Abstract
Background
A set of nozzle systems for proton therapy is now being developed at China Institute of Atomic Energy. To realize the measurement of beam dose, a set of charge measurement electronics is designed, which is used to measure the output charge signal of the ionization chamber integration plane. Also, the charge measurement device can be used to measure the position information of ion chamber strips just by changing certain parameters.
Methods
The device realizes the integration and amplification of charge by IVC102, a precise low-noise integrator, and the DSP controls the ADC to collect and process the data. Modbus communication protocol is used to realize communication with the host computer which makes it possible to read data and set the parameter.
Results
A fixed charge generator is designed to generate 20 pC to 2 nC charge, which is used to test the measuring accuracy of electronics. The results have shown that the accuracy is within 0.62% in the above charge range.
Conclusion
After experimental tests, the front-end electronics meet the design goal and play a certain pre-test and verification role in the dose and beam position monitoring in the proton therapy system.
Similar content being viewed by others
References
T. Zhang, C. Wang, M. Li et al., Developments for 230MeV superconducting cyclotrons for proton therapy and proton irradiation. Nucl. Instrum. Methods Phys. Res. 406, 244–249 (2017)
T. Zhang, J. Lin, Z. Yin et al., Design and construction of the main magnet for a 230-MeV superconducting cyclotron. ITAS 28(99), 1–1 (2017)
T. Zhang, M. Li, C. Wang et al., Investigation and quantitative simulation of beam physics for 230 MeV SC cyclotron under construction at CIAE. Nucl. Instrum. Methods Phys. Res. Sect. B 468, 8–13 (2020)
Z. Yin, B. Ji, X. Fu et al., Design of RF system for cyciae-230 superconducting cyclotron. Nucl. Instrum. Methods Phys. Res. 854(Complete), 25–30 (2017)
J.M. Schippers, Beam delivery systems for particle radiation therapy: current status and recent developments. Rev. Accel. Sci. Technol. 02(01), 179–200 (2009)
B. Marchand, D. Prieels, B. Bauvir, R. Sépulchre, M. Gérard, IBA proton pencil beam scanning: an innovative solution for cancer treatment. In Proceedings of EPAC, (2000), pp. 2539–2541
G. Klimpki, M. Eichin, C. Bula et al., Real-time beam monitoring in scanned proton therapy. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 891(21), 62–67 (2018)
F. Fausti, G. Mazza, S. Giordanengo et al., Single Event Upset tests and failure rate estimation for a front-end ASIC adopted in high-flux-particle therapy applications. Nucl. Instrum. Methods Phys. Res. Sect. A Accel. Spectrom. Detect. Assoc. Equip. 918(21), 54–59 (2019)
S. Lin et al., More than 10 years of experience of beam monitoring with the Gantry 1 spot scanning proton therapy facility at PSI. Med. Phys. 36, 5331–5340 (2009)
M. Torikoshi et al., Beam monitor system for high-energy beam transportation at HIMAC. Nucl. Instrum. Methods Phys. Res. A 435, 326–338 (1999)
T. Furukawa et al., Performance of the NIRS fast scanning system for heavy-ion radiotherapy. Med. Phys. 37, 5672–5682 (2010)
E.S. Diffenderfer, I.I. Verginadis, M.M. Kim et al., Design, implementation, and in vivo validation of a novel proton FLASH radiation therapy system. Int. J. Radiat. Oncol. Biol. Phys. 106(2), 440–448 (2020)
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Huang, P., Yin, ZG., Mu, XE. et al. Design of the front-end electronics for parallel plate ion chamber. Radiat Detect Technol Methods 5, 558–563 (2021). https://doi.org/10.1007/s41605-021-00278-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s41605-021-00278-1