Introduction

The earliest documented pregnancy hormonal signal is human chorionic gonadotropin (HCG). The blastocyst (fertilized egg) start to produce HCG before implantation, then the syncytiotrophoblast generate HCG exponentially after blastocyst implantation [1]. Blood HCG levels are detectable at considerable levels ten days after conception. Between weeks 10 and 11, the placenta generates the most HCG, and then the rate of production drops until week 12 when it stabilizes. HCG facilitates the corpus luteum's rescue and maintains progesterone production [2, 3].

HCG levels are elevated in all cases of gestational trophoblastic diseases (GTDs), making it an ideal tumor marker. GTDs are pregnancy disorders including a spectrum of diseases, ranging from the potentially premalignant hydatidiform mole to the highly aggressive choriocarcinoma. HCG is produced by all trophoblastic tumors, and monitoring of therapy is largely based on the determination of HCG in serum [4].

Total HCG in GTDs is the best example of a tumor marker because it has about 100% sensitivity and specificity for trophoblast malignant tissues. The amount of hydatidiform mole or tumor tissue is directly proportional to the blood level of total HCG. Hyperglycosylated HCG is a definitive tumor marker of invasion and malignancy in invasive moles and choriocarcinomas [4, 5]. High concentrations of HCG variants have been detected in human tumors that do not develop from trophoblastic tissue, including testicular, ovarian, bladder, kidney, lung, breast, and gynecological cancers [6]. These findings imply that HCG variants play a significant role in almost all human malignancies, and the production of specific antibodies against them can be exploited in cancer therapy [7].

The problem with automated laboratory HCG tests using an antibody to βHCG C-terminal peptide as the capture or tracer antibodies is that, it prevents detection of circulating HCG variants lack C-terminal peptide epitope. Using these antibodies severely restrict the specificity of the HCG assays [8]. To measure all HCG variants, the HCG-RIA only requires one polyclonal antibody to core β-subunit-1 or β-subunit-2. The GTDs Centre in the UK continues to monitor patients receiving treatment for GTDs using an in-house radioimmunoassay of this kind [9]. HCG-RIA has the advantage of detecting the major forms of HCG both in serum (i.e., HCG and βHCG) and in urine (HCG, HCGβ, and HCGβcf) in a fairly equimolar fashion [10].

In this study, purified HCG was optimized for labeling with radioactive iodine-125 to produce 125I-HCG tracer with high purity and specific activity, then used to prepare HCG standard solutions. All these reagents along with polyclonal β HCG antibodies can be used to develop a sensitive, precise, and accurate method for HCG estimation as a useful tool for pregnancy monitoring and diagnosis of GTDs.

Experimental

All the chemical reagents are of analytical grade. Chorionic gonadotropin human: C1063; Chorionic Gonadotropin, Human Urine, Standard Grade: 230734; Bovine serum, adult: B9433; Bovine Serum Albumin (BSA): A7888; Freund’s adjuvant complete (FAC): F-5881; Freund’s adjuvant incomplete (FAI): F-5506; polyethylene glycol 8000 (PEG): P4463; chloramine-T: C9887; sodium metabisulphite: S1516; potassium iodide: P2963; Sephadex G-25: G2580; sodium azide: S2002, were purchased from Sigma Chemical Co. USA of high quality and purity; sodium iodide –125 (125I Na) radioactive concentration 3700 MBq/ml was purchased from Izotop, Hungary. Microprotein determination kit: AGAPPE diagnostic, Switzerland GmbH. HCG Kit: AIA-PACK kit, TOSOH, Japan.

Preparation of HCG

HCG; was purified in our labs from pooled urine of pregnant women at the 10th week of gestation with high purity of 98.2% and biological activity (10,811 IU/mg) using alcohol precipitation, anion exchange, and gel chromatography according to our recently published method [11]. Polyclonal βHCG antisera; were produced from rabbits with high titer (1/104) and displacement (84.8%) using purified βHCG subunits according to our recently published method [11].

Preparation of 125I- HCG tracer

The immunologically active HCG molecule is composed of 237 amino acids containing seven tyrosine residues that can be radiolabeled using the optimized chloramine-T technique [12]. The labeling reaction was optimized by investigating some effects such as pH, Chloramine-T content, substrate (HCG) content, and the reaction time. The iodination reactions were carried out in the following sequence (in polystyrene eppendorf tube, 10 µl 0.25 M phosphate buffer pH 7.4, 10 µl HCG containing 2.0 µg (w/v), 10 µl Na125I (500 µCi), the reaction started by adding 5 µl chloramine- T solution containing 25 µg (w/v), then mixed by gentle vortexing for one minute, and the reaction was quenched by adding 6 µl sodium metabisulfite containing 30 µg (w/v), and 10 µl potassium iodide was added as a carrier) [13], then 3 µl was taken from the reaction mixture, and then applied on paper chromatography[14], and placed vertically in saturated jar contain 1 ml 70% methanol as a mobile phase. After approximately one hour, the mobile phase had reached the end line, the paper was allowed to air dry and spliced into 1 cm fractions, which were counted as (CPM) using a gamma counter. The iodination yield was calculated by measuring the radioactivity of all the fragments, dividing the radioactivity (CPM) at the spotting site (radiolabelled protein peak) by the total radioactivity (CPM) on the whole paper (the first peak and the second free iodide peak), and multiplying by 100.

$${\text{Iodination}}\,{\text{ yield}} \% = \frac{{{\text{Sum}}\,{\text{ of}}\, {\text{counts}} \,{\text{of}}\, {\text{radiolabelled}}\, {\text{protein}}\,{\text{ peak}} }}{{{\text{Sum}} \,{\text{of}}\, {\text{total}}\, {\text{counts}}}} \times 100$$
(1)

Factors affecting the iodination of HCG

  1. a.

    Effect of pH The pH value of the labeling reaction was optimized by studying the effect of different pH values (5.5, 6.5, 7.5, 8.5, and 9.5) of 0.25 M phosphate buffer, while all other parameters remained constant. After that, the iodination yield was calculated for each pH value.

  2. b.

    Effect of Chloramine T content The chloramine-T content and molar ratio of Chloramine-T to HCG content was optimized by studying the effect of different chloramine-T concentrations (10, 25, 50, 75, and 100 µg), while all other parameters remained constant. Then, the iodination yield was calculated for each oxidizing agent concentration.

  3. c.

    Effect of Substrate (HCG) content The optimum substrate (HCG) concentration was determined by studying the effect of different HCG concentrations (0.5, 1.0, 2.0, 4.0, and 8.0 µg), all other parameters were held constant. Then the iodination yield was calculated for each substrate concentration.

  4. d.

    Effect of Reaction Time The reaction time of labeling was optimized by repeating the labeling reaction at different times (15, 30, 60, 120, and 300 s), while all other reaction parameters remained fixed, and the iodination yield was calculated for each reaction time.

Purification of the radio-iodinated 125I-HCG tracer

Following labeling reaction optimization, the reaction was carried out at these optimum conditions, and the reaction mixture was immediately applied to a Sephadex-G25 column pre-equilibrated with assay buffer (0.05 M phosphate buffer pH 7.4 containing 0.4% BSA, 0.9% sodium chloride and 0.1% sodium azide) [15]. 0.5 ml fractions were collected at a flow rate of 0.5 ml/min and counted in units of a micro curie (μCi) by using a dose calibrator. In general protein peak is used for the specific activity check, relevance of second peak analysis.

Characterization of the tracer

The prepared 125I-HCG tracer was characterized in terms of radioiodination yield, radiochemical purity, specific activity (µCi/µg), immunoreactivity, and stability.

  1. (a)

    Radioiodination yield (%)

    Following purification of 125I-HCG tracer on a Sephadex-G25 column, radioiodination yield was calculated as a percent of activity (µCi) of labeled HCG (125I-HCG) fractions to the total radioactivity (µCi) of all fractions (125I-HCG + free iodide), multiplied by 100.

  2. (b)

    Radiochemical purity (%)

    The radiochemical purity of the purified 125I-HCG tracer was tested by the paper chromatography method using 70% methanol as a mobile phase. Then, the 1 cm fragments were counted using a gamma counter in units of (CPM). Radiochemical purity was measured as a percent of radioactivity (CPM) of labeled HCG (125I-HCG) to the total radioactivity counted.

  3. (c)

    Immunoreactivity

    Immunoreactivity of the prepared 125I-HCG tracer was tested in terms of maximum binding (B0%) and nonspecific binding (NSB %) at different dilutions of anti βHCG antisera.

  4. (d)

    Specific activity

    The specific activity of the 125I-HCG tracer, the radioactive iodine content per unit mass of substrate (HCG) was measured by multiplying the radioiodination yield (%) by the total activity(µCi) added in the labeling reaction, and dividing the result by the concentration(µg) of the substrate (HCG) added in the reaction mixture [16]. The specific activity of the prepared 125I-HCG tracer was expressed in units of (µCi /µg).

    $$\begin{aligned} {\text{Specific}}{\mkern 1mu} {\text{activity}} = & \left( \begin{gathered} {\text{Radioiodination}}{\mkern 1mu} {\text{yield}}\left( \% \right) \hfill \\ \times {\text{Total}}{\mkern 1mu} {\text{Activity}}{\mkern 1mu} {\text{added}}\left( {\mu Ci} \right) \hfill \\ \end{gathered} \right) \\ & \frac{1}{{{\text{Concentration}}{\mkern 1mu} {\text{of}}{\mkern 1mu} {\text{HCG}}\left( {\mu g} \right)}} \\ \end{aligned}$$
    (2)
  5. (e)

    Tracer stability

    The stability of the prepared 125I-HCG tracer was studied under different conditions (diluted and undiluted), both at different temperatures (− 20 °C, 4 °C and 25 °C). Stability was estimated by measuring the maximum binding (%B0) and non-specific binding (%NSB) through three months of storage period [17].

Preparation of HCG standard solutions

HCG standard grade (3rd IS 75/537) as a lyophilized powder from sigma Aldrich company (2500 IU/vial) was dissolved in 1250 μl of double distilled water containing 0.1% sodium azide and divided into aliquots stored at–20 ºC. The stock HCG solution was estimated in duplicate by commercial EIA (AIA360 automated enzyme immunoassay TOSOH), and the net concentration of stock HCG solution was measured to be 2000 IU/ml. Two sets of standard solutions of HCG were prepared, the first set of standard HCG was prepared in HCG-free bovine serum matrix and the second one was prepared in assay buffer matrix. HCG solution with concentration 4000 mIU/ml was prepared by diluting 2 μl of stock solution with 998 μl of diluting matrix [18]. Furthermore, concentrations (1, 5, 25, 50, 100, 250, 500, 1000, and 2000 mIU/ml) were prepared [19], HCG standards that prepared in both HCG free bovine serum matrix and assay buffer matrix were tested for binding with βHCG specific antisera, then stored refrigerated at 4 ºC or frozen for long time storage at -20 till use.

HCG-RIA optimization

The parameters of the radioimmunoassay system such as incubation time, temperature, sample volume, dilution of βHCG antisera, reaction volume, and separating agents (secondary antibody, non-immunized rabbit serum and PEG 8000), have been optimized by keeping all the parameters constant and focusing the study on each one of them, temperature (4, 25, and 37 °C), incubation time (1, 2, 3, 6, 12 h), sample volume (50,100,200, and 300 µl), reaction volume (300,500 and 1000 µl), secondary antibody (from 1:10 to 1:100), normal rabbit serum (1:100,1:200, 1:300, and 1:400), and PEG 8000 (4, 8, 12%). The optimum parameter is that with the highest binding and displacement percentages were measured.

Assay design

Following optimization, the HCG-RIA was formulated as follows: in a polystyrene test tube, add 100 µl sample or standard, 100 µl βHCG polyclonal antibodies at a dilution of 1:10,000, and 100 µl radiolabeled 125IHCG tracer at a radioactivity count of approximately 20,000 CPM/100 µl. After gentle vortexing and incubation for three hours at room temperature (25 °C), the following separating agents were added: 100 µl secondary antibody at dilution 1:50, 100 µl NRS (non-immunized rabbit serum) at dilution 1:100, and 500 µl PEG 8000 at concentration 6%. Again, gently vortex the tubes, incubate at RT (25 °C) for 30 min, and centrifuged at 4 °C, 5000 rpm for 15 min, the supernatant was discarded by decantation, the tubes were kept in the inverted position on a blotting paper/ tissue, and the tubes were counted for radioactivity as CPM using a gamma counter.

Calculation of the assay output results

After measuring, the total activity that was added to each tube and the bounded activity that counted after decantation. The data of the radioimmunoassay were represented as maximum binding (B0) for zero standard tubes, and known standard binding (Bs).

Standard curve representation

The standard curve for HCG-RIA was represented by plotting the %B/B0 for each HCG standard value (1, 5, 25, 50, 100, 250, 500, 1000, and 2000 mIU/ml) versus the concentration of individual HCG standards [20].

Validation study of HCG–RIA

Radioimmunoassay validation was achieved by measuring some quality control parameters, such as assay sensitivity, cross-reactivity with related glycoproteins FSH, LH, and TSH, intra-assay and inter-assay precision, and accuracy (recovery and dilution), and finally method comparison with a commercial kit.

Statistical methods

CVs and SD were calculated within and between runs as described by Krouwer and Rabinowitz 1984 [21]. The slope of the calibration curve was calculated after the logit-log transformation of % B/B0 as described by Rodbard 1974 [20].

Results and discussion

Optimization of the iodination reaction

HCG was labeled with iodine-125 using the optimized Chloramine T-method, at chloramine T concentration (25 μg), pH (7.5) of 0.25 M phosphate buffer, HCG content (2.0 μg), and reaction time (60 Sec.) the iodination yield obtained was 71.58 ± 0.92% as shown in Fig. 1 and the specific activity was 170.42 ± 1.65 µCi/µg. The purity of the prepared tracer was measured by paper chromatography method, and the radiochemical purity of the tracer calculated to be 99.2 ± 0.05% as shown in Fig. 2.

Fig. 1
figure 1

Radiochemical yield of 125I-HCG. The mixture of the HCG iodination reaction was applied on sephadex G-25 column, eluted with 0.25 M phosphate buffer pH 7.4 and 0.5 ml fractions were collected at flow rate 0.5 ml/min. The radioactivity in the fractions were counted as (µCi) by dose calibrator instrument

Fig. 2
figure 2

Radiochemical purity of 125I-HCG. 3 µl from the fraction no. 7 containing the radiolabelled HCG tracer was applied on paper chromatography, eluted with 70% methanol and the the 1 cm paper fragments were counted as (CPM) using the gamma counter

Characterization of the prepared 125I- HCG tracer

  1. -

    Radiochemical yield When the tracer was purified by gel filtration technique using a Sephadex-G25 column (PD-10), the radioactivity of the first peak represented 71.6% of the iodination process, as shown in Fig. 1.

  2. -

    Radiochemical purity When the tracer was analyzed by paper chromatography technique, the results represented in Fig. 2 showed that the purity of the prepared radio-iodinated HCG tracer is 99.2 ± 0.05%.

  3. -

    Specific activity The radioactivity content per mass unit of labeled HCG tracer was measured to be (170.42 ± 1.65 µCi/µg), indicating that; the prepared tracer with high specific activity can be used in RIA with a sharp radioactive signal that enhances assay sensitivity.

  4. -

    Immunoreactivity The immunoreactivity of the prepared HCG tracer was estimated by examining its binding with the specific anti βHCG polyclonal antisera. Figure 3 shows that the tracer prepared using the chloramine-T method had a high maximum binding percent (%B0 = 84.7%), a high displacement percent (D% = 84.8%), and a very low nonspecific binding percent (%NSB = 2.1%). These results indicate that the prepared tracer is immunologically active and had very specific antibody binding.

    Fig. 3
    figure 3

    Immunoreactivity check for the prepared 125I-HCG tracers

  5. -

    Tracer stability The results of tracer stability studies conducted over two months storage period at different temperatures under diluting and undiluted conditions showed that, the tracer is more stable when stored as a diluted solution in assay buffer matrix at − 20 °C over two months.

As shown in Table 1; the HCG tracer was partially deteriorated when stored undiluted at 4 °C and completely deteriorated when stored for two months at 25°C. On the other hand, when the tracer was preserved at − 20 °C, the binding of the tracer with antibody kept high while the nonspecific binding remained low through the storage period.

Table 1 Effect of storage temperature and dilution on the stability of 125I-HCG Tracer

Preparation of HCG standards

Two types of standards were prepared: The first in an assay buffer matrix, and the second in a bovine serum matrix.

Both standard sets were used in constructing standard curves for HCG-RIA, and the results shown in Fig. 4 demonstrated that the binding of antigen and antibody was greater when assay buffer was used as a matrix than bovine serum matrix.

Fig. 4
figure 4

Comparative study between the locally prepared HCG standards in assay buffer and HCG free bovine serum matrix

From the obtained results in Fig. 4, high binding was observed when assay buffer was used as a matrix for standard HCG preparation. The prepared HCG standard solutions in assay buffer matrix were selected and characterized by measuring each individual standard with commercial (HCG kit, TOSOH). The results of indigenous and commercial HCG standards were statistically analyzed for linear fit as shown in Fig. 5, the result of the correlation coefficient’r’ = 0.999 revealed a strong positive correlation between them.

Fig. 5
figure 5

Linear regression equation and correlation coefficient “r” between indigenous HCGstandards RIA and commercial HCG standard TOSOH

HCG-RIA optimization

The optimization results per each HCG-RIA parameter are listed below: The optimal reaction time was 3 h, and the binding and displacement percentages were 44.9 and 81.5%, respectively. The optimum temperature for HCG-RIA was 25 °C, with binding and displacement percentages of 44.2 and 80.1%, respectively. The optimal sample volume for HCG-RIA was 100 µl, with binding and displacement percentages of 45.0 and 82.4%, respectively. The optimal tracer radioactivity concentration for HCG-RIA was 20 × 103 CPM, with binding and displacement percentages of 41.3 and 79.6%, respectively. The optimal dilution for the goat anti rabbit secondary antibody in the HCG-RIA separation technique was 1:50, with binding and displacement of 50.6% and 84.2%, respectively. In non-immunized rabbit serum, a dilution of 1:100 produced the best results for HCG-RIA. The recommended PEG-8000 content for HCG-RIA was 6%, with binding and displacement percentages of 58.7% and 78.6%, respectively. Additionally, 300 µl of the HCG-RIA reaction volume yielded the greatest outcomes, with binding and displacement percentages of 43.0 and 79.6%, respectively.

Standard curve for HCG-RIA

The values of % B/B0 for each HCG standard were plotted against the concentration of HCG standards on a logit-log sheet. Figure 6 shows a linear inversed line, implying that as the concentration of HCG increases, the binding of radioactive HCG-tracer with the βHCG antibody decreases, and consequently the radioactive count decrease; from this curve, any unknown sample can be measured. The HCG-RIA range of measures (1, 5, 25, 50, 100, 250, 500, 1000, and 2000 mIU/ml) developed in-house was broad and can detect native HCG in samples up to 2000 mIU/ml.

Fig. 6
figure 6

Standard curve of HCG-RIA

HCG-RIA validation

According to Andreasson et al. 2015 [22], the results obtained from locally prepared HCG-RIA were validated by measuring the following statistical parameters:

Sensitivity, the lowest concentration of analyte (HCG) that has been demonstrated to be measurable with acceptable levels of precision and accuracy and can be distinguished statistically from zero standards [23], a detection limit of 0.5 mIU/ml was obtained by assaying 20 replicates of zero standards (Table 2) and the binding % B/B0 at 95% probability of true results by the assembly on HCG-RIA standard curve [24], the sensitivity obtained was excellent when compared with the commercial ELISA and EIA kits.

Table 2 The sensitivity of HCG-RIA

Specificity, the ability of the antibody to recognize an antigen even when the antigen is measured in a medium of similar substances that may be expected to be present [22]. The specificity was tested according to the method of Abraham [25], also called the 50% inhibition method, where specificity was calculated by dividing the concentration of HCG by the concentration of cross-reactant at % B/B0 equal to 50% to HCG standards, and the obtained specificity was 98.8% [11], indicating that the locally prepared polyclonal anti-βHCG specifically binds to HCG and has no cross-reactivity with related glycoproteins.

Precision is the degree to which separate test findings acquired under predetermined conditions accord closely [22]. Three pooled samples of human serum were used to calculate the mean of 10 duplicates in order to estimate the intra-assay (Within-Run) precision for HCG. The three levels of pooled human serum samples had coefficients of variation (% CVs) of 6.3, 6.4, and 5.7%. The statistical analyses were calculated for each sample and the % CVs were 8.3, 7.1, and 7.4% as shown in Table 3. The inter-assay (Run to Run) precision for HCG was derived from the mean of the average of duplicates from 10 independent runs with the same three pooled human sera. The % CV of the intra-assay (within assay variation) should be less than 10%, according to Pillai and Bhandarkar 1998 [26] and El-Kolaly et al. 2001[27], and the % CV of the measured ligand concentration in the case of the inter-assay (between assay variations) should be less than 15% [28]. Collectively the results show that the developed HCG RIA is with acceptable precision.

Table 3 Precision profile for HCG-RIA

Accuracy, the degree of agreement between the measured and true values; in this study, the assay accuracy was evaluated using recovery and dilution tests (parallelism).

Recovery test This test compares the concentrations of three samples of human serum from various diagnostic states before and after the addition of a known quantity of HCG. The current study's HCG-RIA recovery test results varied from 96.4 to 104.3%, as reported in Table 4.

Table 4 Recovery test for HCG-RIA

Dilution test The linearity of the assay was evaluated using the HCG content in three pooled human serum samples at varying dilutions. The outcomes demonstrated that the current study's method for HCG retained good linearity under dilution. The findings of the dilution test, which are displayed in Table 5, ranged from 91.1 to 107.6%. The recovery statistics from the current investigation for HCG are well-concordant with those reported by Pillai and Bhandarkar in 1998 [26] and Mehany et al. in 2007[29], who stated that the recovery in the RIA method should be 100 ± 15%.

Table 5 Dilution test for HCG-RIA

Comparison with reference method

The results of the statistical analysis of the linear correlation between the HCG levels of 50 serum samples measured commercially using the TOSOH immunoassay analyzer and the developed RIA system show that there is a strong linear positive correlation (r = 0.998) between the locally developed HCG-RIA technique and the commercial EIA technique, as shown in Fig. 7.

Fig. 7
figure 7

Linear correlation between results obtained by HCG-RIA method and EIA method

Conclusions

In this research article, purified HCG was radiolabelled with iodine-125 under optimal conditions; 125I-HCG tracer with high yield (71.58 ± 0.92%), high purity (99.2 ± 0.05%), and high specific activity (170.42 ± 1.65 µCi/µg) was obtained. Standard solutions for HCG were prepared and characterized. All HCG-RIA parameters were optimized, formulated, and then characterized by validity testing. The high sensitivity 0.5 mIU/ml of the developed HCG-RIA enables the detection of very low concentrations of HCG in human serum, and the obtained standard curve covers a wide range of detection up to 2000 mIU/ml, the precision and accuracy study revealed a low % CV. The problem of underestimation of HCG measured by automated HCG assays that use monoclonal antibodies for βHCG-CTP could be solved by using polyclonal βHCG antisera along with prepared 125I-HCG tracer and HCG standard solutions in the developed HCG-RIA system. Based on the above study a system can be developed for the estimation of the recently discovered hyperglycosylated-HCG isoform, an invasive molecule which may have a role in spontaneous abortion of pregnancy and the development of several types of human cancers.