Abstract
Support-vector machines (SVMs) can potentially improve patient monitoring during nitrous oxide anaesthesia. By elucidating the effects of low-dose nitrous oxide on the power spectra of multi-channel EEG recordings, we quantified the degree to which these effects generalise across participants. In this single-blind, cross-over study, 32-channel EEG was recorded from 12 healthy participants exposed to 0, 20, 30 and 40% end-tidal nitrous oxide. Features of the delta-, theta-, alpha- and beta-band power were used within a 12-fold, participant-wise cross-validation framework to train and test two SVMs: (1) binary SVM classifying EEG during 0 or 40% exposure (chance = 50%); (2) multi-class SVM classifying EEG during 0, 20, 30 or 40% exposure (chance = 25%). Both the binary (accuracy 92%) and the multi-class (accuracy 52%) SVMs classified EEG recordings at rates significantly better than chance (p < 0.001 and p = 0.01, respectively). To determine the relative importance of frequency band features for classification accuracy, we systematically removed features before re-training and re-testing the SVMs. This showed the relative importance of decreased delta power and the frontal region. SVM classification identified that the most important effects of nitrous oxide were found in the delta band in the frontal electrodes that was consistent between participants. Furthermore, support-vector classification of nitrous oxide dosage is a promising method that might be used to improve patient monitoring during nitrous oxide anaesthesia.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
1 Introduction
Nitrous oxide is a weak anaesthetic gas with hypnotic and analgesic properties, which has strong dissociative effects [1]. It is often used in isolation to induce a relaxed state and provide analgesia in dentistry, obstetrics, or acute trauma [2]. In the operating theatre, nitrous oxide may be combined with other anaesthetic vapours to achieve general anaesthesia [3].
Depth of anaesthesia monitoring is commonly used during surgery for patient safety [4]. Monitoring the anaesthetic depth reduces the risk of patient awareness intraoperatively, and also reduces recovery time, as it mitigates anaesthetic overdosing [5]. These monitors are based on characteristic alterations in electroencephalogram (EEG) recordings induced by anaesthetic drugs. Gamma-amino-butyric-acid (GABA)-ergic drugs are most commonly used for inducing unconsciousness. Therefore, most anaesthetic depth monitors are calibrated for GABA-ergic drugs, like propofol [4], but are insensitive to the effects of nitrous oxide [6,7,8,9]. This insensitivity is likely underpinned by different electrophysiological effects arising from targeting different receptors, as nitrous oxide primarily binds to the N-methyl-D-aspartate (NMDA) receptor [1].
Studies of alterations in the EEG power spectrum during nitrous oxide exposure have produced mixed results. An increase in power in high-frequency bands (beta and gamma) is often reported, with a relative decrease in power in the lower frequency bands (delta and alpha) [10]. This power reduction [11] is localised in the frontal region [12, 13] and most pronounced at higher exposures (i.e., 40 to 60%) [12, 14, 15]. Nitrous oxide also reduces functional connectivity in parietal (at 60% nitrous oxide) and frontal (16 to 30% nitrous oxide) regions [13, 16]. Our own group reported a reduction in temporal complexity in the central region that tracked cognitive impairment caused by nitrous oxide inhalation [17].
Machine-learning classification can analyse EEG recordings and is increasingly used in medicine and biology for behavioural and mental state monitoring [18,19,20]. To do so, labelled EEG recordings are represented in a derived feature space. This representation can be used, first, to infer which features are diagnostic (i.e., which features discriminate EEG recordings of one category from another) and, second, to assess the generalizability of diagnostic features across participants. Support-vector machines (SVMs) are one such approach to classification. SVMs fit a discriminant function to the derived feature space so as to maximise the margin separating EEG recordings from two categories [21]. SVMs offer several advantages to other classification approaches: they often generalise with good accuracy where high-dimensional data are concerned, and SVMs often yield readily interpretable results (i.e., the determination of the diagnostic features) [22], making them algorithmically explainable.
SVM performance relies heavily on the chosen EEG features. A common way of quantifying EEG is by calculating the power in pre-defined frequency bands (delta, theta, alpha and beta) [10]. These features are commonly used in EEG analysis for level of anaesthesia classification as they are easy to calculate, and have a physiological meaning as they are known to be related to certain brain states and to be driven by specific brain regions [23,24,25,26,27], making them biologically explainable. Besides frequency, entropy is also commonly used as a feature input for SVM modelling [24, 26, 28,29,30]. SVM classification of the depth of anaesthesia for all kinds of anaesthetics (excluding nitrous oxide) has been compared with other machine-learning algorithms, like random forest classification, regression, and artificial neural networks, with mixed results [26, 27, 29]. For this reason we will only focus on SVM analysis and will investigate the features selected by the SVMs to help explain the SVM algorithmically and biologically.
This study aimed to analyse previously recorded EEG [17] with SVMs to investigate their use to improve patient monitoring during nitrous oxide anaesthesia. By classifying the effects of low-dose nitrous oxide on the power spectra of multi-channel EEG recordings, we quantified the degree to which these effects generalise across participants.
2 Materials and methods
2.1 Trial design and participants
This secondary analysis study is based on multidose data collected in a single-blind, cross-over trial conducted at the Waikato Clinical School, University of Auckland, in July and August 2018. The original study aim was to expose healthy volunteers to nitrous oxide to develop a quantitative EEG analysis algorithm to detect cognitive impairment in scuba divers. This study was performed in line with the principles of the Declaration of Helsinki. The study protocol was approved by the Health and Disability Ethics Committee, Auckland, New Zealand (16/NTA/93), and was registered with the Australian New Zealand Clinical Trial Registry (U1111-1181-9722, RRID:SCR_002967).
Recruitment and inclusion and exclusion criteria are described in more detail elsewhere [17]. In summary, participants were healthy, certified divers aged between 18 and 60 years. Candidate participants currently using recreational drugs, tobacco, psychoactive medication, excessive alcohol, or over five caffeine-containing beverages a day were excluded. Participants had at least 6 h of sleep, abstained from any caffeinated drink on the day, fasted for 4 h and refrained from alcohol for 24 h prior. All participants provided written informed consent.
2.2 Experimental procedures
Twelve participants visited the facility once. Each experimental session began with a baseline measurement without nitrous oxide, followed by titrated gas mixtures to produce 20, 30 and 40% end-tidal nitrous oxide (in random order) using an anaesthetic circuit. There were 20-min breaks between the nitrous oxide exposures. Each exposure started with a 3–5 min wash-in until stable end-tidal nitrous oxide levels were achieved. The EEG was recorded directly after the wash-in period and at the end of the 10-min exposure with one minute eyes-open and one minute eyes-closed.
The EEG was recorded at 1024 Hz using a portable active electrode 32-channel system (ActiveTwo, BioSemi, Amsterdam, the Netherlands) divided over the scalp based on the international 10–10 system [31] with two additional electrooculogram electrodes placed under the eyes. The offset (impedance equivalent for active systems) was maintained below 25 µV with gelling (SignaGel, Parker Laboratories, Fairfield, NJ, USA).
2.3 EEG pre-processing and feature extraction
EEG pre-processing removed eye movement, muscle and noise artefacts using the Fieldtrip toolbox (version c6d58e9 RRID:SCR_004849) [32]. The data for the eyes-open condition at the end of each exposure was segmented from the continuous recording, re-referenced to the average, de-meaned, de-trended, and resampled to 256 Hz. Line (50 Hz, including higher harmonic) and low-frequency noise (< 1 Hz) were filtered out. Next, independent component analysis was used to filter noise components like eye blinks, high-frequency noise, non-physiological noise, and bad channels. The data were manually inspected for remaining artefacts. Whole two-second epochs containing artefacts were discarded (Matlab code is available at [17]).
Spectral power was estimated with a multitaper Fourier transformation implemented in the Fieldtrip toolbox using a Hanning window from 1 to 30 Hz with 1 Hz increments. The average power per frequency band was calculated for: delta (1–4 Hz), theta (4–8 Hz), alpha (8–14 Hz) and beta (14–30 Hz) for each exposure and channel as 128 features (4 frequency bands × 32 channels) for the SVM classification.
2.4 Initial exploratory analysis
At an early stage of our analysis, we wondered whether a traditional approach to classification via linear dimensionality reduction would provide insight into the nature of the recordings, specifically, whether the high-dimensional feature space (4 frequency bands × 32 channels) comprised a signal enabling classification. We adapted the approach described by Jamieson et al. [33], to explore whether a traditional approach to classification via linear dimensionality reduction would provide insight into the nature of the recordings. We applied principal component analysis to the high-dimensional representation (4 frequency bands × 32 channels) of EEG recordings during 0 and 40% nitrous oxide exposure. This low-dimensional representation of recordings (i.e., principal component scores) was used as input to a simple Fisher’s linear discriminant analysis (i.e., binary classification). We used 12-fold, leave-one-participant-out cross-validation for training and testing (on fold 1, we trained the linear discriminant analysis using data recorded from participants 1 through 11, and then tested the trained linear discriminant analysis using data recorded from participant 12; on fold 2, training used participants 1 through 10 and 12, and testing used participant 11; and so on). To evaluate the performance of the linear discriminant analysis, a confusion matrix was calculated for each leave-one-participant-out fold; overall accuracy was calculated by averaging accuracy across folds. The theoretical baseline of a binary classifier accuracy is 50% correct; we estimated an empirical baseline for our linear discriminant analysis using a shuffle test [34]. To do so, on each of 100 iterations, we repeated the above-described procedure, but first shuffled labels (“0% exposure” or “40% exposure”) on training data (not testing data). The empirical baseline was computed by averaging overall accuracy across iterations.
2.5 SVM classification
EEG recordings during 0 or 40% exposure were classified using an SVM (i.e., binary classification) and 12-fold, leave-one-participant-out cross-validation. We used a linear SVM because preliminary testing indicated a linear SVM’s performance was comparable to that of SVMs involving higher-order polynomial kernels or a Gaussian kernel. Other hyper parameters (box constraint, kernel scale and standardization) appeared to have no influence on model performance, and were left to the default. To evaluate the performance of the SVM, a confusion matrix was calculated for each leave-one-participant-out fold; overall accuracy was calculated by averaging accuracy across folds. We estimated an empirical baseline for the SVM using a shuffle test, similar to the method described in the initial exploratory analysis [34]. We also classified EEG recordings during 0, 20, 30 or 40% nitrous oxide exposure using a “one-versus-one” multi-class SVM model.
To gain a deeper understanding of which EEG features enabled classification, we repeated the binary and multi-class SVM analyses using "lesioned" datasets. Each lesioned dataset comprised only the features of one frequency band (e.g., alpha), or only the features of three frequency bands (e.g., delta, theta, and beta), thus quantifying the role of an individual frequency band (e.g., alpha) in classification performance. Similarly, EEG channels were divided into four regions (frontal, lateral, central and occipital). These regions were used to perform another lesioning scheme to explore the regional variation. The effects of "lesioning" on performance were visualised using violin plots [35].
2.6 Statistical analysis
Model performance is denoted as mean accuracy with a binomial 95%-confidence interval. We compared model performance (either binary or multi-class) of the complete model with the empirical baseline using a two-tailed, paired Wilcoxon signed test, as data were not normally distributed (Shapiro–Wilk test). If significant, we performed a further analysis with the complete model, the empirical baseline model and the lesioned dataset models (e.g., delta, theta, alpha, and beta, and data from three frequency bands) using a Friedman's test for inequivalence between the models. For post-hoc analysis, we used a Nemenyi test for pairwise comparisons of all models, as suggested by Demsar [36, 37]. Differences were regarded as significant at p < 0.05. All analyses were performed using Matlab version 2022a (Mathworks, Natick, MA, USA, RRID:SCR_001622) and are available in Online Appendix 1.
3 Results
In pre-processing, we removed on average 1.7 of every 30 two-second-long EEG samples due to artefacts (5.7%; average across participants). Most samples were removed due to muscle-related artefacts.
3.1 Initial exploratory analysis
The initial exploratory analysis indicated the presence of a signal for classification (Fig. 1). As expected, explained variance (red curve) increased with the number of principal components such that 95% and 99% of the variance was explained by 9 and 19 components. Classifier performance increased with the number of components (black curve), e.g., using 16 components enabled classifier performance of 83.3% correct. This performance was significantly better than chance.
3.2 Binary SVM model
The binary SVM model achieved an average classification accuracy of 92% (95%CI 73% to 99%) compared to the chance-level accuracy calculated with the shuffle test (49%) (p < 0.001). The confusion matrix shows the predicted versus true exposures (Fig. 2).
As expected, all frequency band lesioned models performed worse than the complete model (i.e., the model trained and tested on all frequency-band features; Fig. 3). However, the model containing only the delta-band features showed reasonably good accuracy (79%) compared to the complete model. In contrast, the models trained and tested using only theta-, alpha- or beta-band features showed much-reduced accuracy (54, 54 and 58%, respectively). The Friedman's test showed a statistically significant difference between the performance of models (X2(9) = 41.05, p < 0.001). Post-hoc analysis revealed that the models trained and tested, which did not contain the alpha- or theta-band (~ alpha and ~ theta) features, were significantly better than the chance-level accuracy (Fig. 4).
In most cases, the lesioning of electrode regions only marginally reduced the accuracy of the models. The exceptions are the model containing only the lateral electrode features (67%) and the model not containing the frontal electrode features (~ frontal; 75%; Fig. 5). The Friedman's test showed a significant difference between the performance of models (X2(9) = 37.47, p < 0.001). Post-hoc analysis revealed that the models trained and tested, which did not contain the lateral or occipital electrode (~ lateral and ~ occipital) features, were significantly better than the chance-level accuracy (Fig. 6).
3.3 Multi-class SVM model
The multi-class SVM model was significantly better, with an average classification accuracy of 52% (95%-CI 37% to 67%), compared to chance-level accuracy calculated with the shuffle test (26%, p = 0.01). The confusion matrix shows the predicted versus true exposures. Sensitivity (3/12 = 25%) was lowest for the 30% exposure; which the model tended to confuse with the 0 and 40% exposures (Fig. 7).
As expected, all frequency band lesioned multi-class models performed worse than the complete model (Fig. 8). The models trained and tested using only the theta- (25%) or alpha-band (27%) features performed worse than the complete model. The fold with 0% accuracy of the complete model is of the same participant as the 0% accuracy fold of the delta-only lesioning. The Friedman’s test showed a statistically significant difference between the performance of models (X2(9) = 25.43, p = 0.003). Post-hoc analysis revealed that only the model which did not contain the theta-band features (~ theta), was significantly better than chance-level accuracy based on the shuffle test (Fig. 9).
The regional lesioning of features only marginally reduced the accuracy, except for the model only containing the frontal electrode features (Fig. 10). Friedman's test and post-hoc analysis did not show statistically significant differences between the lesioned models (Fig. 11).
4 Discussion
This study aimed to investigate the use of SVMs to improve patient monitoring during nitrous oxide anaesthesia. By classifying whether low-dose nitrous oxide affected certain frequency bands of the power spectra of multi-channel EEG recordings, and whether these effects generalised across participants. We evaluated two SVMs: a binary SVM and a multi-class SVM. Both identified the same frequency bands and electrode regions. The binary model, which classified 0 and 40% nitrous oxide exposures, identified that the absence of the theta- or alpha-band or lateral or occipital electrodes did not significantly change the classification accuracy. The multi-class model classified 0, 20, 30 and 40% nitrous oxide exposure and found similar results for the theta band, i.e., the most important effects of nitrous oxide were found in both extremes of the frequency range (decreased power in the delta band and increased power in the beta band) in the frontal and central electrodes.
In the initial exploratory analyses, we showed that the principal component analysis with linear discriminant analysis method proved to have a viable signal in the dataset that could be classified. The subsequent SVM parameter optimisation showed that the linear model was comparable to the higher polynomial kernels. Given that the principal component analysis with linear discriminant analysis is also a linear method, it is not surprising that the SVM linear model had a similar performance.
The performance accuracy of the binary (92%) and multi-class (52%) SVMs is slightly lower than the accuracies of SVM models in other anaesthetics [24,25,26,27,28,29,30]. This could be because nitrous oxide produces less pronounced EEG effects. However, a more likely explanation is that most other studies classified over a larger dosage range, up to dosages that caused unconsciousness [23, 24, 28,29,30]. Our study only used low-dose nitrous oxide, enough to produce analgesia but without causing unconsciousness.
4.1 Frequency power lesioning
The frequency power analysis performed in the original study found a power increase in the beta band and a relative decrease in the delta band, most prominent in the frontal region and most visible in the 20% and 40% exposures [17]. These changes might help explain the frequency band and electrode regional lesioning results, which found that delta and beta in the frontal and central regions supported classification.
The alterations in the power spectra of EEG recordings are inconsistent with some previous descriptions of changes in EEG caused by increased dosage of nitrous oxide [10, 13] but consistent with others [11, 12, 38]. These differences might be due to variations in dosage, administration method, and exposure time or differences in the pre-processing of the EEG signals. The studies cited above have contributed to the debate about the importance of alpha-band power to characterise the effects of nitrous oxide. Our study has shown that features based on the alpha frequency band do not contribute to classification accuracy, but did illustrate the importance of the delta band.
4.2 Binary versus multi-class model
The binary and multi-class analysis indicated that the theta and alpha frequency bands are of no interest in distinguishing nitrous oxide dosage. Moreover, the multi-class model confirmed the unimportance of the theta-band. The difference between binary and multi-class analyses could be due to the increased complexity of the multi-class model because it needs to classify smaller differences. Compared to the binary model, which only had to classify between 0 and 40% nitrous oxide.
4.3 Strengths and limitations
Our study had two strengths in particular. First, the dataset contained four nearby levels of low-dose nitrous oxide, which is a challenging task to assign to any classification model. Secondly, the SVM machine-learning algorithm we chose is computationally fast and robust with high-dimensional data, making it easy to implement into a clinical monitor.
Our study also had some limitations. First, we recruited only 12 healthy participants, which may limit the external validity of our results (i.e., the ability of these results to generalise to other, larger cohorts with different health profiles). However, we note the good internal validity of our models; that is, our models performed well within a 12-fold, leave-one-participant-out cross-validation framework. Additional data and analysis might further improve the algorithm and confirm these findings. Second, the data analysis focused only on SVM, as it is a robust method with the benefit of minimising overfitting on the small available dataset. Furthermore, only power spectral features were used to train the SVM, as they are biologically explainable. Future research could extend into other machine learning algorithms and quantitative EEG metrics like entropy as features, as they could provide novel insights and be compared against each other. Third, we recorded EEG using 32 electrodes covering the whole brain, and recordings on all electrodes were incorporated into the model as a feature. To translate our findings into an anaesthetic monitoring device, the number of electrodes would need to be reduced to facilitate day-to-day use (e.g., similar to the bispectral index ‘BIS’ monitor). The regional lesioning indicated that the frontal and central regions are important, which is promising for future incorporation into a clinical monitor, as most commonly, only a few electrodes in the frontal region are used clinically. Fourth, we studied exposure to nitrous oxide only. This design allowed us to achieve one of our study aims, which is to isolate the effects of nitrous oxide on EEG recordings. However, to fully understand its translational potential, the model needs to be evaluated on clinical EEG recordings from patients who are administered nitrous oxide in combination with other more potent anaesthetic vapours, which is common clinical practice. Last, in our analysis, we used a combination of manual and automatic artefact rejection techniques before the SVM analysis. This is not possible in a real-time clinical monitor. Hence, the SVM analysis in real-time might be complicated by noise. Comparisons with other machine learning algorithms can include sensitivity to noise analysis.
5 Conclusion
This is the first study using a support-vector machine to classify nitrous oxide dosage. We report promising accuracies and generalizability across participants and have identified the delta band in the frontal region as a potentially useful component of multi-channel EEG power spectra. Support-vector classification of nitrous oxide dosage might enable patient monitoring during anaesthesia. This would require further analysis of mixed anaesthetic agent EEG data.
Data availability
The data supporting this study's findings are available from the corresponding author upon reasonable request. EEG analysis code is available online in previous publications.
References
Brown EN, Purdon PL, Van Dort CJ. General anesthesia and altered states of arousal: a systems neuroscience analysis. Annu Rev Neurosci. 2011;34:601–28. https://doi.org/10.1146/annurev-neuro-060909-153200.
O’Sullivan Í, Benger J. Nitrous oxide in emergency medicine. Emerg Med J. 2003;20:214–7. https://doi.org/10.1136/emj.20.3.214.
Myles PS, Leslie K, Silbert B, Paech MJ, Peyton P. A review of the risks and benefits of nitrous oxide in current anaesthetic practice. Anaesth Intensive Care. 2004;32:165–72. https://doi.org/10.1177/0310057X0403200202.
Marchant N, Sanders R, Sleigh J, Vanhaudenhuyse A, Bruno M-A, Brichant JF, Laureys S, Bonhomme V. How electroencephalography serves the anesthesiologist. Clin EEG Neurosci. 2014;45:22–32. https://doi.org/10.1177/1550059413509801.
Kaiser HA, Hight D, Avidan MS. A narrative review of electroencephalogram-based monitoring during cardiovascular surgery. Curr Opin Anesthesiol. 2020;33:92–100. https://doi.org/10.1097/ACO.0000000000000819.
Rampil IJ, Kim JS, Lenhardt R, Negishi C, Sessler DI. Bispectral EEG index during nitrous oxide administration. Anesthesiology. 1998;89:671–7. https://doi.org/10.1097/00000542-199809000-00017.
Barr G, Jakobsson JG, Öwall A, Anderson RE. Nitrous oxide does not alter bispectral index: Study with nitrous oxide as sole agent and as an adjunct to iv anaesthesia. Br J Anaesth. 1999;82:827–30. https://doi.org/10.1093/bja/82.6.827.
Anderson RE, Jakobsson JG. Entropy of EEG during anaesthetic induction: a comparative study with propofol or nitrous oxide as sole agent. Br J Anaesth. 2004;92:167–70. https://doi.org/10.1093/bja/aeh036.
Voss L, Sleigh J. Monitoring consciousness: the current status of EEG-based depth of anaesthesia monitors. Best Pract Res Clin Anaesthesiol. 2007;21:313–25. https://doi.org/10.1016/j.bpa.2007.04.003.
Purdon PL, Sampson A, Pavone KJ, Brown EN. Clinical electroencephalography for anesthesiologists. Anesthesiology. 2015;123:937–60. https://doi.org/10.1097/ALN.0000000000000841.
Foster BL, Liley DTJ. Nitrous oxide paradoxically modulates slow electroencephalogram oscillations: implications for anesthesia monitoring. Anesth Analg. 2011;113:758–65. https://doi.org/10.1213/ANE.0b013e318227b688.
Foster BL, Liley DTJ. Effects of nitrous oxide sedation on resting electroencephalogram topography. Clin Neurophysiol. 2013;124:417–23. https://doi.org/10.1016/j.clinph.2012.08.007.
Pelentritou A, Kuhlmann L, Cormack J, Mcguigan S, Woods W, Muthukumaraswamy S, Liley D. Source-level cortical power changes for xenon and nitrous oxide—induced reductions in consciousness in healthy male volunteers. Anesthesiology. 2020;132:1017–33. https://doi.org/10.1097/aln.0000000000003169.
Pelentritou A, Kuhlmann L, Cormack J, Woods W, Sleigh J, Liley D. Recording brain electromagnetic activity during the administration of the gaseous anesthetic agents xenon and nitrous oxide in healthy volunteers. J Vis Exp. 2018;131:e56881. https://doi.org/10.3791/56881.
Kuhlmann L, Liley DTJ. Assessing nitrous oxide effect using electroencephalographically-based depth of anesthesia measures cortical state and cortical input. J Clin Monit Comput. 2018;32:173–88. https://doi.org/10.1007/s10877-017-9978-1.
Kuhlmann L, Foster BL, Liley DTJ. Modulation of functional EEG networks by the NMDA antagonist nitrous oxide. PLoS ONE. 2013;8:e56434. https://doi.org/10.1371/journal.pone.0056434.
Vrijdag XCE, van Waart H, Mitchell SJ, Sleigh JW. An electroencephalogram metric of temporal complexity tracks psychometric impairment caused by low-dose nitrous oxide. Anesthesiology. 2021;134:202–18. https://doi.org/10.1097/ALN.0000000000003628.
Hosseini M-P, Hosseini A, Ahi K. A review on machine learning for EEG signal processing in bioengineering. IEEE Rev Biomed Eng. 2021;14:204–18. https://doi.org/10.1109/RBME.2020.2969915.
Müller K-R, Tangermann M, Dornhege G, Krauledat M, Curio G, Blankertz B. Machine learning for real-time single-trial EEG-analysis: from brain–computer interfacing to mental state monitoring. J Neurosci Methods. 2008;167:82–90. https://doi.org/10.1016/j.jneumeth.2007.09.022.
Saeidi M, Karwowski W, Farahani FV, Fiok K, Taiar R, Hancock PA, Al-Juaid A. Neural decoding of EEG signals with machine learning: a systematic review. Brain Sci. 2021;11:1525. https://doi.org/10.3390/brainsci11111525.
James G, Witten D, Hastie T, Tibshirani R. An introduction to statistical learning. New York: Springer; 2021.
Chorowski J, Wang J, Zurada JM. Review and performance comparison of SVM- and ELM-based classifiers. Neurocomputing. 2014;128:507–16. https://doi.org/10.1016/j.neucom.2013.08.009.
Casey CP, Tanabe S, Farahbakhsh Z, Parker M, Bo A, White M, Ballweg T, Mcintosh A, Filbey W, Saalmann Y, Pearce RA, Sanders RD. Distinct EEG signatures differentiate unconsciousness and disconnection during anaesthesia and sleep. Br J Anaesth. 2022;128:1006–18. https://doi.org/10.1016/j.bja.2022.01.010.
Shalbaf A, Shalbaf R, Saffar M, Sleigh J. Monitoring the level of hypnosis using a hierarchical SVM system. J Clin Monit Comput. 2020;34:331–8. https://doi.org/10.1007/s10877-019-00311-1.
Halder S, Juel BE, Nilsen AS, Raghavan LV, Storm JF. Changes in measures of consciousness during anaesthesia of one hemisphere (Wada test). Neuroimage. 2021;226:117566. https://doi.org/10.1016/j.neuroimage.2020.117566.
Tacke M, Kochs EF, Mueller M, Kramer S, Jordan D, Schneider G. Machine learning for a combined electroencephalographic anesthesia index to detect awareness under anesthesia. PLoS ONE. 2020;15:e0238249. https://doi.org/10.1371/journal.pone.0238249.
Nguyen-Ky T, Tuan HD, Savkin A, Do MN, Van NTT. Real-time EEG signal classification for monitoring and predicting the transition between different anaesthetic states. IEEE Trans Biomed Eng. 2021;68:1450–8. https://doi.org/10.1109/TBME.2021.3053019.
Ra JS, Li T, Li Y. A novel spectral entropy-based index for assessing the depth of anaesthesia. Brain Informatics. 2021;8:10. https://doi.org/10.1186/s40708-021-00130-8.
Liu Q, Ma L, Fan S-Z, Abbod MF, Shieh J-S. Sample entropy analysis for the estimating depth of anaesthesia through human EEG signal at different levels of unconsciousness during surgeries. PeerJ. 2018;6:e4817. https://doi.org/10.7717/peerj.4817.
Nicolaou N, Houris S, Alexandrou P, Georgiou J. Entropy measures for discrimination of awake vs anaesthetized state in recovery from general anesthesia. In: 2011 Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2011: pp. 2598–2601. https://doi.org/10.1109/IEMBS.2011.6090717
Klem GH, Lüders HO, Jasper HH, Elger C. The ten–twenty electrode system of the International Federation of Clinical Neurophysiology. Electroencephalogr Clin Neurophysiol. 1999;52:3–6.
Oostenveld R, Fries P, Maris E, Schoffelen J-M. FieldTrip: open source software for advanced analysis of MEG, EEG, and invasive electrophysiological data. Comput Intell Neurosci. 2011;2011:1–9. https://doi.org/10.1155/2011/156869.
Andrew R, Jamieson Maryellen L, Giger K, Drukker H, Li Y, Yuan N, Bhooshan (2010) Exploring nonlinear feature space dimension reduction and data representation in breast CADx with Laplacian eigenmaps and -SNE. Med Phys 37(1):339–351. https://doi.org/10.1118/1.3267037
Etienne C, Karim J (2015) Exceeding chance level by chance: The caveat of theoretical chance levels in brain signal classification and statistical assessment of decoding accuracy. J Neurosci Meth 250:126–136. https://doi.org/10.1016/j.jneumeth.2015.01.010
Bechtold B. Violin plots for Matlab. 2016. Github Proj. https://doi.org/10.5281/zenodo.4559847.
Demšar J. Statistical comparisons of classifiers over multiple data sets. J Mach Learn Res. 2006;7:1–30.
Cawley GC. nemenyi-test. https://github.com/iceshzc/nemenyi-test/blob/master/criticaldifference.m (2008). accessed 8 Aug 2022
Pavone KJ, Akeju O, Sampson AL, Ling K, Purdon PL, Brown EN. Nitrous oxide-induced slow and delta oscillations. Clin Neurophysiol. 2016;127:556–64. https://doi.org/10.1016/j.clinph.2015.06.001.
Acknowledgements
We are grateful to all participants that participated in this study. Furthermore, we would like to acknowledge the staff of the Department of Anaesthesia at the Waikato Clinical School for their support during the data collection. We thank Andreas Kempa-Liehr for his comments on an early version of the manuscript.
Funding
Open Access funding enabled and organized by CAUL and its Member Institutions. This work has been supported by funding from the Office of Naval Research Global (ONRG), United States Navy (Grant No. N62909-18-1-2007).
Author information
Authors and Affiliations
Contributions
XCV, LEH, EIT, HvW, SJM and JWS designed the study. XCV and HvW collected data. XCV, LEH, EIT and JWS analysed and interpreted the data. XCV and LEH drafted the manuscript and created the figures. All authors reviewed and revised the manuscript. All authors have approved the final version of the manuscript and agree to be accountable for all aspects of the work. All persons designated as authors qualify for authorship, and all those who qualify for authorship are listed.
Corresponding author
Ethics declarations
Conflict of interest
The authors have no relevant financial or non-financial interests to disclose.
Ethical approval
This study was performed in line with the principles of the Declaration of Helsinki. Approval was granted by the Health and Disability Ethics Committee, Auckland, New Zealand (8/9/2016, reference 16/NTA/93), and was registered with the Australian New Zealand Clinical Trial Registry (ANZCTR: U1111-1181-9722) on 12/03/2018. Written informed consent was obtained from all individual participants included in the study.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.