Abstract
Consistent mechanical performance from implantation through healing and scaffold degradation is highly desired for tissue-regenerative scaffolds, e.g. when used for vascular grafts. The aim of this study was the paired in vivo mechanical assessment of biostable and fast degrading electrospun polyester-urethane scaffolds to isolate the effects of material degradation and tissue formation after implantation. Biostable and degradable polyester-urethane scaffolds with substantial fibre alignment were manufactured by electrospinning. Scaffold samples were implanted paired in subcutaneous position in rats for 7, 14 and 28 days. Morphology, mechanical properties and tissue ingrowth of the scaffolds were assessed before implantation and after retrieval. Tissue ingrowth after 28 days was 83 ± 10% in the biostable scaffold and 77 ± 4% in the degradable scaffold. For the biostable scaffold, the elastic modulus at 12% strain increased significantly between 7 and 14 days and decreased significantly thereafter in fibre but not in cross-fibre direction. The degradable scaffold exhibited a significant increase in the elastic modulus at 12% strain from 7 to 14 days after which it did not decrease but remained at the same magnitude, both in fibre and in cross-fibre direction. Considering that the degradable scaffold loses its material strength predominantly during the first 14 days of hydrolytic degradation (as observed in our previous in vitro study), the consistency of the elastic modulus of the degradable scaffold after 14 days is an indication that the regenerated tissue construct retains its mechanical properties.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Regenerative medicine has emerged as one of the most dynamic drivers in the development of advanced engineered biomaterial solutions for tissue engineering applications [1, 2]. One crucial element in regenerative medicine are scaffolds that facilitate and guide the regeneration of biological tissues according to the biophysical requirements of the specific application.
The ideal conduit for arterial replacement or bypass remains autologous grafts, i.e. the patient’s own artery or vein. Autologous grafts are, however, often unavailable due to either disease or previous use for bypass grafting [3,4,5,6,7]. Whereas currently available vascular prostheses made from polyethylene terephthalate (PET) such as Dacron®, and expanded polytetrafluoroethylene (ePTFE) perform well as large-calibre replacements, their long-term patency is discouraging in small to medium graft (< 7 mm) applications [3, 4, 8]. This limitation is mainly a result of a poor healing process with surface thrombogenicity due to the lack of endothelial cells, as well as anastomotic intimal hyperplasia [5, 9]. Although attempts have been made to improve these grafts by means of various coatings, the results are not conclusive [5, 10,11,12].
Tissue engineering and regeneration has provided a promising avenue for the development of vascular grafts. The ideal scaffold facilitates tissue regeneration such that the new tissue constructs mimics biologically and mechanically the healthy vascular tissue [2]. For small-to-medium diameter tissue engineered or tissue regenerative vascular grafts, a key factor for long term success is porosity [8]. The complete healing of the porous vascular graft implies interconnected pores to permit transmural tissue ingrowth [8, 13, 14].
Several approaches have been used to obtain porous polymeric scaffolds for vascular grafts, including particulate leaching [14,15,16], thermally induced phase separation [17, 18]. Another method of introducing porosity in vascular scaffolds is electrospinning [19,20,21,22,23,24]. Here, the adjustment of process parameters allows controlling the degree of fibre alignment in the resulting polymeric network [25, 26]. Fibre alignment strongly determines the mechanical properties of the scaffold [9]. Random fibre alignment leads to mechanical isotropy of the scaffold, which exhibits similar properties in different directions. In contrast, the predominant alignment of fibres in one direction introduces mechanical anisotropy [26].
Electro-spun scaffolds with a high degree of fibre alignment exhibit the highest elastic modulus, i.e. stiffness, in direction of the predominant fibre alignment and the lowest elastic modulus perpendicular to the fibre alignment. Mechanical anisotropy can offer a useful tool to tailor the structural properties of the scaffold but also adds complexity to the design process. The latter needs to be addressed adequately, in particular for biodegradable scaffolds and considering that tissue regenerating in the scaffold will contribute to structural and mechanical properties of the regenerating tissue construct.
Scaffold degradation and tissue incorporation are transient processes, and gradual change of the mechanical properties of the scaffold-tissue construct is expected until degradation and tissue ingrowth are complete. Biomechanical properties of scaffolds were predominantly obtained prior to tissue culturing or implantation [19, 22, 27,28,29,30] whereas the changes in mechanical properties due to in situ tissue ingrowth after implantation in fast-degrading scaffolds have received very little attention.
In previous studies, we separately quantified the effect of in vitro degradation [31] and in situ tissue ingrowth in a biostable scaffold [32] on the scaffolds’ mechanical properties. The present study aimed at investigating the combined effects of concurrent in situ tissue ingrowth and scaffold degradation on the mechanical properties of a fast-degrading electrospun polyester-urethane scaffold. Matched implants of the polyester-urethane scaffold with a fast-degrading and a non-degrading formulation, respectively, served as internal controls.
2 Materials and methods
2.1 Material
DegraPol® (ab medica S.p.A, Cerro Maggiore, MI, Italy), a biodegradable polyester-urethane that consists of poly(ε-caprolactone-co-glycolide)-diol soft segments and poly(3-(R-hydroxybutyrate)-co-(ε-caprolactone))-diol hard segments was used. Both polymer segments are biodegradable and their degradation products are non-toxic [33]. The mechanical properties of the polymer can be modulated by adjusting the ratio of the hard and soft segments. Changing the ratio of ε-caprolactone to glycolide affects the degradation characteristics. DegraPol® DP0 and DP30 have ε-caprolactone-to-glycolide ratio of 100:0 and 70:30 respectively, and both have a hard-to-soft segment ratio of 40:60 (unpublished data).
2.2 Electro-spinning and sample preparation
DegraPol® DP0 and DP30 were each dissolved in chloroform at room temperature with subsequent sonication in distilled water at 37 °C for 90 min to obtain polymer solutions of 24% by weight concentration. The DegraPol® solution was electro-spun from a hypodermic needle (SE400B syringe pump, Fresenius, Bad Homburg, Germany) onto a rotating and bi-directionally translating tubular target (hypodermic tubing, Small Parts, Loganport, IN, USA) using a custom-made rig. The spinning parameters were: solution flow rate of 4.8 mL/h, target outer diameter of 25 mm, target rotating speed of 9600 rpm, target translational speed of 2.6 mm/min, potential difference of 15 kV, and source-target distance of 250 mm. After completion of the spinning process, the electro-spun structure on the target hypotube was submersed in ethanol for 5 min, removed from the mandrel and dried under vacuum (Townson & Mercer Ltd, Stretford, England; room temperature, 90 min) and trimmed on either end to discard regions with decreasing wall thickness. Tubular scaffolds were cut into rectangular samples with the longer edge aligned with either the circumferential (C) or axial (A) direction of the tube. Samples underwent standard ethylene oxide sterilisation (55 °C, 60% relative humidity, 12 h) and were subsequently placed in vacuum for 6 h to allow removal of ethylene oxide.
2.3 Physical characterisation of scaffold before implantation
Physical characterisation included microscopic analysis of fibre diameter and alignment, measurement of scaffold wall thickness and porosity as described previously [32].
2.3.1 Fibre diameter and alignment
Scanning electron microscope (SEM) images were obtained (Nova NanoSEM 230, FEI, Hillsboro, OR, US) of samples after sputter-coating with gold (Polaron SC7640, Quorum Technologies, East Grinstead, UK). The fibre diameter was measured on ×750 SEM micrographs with Scion Image (Scion Corporation, Frederick, USA) (n = 3 samples; 4 images per sample, i.e. one image from two different locations of each sample surface; ten measurements per image). Fibre alignment was quantified on ×100 SEM images using Fiji [34], by means of Fourier component analysis with the Directionality plug-in (written by JeanYves Tinevez) as described by Woolley et al. [35]. The analysis provides a dispersion factor as a measure of alignment of the fibres with smaller dispersion values indicating higher degree of alignment.
2.3.2 Sample dimensions
Width, length and thickness of scaffold samples were measured on macroscopic images obtained with a Leica DFC280 stereo microscope using Leica IM500 imaging software (Leica Microsystems GmbH, Wetzlar, Germany).
2.3.3 Scaffold porosity
Scaffold porosity, P, was determined by hydrostatic weighing typically employed for density determination. The porosity of fibrous networks is described as volumetric ratio
and when expressing volumes as masses in the form of
Introducing measurable quantities mwet, mdry and msubmerged provides
with mdry as the mass of the dry scaffold, msubmerged as the mass of the scaffold submerged in ethanol, and mwet as the scaffold mass after removal from the ethanol while ethanol is retained in the pores. The porosity was measured with n = 1 per spun scaffold.
2.4 Molecular weight
Weight-average molar weight, MW, during degradation was determined by gel permeation chromatography (GPC) using tuneable absorbance detector (Waters 486, set to 260 nm), high-performance liquid chromatography pump (Waters 510), differential refractometer (Waters 410) and auto sampler (Spectra Series AS100, Thermo Separations). Five columns (Styragels HR1 through HR6, Waters) and a pre-column filter were used at 30 °C. Polystyrene standards were used for calibration. Samples were dissolved in tetrahydrofuran (1 mg/ml, 37 °C, sonication for 20 min). A sample volume of 180 µl was injected with a flow rate of 1.06 ml/min.
2.5 Study design
This study used a subcutaneous rat model with implant durations of 7, 14 and 28 days and n = 10 animals per each time point. Each animal received six implants, namely thee biostable DP0 scaffold samples and three biodegradable DP30 scaffold samples.
2.6 Implantation procedure
All animal experiments were authorised by the Faculty of Health Sciences Research Ethics Committee of the University of Cape Town and were performed in accordance with the National Institutes of Health (NIH, Bethesda, MD) guidelines.
Male Wistar albino rats with body mass 200–250 g were used. Anaesthesia was induced by placing the animal in an inhalation chamber with an air flow of 5% isoflurane for 5 min. The animals were shaved and sterilized with iodine in the area of the incision. Anaesthesia was maintained by nose cone delivery of 1.5% isoflurane at an oxygen output of 1.5 L/min at 1 bar and 21 °C. The body temperature of the animals was maintained throughout the surgical procedure by placing the animal on a custom-made heating pad at 37 °C.
Six longitudinal incisions of 1 cm (three on either side of the dorsal midline) were made and subcutaneous pockets of 2 cm depth bluntly dissected. Scaffold samples were placed within the pockets and incisions closed with silk 4/0 (Ethicon Inc, Somerville, NJ) interrupted sutures in a subcuticular fashion.
2.7 Implant retrieval
Anaesthesia was induced by placing the animal in an inhalation chamber with an air flow of 5% isoflurane for 5 min. Rats were euthanized by inhalation of 5% halothane in air followed by a cardiac injection of 1 ml saturated KCl solution. The samples with surrounding tissue were excised. Samples for mechanical testing were submerged in phosphate buffered saline (PBS) solution and stored at 4 °C. Samples for histological analysis underwent tissue fixation in 10% formalin (Sigma Aldrich, Steinheim, Germany) for 48 h and transferred into 70% ethanol solution for storage.
2.8 Mechanical testing
Prior to testing, the thickness, length and width of retrieved samples were determined with a calliper. Tensile testing was performed within 12 h of implant retrieval with samples submerged in PBS at 37 °C (Instron 5544, 10 N load cell; Instron, Norwood, USA). Samples were fixed using custom-made clamps resulting in a gauge length of approximately 10 mm. The direction of tensile load represented either circumferential (C samples) or axial direction (A samples) of the original tubular scaffold.
Pre-implant samples (T = 0 days) underwent a loading protocol comprising either (1) one initial extension to 12% strain, five cycles between 12% and 8% strain, and a final extension to failure or maximum force of 7.4 N (all at strain rate of 9.6%/min); or (2) one single extension to 16% strain at 9.6%/min. Retrieved implant samples were not subjected to loading cycling but underwent one single extension only (to 16% strain at 9.6%/min) to avoid inducing structural damage in the degraded samples at the latter stages of the study. Stress–strain data obtained for pre-implant scaffold samples with and without load-cycling was used to derive a scaling function SC(ε) = σCycled(ε)/σNoncycled(ε) that quantifies the mechanical effect of the load-cycling. Stress–strain data obtained without load cycling on retrieved scaffold samples were adjusted using SC for each test and average data calculated. Characteristic mechanical measures were derived from these data, namely stress σ12% at 12% strain and σ16% at 16% strain, and elastic modulus E6% at 6% strain and E12% at 12% strain. (Note: Stress and strain always refers to nominal stress and Cauchy strain, respectively.)
2.9 Histology, light microscopy and assessment of tissue ingrowth
Fixed samples underwent tissue processing and paraffin embedding. Sections of 3 μm thickness were prepared from mid sample regions and stained with haematoxylin and eosin (H&E) for nuclei and cytoplasm. Microscopic images were acquired with a Nikon Eclipse 90i microscope with digital camera DXM-1200C (Nikon Corporation, Tokyo, Japan).
Tissue ingrowth was quantified by morphometric image analysis classifying areas as open space or tissue (VIS Visiopharm Integrator System, Visiopharm, Hørsholm, Denmark). A region of interest (ROI) in the image was segmented in tissue (nuclei, cytoplasm and extracellular matrix) and open space with an untrained k-means clustering technique. The algorithm classified scaffold fibres as open space since the polymer did not stain. This was corrected by adjusting for scaffold porosity.
2.10 Statistical analysis
For quantitative data, one-way ANOVA was performed when more than two groups were compared, using Tukey HSD post hoc analysis with p < 0.05 indicating statistical significance. Data are provided as mean values ± standard deviation.
3 Results
SEM micrographs of the electrospun DegraPol® DP0 and DP30 scaffolds are shown in Fig. 1. Fibre diameter (13.0 ± 2.2 µm vs 12.7 ± 4.1 µm) and scaffold porosity (74.0 ± 3.2% vs 77.0 ± 2.7%) agreed reasonably well between DP0 and DP30. The fibre dispersion was lower (i.e. fibre alignment was higher) in DP0 scaffolds (9.6° ± 2.0°) compared to DP30 scaffolds (17.8 ° ± 4.6°, p < 0.05). Scaffolds of both materials exhibited some degree of fibre merging (Fig. 1 bottom row). Morphological scaffold data and dimensions of the scaffold samples for implantation are summarised in Table 1.
Weight-averaged molar weight of DegraPol® DP30 decreased significantly from 77.6 ± 5.7 kDa at T = 0 days to 26.1 ± 6.7 kDa at T = 28 days, see Fig. 2. Approximately half of this decrease occurred within the first 7 days (51%), whereas the decrease continued more gradually (T = 7 to 14 days: 19%, T = 14 to 21 days: 16%, T = 21 to 28 days: 14%). Biostable DegraPol® DP0 was analysed at T = 0 and 28 days only (n = 1) and did not exhibit a change in MW.
Tissue ingrowth after 28 days was slightly higher, although not statistically significant, in DegraPol® DP0 scaffolds (83 ± 10%) compared to DegraPol® DP30 scaffolds (77 ± 4%). The progress of tissue ingrowth throughout the study is illustrated in representative micrographs of H&E sections of the scaffolds and quantified in Fig. 3.
Figure 4a shows stress–strain data for axial and circumferential directions of pre-implant scaffold samples with and without load cycling. The resulting axial and circumferential stress scaling functions SC(ε) are illustrated in Fig. 4b. These functions quantify the impact of the load cycling for strains up to 16%. Disregarding data for strain below 2% (due to low signal-to-noise ratio), it was observed for axial DegraPol® DP0 scaffold samples that load cycling reduced the stress, σCycled, to 0.43% of the stress of non-cycled scaffolds, σNoncycled, for strain, ε, between 2% and 12%. For ε ≥ 12%, load cycling decreased the stress to σCycled = 0.33 σNoncycled, indicating further weakening of the scaffold. For circumferential DegraPol® DP0 samples, σCycled gradually increased from 0.20 σNoncycled at ε = 2% to 0.59 σNoncycled at ε = 12% strain, and decreased to 047 σNoncycled at ε = 16%. For DegraPol® DP30 scaffold, load cycling resulted in stress between σCycled = 0.80 σNoncycled at ε = 2% and σCycled = 0.98 σNoncycled at ε = 12% and 16% in axial scaffold direction, compared to σCycled = 0.26 σNoncycled at ε = 2%, 0.93 σNoncycled at ε = 12%, and 0.97 σNoncycled at ε = 16% in circumferential scaffold direction.
Figure 5 illustrates stress–strain data obtained without load cycling as measured for DegraPol® DP0 and DP30 scaffolds in circumferential and axial directions at the various time points of the study. Stress–strain data after incorporation of the effect of load cycling using the stress scaling function SC(ε) are provided in Fig. 6; the mechanical characteristics σ12%, σ16%, E6% and E12% are shown in Fig. 7. The changes of stress and elastic modulus from 0 to 14 and from 14 to 28 days are summarized in Table 2.
4 Discussion
The current study facilitated the paired in vivo assessment of the mechanics of biostable and degradable electrospun scaffolds based on the same polyester-urethane formulation to isolate the effects of material degradation and tissue formation after implantation. Biostable and biodegradable polyester-urethane scaffolds were implanted in the same animal.
The mechanical properties of biostable DegraPol® DP0 and degradable DegraPol® DP30 were found to be in the same range, although different responses were observed over implantation time. For DegraPol® DP0, circumferential data, but not axial, showed a significant increase from T = 7 to 14 days, and a significant decrease from T = 14 to 28 days in all characteristic measures, namely stress at 12% and 16% strain, and elastic modulus at 6% and 12% strain (Table 2). DegraPol® DP30 scaffolds exhibited a significant increase in stiffness in vivo. Whereas E6% did not change, E12% increased between T = 7 to 14 days by 50% (p < 0.05) for circumferential samples and by 213% for axial samples. This increase in elastic modulus was not reflected in the stress of circumferential samples but was for axial samples. Stress at 16% remained constant between T = 0 and 7 days but increased by 296% at T = 14 days (p < 0.05) and remained at that level at T = 28 days. The increase in elastic modulus after T = 7 days was linked to tissue ingrowth. Although there was approximately 30% ingrowth by 7 days, it was not yet fully interconnected throughout the scaffold, and thus did not contribute much to the overall mechanics. As tissue continued to populate the scaffold, regions with neo-tissue interconnected and developed greater structural strength.
Consistent mechanical performance from implantation through healing and scaffold degradation is highly desired for tissue-regenerative scaffolds, e.g. when used for vascular grafts. In the early phase after implantation, the circumferential stiffness in the scaffolds is based predominantly on fibres. As scaffold fibres degrade, the scaffold stiffness will decrease, unless the tissue balances out the difference. The consistency observed in circumferential stiffness of DegraPol® DP30 scaffold was thus of specific importance. Our previous in vitro study showed that the loss in material strength of DegraPol® DP30 predominantly occurred in the first 14 days [31], which correlated with decrease in molecular weight by T = 14 days. Assuming a similar trend for the scaffold in vivo, it holds that the tissue incorporation was contributing to the overall stiffness and balancing out the loss due to degradation. The stability in circumferential mechanical response contrasts, however, with the results by Teutelink et al. [36] where the grafts not implanted were significantly stiffer than the neo-vessels explanted after 90 days.
DegraPol® DP0 and DP30 scaffolds both featured a substantial degree of fibre alignment. However, DegraPol® DP0 scaffolds showed higher fibre alignment than DegraPol® DP30 scaffolds (p < 0.05), with high goodness values for the Gaussian fits of the fibre dispersion data in both cases (Table 1).
Tissue ingrowth was similar in DegraPol® DP0 and DP30 scaffolds, although it appeared to be slightly slower in the latter (Fig. 3). Levels of tissue ingrowth increased significantly between all subsequent time points of the study for both DegraPol® DP0 and DP30 scaffolds (p < 0.05) except the increase between T = 14 and 28 days for DegraPol® DP30.
In order to avoid inducing premature damage in DegraPol® scaffold samples at later time points of the study, load cycling was not included in the protocol for mechanical characterisation. However, the effect of load cycling was characterised for pre-implant scaffolds and applied to scaffolds retrieved at the various time points. DegraPol® DP0 scaffolds displayed a higher sensitivity to load cycling than the DP30 scaffolds. For DP0 scaffolds, the stress scaling parameter SC ranging between 0.20 and 0.59 indicates that load cycling reduced stress in the scaffold by at least 41%. In addition, the decrease in stress σNoncycled (Fig. 4a) bottom left panel) and stress scaling parameter SC (Fig. 4b left panel) for strain above 12% indicates damage induced in the DegraPol® DP0 scaffold due to load cycling) both in direction of and transverse to the fibre alignment. For DegraPol® DP30, SC ranged between 0.26 and 0.98 across the strain range studied. The DegraPol® DP0 scaffolds showed a more substantial weakening due to pre-cycling when compared to the DegraPol® DP30 scaffolds.
DegraPol® DP0 scaffolds were significantly more aligned than DegraPol® DP30, with a mean dispersion angle of less than 10°, compared to > 17° for DegraPol® DP30. DegraPol® DP30 scaffolds also showed more fibre merging than DegraPol® DP0, another possible factor. Lee et al. [37] reported similar behaviour in Pellethane® electrospun meshes. They indicated that a higher degree of fibre merging resulted in better distribution of load compared to an otherwise similar structure with lesser fibre merging. Low degrees of fibre merging led to a structure in which the load was not as readily distributed among fibres, resulting in earlier onset of failure of individual fibres, and consequently the entire mesh. This correlated with the observation that DegraPol® DP30 fibres displayed more merging and higher tolerance to cyclic loading than the DegraPol® DP0 scaffolds. In addition, higher alignment led to individual fibres experiencing higher strains at low scaffold strain and thus failing earlier, as the fibres were already straight and thus did not need to first straightened out [38]. Again, the scaffold morphology correlated with tensile test results, with the higher aligned DegraPol® DP0 scaffolds incurring more damage at lower strains, as reflected by the decrease in stress post 12% strain.
5 Conclusions
Considering our previous in vitro observation that the DegraPol® DP30 scaffold loses its material strength predominantly during the first 14 days of hydrolytic degradation, the consistency of the elastic modulus of the degradable scaffold between 14 and 28 days in the current study is an indication that the regenerated tissue construct retains its mechanical properties. The strain stiffening of the tissue constructs with DegraPol® DP30 scaffold from 7 days of implantation onwards (evidenced through the increase of elastic modulus from 6 to 12% strain) is an additional advantage and mechanical behaviour observed in soft biological tissues such as blood vessels. For the application of the scaffold as tissue regenerative vascular grafts, further research is required to indicate whether similar results will be obtained in the circulatory implant position.
Data availability
Quantitative data for fibre diameter, fibre alignment, tissue ingrowth and mechanical characterisation are available on ZivaHUB (http://doi.org/10.25375/uct.9892367).
References
Williams DF (2006) To engineer is to create: the link between engineering and regeneration. Trends Biotechnol 24(1):4–8. https://doi.org/10.1016/j.tibtech.2005.10.006
Furth ME, Atala A, Van Dyke ME (2007) Smart biomaterials design for tissue engineering and regenerative medicine. Biomaterials 28(34):5068–5073
Ratcliffe A (2000) Tissue engineering of vascular grafts. Matrix Biol 19(4):353–357
Chlupac J, Filova E, Bacakova L (2009) Blood vessel replacement: 50 years of development and tissue engineering paradigms in vascular surgery. Physiol Res 58(Suppl 2):S119–S139
Berglund JD, Galis ZS (2003) Designer blood vessels and therapeutic revascularization. Br J Pharmacol 140(4):627–636. https://doi.org/10.1038/sj.bjp.0705457
Wang X, Lin P, Yao Q, Chen C (2007) Development of small-diameter vascular grafts. World J Surg 31(4):682–689. https://doi.org/10.1007/s00268-006-0731-z
Teo W-E, Ramakrishna S (2009) Electrospun nanofibers as a platform for multifunctional, hierarchically organized nanocomposite. Compos Sci Technol 69(11–12):1804–1817. https://doi.org/10.1016/j.compscitech.2009.04.015
Zilla P, Bezuidenhout D, Human P (2007) Prosthetic vascular grafts: wrong models, wrong questions and no healing. Biomaterials 28(34):5009–5027
Limbert G, Omar R, Krynauw H, Bezuidenhout D, Franz T (2016) The anisotropic mechanical behaviour of electro-spun biodegradable polymer scaffolds: experimental characterisation and constitutive formulation. J Mech Behav Biomed Mater 53:21–39. https://doi.org/10.1016/j.jmbbm.2015.07.014
Ritter EF, Fata MM, Rudner AM, Klitzman B (1998) Heparin bonding increases patency of long microvascular prostheses. Plast Reconstr Surg 101(1):142–146
Gosselin C, Vorp DA, Warty V, Severyn DA, Dick EK, Borovetz HS, Greisler HP (1996) ePTFE coating with fibrin glue, FGF-1, and heparin: effect on retention of seeded endothelial cells. J Surg Res 60(2):327–332. https://doi.org/10.1006/jsre.1996.0052
Herring M, Smith J, Dalsing M, Glover J, Compton R, Etchberger K, Zollinger T (1994) Endothelial seeding of polytetrafluoroethylene femoral popliteal bypasses: the failure of low-density seeding to improve patency. J Vasc Surg 20(4):650–655
Mooney DJ, Baldwin DF, Suh NP, Vacanti JP, Langer R (1996) Novel approach to fabricate porous sponges of poly(-lactic-co-glycolic acid) without the use of organic solvents. Biomaterials 17(14):1417–1422
Pennel T, Zilla P, Bezuidenhout D (2013) Differentiating transmural from transanastomotic prosthetic graft endothelialization through an isolation loop-graft model. J Vasc Surg 58(4):1053–1061
Murphy WL, Dennis RG, Kileny JL, Mooney DJ (2002) Salt fusion: an approach to improve pore interconnectivity within tissue engineering scaffolds. Tissue Eng 8(1):43–52. https://doi.org/10.1089/107632702753503045
Jun H-W, West JL (2005) Endothelialization of microporous YIGSR/PEG-modified polyurethaneurea. Tissue Eng 11(7–8):1133–1140. https://doi.org/10.1089/ten.2005.11.1133
Nam YS, Park TG (1999) Biodegradable polymeric microcellular foams by modified thermally induced phase separation method. Biomaterials 20(19):1783–1790. https://doi.org/10.1016/s0142-9612(99)00073-3
Nam YS, Park TG (1999) Porous biodegradable polymeric scaffolds prepared by thermally induced phase separation. J Biomed Mater Res 47(1):8–17. https://doi.org/10.1002/(sici)1097-4636(199910)47:1%3c8:aid-jbm2%3e3.0.co;2-l
Bergmeister H, Schreiber C, Grasl C, Walter I, Plasenzotti R, Stoiber M, Bernhard D, Schima H (2013) Healing characteristics of electrospun polyurethane grafts with various porosities. Acta Biomater 9(4):6032–6040. https://doi.org/10.1016/j.actbio.2012.12.009
Hasan A, Memic A, Annabi N, Hossain M, Paul A, Dokmeci MR, Dehghani F, Khademhosseini A (2014) Electrospun scaffolds for tissue engineering of vascular grafts. Acta Biomater 10(1):11–25. https://doi.org/10.1016/j.actbio.2013.08.022
Hu J-J, Chao W-C, Lee P-Y, Huang C-H (2012) Construction and characterization of an electrospun tubular scaffold for small-diameter tissue-engineered vascular grafts: a scaffold membrane approach. J Mech Behav Biomed Mater 13:140–155. https://doi.org/10.1016/j.jmbbm.2012.04.013
McClure MJ, Sell SA, Simpson DG, Walpoth BH, Bowlin GL (2010) A three-layered electrospun matrix to mimic native arterial architecture using polycaprolactone, elastin, and collagen: a preliminary study. Acta Biomater 6(7):2422–2433. https://doi.org/10.1016/j.actbio.2009.12.029
Hong JK, Madihally SV (2011) Next generation of electrosprayed fibers for tissue regeneration. Tissue Eng B Rev 17(2):125–142. https://doi.org/10.1089/ten.teb.2010.0552
Vatankhah E, Semnani D, Prabhakaran MP, Tadayon M, Razavi S, Ramakrishna S (2014) Artificial neural network for modeling the elastic modulus of electrospun polycaprolactone/gelatin scaffolds. Acta Biomater 10(2):709–721. https://doi.org/10.1016/j.actbio.2013.09.015
Wu H, Fan J, Chu C-C, Wu J (2010) Electrospinning of small diameter 3-D nanofibrous tubular scaffolds with controllable nanofiber orientations for vascular grafts. J Mater Sci Mater Med 21(12):3207–3215. https://doi.org/10.1007/s10856-010-4164-8
Ayres CE, Bowlin GL, Pizinger R, Taylor LT, Keen CA, Simpson DG (2007) Incremental changes in anisotropy induce incremental changes in the material properties of electrospun scaffolds. Acta Biomater 3(5):651–661
de Valence S, Tille JC, Giliberto JP, Mrowczynski W, Gurny R, Walpoth BH, Möller M (2012) Advantages of bilayered vascular grafts for surgical applicability and tissue regeneration. Acta Biomater 8(11):3914–3920. https://doi.org/10.1016/j.actbio.2012.06.035
Zhu M, Wang K, Mei J, Li C, Zhang J, Zheng W, An D, Xiao N, Zhao Q, Kong D, Wang L (2014) Fabrication of highly interconnected porous silk fibroin scaffolds for potential use as vascular grafts. Acta Biomater 10(5):2014–2023. https://doi.org/10.1016/j.actbio.2014.01.022
Henry JJD, Yu J, Wang AJ, Lee R, Fang J, Li S (2017) Engineering the mechanical and biological properties of nanofibrous vascular grafts for in situ vascular tissue engineering. Biofabrication. https://doi.org/10.1088/1758-5090/aa834b
Jirofti N, Mohebbi-Kalhori D, Samimi A, Hadjizadeh A, Kazemzadeh GH (2018) Small-diameter vascular graft using co-electrospun composite PCL/PU nanofibers. Biomed Mater 13(5):055014. https://doi.org/10.1088/1748-605x/aad4b5
Krynauw H, Bruchmüller L, Bezuidenhout D, Zilla P, Franz T (2011) Degradation-induced changes of mechanical properties of an electro-spun polyester-urethane scaffold for soft tissue regeneration. J Biomed Mater Res B 99B(2):359–368. https://doi.org/10.1002/jbm.b.31907
Krynauw H, Buescher J, Koehne J, Verrijt L, Limbert G, Davies NH, Bezuidenhout D, Franz T (2019) Tissue ingrowth markedly reduces mechanical anisotropy and stiffness in fibre direction of highly aligned electrospun polyurethane scaffolds. bioRxiv:779942
Milleret V, Simonet M, Bittermann AG, Neuenschwander P, Hall H (2009) Cyto- and hemocompatibility of a biodegradable 3D-scaffold material designed for medical applications. J Biomed Mater Res B Appl Biomater 91(1):109–121. https://doi.org/10.1002/jbm.b.31379
Schindelin J, Arganda-Carreras I, Frise E, Kaynig V, Longair M, Pietzsch T, Preibisch S, Rueden C, Saalfeld S, Schmid B, Tinevez J-Y, White DJ, Hartenstein V, Eliceiri K, Tomancak P, Cardona A (2012) Fiji: an open-source platform for biological-image analysis. Nat Methods 9(7):676–682
Woolley AJ, Desai HA, Steckbeck MA, Patel NK, Otto KJ (2011) In situ characterization of the brain–microdevice interface using device capture histology. J Neurosci Methods 201(1):67–77. https://doi.org/10.1016/j.jneumeth.2011.07.012
Teutelink A, van der Laan MJ, Milner R, Blankensteijn JD (2003) Fabric tears as a new cause of type III endoleak with Ancure endograft. J Vasc Surg 38(4):843–846
Lee K, Lee B, Kim C, Kim H, Kim K, Nah C (2005) Stress-strain behavior of the electrospun thermoplastic polyurethane elastomer fiber mats. Macromol Res 13(5):441–445. https://doi.org/10.1007/bf03218478
Mauck RL, Baker BM, Nerurkar NL, Burdick JA, Li WJ, Tuan RS, Elliott DM (2009) Engineering on the straight and narrow: the mechanics of nanofibrous assemblies for fiber-reinforced tissue regeneration. Tissue Eng B Rev 15(2):171–193. https://doi.org/10.1089/ten.TEB.2008.0652
Acknowledgements
The authors thank ab medica S.p.A for donating the DegraPol® material for this study. ETH Zurich and the University of Zurich are owners and ab medica S.p.A is exclusive licensee of all Intellectual Property Rights of DegraPol®.
Funding
This study was supported financially by the National Research Foundation of South Africa and the South African Medical Research Council. HK received a matching dissertation grant from the International Society of Biomechanics. JK acknowledges financial support from the Deutsche Herzstiftung (Jahresstipendium S/08/10).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflicts of interest
The authors declare that they do not have conflicts of interests.
Ethical approval
All animal experiments were authorised by the Faculty of Health Sciences Research Ethics Committee of the University of Cape Town and were performed in accordance with the National Institutes of Health (NIH, Bethesda, MD) guidelines.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Krynauw, H., Omar, R., Koehne, J. et al. Electrospun polyester-urethane scaffold preserves mechanical properties and exhibits strain stiffening during in situ tissue ingrowth and degradation. SN Appl. Sci. 2, 953 (2020). https://doi.org/10.1007/s42452-020-2764-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s42452-020-2764-6