An MLC-based version for the ecliptic method for the determination of backscatter into the beam monitor chambers in photon beams of medical accelerators

  • Technical Paper
  • Published:
Australasian Physical & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

A very simple method to measure the effect of the backscatter from secondary collimators into the beam monitor chambers in linear accelerators equipped with multi-leaf collimators (MLC) is presented here. The backscatter to the monitor chambers from the upper jaws of the secondary collimator was measured on three beam-matched linacs by means of three methods: this new methodology, the ecliptic method, and assessing the variation of the beam-on time per monitor unit with dose rate feedback disabled. This new methodology was used to assess the backscatter characteristics of asymmetric over-traveling jaws. Excellent agreement between the backscatter values measured using the new methodology introduced here and the ones obtained using the other two methods was established. The experimental values reported here differ by less than 1 % from published data. The sensitivity of this novel technique allowed differences in backscatter due to the same opening of the jaws, when placed at different positions on the beam path, to be resolved. The introduction of the ecliptic method has made the determination of the backscatter to the monitor chambers an easy procedure. The method presented here for machines equipped with MLCs makes the determination of backscatter to the beam monitor chambers even easier, and suitable to characterize linacs equipped with over-traveling asymmetric secondary collimators. This experimental procedure could be simply implemented to fully characterize the backscatter output factor constituent when detailed dosimetric modeling of the machine’s head is required. The methodology proved to be uncomplicated, accurate and suitable for clinical or experimental environments.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

Notes

  1. Linacs located at CSSSG-Hôpital de Gatineau, Gatineau QC, Canada.

References

  1. Kubo H (1989) Telescopic measurement of backscattered radiation from secondary collimator jaws to a beam monitor chamber using a pair of slits. Med Phys 16:295

    Article  CAS  PubMed  Google Scholar 

  2. Huang PH, Chu J, Bjärngard BE (1987) The effect of collimator backscatter radiation on photon output of linear accelerators. Med Phys 14:268

    Article  CAS  PubMed  Google Scholar 

  3. Liu HH, Mackie TR, McCullough EC (2000) Modeling photon output caused by backscattered radiation into the monitor chamber from collimator jaws using a Monte Carlo technique. Med Phys 27:737

    Article  CAS  PubMed  Google Scholar 

  4. Sanz DE, Alvarez GD, Nelli FE (2007) Ecliptic method for the determination of backscatter into the beam monitor chambers in photon beams of medical accelerators. Phys Med Biol 52:1647

    Article  PubMed  Google Scholar 

  5. Zhu TC, Ahnesjö A, Lam KL, Li XA, Ma CMC, Palta JR, Sharpe MB, Thomadsen B, Tailor RC (2009) Report of AAPM Therapy Physics Committee Task Group 74: in-air output ratio, Sc, for megavoltage photon beams. Med Phys 36:5261

    Article  PubMed  Google Scholar 

  6. Cranmer-Sargison G, Weston S, Evans JA, Sidhu NP, Thwaites DI (2011) Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors. Med Phys 38:6592

    Article  CAS  PubMed  Google Scholar 

  7. Ahnesjö A, Knoos T, Montelius A (1992) Application to the convolution method for calculation of output factors for therapy photon beams. Med Phys 19:295

    Article  PubMed  Google Scholar 

  8. Eclipse algorithms reference guide: chapter 4 photon beam source model, Varian medical systems, Inc. (2011)

  9. Zavgorodni S, Alhakeem E, Townson R (2014) Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling. Phys Med Biol 59:911

    Article  PubMed  Google Scholar 

  10. Yang Y, Xing L, Boyer AL, Song Y, Hu Y (2002) A three-source model for the calculation of head scatter factors. Med Phys 29:2024

    Article  PubMed  Google Scholar 

  11. Sharpe MB, Jaffray DA, Battista JJ, Munro P (1995) Extrafocal radiation: a unified approach to the prediction of beam penumbra and output factors for megavoltage x-ray beams. Med Phys 22:2065

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

I would like to thank Dr. Sergei Zavgorodni for kindly giving permission for using the data on backscatter measured by his team and for providing valuable comments on this manuscript. I want to express gratitude also to Ms. Debbie Wright for her valuable help in revising this document, and to my daughters Franca Nelli and Anna Nelli for their collaboration in producing some of the figures used here.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Flavio Enrico Nelli.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nelli, F.E. An MLC-based version for the ecliptic method for the determination of backscatter into the beam monitor chambers in photon beams of medical accelerators. Australas Phys Eng Sci Med 39, 221–228 (2016). https://doi.org/10.1007/s13246-015-0413-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13246-015-0413-y

Keywords

Navigation