Introduction

In recent years, elemental bioimaging has gained more and more importance owing to the increasing importance of life sciences and to instrumental improvements [13]. One of the most important techniques for this purpose is the combination of laser ablation (LA) and inductively coupled plasma mass spectrometer (ICP-MS), either equipped with a quadrupole (QP) or with a sector field (SF) mass analyzer [4, 5]. A spatial resolution down to 2–4 μm for commercially available instruments and limits of detection in the low ng/g range may be achieved owing to the high sensitivity and low background noise for detecting most elements of the periodic table. LA-ICP-MS has therefore become a versatile tool for the life sciences [68].

However, there are still major challenges remaining for this technology: spatial resolution still needs to be improved, and increased sensitivity is required to sustain low limits of detection for decreasing amounts of sample. Once the sample is ablated and subsequently transported by a carrier gas (He and Ar) to the ICP-MS for vaporization, atomization, and ionization, it cannot be analyzed a second time. This is particularly problematic if the investigated analyte is not free from spectral polyatomic interferences deriving from the plasma gas or when drift effects occurred during analysis, although interferents may be utilized as internal standards in order to overcome drift effects [9, 10].

Thus, it would be superior to acquire signals of all elements, or at least all elements that could be of further interest, in one single run and to have a full mass spectrum available for each acquired pixel [11]. In the field of “organic” mass spectrometric imaging, with matrix assisted laser desorption ionization (MALDI) as its common ionization technique, this is achieved for example by utilization of Orbitrap [12] or time of flight (ToF) [13] mass analyzers. However, these techniques suffer from a lack of simple quantitation capabilities compared with ICP-based approaches. Nevertheless, “inorganic” mass spectrometry struggles to cope with short transient signals, and different dedicated approaches have been proposed to overcome this issue in LA-ICP-MS [11, 1416].

Single-collector sector field instruments in the straight or reversed Nier-Johnson geometry are limited in their scan speed across the entire mass region of interest owing to the settling time of the utilized electromagnet and the field regulator. A fast scan of the electrostatic analyzer (ESA) in conjunction with changes in the acceleration voltage is only possible in a mass range of about +30% from the adjusted m/z of the magnet. This combination does not show better sensitivity compared with modern quadrupole instruments. Instruments using a line of direct-charge-detectors (DCD) or Faraday-strip array detectors in the Mattauch-Herzog geometry are, in theory, capable of recording the complete “inorganic” m/z range simultaneously [6, 17]. However, these techniques suffer from comparably poor limits of detection (LOD) and a lack of robustness. Therefore, multi-collector instruments remain the gold standard for high-precision isotopic ratio determinations [1821].

For this reason, another fast scanning technique, widely applied in “organic” mass spectrometry, was evaluated as a detection system for LA-ICP-MS in 1996 [22]; early commercial systems could not compete with quadrupole systems with respect to detection power and duty cycle. Günther et al. [23] recently published several articles about the construction and application of an ICP-TOF-MS instrument. It is mainly used for the analysis of nanoparticles and overcomes the typical issues of former instrument generations, including the inferior time resolution, which is due to the necessity to sum hundreds of spectra to improve the signal to noise ratio (S/N). Bandura et al. described the development and application of a recently commercialized ICP-ToF-MS for the analysis of single cells tagged with rare earth elemental labels (CyTOF). The instrument is able to acquire a mass spectrum from m/z 125 to 215 in 13 μs, resulting in a scan speed of 76.8 MHz [24].

For the analysis of nanoparticles in the “single particle mode,” the sample is highly diluted, and individual particles are recorded as spikes. The number of spikes is proportional to the concentration of particles, and the intensity of each spike corresponds to the particle size. For this work, short dwell times in the range of a few milliseconds are required, whereas the visualization of the generated ion cloud calls for dwell times of 100 μs or less [25, 26]. Even if this is possible with some commercially available instruments, the delay between consecutive scan cycles prevents its application. Therefore, oscilloscopes have been used to read out the signals of the secondary electron multiplier detectors [27]. The same considerations are valid for single-shot LA experiments [28]. For the hyphenation of gas chromatography (GC) and MS, the use of the second stability region of the Mathieu diagram has been discussed, allowing scan speeds of up to 1000 Hz for an m/z region of 80, but this suffers from poor sensitivity compared with region one and mass aliasing [29]. Hence, short transient signals remain a challenge for elemental mass spectrometry [3032].

In this work, a new approach is presented to close the gap between detection schemes used for “inorganic” and “organic” mass spectrometry: data acquisition by continuous sweeping or fast jumping over the complete m/z range of interest. This technique is widely applied in quadrupole-based ICP-MS instead of the conventional peak-hopping mode with the goal to reveal spectral interferences and unexpected matrix components as well as for semiquantitative data evaluation. Here, the sensitivities for all analytes present in the external calibration are plotted against the corresponding m/z value after correction for isotopic abundances. This fast scan provides several other unique benefits, especially for transient signals generated by hyphenated techniques, including LA-ICP-MS, considering that the settle time of the quadrupole depends on the width of the mass jump over the mass region of interest. Further advantages are the enhancement of sensitivity by summation of mass channels as well as the possibility of generating mass ratio and interferences plots. These will be discussed and utilized for different applications, including the analysis of dried droplets and thin tissue slices of animal or human origin. Additionally, a simple Visual Basic for Applications (VBA) macro is presented for data handling.

Experimental

Instrumentation

An inductively coupled plasma mass spectrometer of the most recent generation (iCAP Qc; Thermo Fisher Scientific, Bremen, Germany), equipped with a quadrupole mass analyzer and a flatapole-based collision/reaction cell (QCell) was used throughout all studies. Owing to the possible formation of unwanted polyatomic spectral interferences, the high sensitivity kinetic energy discrimination mode (KEDS) of the ICP-MS system was utilized with an energy barrier of 3 V between the cell and the quadrupole mass analyzer. The instrument was operated at a power of 1550 W (free-running generator), a cool gas flow of 14 L/min, and an auxiliary gas flow of 0.5–0.8 L/min. The sampling depth was set to 5–7 mm. Helium (99.999% purity) served as collision gas for all studies involving dried coffee and whole blood samples, and a mixture of 6% (v/v) H2 in He (99.999% purity) was used for all other experiments. The cell gas flow and the lenses in front and behind the collision/reaction cell were adjusted on a daily basis using a Co standard solution with a concentration of 1 μg/L in 2% (w/w) HNO3 and 0.5% (w/w) HCl. The cell gas flow was set to give a 59Co/35Cl16O ratio >25. This resulted in flow values of 4.2 mL/min for pure He and 5.9 mL/min for the gas mixture.

The utilized laser ablation system (LSX-213; Teledyne CETAC Technologies Inc., Omaha, NE, USA) was based on a frequency-quintupled, Q-switched Nd:YAG laser and coupled to the ICP-MS by means of a PTFE-coated Tygon tube.

Critical laser parameters, including laser energy and shot repetition rate, were adjusted according to the utilized scan rate in order to ablate the investigated material quantitatively in a multiline scan with no gap between the lines. Detailed parameters are given in the corresponding sections. The scan speed of the ablation stage was set to produce one pixel per second in the final image to simplify image processing and to improve comparability between different studies. Only for the coffee droplet studies, the scan speed was adjusted based on the set dwell time to show the influence of the quadrupole settling. Special attention was paid to the formation of debris, as a re-ablation would distort the obtained elemental images. A cylindrical ablation chamber with sample insert was used to reduce the inner volume of the cell to approximately 25 cm3, resulting in combined washout and transport times in the range of 1.8–2.2 s, whereas the washout time of the cell itself was in the range of 0.9 s. To achieve these values, a He flow of 0.7 L/min was used to flush the generated aerosol from the ablation cell to the ICP-MS. Coupling of the instruments was performed using a home-built dual sample inlet in which the aerosol from the laser ablation is mixed with the tertiary aerosol, generated by a μFlow-PFA nebulizer (Elemental Scientific Inc., Omaha, NE, USA). The cyclonic quartz glass spray chamber was cooled to 2.7 °C. This dual mode approach has two major advantages. First, it facilitates the tuning procedure: the x- and y-position of the torch, the nebulizer gas flow, the potential of the second extraction lens, and the lens in front of the QCell were tuned on a daily basis by introducing a multi-element standard with a sensitivity for 1 μg/L of at least 350,000 cps, an oxide ratio of <0.01 (measured as 140Ce oxide), and a doubly charged ratio of <0.03 (measured on 137Ba), with all other gas flows set as described. Typical nebulizer flow values were in the range of 0.7 L/min for a 3.5 mm injector pipe. In each case, Ni sampler and skimmer cones were used, the latter equipped with an insert length of 2.8 mm. Second, the dual mode approach allows for a constant monitoring of the instrument’s sensitivity. For all studies, the relative standard deviations (RSDs) were well below 5% for the complete runs.

Macroinstruction files were written using Microsoft (Redmond, WA, USA) Excel 2013 VBA and Microsoft Visual Studio Professional 2013 in order to handle the raw data. The ICP-MS was controlled using the Qtegra ISDS 1.7 software (Thermo Fisher Scientific, Bremen, Germany). The exported CSV files were transformed into images using ImageJ 1.46r (Wayne Rasband, http://imagej.nih.geov/ij). Histograms were plotted using Origin Pro 8 SR2 (OriginLab, Northampton, MA, USA).

Materials

Elemental standards (1000 mg/L) were ordered from VHG Labs (Manchester, NH, USA). Nitric acid (suprapure, 65%) was purchased from Merck (Darmstadt, Germany). Paraplast X-TRA (melting point 52°C) and Surgipath Paraplast (melting point 56°C) were purchased from Leica Biosystems (Wetzlar, Germany). Roti-Histofix (4% formaldehyde in phosphate buffer) was obtained from Carl Roth (Karlsruhe, Germany). Superfrost Plus adhesive microscope slides were purchased from Thermo Fisher Scientific (Waltham, MA, USA). Isobutanol was obtained from Amresco (Solon, OH, USA) and ethanol from AppliChem (Darmstadt, Germany). For lymph node tissue section preparation, formaldehyde was purchased from H. Möller (Steinfurt, Germany), ethanol and xylene from Kassner and Sasse (Dortmund, Germany), Certistain eosin Y C.I. from Merck. Water was freshly purified through an Aquatron Water Stills purification system model A4000D (Barloworld Scientific, Nemours Cedex, France). All chemicals were used in the highest purity available.

FVB/N HanHsd mice (Harlan Laboratories, Horst, The Netherlands) were grown in the animal facilities of the University Hospital of Münster. Cisplatin [cis-diamminedichloridoplatinum(II), CDDP] treatment was carried out with a 8 × 4 mg/kg delivery over 8 wk. The animals were sacrificed 7 wk later. Experiments were approved by a governmental committee on animal welfare (LaNUV, permission number A 140.2011) and performed in accordance with national animal protection guidelines. Perfusion, fixation, staining, and embedding of the tissue sample before sectioning to 3 μm was done as described recently [33]. Lymph node samples were obtained from anonymized human patients as a waste product after surgeries and prepared to 8-μm-thick sections as described elsewhere [33]. Human whole blood was taken from a healthy volunteer.

Results and Discussion

General Considerations

The spatial resolution of an image is adjustable for the y-direction via the laser spot diameter focused onto the sample. However, the maximum feed rate of the laser beam in the x-direction in a line-by-line scan is typically limited by the washout time of the ablation cell and the scan speed of the quadrupole mass analyzer. Currently, the maximum number of pixels that can be recorded per second seems to be in the region of 4 Hz [34]. This also means that if one is interested in only one analyte, the dwell time of the quadrupole is adjusted in this case to f = 0.25 s, whereas the settle times of the quadrupole are not taken into account. Thus, the dwell time is set to 1/f. The electronics of commercially available quadrupole systems are able to scan the mass analyzer much faster than this and, hence, not only one but a variety of analytes of interest are typically selected for elemental bioimaging studies. Thus, the equation can be adapted to 1/(j*c*f), where j is the number of analytes and c is the number of channels for one mass peak. Typically, not more than 10 analytes are selected, even if more analytes would be favorable owing to the destructive nature of the LA technique. This results in a dwell time of 25 ms in the example mentioned above. However, with an increasing number of analytes, the duty cycle of the quadrupole is increasingly important. As the settle time, which is necessary for stable operation, depends on the width of the mass jump, the total scan time for, e.g., 100 channels, is not always the same, but rather depends on the selected mass region. Thus, the optimum duty cycle for a given number of mass channels is achieved if the mass window is comparably narrow. Typical values are in the range of not more than 0.2–0.5 u distance between consecutive mass channels, depending on the instrument. If one would select all isotopes that could be of particular importance for an elemental bioimaging study, most of the analysis time would be wasted to adjust the correct m/z value. This lowers the effective time for ion collection and, accordingly, the LODs because of poor counting statistics. Moreover, the possibility of false positive results increases because of spikes of different origin. The reason for this is that the displayed unit of the intensities is counts per second (cps), and the raw intensity within one dwell time is multiplied by the reciprocal of the dwell time. Additionally, unidentified spectral interferences may occur. Hence, a fast survey scan acquisition of the complete m/z region with fast jumping between the mass channels could circumvent these issues. The principle of this approach, which can be used on nearly every commercially available quadrupole ICP-MS instrument, is shown in Figure 1. It virtually adds another dimension to elemental bioimaging: the m/z ratio in addition to x- and y-position and the intensity of the signal. By plotting the complete m/z region of interest, one has a powerful tool to reveal spectral interferences and a proof for the identity of the analyte by its isotopic abundances.

Figure 1
figure 1

Schematic depiction of the fast survey scan acquisition approach: the sample is ablated in a line-by-line scan, whereas the produced dry aerosol is subsequently transported to the ICP-MS system equipped with a quadrupole mass analyzer. Owing to a fast survey scan acquisition, a complete mass spectrum for each pixel is obtained

Data Processing and Sensitivity Improvement

To prove if the fast scanning approach is able to provide accurate results without image distortion, a coffee droplet was spotted on a microscopic slide. After drying, the droplet was ablated in 59 line scans with a gap between the lines of 0 μm, a spot size of 25 μm, a laser scan rate of 50 μm/s, and a laser energy of 2.9 mJ/shot using a repetition rate of 20 Hz. The transient survey scan covered a mass range from 22.1 to 240 u with a distance between consecutive mass channels of 0.2 u. Each channel was scanned with a dwell time of 500 μs. Thus, a number of 1089 channels should be scanned within 544.5 ms. However, owing to the settling time of the quadrupole, the duty cycle needed to be taken into account, which equaled 37.8% in this experiment (calculated by the exact start and stop times of each line scan) and, hence, 1.44 s were needed for one scan. Figure 2 shows an example of a mass spectrum as the sum of intensities of one single line, whereas Figure 3 shows a summary of the obtained data for potassium. As can be readily seen, the image of the droplet is stretched in the x-direction because of the prolonged acquisition time, whereas in the y-direction, an accurate distance was obtained (Figure 3a). Nevertheless, refitting the size of the pixels by a factor of 0.378 in the x-direction gave a correct representation. This shows that the settling time of the quadrupole is the only effect, which needs to be taken into account for the conversion of the time scale to the distance scale (Figure 3d). This statement was proven by comparison to a fluorescence image of the same droplet (Figure 3c) showing the same width and height as well as the same structural details completely as the potassium ion image.

Figure 2
figure 2

Mass spectrum as the sum of intensities obtained after the ablation of one single line of a dried coffee droplet. As the complete mass range of interest was recorded, the identity of the analytes can be proven by their isotopic pattern

Figure 3
figure 3

39K trace of a coffee droplet: image of central mass channel (a) and summation over 5 mass channels (b), both without correction for distortion, which has its origin in the prolonged scan time of the quadrupole compared with the adjusted value. Fluorescence image of the coffee droplet (c), which is in excellent agreement with the duty cycle corrected ion image (d). Distortion of the obtained images was overcome by calculation of the corresponding duty cycle, whereas sensitivity was improved by summation of all mass channels for one isotope. In (d), a bilinear interpolation, in which pixels with the mean value of the adjacent ones are added for both directions, was additionally used during distortion correction

The obtained sensitivity is directly proportional to the contrast of the ion images. The novel imaging technique allows adding up the different mass channels for one mass peak, e.g., five channels in this example. The approach is commonly performed on SF-instruments, thereby virtually increasing their sensitivity for every E-scan. As can be seen in Figure 3, the same effect is observed for this approach. After summation of all mass channels in the range of 38.5–39.5 u and considering that in normal resolution mode the peak width at 10% peak height (R = 300) was adjusted to be < 0.7 u, the sensitivity is significantly increased (Figure 3a → Figure 3b). This provides an enhanced contrast of the final image and removes false pixel colors originating from spikes in the raw data.

Another critical aspect in recording transient signals by sequential scans is the spectral skew: when a low mass analyte like lithium is only present in the last part of an ablated voxel, which is the volume equivalent of a pixel, or a high mass analyte like uranium is only present in the first part, they may not be detected, as each voxel is scanned beginning at the lower mass side [14]. This is also valid for imaging experiments involving more than one analyte, but becomes more significant when the number of monitored analytes increases. It is therefore of particular importance to set the laser beam diameter to a sufficiently small size to represent the structure of the investigated sample. For a spiked droplet shown in Figure 3, this was already achieved using a spot size of 25 μm. Adjacent pixels show only slight variations in terms of intensity. Hence, there is no evidence that spectral skew obstructs the correct representation of elemental distributions inside a sample.

Limits of Detection, Accuracy, and Precision in Isotopic Ratio Analyses

To investigate the effect of shortened dwell times and mass channel summation in detail, different scan parameters were tested during the ablation of NIST SRM 612–trace elements in glass. Titanium (scanned at m/z 49) was chosen as a model analyte with low isotopic abundance and high blank values, whereas thorium served as high mass analyte with superior LOD.

Owing to calculation of the intensity as counts per second, the gas blank values as well as the obtained intensities for the ablation of the standard do not vary significantly from each other for the different investigated scan settings. This is demonstrated in Table 1. The relative standard deviation (RSD) is proportional to the number of runs, when calculated for a fixed scan duration of 60 s. Summation of three mass channels (spacing = 0.1 u) does not influence the limits of detection, which were calculated according to the 3σ-criterion. This is because independent of the number of mass channels, the number of recorded data points and, in good approximation, the ion current hitting the secondary electron multiplier, stays the same. Hence, the results are in accordance to Poisson (counting) statistics, which can be seen by the constancy in the calculated statistical factor, which is the square root of the raw intensities times the number of scans divided by the calculated LOD. In general, imaging based on fast survey scan acquisition degrades the obtainable LODs by approximately one order of magnitude in a worst case scenario. However, this does not limit the applicability significantly when taking the current instrumental capabilities into account. Nevertheless, the results provide further evidence for the validity of mass trace summation, as sensitivity and, hence, image contrast is improved. For thorium, LODs could not be calculated, as the gas blank values for the different scan settings always equaled zero. The standard deviations for the ablation of the glass SRM were always in the same range (approximately 30%) and, hence, the fast scanning imaging approach over a wide mass range here delivered the same results compared with conventional imaging experiments monitoring only one isotope. However, the wider mass window enabled the time-resolved monitoring of oxide formation (about 5%) and thus further experiments were conducted to show the ability to obtain accurate and precise results.

Table 1 Experimental limits of detection using different scan settings for 49Ti in NIST SRM 612 – Trace elements in glass (reference value = 50.1 ± 0.8 mg/kg). The observed RSDs increase with the number of runs conducted within a defined time interval. The number of scanned channels per mass, here with a spacing of 0.1 u, virtually increases the sensitivity, but not the precision. Hence, the achievable LODs are in the same range for experiments involving 1 or 3 channels. The presented findings are in good agreement to Poisson statistics. [35] A calculated statistical factor (square root of the raw intensities times the number of scans divided by the calculated LOD) is almost constant for all settings

Figure 4 shows an overview of the data, which were obtained by ablation of a dried human whole blood droplet, spiked with 10 mg/L lead. Here, the laser parameters were set to a spot size of 25 μm, a shot repetition rate of 20 Hz, a scan speed of 25 μm/s, and a laser energy of 1.95 mJ/shot. Again, the “coffee ring” structure can be seen, e.g. for m/z 206 in (Figure 4a). However, as the mass range from 200.1 to 210.0 u was scanned with a spacing of 0.05 u between the mass channels and a dwell time of 0.005 s, a region comprising all Pb isotopes could be used for data evaluation. The duty cycle for this experiment was calculated to be 71.4% attributable to the smaller investigated mass region and the smaller spacing between the 198 recorded mass channels (about 14,256 data points per line). As before, a VBA software macro was utilized to extract the m/z traces of interest from the raw data and to add up all mass channels for one isotope. In addition, the macro is able to display the recorded isotopic pattern and to divide the results in order to generate isotopic ratio maps as in Figure 3b. This can be found in the Supplementary Information.

Figure 4
figure 4

Results of the ablation of a dried human whole blood droplet in 56 line scans spiked with 10 mg/L lead: qualitative data for m/z 205.5–206.5 (a), isotopic ratio map of 208Pb/206Pb (b), and histogram of the isotopic ratio distribution (green bars, 100 center aligned bins, each 0.02), Lorentz fitting (black line) with expected natural abundance (red line) (c)

It can be seen that not only the acquisition of an isotopic mass can be used to investigate the concentration depending distribution of trace elements but, furthermore, the isotopic pattern helps to gain more evidence about the identity of the analyte by comparing the found isotopic abundance pattern to the expected one.

However, inorganic fast survey scan acquisition imaging is also able to use the interferents as internal standards. It is even possible to do a mathematical interference correction for each pixel (i.e., to subtract the interferent signal from the analyte signal. The isotopic ratios for the dried blood sample found were plotted as a histogram with 100 bins, each with a width of 0.02. It shows how often an isotopic ratio was found for the complete image. The bars were subsequently fitted with a Lorentz curve and the maximum was compared with the natural abundance. As can be seen in Figure 3c, the found ratio was in good agreement with the expected value. The width of the graph is comparably narrow and provides accurate results.

Applicability to Thin Tissue Sections

To investigate more challenging samples and to prove the capability of the fast scanning imaging approach, thin tissue slides were analyzed. Figure 5 shows the isotopic pattern, which was achieved after cumulating all intensities of a single line scan over a mouse kidney after Cisplatin treatment (Figure 5a). The very good agreement between the measured and the calculated isotopic pattern is obvious (Figure 5b). The scan range was 190.1–200.0 u with a spacing of 0.05 u and a dwell time of 0.005 s. Even the minor isotope 192Pt, which has a natural abundance of 0.79%, can be detected. In addition, a meaningful Pt distribution map was obtained using the fast survey scan acquisition with a laser spot diameter of 50 μm, a scan speed of 50 μm/s, a repetition rate of 20 Hz, and a laser energy of 1.95 mJ/shot. In the generated image for 195Pt (m/z 194.5–195.5 u, sum of 19 mass channels) of the 3 μm thick kidney section, Pt is distributed heterogeneously in the medulla inner part of the kidney, whereas an accumulation can be seen in the cortex. This result is in excellent agreement to literature, as the cortex contains the proximal tubules. They are specifically damaged after a CDDP treatment [32, 36, 37]. Additionally, an eosin Y (disodium tetrabromofluorescein) stained human lymph node was analyzed using the same instrumental conditions. This staining is commonly applied in histology to improve visualization of the thin tissue slices for microscopy. It dyes cytoplasmic material in red color [38]. Moreover, it has been recently of interest in elemental bioimaging as a reference marker for structural information [8, 33, 39].

Figure 5
figure 5

Ion image for 195Pt of mouse kidney after Cisplatin treatment (a) and cumulated isotopic pattern (black line) with simulated natural abundance (red bars) for a single line scan (b)

In Figure 6, the extracted isotopic images for 65Cu (m/z 64.5-65.5) and 79,81Br (m/z 78.5–79.5 + 80.5–81.5, respectively) for a human lymph node sample sectioned at 8 μm are shown. The sample was ablated using a spot size of 50 μm, a laser scan speed of 50 μm/s, a shot repetition rate of 20 Hz, and a laser energy of 2.1 mJ/shot. In this case, the mass region from 43.0 to 92.0 u was investigated with a spacing of 0.1 u and a dwell time for each of 490 channels of 0.002 ms. The duty cycle equaled 59.1%. The image for Cu (Figure 6a) and Br (Figure 6b) represent the sample structure well.

Figure 6
figure 6

Images of a human lymph node sample sectioned at 8 μm and ablated with a laser spot diameter of 50 μm , showing the distribution of 65Cu (m/z 64.5–65.5) and 79,81Br (m/z 78.5–79.5, 80.5–81.5, respectively)

Conclusion

Fast survey scan acquisition was shown to be a valuable addition to the toolbox of elemental bioimaging. Advantages, including the possibility to obtain more information per image, the option to sum up mass channels, the reduction of the amount of spikes, and the increase of the sensitivity by a factor of five or even more are presented. In addition, the capability of this method to generate spatially resolved isotopic abundance maps and isotopic patterns for certain regions of the investigated samples are shown for murine kidney and human lymph node samples.