1 Introduction

Owing to their various advantages such as small size, easy fabrication, low cost, small sample volumes, and no need for a pump, microfluidic paper-based analytical devices (μPADs) have recently attracted the attention of researchers and industry (Martinez et al. 2007, 2010). They are particularly suited for point-of-care testing (POCT) applications. The advantage of POCT is that minimal effort and manipulation is required by the user. Various quantitative μPADs based on colorimetric, fluorescent, and electrochemical methods have been reported as POCT devices (Cheng et al. 2010; Nie et al. 2010; Mahato et al. 2017; Iwasaki et al. 2020). However, general immunoassays require multiple reagents to be sequentially introduced into the device, which increases the number of manipulation steps rendering the device unsuitable as a POCT. To solve this problem, several flow control methods have been reported (Li et al. 2008, 2013; Fu et al. 2011; Chen et al. 2012; Houghtaling et al. 2013; Shin et al. 2014; Toley et al. 2015). The method of sequentially injecting multiple buffers into the analysis region of a µPAD using the time delay caused by the difference in paper length and the ease of flow of the porous membrane has been widely adopted (Fu et al. 2011; Toley et al. 2013; Shin et al. 2014). Although this method allows multiple reagents that are simultaneously injected into different inlets to sequential flow to the analytical region without the need for additional equipment, the design of the μPADs requires complex calculations (Fu et al. 2011). A switchable valve that could control the opening and closing of the valve by connecting and disconnecting the separated papers in the middle of the paper channel was reported (Li et al. 2008, 2013, 2017; Giokas et al. 2014; Toley et al. 2015; Tu et al. 2021). However, these mechanical valves have a complicated three-dimensional structure and need to be manually switched and manipulated. Other interesting switches and valves have also been reported (Chen et al. 2012; Houghtaling et al. 2013). However, nearly all valves developed to date are passive, open and close only when a certain fluid level is reached, can be used only once, or require the dissolution of unnecessary reagents, such as surfactants; therefore, they are not practical for use in for complex immunoassays.

A controllable valve based on a polyvinylidene difluoride (PVDF) porous membrane functionalized with poly(N-isopropylacrylamide) (PNIPAAm), which is a thermoresponsive polymer that changes its hydrophilic properties at approximately 32 °C, has been reported (Iwasaki et al. 2019). PNIPAAm swells when it becomes hydrophilic at temperatures below 32 °C, and shrinks when it becomes hydrophobic at temperatures above 32 °C owing to dehydration. Therefore, at temperatures below 32 °C, the pores of the PVDF membrane are filled with swollen PNIPAAm; thus, the valve membrane closes to liquids and proteins (Fig. 1). In contrast, when the pores of the valve membrane are occupied by shrunken PNIPAAm, liquids and proteins can pass through the valve membrane. Further, with small amounts of PNIPAAm in the valve membrane, the pores are not completely filled and proteins may pass through, while excessive PNIPAAm in the valve membrane causes the pores to remain closed even at temperatures above 32 °C, resulting in an interrupted flow through the valve membrane. Therefore, the amount of PNIPAAm in the thermoresponsive valve is an important factor. We previously performed the free radical polymerization of PNIPAAm using N,N-methylenebisacrylamide (BIS), ammonium persulfate (APS), and N,N,N',N'-tetramethylethylenediamine (TEMED) (Iwasaki et al. 2019). Upon adding a drop of TEMED, polymerization rapidly progressed from the drop position through the entire monomer solution. However, the extent of polymerization was uneven, which may be attributed to the slow diffusion of TEMED in the porous membrane. On the other hand, in plasma-induced graft polymerization, radicals that initiate polymerization in a chain are homogeneously generated on the surface of the substrate by plasma treatment. Therefore, it is expected that uniformly polymerized PNIPAAm in a porous membrane is obtained by plasma-induced graft polymerization. Therefore, in this study, we performed plasma-induced graft polymerization of NIPAAm to functionalize the PVDF porous membranes, and obtain a good quality thermoresponsive valve membrane. In addition, we investigated the relationship between the amount of PNIPAAm and valve performance by varying the polymerization conditions, and evaluated the responsiveness of the valve by adding a dynamic temperature change.

Fig. 1
figure 1

Conceptual schematic of the thermoresponsive valve membrane. When the temperature is below the lower critical solution temperature (LCST; 32 °C), poly N-isopropylacrylamide (PNIPAAm) prevents the passage of proteins by swelling and filling the pores. When the temperature is above the LCST, proteins are able to flow through the pores because PNIPAAm shrinks owing to dehydration

2 Experimental

2.1 Reagents

Polyvinylidene fluoride (PVDF) membranes (Durapore® membrane filter, HVLP02500), nitrocellulose membranes (Hi-Flow™ Plus HF120), and absorbent pads (SureWick®) were purchased from Merck Millipore Ltd. (County Cork, Republic of Ireland). NIPAAm, BIS, and APS were purchased from Sigma-Aldrich (Tokyo, Japan). Polydimethylsiloxane (PDMS, SILPOT 184) was purchased from Dow Corning Toray Co., Ltd. (Tokyo, Japan). We used ultrapure water with a conductivity of 18.2 MΩ·cm in all experiments.

2.2 Preparation of a thermoresponsive valve membrane

Plasma-induced graft polymerization of NIPAAm onto PVDF membranes can be performed by immersing PVDF in NIPAAm solution after generating radicals on PVDF by plasma irradiation. Because the polymerization reaction is inhibited by oxygen, it is necessary to immerse the PVDF in the NIPAAm solution without exposing it to the atmosphere after plasma irradiation. We assembled a plasma-induced graft polymerization system that can perform these processes (Fig. 2). The system consists of a glass nipple, which is the plasma production area, a glass bell jar to hold the NIPAAM solution, a gate valve to separate the NIPAAM solution and plasmas, a vacuum pump series consisting of a turbo molecular pump and a diaphragm pump, an argon gas flow control system, a push rod to push out the PVDF membrane, and a radio frequency (RF) power supply to produce plasmas.

Fig. 2
figure 2

Photograph (a) and schematic (b) of the plasma graft polymerization apparatus. RF radio frequency power source, TMP turbo molecular pump, DP diaphragm pump, MFC mass flow controller

Using this system, a thermoresponsive valve membrane was prepared according to the following protocol: the NIPAAm monomers were purified by recrystallization from hexane, and the purified NIPAAm monomers were dissolved in a 50/50 vol% methanol/water solution. The concentration of the NIPAAm monomer ranged from 2.0 to 3.5 wt%. BIS and APS were dissolved in the NIPAAm solution with the respective molar ratios to the NIPAAm monomers of 0.5 and 1%. Then, the NIPAAm solution was poured into a glass bell jar. The PVDF membrane was placed in the glass nipple to ensure it was located at the center of the coil wire. The glass nipple and glass bell jar containing the PVDF membrane and NIPAAm solution, respectively, were evacuated with a diaphragm pump to degas the NIPAAm solution. Next, the gate valve connecting the glass nipple and the glass bell jar was closed, and the glass nipple was evacuated using a turbo molecular pump. Once the vacuum reached below 10−5 Torr, argon gas was introduced into the glass nipple at a flow rate of 30 sccm, and the PVDF membrane was subjected to 20 W inductively coupled plasma for 30 s. Then, the supply of argon gas and all vacuum pumping systems were stopped, the gate valve was opened, a push rod was used to push the PVDF membrane into the NIPAAm solution, followed by closing the gate valve. Then, the gate valve and the glass bell jar containing the mixture of the NIPAAm solution and PVDF membrane were removed from the apparatus and incubated at 30 °C to initiate NIPAAm polymerization. After incubation, the PVDF-PNIPAAm membranes were immersed in ultrapure water and washed by shaking for more than 3 h. The polymerization time (incubation period) was varied from 4 to 30 h, and the valve performance of each membrane was compared.

2.3 Membrane characterization

The surface of the PVDF-PNIPAAm membrane was characterized by Fourier-transform infrared spectroscopy in attenuated total reflectance mode (ATR-FTIR) with a germanium crystal (FT/IR-6300, Jasco Co., Japan) at room temperature. The sample surface was pressed onto the Ge crystal with the active surface. The FTIR spectra were acquired in the range of 4000–350 cm−1 using 16 scans at a resolution of 4 cm−1.

The microstructure of the thermoresponsive valve membrane was investigated using scanning electron microscopy (SEM; JSM-6701F, JEOL Ltd. Tokyo, Japan) in the open and closed states. The membranes were immersed in water at 25 and 40 °C overnight to ensure that the valves were in the closed and open states, respectively. The membranes were removed from the water, immediately frozen by dipping in liquid nitrogen, and then freeze-dried (FD-1000; Tokyo RIKAKIKAI Co., Ltd., Tokyo, Japan). Prior to SEM observation, the membranes in the open and closed states were coated with a thin gold film using a desktop quick coater (SC-701HMCII; Sanyu Electron Co., Ltd., Tokyo, Japan).

2.4 Evaluation of valve performance in paper microfluidic channel

We evaluated the valve performance of the fabricated thermoresponsive valve using a three-dimensionally stacked paper microfluidic channel (Fig. 3). Two nitrocellulose membranes (Bottom: 50 mm × 5 mm; top: 25 mm × 5 mm) (HF120, Merck Ltd.) were stacked in a T-shape on top of each other, with the thermoresponsive valve membrane (7 mm × 7 mm) sandwiched between the membranes at the junction. All membranes were cut to their respective sizes using a laser cutting machine (Blaster; Hotproceed Inc., Fukuoka, Japan). The stacked T-shaped paper microfluidic channel was covered with polymethylmethacrylate (PMMA) housings. The junction was gently and uniformly pressed using an elastic polydimethylsiloxane block with a torque wrench of 1 N cm to ensure contact for ease of flow (Iwasaki et al. 2015). One edge of the bottom nitrocellulose membrane was held in place using a PMMA jig with an inlet chamber. Absorbent pads were attached to the other edge of the bottom nitrocellulose membrane and the upper edge of the top nitrocellulose membrane. A small Peltier element was placed under the junction as a heating element. This Peltier element, controlled by a microcomputer (Arduino Due), can regulate the temperature of the valve. The open and closed states of the valve membrane can be verified by observing whether the green dye flows to the top nitrocellulose membrane, which is oriented horizontally.

Fig. 3
figure 3

Schematic of the experimental setup for the valve performance test

A green dye was used to visualize the flow in the microfluidic channel. First, 50 µL of the green dye solution was injected into the inlet chamber of the bottom nitrocellulose membrane (Fig. 3). Once the inlet chamber was empty, an additional 50 µL of green dye solution was injected. The valve membrane was heated and cooled after a certain time following the injection of the green dye, and the opening and closing behavior of the valve was evaluated.

3 Results and discussion

3.1 SEM observation

A thermoresponsive valve membrane fabricated by polymerization for 20 h with 2.0 wt% NIPAAm monomer and an untreated PVDF membrane were observed by SEM (Fig. 4). The PVDF-PNIPAAm membrane showed polymer adhesion on its surface when compared with the untreated PVDF membrane. The microstructures of open and closed thermoresponsive valve membranes prepared using different polymerization times were also observed by SEM (Fig. 5). For the polymerization time of 15 h, several uniformly distributed pores at both 20 and 40 °C were observed in the membrane. It was also confirmed that the number of pores filled with PNIPAAm proportionally increased with the polymerization time at 20 °C, while they remained open (unfilled) at 40 °C regardless of polymerization time (15, 20, and 30 h). These results show that the uniform polymerization of NIPAAm on the porous membranes was successful, which was difficult to achieve by free radical polymerization in our previous work (Iwasaki et al. 2019). Further, the closing and opening of the pores by swelling and shrinking, respectively, was observed.

Fig. 4
figure 4

Scanning electron microscopy images of an untreated PVDF membrane (a) and a valve membrane fabricated by polymerization for 20 h with 2.0 wt% NIPAAm monomer (b)

Fig. 5
figure 5

Scanning electron microscopy images of the valve membranes polymerized with 2 wt% NIPAAm monomer concentration for different polymerization times (15, 20, and 30 h) at swelling (20 °C) and shrinking (40 °C) states. Scale bars represent 10 µm

3.2 ATR-FTIR spectroscopy

The results of the ATR-FTIR measurements of a PNIPAAm sheet and the PVDF membranes before and after graft polymerization for 30 h at 2 wt% NIPAAm monomer concentration are shown in Fig. 6. PNIPAAm sheets were prepared using the method described in the Supporting Information (SI) for FTIR spectral comparison. Absorption peaks at 835, 878, 1173, 1233, and 1400 cm−1 were observed for both membranes, which correspond to the CF2 symmetric stretching, CH2 bending, CF2 antisymmetric stretching, CH stretching, and CH2 bending, respectively. The PVDF-PNIPAAm membrane showed new absorption peaks at 1456, 1540 and 1645 cm−1, which correspond to CH2 bending, N–H stretching and the second amide C=O stretching of the O=C–NH groups in the PNIPAAm chain (Hirashima et al. 2005; Yu et al. 2011; Xiao et al. 2014), respectively. This result shows that NIPAAm was successfully polymerized on the PVDF membranes.

Fig. 6
figure 6

Fourier transform infrared spectra of the PVDF membrane, PNIPAAm, and the PVDF-PNIPAAm membrane polymerized with 2 wt% NIPAAm monomer concentration for 30 h

3.3 Valve function of thermoresponsive valve membrane

We compared the behaviors of the opening action of valve membranes prepared using different polymerization times. The Peltier element was heated to 35 °C 240 s after the first injection of the green dye. The thermoresponsive valve membrane showed three patterns of valve performance depending on the polymerization conditions of PNIPAAm. Time-lapse images of the typical flow state for each pattern obtained using the valve membranes fabricated with different polymerization times at 2 wt% NIPAAm monomer concentration are shown in Fig. 7. Videos of these experiments are also available in the SI. The green contrast in Fig. 7 and the videos noticeably increased and became quite visible. For the membrane with a polymerization time of 4 h (Case 1), the green dye passed through the valve membrane and flowed into the top membrane in the horizontal direction prior to heating because the valve membrane remained open even at room temperature. For the polymerization time of 22 h (Case 2), the green dye did not pass through the valve membrane before heating. Approximately 20 s after heating was initiated, the green dye began to flow into the top membrane. Finally, the green dye reached the absorbent pad located at the right edge of the top membrane 240 s after heating was initiated, indicating a good valving performance. For the longer polymerization time of 30 h (Case 3), the green dye did not pass through the valve membrane before heating, as in Case 2. However, even after 240 s of heating, there was no sign of the green dye permeating the valve membrane and flowing into the top membrane. The green dye was only observed to flow to the right edge of the top membrane after 1200 s of heating.

Fig. 7
figure 7

Time-lapse photographs of the green dye flowing through a T-shaped paper microfluidic channel in three typical examples of valve membranes fabricated with 2 wt% NIPAAm monomer concentration with different polymerization times (4, 22, 30 h). The valve membrane was heated from room temperature (RT) to 35 °C 240 s after heating was initiated using a Peltier element placed directly under the valve section (Fig. 1). The open and closed states of the valve membrane can be verified by observing whether the green dye flows to the top nitrocellulose membrane, which is oriented horizontally. In this image, the green contrast noticeably increased and became quite visible

The results of the valve performance test of all the valve membranes and the weight of PNIPAAm obtained from the weight difference of the membranes measured before and after graft polymerization are plotted in Fig. 8. The results of the valve performance test are divided into cases 1, 2, and 3, and are plotted with a cross, circle, and triangle markers, respectively. Since the weight change before and after graft polymerization is minimal, only the PNIPAAm weight should be taken as a reference. Although there was some variation in the PNIPAAm weight and valve performance test results, the weight tended to increase with increasing polymerization time. As for the valve performance, the short polymerization time tended to be case 1, the polymerization time from 20 to 22 h tended to be case 2, and the polymerization time of 30 h was case 3. In addition, the extent of PNIPAAm polymerization also increased depending on the monomer concentration. With regard to monomer concentration, less than 2.2 wt% of NIPAAm monomer concentration exhibited the case 1 scenario, 2.3–3.0 wt% exhibited the case 2 scenario, and 3.5 wt% exhibited the case 3 scenario. These results indicate that the amount of PNIPAAm can be controlled by the polymerization time and monomer concentration. Further, with small amounts of PNIPAAm, the valve cannot be closed, while with large amounts, poor flow is observed despite an open valve, as expected.

Fig. 8
figure 8

Relationship between the weight of PNIPAAm functionalized on the porous membrane and a polymerization time at 2.0 wt% NIPAAm monomer concentration, and b NIPAAm monomer concentration at 5 h polymerization time. The results of the valve function test are represented by the shape of the marker; case 1 (cross): the valve did not close even at room temperature; case 2 (circle): the valve showed a good valving performance; case 3 (triangle): poor permeability was observed after valve opening

In comparison with the results of the valve performance test and SEM observation, for the polymerization time of 15 h, it can be understood that the flow could not be stopped because the pores were barely closed. Although only some pores were closed in the case of 20 and 30 h, they were sufficient to stop the flow. At 40 °C, the pores were open in both cases (20 and 30 h polymerization time); however, at 30 h, the pores seemed to be filled more. This slight difference may have prevented a free flow during heating. Thus, it is considered that the functioning of the valve is not dependent on whether the pores are completely filled, but rather on the extent of open valves. Although it seems difficult to control the degree of opening, the optimal valve performance was obtained with a monomer concentration in the range 2.3–3.0 wt% forr 5 h of polymerization.

We further evaluated whether the valve could be closed from an open state using the same T-type microfluidic channel. In this experiment, the PVDF-PNIPAAm membrane polymerized for 5 h with 2.5 wt% NIPAAm monomer was used. Heating was initiated 180 s after the injection of the green dye, cooling was initiated at 330 s, and heating was performed once more at 540 s (Fig. 9). A video of this experiment is available in the SI. In this figure, the green contrast remarkably increased and became quite visible. The green dye could not penetrate the valve membrane and only flowed to the lower channel. However, when the valve was heated, the dye permeated the valve membrane and flowed to the upper channel (in the horizontal direction). When the valve membrane was cooled, the lateral flow ceased. The flow remained stopped as long as the valve was maintained below the LCST. When the valve was reheated, the dye flow restarted. These results show that the temperature-responsive valve could be opened and closed reversibly by temperature control.

Fig. 9
figure 9

Time-lapse photograph of the green dye flowing through a T-shaped paper microfluidic channel using valve membranes fabricated by polymerization for 5 h with 2.5 wt% NIPAAm monomer. The valve membrane was heated from room temperature (RT) to 35 °C 180 s after the green dye was injected, cooled from 330 to 540 s, and then heated once more from 540 to 660 s. In this image, the green contrast noticeably increased and became quite visible

4 Conclusions

In the present study, we investigated plasma-induced graft polymerization as a method to fabricate a thermoresponsive valve for µPADs by surface treatment of a PVDF porous membrane with PNIPAAm. The apparatus for plasma-induced graft polymerization was fabricated, and the polymerization conditions were varied. Surface analysis by ATR-FTIR confirmed that the fabricated thermoresponsive valves were successfully coated with PNIPAAm. SEM observation confirmed that the pores of the valve membranes opened and closed in response to temperature and that the number of pores filled with PNIPAAm proportionally increased with the polymerization time. Subsequently, a valve performance test was conducted to determine the conditions required for an optimal valve performance. The thermoresponsive valve fabricated by polymerization with monomer concentration ranging from 2.3 to 3.0 wt% for 5 h of polymerization or 2.0 wt% for 20–22 h of polymerization showed a good valve performance; they stopped the flow at room temperature and opened to allow the flow within 20 s after heating was initiated using a Peltier element located just under the valve. We also demonstrated that this temperature-responsive valve can be reversibly opened and closed via repeated valve heating/cooling experiments. Although a detailed evaluation of the fluid behavior needs to be conducted, we showed that the thermoresponsive valve fabricated by plasma-induced graft polymerization may potentially be used to control the flow of multiple reagents in µPADs. Future investigations may be conducted to demonstrate the effectiveness of our thermoresponsive valve using complex analytical techniques, such as immunoassays with paper analysis chips.