Background & Summary

The human lung is among the largest solid organs in the human body. Traditionally, studies of lung microanatomy at the organ scale require lengthy operations in targeted sampling, tissue preparation, histological staining, and sectioning1,2. Nowadays, ex vivo clinical evaluations of whole lung microstructures are carried out without sectioning using absorption-contrast micro-CT at around 100 μm voxel size. A limited area of the lung may then be selected for imaging at higher resolution using histology3,4,5. X-ray phase-contrast imaging6,7 provides higher sensitivity and contrast than laboratory micro-CT8. Compared with optical virtual histology9, X-ray phase contrast from free-space propagation requires no imaging optics and, at the same time, removes the need for laborious tissue clearing and fluorescent labeling that are essential for optical imaging10. The compatibility of X-ray phase-contrast imaging with existing X-ray sources will facilitate its gradual adoption and transition from preclinical research to clinical diagnostics6,11,12. At synchrotron facilities, systematic upgrades13,14 in the X-ray source and imaging techniques over the past decades provide the means to tackle biological questions across meaningful scales and resolution11,15,16,17,18,19,20. Although synchrotron-based X-ray imaging can access finer anatomical detail than laboratory micro-CT19,21,22,23, many bioimaging scenarios require further upscaling of the imaging throughput and accommodation of large sample size while maintaining microscopic resolution24,25.

Thanks to the high X-ray photon flux and spatial coherence achieved at modern fourth-generation synchrotron sources and careful design of the sample preparation and imaging protocol, it is now possible to image complete, large, partially dehydrated human organs in their entirety at micrometer resolution using hierarchical phase-contrast tomography (HiP-CT)26. The technique integrates a multiscale imaging workflow4,27,28,29,30 into a single setup, utilizing propagation phase contrast obtained from high-energy, polychromatic X-rays and tunable detection settings. Therefore, scanning an entire human organ (with a size of 5–30 cm in each dimension) at multiple resolutions can be executed without dissecting the sample or requiring to transport to different instrument locations or facilities27,28,30. HiP-CT features a customized flat-field correction, an attenuation scanning protocol, along with an efficient tomographic sampling and stitching pipeline to cover large, soft-tissue organs entirely, without staining27,31 or clearing10. The integrated single-modality, multiscale imaging approach of HiP-CT26 ensures a simplified image registration procedure thanks to the consistent tissue contrast across lengthscales. Its imaging protocol takes inspiration from existing multiscale approaches4,18,19,29, starting with a two-step tomographic sampling of the whole organ (full-field tomography), followed by progressive zoom-ins to selected features of the microanatomy through local tomographies at various finer resolutions compatible with the relevant anatomical context. HiP-CT requires the sample such as a soft-tissue organ to be embedded in 70% ethanol solution in water and immobilized with agar blocks throughout imaging (see Fig. 1a,b). The flat-field correction takes reference from a separate container (reference jar) of the same size as the sample jar to enhance the soft tissue contrast (see Fig. 1c). We provide here the dataset of an intact human left lung imaged by HiP-CT at 25.08 μm voxel size (full organ, see Fig. 1d) and at 6.05–6.5 μm and 2.45–2.5 μm voxel sizes for various local volumes of interest (VOIs) accomplished by optimization of the incident X-ray spectrum, propagation distance, scintillator thickness, and the coupling optics before the detector (see Methods). The X-ray imaging experiments were carried out at the European Synchrotron Radiation Facility (ESRF) BM05 beamline using the recently upgraded fourth-generation extremely brilliant X-ray source (ESRF-EBS)32,33.

Fig. 1
figure 1

(a) A human left lung in lateral (left) and medial (right) view. (b) The instillation-fixed whole lung is mounted in a sealed, size-compatible, plastic cylindrical PET jar (140 mm in diameter, 260 mm in height) filled with 70% ethanol solution and agar blocks. (c) Sketch of the HiP-CT imaging setup using propagation phase contrast from bright and coherent X-rays at ESRF BM05 beamline. The reference jar contains the same embedding medium as the sample jar. The incident X-ray energy is adjusted to 70–85 keV through filters depending on the resolution requirement. (d) Volume rendering of the whole left lung imaged at 25.08 μm voxel size using HiP-CT in lateral (left) and medial (right) views. Major anatomical features labeled in a,b,d are the (1) left upper lobe, (2) interlobar fissure, (3) left lower lobe, (4) left mainstem bronchus.

Methods

Lung preparation and mounting

The entire left lung (see Fig. 1a) was harvested from a body donor, a 94-year-old woman who succumbed to natural causes, with the medical record provided in Table 1. Body donation was based on free consent by the donor antemortem. The relevant postmortem medical procedures were carried out at Laboratoire d’Anatomie des Alpes Françaises (LADAF) according to the Quality Appraisal for Cadaveric Studies scale recommendations34. All dissections respected the memory of the deceased. The protocols for transport and imaging were approved by the French legislation for body donation. The body of the deceased donor was embalmed and the lung preparations were carried out at ~36 hours postmortem. The lung was instilled through the trachea with a 4% formalin solution using 30 cm of water column positive pressure. The trachea was then ligatured to maintain the inflated configuration in order to fix the lungs in a non-collapsed state. The body was then kept at 4 °C for 3 days before the dissection. Once removed, the lungs were immersed in 4% formalin solution for 3 more days. Afterwards, it was successively immersed in ethanol solutions (odorless bioethanol, Cheminol, France) with increasing concentration up to 70% (volume fraction). The lung was kept inflated during ethanol dehydration by repeatedly pushing the solution through its main bronchus with a syringe. The significantly lower density of ethanol (789 kg/m3) compared with water (1000 kg/m3) provides a high base contrast for soft tissues35,36.

Table 1 Medical record of the body donor.

We used a PET (polyethylene terephthalate) jar of comparable size to the lung for X-ray imaging due to its commercial availability (3600 mL Sweep Jars with Cap, Medline Scientific, UK), high radiation tolerance37 and optical transparency in assisting sample alignment and assessment of sample condition during imaging. To secure the lung tightly in place and prevent it from touching the container edges on all sides, we prepared agar (Agar-Agar powder, from wild red seaweed, Nat-Ali, France) blocks in ~1 cm3-sized cubes and stacked them at the bottom of the jar and around the organ to firmly embed the lung. The procedure for the agar preparation has been thoroughly described preivously26. The gaps between the small agar blocks provide the escape routes for residual gas removal. The sample mounting procedure involves alternated filling of the agar-ethanol mixture and gentle vacuum degassing to minimize the existing microbubbles from dissolved air in the solution environment and within the organ, thereby eliminating their interference with imaging. The degassing procedure used a membrane pump to directly pump38 above the PET sample jar with the lid open in a sealed vacuum glass dryer. Prior to imaging, the PET jar containing the lung, ethanol solution, and agar embedding was placed in a custom-made sample holder to connect to the rotation stage at the synchrotron beamline26.

Synchrotron X-ray imaging and reconstruction

The implementation and capabilities of HiP-CT have been described in detail in a separate publication26. Here, we describe the settings used for lung imaging. All X-ray imaging experiments were carried out at the ESRF bending magnet beamline BM0539. The polychromatic synchrotron beam produced at the beamline was passed through a set of filters and then directly used for imaging without additional X-ray optics. The voxel size is effectively controlled by the adjustable visible-light imaging optics situated after the LuAG:Ce (cerium-doped lutetium aluminium garnet) X-ray scintillator (custom-made by Crytur, Czechia) and before the sCMOS (scientific complementary metal-oxide-semiconductor) light sensor (PCO edge 4.2 CLHS, PCO Imaging, Germany). Specifically, the imaging optics include the dzoom (“demagnifying zoom”) and zoom lenses, which cover the ranges of 6.5–25.5 μm and 1.3–6.3 μm, respectively. Because the synchrotron beam size (with usable area 50 mm × 4 mm at BM05) is considerably smaller than the size of the human left lung (container size 260 mm height, up to 140 mm width at the widest), imaging the entire left lung at 25.08 μm voxel size required stitching together multiple subscans. We used the half-acquisition (or half-object acquisition)40 method developed at ESRF for imaging the VOIs at 6.5 μm and 2.5 μm in voxel size. For the entire lung, we developed a quarter-acquisition method26 that includes the half-acquisition in combination with an annular scan to cover its complete horizontal extent (see Fig. 2).

Fig. 2
figure 2

Synchrotron-based hierarchical phase-contrast tomography (HiP-CT) at multiple lengthscales and their associated data acquisition and image reconstruction pipelines. The full-field tomography data at 25.08 μm voxel size are processed with pipeline (a). The local tomography data for volumes of interest (VOIs) at 6.05–6.5 μm (level 1) and 2.45–2.5 μm (level 2) voxel size are processed with pipeline (b). (c) A schematic illustrating the relationships of the various cylindrical volumes imaged with HiP-CT. The triple arrows in the pipeline before merging the subscans indicate that the same procedure is carried out on each subscan.

Data processing of the measured X-ray projections consists of three stages, pre-reconstruction, reconstruction and post-reconstruction, which are illustrated in separate rows in Fig. 2. Ring artifacts from the detectors are corrected in two steps: (1) Before reconstruction, the mean of the projections is subtracted from the projections to remove the rings with constant intensity rings; (2) After reconstruction, the residual inhomogeneous intensity rings were removed using the polar transform combined with linear motion blurring filter41. Tomographic reconstruction employs the phase and amplitude estimates obtained from Paganin’s method42, followed by a 2D unsharp mask of the retrieved phase maps as input for the filtered backprojection algorithm. These reconstruction steps are implemented in PyHST243. Eventually, the processed volumes are converted to a 16-bit format and binned further to produce the datasets described in Tables 23. The reconstruction and postprocessing steps are illustrated for the three types of imaged sample volumes, respectively, in Fig. 2. We summarize below the imaging and reconstruction protocols for the human lung at each imaged resolution including the key parameters.

  • Full-field tomography (the whole organ at 25.08 μm voxel size, see Fig. 2a,c): The detected X-ray energy averaged at ~93 keV after filters and sample, the propagation distance was 3475 mm, the scintillator thickness was 2 mm. In total, two sets of 9990 projections were measured by the quarter-acquisition method26 with an offset of 800 pixels for the half-acquisition. A step size of 2.2 mm in the vertical (z) direction was used to cover the height of the sample jar with a total of 98 quarter-acquisition subscans. Radiographic stitching was first carried out to recover a half-acquisition scan40 before the reconstruction.

  • Local tomography of level 1 VOI (6.5 μm and 6.05 μm voxel size, see Fig. 2b,c): The detected X-ray energy averaged at ~88 keV (~89 keV) after filters and sample, the propagation distance was 3500 mm (3475 mm), the scintillator thickness was 1 mm (2 mm) for the VOIs with 6.5 μm (6.05 μm) voxel size. In total, 6000 projections were measured by the half-acquisition method with an offset of 900 pixels. A step size of 2.2 mm in the vertical direction was used to cover the height of the VOIs.

  • Local tomography of level 2 VOI (2.5 μm and 2.45 μm voxel size, see Fig. 2b,c): The detected X-ray energy averaged at ~77 keV (~79 keV) after filters and sample, the propagation distance was 1440 mm (1500 mm), the scintillator thickness was 0.25 mm (0.2 mm) for the VOIs with 2.5 μm (2.45 μm) voxel size. In total, 6000 projections were measured by the half-acquisition method with an offset of 900 pixels. A step size of 1.5 mm in the vertical direction was used to cover the height of the VOIs.

Table 2 Volumes of interest and their anatomical references to the whole human left lung sample.
Table 3 Details of the hierarchical X-ray phase-contrast tomography data for a human left lung.

Volume selection and anatomical reference

Besides the full-field tomography of the entire lung, subsequent smaller VOIs were selected with representative features and imaged with local tomography at higher resolution, including 6.5 μm (5 locations) and 6.05 μm (3 locations) for level 1 and 2.5 μm (5 locations) and 2.45 μm (3 locations) for level 2 VOIs, respectively. All VOIs have a cylindrical field of view around the rotation axis after removing the boundary artifacts from the local tomographic reconstruction. To obtain the displacements and rotations, the VOIs are spatially registered to the whole lung data by hand in VGStudio Max (version 3.4) and the procedure to apply them is described in Usage Notes. The sizes of the VOIs, their displacements, and rotations with respect to the center of the whole lung data are listed in Table 2 and illustrated in Fig. 3a–d. In addition, we provide brief anatomical references to the VOI spatial locations in Table 2 with respect to the whole lung data at 25.08 μm. To retain traceable data provenance, we keep the same alphanumeric labels of the VOIs as used in the original experiments. Figure 3e visualizes two selected VOIs in the lower lobe of the lung.

Fig. 3
figure 3

Exploration of the HiP-CT dataset of a human left lung. (ad) Spatial correspondences of the measured cylindrical VOIs at different resolutions within the entire left lung. For each set of VOIs, both the medial (left) and sagittal (right) views are shown. The VOI label corresponds to the assignment in Table 2. (e) Renderings of two imaged VOIs with 6.5 μm and 2.5 μm voxel sizes, respectively. (f) From the whole lung and local tomography data, we visualize the anatomical detail of a partially calcified, spiculated lung nodule in the apical region of the lung on multiple lengthscales. The interlobular septa and perilobular vasculature of the secondary pulmonary lobules (SPLs) are depicted.

Data Records

We provide the volumetric data after reconstruction and post-processing as greyscale (16-bit) 2D image slices in the JPEG2000 format stored in zipped folders. The compression level of JPEG2000 is carefully chosen to ensure minimal difference from the original TIFF-formatted data when they are used for feature quantification or image segmentation. We list the details of the deposited data in Table 3. All data have been deposited at an ESRF data repository (https://human-organ-atlas.esrf.eu/explore/LADAF-2020-27/left%20lung) with digital object identifiers (DOIs) assigned to each scanned volume as listed in Table 4. Each DOI refers to a volume at full resolution and its binned versions. For all VOIs measured by local tomography, including those with voxel sizes of 6.5 μm44,45,46,47,48, 6.05 μm49,50,51, 2.5 μm52,53,54,55,56, and 2.45 μm57,58,59, both the full resolution data (Binning = 1) and the 2 × binned version (Binning = 2) are provided, while for the whole lung data60, the 4 × binned version (Binning = 4) is also provided. The metadata information in Table 2 is also provided in the corresponding text file contained in each data deposit. The landing page associated with each DOI contains extended information on the parameters of the X-ray beamline for phase-contrast tomography, experimental scanning protocol, and data processing procedures.

Table 4 Information about the data records.

Technical Validation

Although the radiation dose in HiP-CT scans is well below the tissue damage threshold26, due to radiation-induced bubble formation, which only appeared after multiple high-resolution, local tomography scans, the sample went through re-degassing before the remaining measurements were made. The bubbles largely come from solvent vaporization61 but don’t cause visible radiation damage as shown by histology26. However, a consequence of re-degassing is that not all of the VOIs have been imaged consecutively during the same beamtime. In the course of re-degassing, the sample was kept in the container to maintain its original position. The sample jar was then placed into the synchrotron X-ray beamline for further imaging. During the process, care was exercised such that the VOIs scanned before and after re-degassing can be registered to the whole volume without large deformation.

In the imaged volumes, contrast is produced by the local density differences between the lung tissue constituents and the void spaces of the airways, alveoli, and blood vessels filled with ethanol solution (see Fig. 3e,f). Within the whole lung data at a voxel size of 25.08 μm, the interlobular septa, the boundaries of the secondary pulmonary lobules62,63 and the perilobular vasculature, are clearly visible (see Fig. 3f). At high spatial resolution, the local density difference increasingly becomes the dominant contributor to image contrast in VOIs26. The consistent contrast across lengthscales provides detailed information for the study of lung morphology for the healthy individual or as a control.

Regarding the total lung capacity for the individual, postmortem estimation is severely hampered by (1) the inability to directly measure anatomical dead space and (2) complex functional interactions between skeletal thorax, diaphragm configuration or tonus and pleural space, which are not reconstructible postmortem. Nevertheless, since the lung was inflated with near-normal pressure at fixation, our best assumption for near-normal inspiration inflation is pressure-controlled inflation. In addition, despite the great efforts in maintaining the integrity of the organ during preparation and scanning, in some subpleural areas of the lung, we have noticed a slight compression. We attribute the potential causes of these features to the following: (1) The formalin used in fixation may have difficulties in reaching the more peripheral lung areas. (2) The tight fit of the agar blocks around the lung. (3) A sign of minimal parabronchial and/or subpleural emphysema due to aging5.

Usage Notes

The multiscale healthy human lung data presented here have been used as clinical control data in studies comparing damage within the lung microstructure due to Covid-19 infection26. The individual VOIs are deposited as 2D image slices perpendicular to the rotation axis (z in Fig. 1) in the tomography geometry. These images may be directly loaded into any typical image processing software for viewing or further quantification. To align the VOIs to the whole lung data, the following transform should be applied to the VOI,

$$I{\prime} (x,y,z)=T(dx,dy,dz){R}_{z}({\theta }_{z})I(x,y,z).$$
(1)

Here I' and I are intensity-valued volumetric data, T is the 3D translation operator, and Rz the 3D rotation operator around the z axis (see Fig. 1b,c). The displacement vector (dx, dy, dz) and z rotation angle θz for each VOI is listed in Table 2. The default greyscale ranges of the images are set with an intensity margin to avoid saturation. Viewing directly by eye may require threshold adjustment.