 6B,C).
From the averaged and de-noised spectrum per donor, we calculated the following RS properties from peak intensities (Supp. Mater. Fig. 6B,C): mineral-to-matrix ratio (ν1PO4/Amide I, ν1PO4/Proline, ν1PO4/Amide III and ν1PO4/CH2-wag), Type-B carbonate substitution (CO3/ν1PO4), crystallinity (the inverse of the line-width of the ν1PO4 peak at half the height from baseline or half-maximum; 1/FWHM), collagen crosslinks/matrix maturity ratio (calculated as the intensity at ~1670 cm−1 per intensity at ~1690 cm−1 or I1670/I1690), and the newly developed I1670/I1610 and I1670/I1640 ratio associated with collagen conformational change24. The locations of Amide I shoulders were not fixed at set wavenumbers but rather identified directly from the local minima of each second derivative (Fig. 2B). We also fitted four sub-peaks within the Amide I band using Gaussian-Lorentzian mixed functions. Briefly, each sub-band with a Gauss/Lorentzian mixture of 50%/50% was first centered on the 4 wavenumber locations that were identified by the second derivative analysis. Then, a non-linear constrained optimization algorithm in MATLAB determined the final wavenumber location of each peak within a ± 5 cm−1 window as well as the final mixture for each sub-band that minimized the weighted root mean square error between the experimental data and the sum of the 4 sub-peaks (Supp. Mater. Fig. 7)52. Finally, the Amide I sub-band ratios (A1670/A1610, A1670/A1640, A1670/A1690) were the respective sub-band area ratios.
We further processed our raw data from research-grade, commercial Raman micro-spectroscopy using three additional methods: (1) only 5th order polynomial fluorescence subtraction (i.e., without a 5-point linear baseline correction) and then calculating ratios based on the intensity of the peaks, (2) same 5th order polynomial fluorescence subtraction and then calculating area ratios in which the integration range of each band was modified from those used in Hammond et al.60 to match our Raman spectral acquisitions (Supp. Mater. Fig. 6D), and (3) calculated the integrated area ratios (same range used for each peak) from our initial spectra (5th order polynomial fluorescence subtraction with a 5-point linear baseline correction, Supp. Mater. Fig. 6B,C). Noise filtering did not vary among the four approaches. To determine whether the number of spectra analyzed affected correlations, 9 randomly chosen spectra out of the 16 acquired from the polished longitudinal surface were averaged prior to RS property calculation. Lastly, RS properties were determined from individual spectrum per donor and then averaged (9 or 32 spectra).
Statistical analysis
All statistical analyses were performed using STATA 12 statistical software (StataCorp LP, College Station, TX, USA) in which bootstrapping (1000 replicates) was used to generate the p-values since the normality assumption in regression analysis did not hold for most properties. Linear correlations between the Raman properties or vBMD and fracture toughness properties or age were determined at a significance level of 0.05. As described in our previous studies57, analysis of covariance was used to determine whether sex was a significant covariate of age in the explanation of each bone property (Supp. Mater. Table 3 and Supp. Mater. Fig. 1). Next, age, vBMD, and Raman properties were considered as independent predictors in general linear models with the fracture toughness properties as the dependent variables. The interaction terms and then the independent variables were removed in a step-wise backwards manner to determine which combination of properties best explained the variance in fracture toughness properties (i.e., highest adjusted R2).
Ethical approval
Upon review of the project, the IRB at the corresponding author’s institution deemed the use of cadaveric tissue as non-human subjects research. All donors were de-identified by the allograft banks or donor services.
Data Availability
All Raman spectroscopy data, mechanical data, and micro-computed tomography data that were used in the correlation analyses are available from the corresponding author upon request.
References
Unal, M., Creecy, A. & Nyman, J. S. The Role of Matrix Composition in the Mechanical Behavior of Bone. Current Osteoporosis Reports 16, 205–215, https://doi.org/10.1007/s11914-018-0433-0 (2018).
McCalden, R. W., McGeough, J. A. & Barker, M. B. Age-related changes in the tensile properties of cortical bone. The relative importance of changes in porosity, mineralization, and microstructure. JBJS 75, 1193–1205 (1993).
Wang, X., Shen, X., Li, X. & Agrawal, C. M. Age-related changes in the collagen network and toughness of bone. Bone 31, 1–7 (2002).
Currey, J. D., Brear, K. & Zioupos, P. The effects of ageing and changes in mineral content in degrading the toughness of human femora. Journal of biomechanics 29, 257–260 (1996).
Yerramshetty, J. S. & Akkus, O. The associations between mineral crystallinity and the mechanical properties of human cortical bone. Bone 42, 476–482 (2008).
Wang, X. et al. Effect of collagen denaturation on the toughness of bone. Clinical Orthopaedics and Related Research® 371, 228–239 (2000).
Granke, M., Makowski, A. J., Uppuganti, S., Does, M. D. & Nyman, J. S. Identifying novel clinical surrogates to assess human bone fracture toughness. Journal of Bone and Mineral Research 30, 1290–1300 (2015).
Dalén, N., Hellström, L.-G. & Jacobson, B. Bone mineral content and mechanical strength of the femoral neck. Acta Orthopaedica Scandinavica 47, 503–508 (1976).
Mayhew, P. M. et al. Relation between age, femoral neck cortical stability, and hip fracture risk. The Lancet 366, 129–135 (2005).
Singer, K., Edmondston, S., Day, R., Breidahl, P. & Price, R. Prediction of thoracic and lumbar vertebral body compressive strength: correlations with bone mineral density and vertebral region. Bone 17, 167–174 (1995).
Vesterby, A. et al. Biologically meaningful determinants of the in vitro strength of lumbar vertebrae. Bone 12, 219–224 (1991).
Manhard, M. K., Nyman, J. S. & Does, M. D. Advances in imaging approaches to fracture risk evaluation. Translational Research 181, 1–14 (2017).
Johannesdottir, F., Allaire, B. & Bouxsein, M. L. Fracture Prediction by Computed Tomography and Finite Element Analysis: Current and Future Perspectives. Current osteoporosis reports, 1–12 (2018).
Kanis, J. A., McCloskey, E., Johansson, H., Oden, A. & Leslie, W. D. FRAX® with and without bone mineral density. Calcified Tissue Int 90, 1–13 (2012).
Donnelly, E. Methods for assessing bone quality: a review. Clinical Orthopaedics and Related Research® 469, 2128–2138 (2011).
Mandair, G. S. & Morris, M. D. Contributions of Raman spectroscopy to the understanding of bone strength. BoneKEy Reports 4 (2015).
Unal, M., Yang, S. & Akkus, O. Molecular spectroscopic identification of the water compartments in bone. Bone 67, 228–236, https://doi.org/10.1016/j.bone.2014.07.021 (2014).
Unal, M. & Akkus, O. Raman spectral classification of mineral- and collagen-bound water’s associations to elastic and post-yield mechanical properties of cortical bone. Bone 81, 315–326, https://doi.org/10.1016/j.bone.2015.07.024 (2015).
Buckley, K. et al. Towards the in vivo prediction of fragility fractures with Raman spectroscopy. Journal of Raman Spectroscopy 46, 610–618 (2015).
Diez-Perez, A. et al. Recommendations for a standard procedure to assess cortical bone at the tissue-level in vivo using impact microindentation. Bone reports 5, 181–185 (2016).
Manhard, M. K. et al. MRI-derived bound and pore water concentrations as predictors of fracture resistance. Bone 87, 1–10 (2016).
Paschalis, E., Gamsjaeger, S. & Klaushofer, K. Vibrational spectroscopic techniques to assess bone quality. Osteoporosis Int, 1–17 (2017).
Paschalis, E. et al. Spectroscopic Characterization of Collagen Cross‐Links in Bone. Journal of Bone and Mineral Research 16, 1821–1828 (2001).
Unal, M., Jung, H. & Akkus, O. Novel Raman Spectroscopic Biomarkers Indicate That Postyield Damage Denatures Bone’s Collagen. Journal of Bone and Mineral Research 31, 1015–1025, https://doi.org/10.1002/jbmr.2768 (2016).
Morris, M. D. & Mandair, G. S. Raman assessment of bone quality. Clinical Orthopaedics and Related Research® 469, 2160–2169 (2011).
Akkus, O., Adar, F. & Schaffler, M. B. Age-related changes in physicochemical properties of mineral crystals are related to impaired mechanical function of cortical bone. Bone 34, 443–453, https://doi.org/10.1016/j.bone.2003.11.003 (2004).
Bi, X. et al. Raman and mechanical properties correlate at whole bone-and tissue-levels in a genetic mouse model. Journal of biomechanics 44, 297–303 (2011).
Makowski, A. J. et al. Applying Full Spectrum Analysis to a Raman Spectroscopic Assessment of Fracture Toughness of Human Cortical Bone. Applied spectroscopy 71, 2385–2394, https://doi.org/10.1177/0003702817718149 (2017).
Matousek, P. et al. Noninvasive Raman spectroscopy of human tissue in vivo. Applied spectroscopy 60, 758–763 (2006).
Draper, E. R. et al. Novel Assessment of Bone Using Time‐Resolved Transcutaneous Raman Spectroscopy. Journal of Bone and Mineral Research 20, 1968–1972 (2005).
Schulmerich, M. V. et al. Transcutaneous Raman spectroscopy of murine bone in vivo. Applied spectroscopy 63, 286–295 (2009).
Okagbare, P. I., Morris, M. D., Begun, D., Goldstein, S. A. & Tecklenburg, M. Noninvasive Raman spectroscopy of rat tibiae: approach to in vivo assessment of bone quality. Journal of biomedical optics 17, 090502 (2012).
Maher, J. R., Inzana, J. A., Awad, H. A. & Berger, A. J. Overconstrained library-based fitting method reveals age-and disease-related differences in transcutaneous Raman spectra of murine bones. Journal of biomedical optics 18, 077001 (2013).
Demers, J.-L. H., Esmonde-White, F. W., Esmonde-White, K. A., Morris, M. D. & Pogue, B. W. Next-generation Raman tomography instrument for non-invasive in vivo bone imaging. Biomedical optics express 6, 793–806 (2015).
Buckley, K. et al. Measurement of abnormal bone composition in vivo using noninvasive Raman spectroscopy. IBMS BoneKEy 11, https://doi.org/10.1038/bonekey.2014.97 (2014).
Nyman, J. S., Granke, M., Singleton, R. C. & Pharr, G. M. Tissue-level mechanical properties of bone contributing to fracture risk. Current osteoporosis reports 14, 138–150 (2016).
Launey, M. E., Buehler, M. J. & Ritchie, R. O. On the Mechanistic Origins of Toughness in Bone. Annual Review of Materials Research 40, 25–53, https://doi.org/10.1146/annurev-matsci-070909-104427 (2010).
Fritsch, A., Hellmich, C. & Dormieux, L. Ductile sliding between mineral crystals followed by rupture of collagen crosslinks: experimentally supported micromechanical explanation of bone strength. Journal of theoretical biology 260, 230–252 (2009).
Gupta, H. et al. Intrafibrillar plasticity through mineral/collagen sliding is the dominant mechanism for the extreme toughness of antler bone. Journal of the mechanical behavior of biomedical materials 28, 366–382 (2013).
Zimmermann, E. A. et al. Age-related changes in the plasticity and toughness of human cortical bone at multiple length scales. Proceedings of the National Academy of Sciences 108, 14416–14421 (2011).
Diab, T. & Vashishth, D. Effects of damage morphology on cortical bone fragility. Bone 37, 96–102 (2005).
Nalla, R. K., Kruzic, J. J., Kinney, J. H. & Ritchie, R. O. Mechanistic aspects of fracture and R-curve behavior in human cortical bone. Biomaterials 26, 217–231 (2005).
Koester, K. J., Ager Iii, J. & Ritchie, R. The true toughness of human cortical bone measured with realistically short cracks. Nature materials 7, 672 (2008).
Peterlik, H., Roschger, P., Klaushofer, K. & Fratzl, P. From brittle to ductile fracture of bone. Nature materials 5, 52 (2006).
Flanagan, C. D., Unal, M., Akkus, O. & Rimnac, C. M. Raman spectral markers of collagen denaturation and hydration in human cortical bone tissue are affected by radiation sterilization and high cycle fatigue damage. J Mech Behav Biomed Mater 75, 314–321, https://doi.org/10.1016/j.jmbbm.2017.07.016 (2017).
McNerny, E., Gong, B., Morris, M. D. & Kohn, D. H. Bone Fracture Toughness and Strength Correlate with Collagen Cross‐Link Maturity in a Dose‐Controlled Lathyrism Mouse Model. Journal of Bone and Mineral Research 30, 455–464 (2015).
Ager, J. W., Nalla, R. K., Breeden, K. L. & Ritchie, R. O. Deep-ultraviolet Raman spectroscopy study of the effect of aging on human cortical bone. Journal of biomedical optics 10, 034012–0340128 (2005).
Buckley, K., Matousek, P., Parker, A. W. & Goodship, A. E. Raman spectroscopy reveals differences in collagen secondary structure which relate to the levels of mineralisation in bones that have evolved for different functions. Journal of Raman Spectroscopy 43, 1237–1243 (2012).
Gong, B., Oest, M. E., Mann, K. A., Damron, T. A. & Morris, M. D. Raman spectroscopy demonstrates prolonged alteration of bone chemical composition following extremity localized irradiation. Bone 57, 252–258 (2013).
Barth, H. D. et al. Characterization of the effects of x-ray irradiation on the hierarchical structure and mechanical properties of human cortical bone. Biomaterials 32, 8892–8904 (2011).
Creecy, A. et al. Low bone toughness in the TallyHO model of juvenile type 2 diabetes does not worsen with age. Bone 110, 204–214, https://doi.org/10.1016/j.bone.2018.02.005 (2018).
Unal, M. et al. Assessing Glycation-mediated Changes in Human Cortical Bone with Raman Spectroscopy. Journal of Biophotonics Accepted Author Manuscript, https://doi.org/10.1002/jbio.201700352 (2018).
Makowski, A. J., Patil, C. A., Mahadevan-Jansen, A. & Nyman, J. S. Polarization control of Raman spectroscopy optimizes the assessment of bone tissue. Journal of biomedical optics 18, 055005 (2013).
Pence, I. & Mahadevan-Jansen, A. Clinical instrumentation and applications of Raman spectroscopy. Chemical Society Reviews 45, 1958–1979 (2016).
Burr, D. The contribution of the organic matrix to bone’s material properties. Bone 31, 8–11 (2002).
Zioupos, P. Ageing human bone: factors affecting its biomechanical properties and the role of collagen. Journal of Biomaterials Applications 15, 187–229 (2001).
Granke, M., Makowski, A. J., Uppuganti, S. & Nyman, J. S. Prevalent role of porosity and osteonal area over mineralization heterogeneity in the fracture toughness of human cortical bone. Journal of biomechanics 49, 2748–2755 (2016).
Willett, T. L., Dapaah, D. Y., Uppuganti, S., Granke, M. & Nyman, J. S. Bone collagen network integrity and transverse fracture toughness of human cortical bone. Bone (2018).
Katsamenis, O. L., Jenkins, T. & Thurner, P. J. Toughness and damage susceptibility in human cortical bone is proportional to mechanical inhomogeneity at the osteonal-level. Bone 76, 158–168 (2015).
Hammond, M. A., Gallant, M. A., Burr, D. B. & Wallace, J. M. Nanoscale changes in collagen are reflected in physical and mechanical properties of bone at the microscale in diabetic rats. Bone 60, 26–32 (2014).
Acknowledgements
Funds from the National Institute of Arthritis and Musculoskeletal and Skin Diseases (AR063157 and AR073133) and the Department of Veterans Affairs, Veterans Health Administration, Office of Research and Development (1I01BX001018) supported the present work. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health or other funding agencies.
Author information
Authors and Affiliations
Contributions
M.U., O.A., J.S.N. conceived the study. M.U. and S.T. acquired the Raman data, while S.U. was responsible for the fracture toughness tests and micro-computed tomography evaluations. M.U. and J.S.N. wrote the paper with input from O.A. A.M.J. oversaw the development of fiber optic probe-based Raman system. All authors have read and approved this paper.
Corresponding author
Ethics declarations
Competing Interests
The authors do not have any direct financial conflict of interest pertaining to the work reported herein. However, J.S.N. and A.M.-J. recently received a grant to pursue translating Raman spectroscopy toward the clinical assessment of bone quality. J.S.N. also serves on the scientific advisory board, without compensation, of ActiveLife Scientific, Inc., the maker of a device that directly assesses bone quality. His lab also received funds from Safoni Genzyme for work unrelated to the present content. The authors have no other non-financial competing interests that directly pertain to the reported findings.
Additional information
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Unal, M., Uppuganti, S., Timur, S. et al. Assessing matrix quality by Raman spectroscopy helps predict fracture toughness of human cortical bone. Sci Rep 9, 7195 (2019). https://doi.org/10.1038/s41598-019-43542-7
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-019-43542-7
- Springer Nature Limited
This article is cited by
-
Spatiotemporal Characterization of Microstructure Morphology, Mechanical Properties and Bone Remodeling of Rat Tibia Under Uniaxial Compressive Overload Loading
Annals of Biomedical Engineering (2024)
-
Fracture Toughness: Bridging the Gap Between Hip Fracture and Fracture Risk Assessment
Current Osteoporosis Reports (2023)
-
Resveratrol prevents ovariectomy-induced bone quality deterioration by improving the microarchitectural and biophysicochemical properties of bone
Journal of Bone and Mineral Metabolism (2023)
-
Age related changes of rib cortical bone matrix and the application to forensic age-at-death estimation
Scientific Reports (2021)
-
Clinical Importance of Bone Matrix Damage Mechanisms for Fracture Prevention
Current Osteoporosis Reports (2021)