Abstract
Objective
To develop two spiral-based bSSFP pulse sequences combined with L + S reconstruction for accelerated ungated, free-breathing dynamic cardiac imaging at 1.5 T.
Materials and methods
Tiny golden angle rotated spiral-out and spiral-in/out bSSFP sequences combined with view-sharing (VS), compressed sensing (CS), and low-rank plus sparse (L + S) reconstruction were evaluated and compared via simulation and in vivo dynamic cardiac imaging studies. The proposed methods were then validated against the standard cine, in terms of quantitative image assessment and qualitative quality rating.
Results
The L + S method yielded the least residual artifacts and the best image sharpness among the three methods. Both spiral cine techniques showed clinically diagnostic images (score > 3). Compared to standard cine, there were significant differences in global image quality and edge sharpness for spiral cine techniques, while there was significant difference in image contrast for the spiral-out cine but no significant difference for the spiral-in/out cine. There was good agreement in left ventricular ejection fraction for both the spiral-out cine (− 1.6 \(\pm\) 3.1%) and spiral-in/out cine (− 1.5 \(\pm\) 2.8%) against standard cine.
Discussion
Compared to the time-consuming standard cine (~ 5 min) which requires ECG-gating and breath-holds, the proposed spiral bSSFP sequences achieved ungated, free-breathing cardiac movies at a similar spatial (1.5 × 1.5 × 8 mm3) and temporal resolution (36 ms) per slice for whole heart coverage (10–15 slices) within 45 s, suggesting the clinical potential for improved patient comfort or for imaging patients with arrhythmias or who cannot hold their breath.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In clinical practice, cardiac magnetic resonance images are often acquired during a breath-hold using an electrocardiography-gated (ECG-gated) balanced steady-state free precession (bSSFP) pulse sequence with segmented Cartesian readouts [1]. While producing images with high image quality and spatiotemporal resolution, this conventional method is time-consuming, typically taking 5–6 min to cover the whole left ventricle, and may be contaminated by ECG-gating and breathing artifacts. Thus, dynamic imaging that does not rely on cardiac gating or breath-holding may be advantageous for patients with arrhythmias or who have difficulties in holding their breath [2,3,4].
Compressed sensing (CS) [5] has been widely used to accelerate data acquisition in dynamic imaging. By exploiting extensive spatiotemporal data redundancy, this technique allows highly undersampled MRI data to be reconstructed without loss of information. Moreover, CS can be extended to further increase imaging speed, for example, by combining it with parallel imaging [6, 7] (PI) to exploit joint sparsity across images captured by multiple coils (e.g., k-t SPARSE-SENSE [8,9,10]), or by combining it with low-rank matrix completion to enable reconstruction of a matrix with missing entries under low-rank and incoherent conditions (e.g., L&S [11], L + S [12]). To date, the L + S model has been investigated in several applications, including cardiac imaging [12, 13], speech imaging [14], and temperature imaging [15], for its ability to achieve good image quality with high acceleration rates.
Since CS-based methods require a sparse representation, incoherent aliasing artifacts are often generated via random under-sampling of Cartesian k-space or the use of non-Cartesian sampling patterns [5]. The achievable incoherence from a 2D Cartesian k-space trajectory, however, is relatively low when compared to that from non-Cartesian trajectories. Therefore, non-Cartesian bSSFP sequences, such as radial- or spiral-based trajectories, may improve dynamic cardiac MRI using intrinsic variable density trajectories. Several investigators have demonstrated that radial [10, 16,17,18] and spiral-out [19,20,21] sampling patterns inherently generate incoherent aliasing artifacts and can achieve high acceleration capability when combined with advanced acceleration methods, both due to their advantageous time-efficiency and inherent robustness to flow and motion artifacts.
In this work, two optimized spiral-based (spiral-out and spiral-in/out) bSSFP pulse sequences were designed and combined with the reconstruction methods to produce highly accelerated cardiac MRI under ungated free-breathing conditions [22, 23]. A comparison of these combinations was then conducted in normal volunteers. Finally, the proposed methods were validated against a standard ECG-gated breath-hold cine sequence under whole heart coverage short-axis views.
Methods
Pulse sequence design
Figure 1a shows the timing diagram of dynamic spiral-based cine sequences within a single scan. Two low spatial resolution datasets were acquired before the bSSFP module using spectral-spatial [24] RF water excitation pulses and two identical, single-shot fully sampled spiral-out arms (green boxes) with two TEs (\(\Delta TE=1 ms\)). A field map with 9 × 9 mm2 in-plane resolution was then generated, followed by means of the maximum likelihood algorithm to estimate a linear map \(f(x,y)\), using the equation below,
where \({f}_{0}\) is the center frequency, and \(\alpha ,\beta\) are the gradients along x and y axis, respectively. Coil sensitivity maps were also calculated from this fully sampled field map dataset using ESPIRiT [25]. Dummy cycles (yellow boxes) with a total number of 100 TRs were then used to approach steady-state magnetization [26]. Immediately after the pre-scan, dynamic data with a total of 320 spiral arms per slice were collected using a tiny golden angle rotation (blue boxes). Interleaved, rotated spiral-out and spiral-in/out readouts were both evaluated in this study.
For spiral-out gradients, we generated a spiral-out arm constrained by the gradient amplitudes and slew rates in a minimum time [27], along with the following rewinder gradients with zeroth and first-order moments nulling via triangular gradients [20], as shown in Fig. 1d. The rewinder gradient was designed to overlap with the next slice-selection prewinder to shorten the minimum TR. The resulting spiral-out readout was 1.28 ms in length. Figure 1e shows the gradients for a spiral-in/out readout using the algorithm described in Ref. 19. Based on the symmetry of the readout (the preceding time-reversed spiral-in arm and the following spiral-out arm), the spiral-in/out prewinder and rewinder waveforms were designed to overlap with the current slice-selection rewinder and the next slice-selection prewinder gradients, respectively. This property enables a shorter TR or a longer spiral readout compared to the spiral-out trajectory in which the overlap is only feasible on one side. Here, we fixed the TR and chose a longer spiral-in/out readout in our experiments. The resulting spiral-in/out readout was 2.04 ms in length.
A variable sampling density spiral (VDS) scheme using dual-density from 1.5 × to 0.3 × Nyquist was designed for the spiral-out trajectories and from 2.3 × to 0.4 × Nyquist for the spiral-in/out trajectories, both containing a total of 64 arms. The radius of the density transition was set to be one third of the k-space radius. Among interleaves, tiny golden ratio angle increments were used for both spiral-out and spiral-in/out cines to reduce phase disturbances induced by eddy currents in bSSFP acquisition. Tiny golden angles were defined using the golden ratio \(\tau =(1+\surd 5)/2\) and the sequence
for symmetric center-through trajectories such as spiral-in/out [28]. For asymmetric trajectories such as spiral-out, the angle sequence was
Here we used the golden angle of 23.63 \(^\circ\) and 47.26 \(^\circ\) for the spiral-in/out readout and the spiral-out readout, respectively. Eight sequential arms were combined to reconstruct each cardiac frame with a nominal acceleration ratio of eight. The actual acceleration ratios, after considering the variable dual-density sampling, were ~ 5.3 × and ~ 3.5 × at the center of k-space and ~ 26.7 × and ~ 20 × at the outer of k-space for spiral-out and spiral-in/out, respectively. The number of slices varies from 10 to 15 depending on the size of the heart. The total acquisition time was 2–3 s per slice in vivo experiments. Other sequence parameters were given in Table 1.
Analysis of point spread functions
Point spread functions (PSFs) of spiral-out and spiral-in/out sampling patterns were both computed for each time frame with eight spiral arms per frame. The maximum intensity projection (t-MIP) [29] over the time domain was computed from each series of a total of 80 PSFs. The sidelobe-to-peak ratio (SPR) [5] was then calculated for each trajectory using Eq. [4] as follows:
System imperfections
In spiral imaging, k-space infidelity may cause image blurring or distortion [19, 30, 31]. In this paper, anisotropic gradient delays and eddy current coefficients were calculated for each axis to estimate the actual k-space trajectories, for both the spiral-out and spiral-in/out readouts, using a model-based trajectory estimation method [19, 31]. Off-resonance effects may also cause severe banding artifacts in bSSFP data acquisition and image blurring during the spiral readout. In this work, a vendor provided cardiac shim package was used to improve field homogeneity, and time-optimized gradients were designed to achieve a TR of less than 4.5 ms for both spiral bSSFP cine sequences to reduce the banding artifacts. Furthermore, a linear off-resonance correction which was estimated from the acquired low-resolution B0 map was performed to compensate for the phase error during the readout [32].
Image reconstruction
The image reconstruction pipeline is shown in Fig. 1b, c. The L + S reconstruction was implemented using Eq. [5] as follows:
where E was a multi-coil encoding operator, including a non-uniform FFT (NUFFT) kernel [33, 34] and the coil sensitivity maps. L and S were the low-rank and sparse components, respectively. d was the undersampled spiral data, and T was a temporal total-variation (TTV) operator as the sparsifying transform [12]. The regularization parameters for the low-rank term (\({\lambda }_{L}\)) and the sparse term (\({\lambda }_{S}\)) were adjustable parameters to weight these terms (L and S) relative to data consistency, which were determined by comparing the resulting image quality for a series of testing values. For in vivo datasets, parameters were chosen based on the qualitative assessment of the residual aliasing artifacts and visual temporal blurring (see Supporting Fig. 1).
For comparison, the CS method was implemented by a similar methodology for parameter selection and by directly enforcing the sparsity on the original matrix, using the following Eq. [6]:
where M was the cardiac image series to be reconstructed in x–y-t space, and T was the TTV operator used as the only regularization parameter. A view-sharing (VS) reconstruction method was also used for reconstruction comparisons [35].
All computations were performed offline in MATLAB (R2022b software; MathWorks, Natick, MA). The NUFFT operator was utilized for direct spiral image reconstruction. A proximal optimized gradient method (POGM) [36] was used for fast L + S iterative convergence, while a nonlinear iterative conjugate gradient algorithm was used for solving CS optimization problem. The parameters \({\lambda }_{L}\) of 0.05 and \({\lambda }_{S}\) of 0.0005 were determined for the in vivo L + S images, while a \({\lambda }_{S}\) of 0.0006 was chosen to reconstruct the in vivo CS images.
Simulation
Simulations were performed using the digital cardiac phantom MRXCAT [37]. A series of fully sampled complex-valued cardiac frames were retrospectively undersampled using two aforementioned spiral-based under-sampling patterns (both with eight-arms). Simulation parameters included: image matrix 220 \(\times\) 220, 80 frames under free-breathing condition, 16 coils, Gaussian noise with a SNR of 15 db. Image reconstruction was performed using multi-coil CS, and L + S with regularization parameters selected by evaluating the normalized root mean square error (NRMSE) and structural similarity index (SSIM) [38]. NRMSE and SSIM values were then calculated for both of these two spiral trajectories along with reconstruction methods, with the fully sampled image as a reference.
In vivo experiments
All in vivo studies were performed on a 1.5 T scanner (MAGNETOM Avanto, Siemens Healthcare, Erlangen, Germany) with a 32-channel anterior–posterior surface coil array. Seven healthy volunteers (five males and two females) participated in this work with written informed consent prior to the experiment. Whole-heart coverage short-axis views were imaged by these two spiral cine sequences under an ungated, free-breathing condition, and by the standard cine technique during several breath-holds as a reference. For each set of experiments, the three pulse sequences were run consecutively at the same image plane with the same cardiac shimming settings. To be noted, the standard cine was performed using a retrospective ECG-gated bSSFP Cartesian sequence. FOV was set to 330 × 260 mm2, with a 40 ms temporal resolution and 1.50 × 1.50 × 8 mm3 spatial resolution. Vendor-provided GRAPPA was used for two-fold acceleration, resulting in a total scan time (also referred to as breath-hold time) of 10–12 s per slice. A set of 25 cardiac frames were reconstructed and interpolated based on the ECG signal for each slice. Other sequence parameters can be found in Table 1.
Image assessment
A blind rating of the diagnostic quality of cine datasets (n = 7) was performed by two cardiologists, with a 1 (nondiagnostic)–5 (excellent) scale. Differences in qualitative image quality ratings for dynamic spiral bSSFP cines along with image reconstruction methods were analyzed, followed by pairwise comparisons using the Wilcoxon signed-rank test. Furthermore, to validate the proposed two spiral cine sequences with the L + S reconstruction method against the standard breath-hold Cartesian cine sequence, short-axis movies from all the subjects were then rated and compared, as well.
Measurement of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in images reconstructed using CS-based methods is not straightforward because of the uneven distribution of noise. Thus, a quantitative calculation of image contrast between the blood pool and the myocardium was performed, and a method similar to that described in Ref 17 and 21 was used to measure the average blood pool and myocardial signal intensity in the end-diastolic midventricular image for each subject. The maximum gradient of signal intensities across the border of the left ventricular (LV) septum was also measured to calculate the edge sharpness (mm−1) in every cardiac phase of the midventricular orientation. Differences in image contrast and mean edge sharpness for the two spiral cine techniques with L + S reconstruction and the standard cine sequence were analyzed, and pairwise comparisons were then implemented via the Wilcoxon signed-rank test.
Ejection fraction calculation
The left ventricular end-diastolic volume (LV EDV) and end-systolic volume (LV ESV) were measured, and the left ventricular ejection fraction (LV EF) was then calculated for the dynamic spiral cines and the standard cine. The agreements in LV EF between the proposed methods and the reference were calculated using a Bland–Altman analysis.
Results
Analysis of PSFs over time
Figure 1f shows the signal intensities across a line over the central region of the t-MIPs from the spiral-out cine and spiral-in/out cine. For spiral-in/out cine, the maximum aliasing (SPR: 0.074) is lower within a circular region near the PSF peak, compared to that from spiral-out cine (SPR: 0.098). For a fixed TR, a longer spiral-in/out readout shows a higher level of incoherence than the spiral-out readout, which results in improved characteristics for sparse image reconstruction.
Simulation
Figure 2 shows the numerical phantom images and the x-t profiles reconstructed from the spiral-out and spiral-in/out readouts using CS and L + S reconstruction methods, and the fully sampled image as the reference. The results indicate that all methods lead to a high SSIM value (> 0.90) and a low NRMSE value (< 3%). Among these methods, the spiral-in/out cine with L + S method has the lowest-intensity image artifacts (red arrows) and maintains the best temporal information (green arrows). The same result can also be found from the in vivo midventricular cardiac images, movies, and the corresponding image ratings (see Supporting Fig. 2, Online Resource 2, and Supporting Fig. 3, respectively).
In vivo cine images
Figure 3 shows images from one healthy volunteer using spiral-out cine (top) and spiral-in/out cine (middle) with L + S reconstruction, and the standard Cartesian cine (bottom). End-diastolic and end-systolic images are shown in the first and second columns, respectively. Fine papillary muscles (red arrows) are clearly visualized in ventricles in images from the dynamic spiral cines. Green arrows pointing to x-t profiles demonstrate the preserved temporal fidelity of these two spiral cine techniques with a slight difference compared to that of the standard Cartesian cine, potentially due to the through-plane respiratory movement in the dynamic techniques. The corresponding videos can be found in Online Resource 3.
Figure 4 shows representative sets of short-axis cine images from another healthy volunteer, using dynamic spiral-out and spiral-in/out bSSFP sequences with L + S reconstruction, and the breath-hold reference. Both the spiral-based sequences produce overall good diagnostic image quality (rating scores > 3), while the spiral-in/out cine shows a visually better image contrast between the blood pool and the myocardium than the spiral-out cine. The standard breath-hold images (bottom row) show blurring artifacts, potentially because of a poor ECG-trigger signal. The corresponding videos are in Online Resource 4.
Image assessment
Figure 5a shows that the global image quality of the standard cardiac gated images was 4.4 \(\pm\) 0.6, compared to 3.8 \(\pm\) 0.8 for spiral-out bSSFP cine with the L + S method and 3.9 \(\pm\) 0.7 for spiral-in/out bSSFP cine with the L + S method. The total dataset (n = 7) included three short-axis (apex, middle, base) slices per subject. Error bars represent the standard deviations. All cardiac images were diagnostic (score > 3), while the differences in image quality between the spiral-out cine and the standard cine, and between the spiral-in/out cine and the standard cine were both significant (p < 0.05). Figure 5b shows the measured image contrast between the blood and myocardium for spiral-out bSSFP cine, spiral-in/out bSSFP cine, and standard Cartesian bSSFP cine. Error bars represent standard deviations. The image contrast was significantly higher (p < 0.05) for the standard cine (4.1 ± 0.5) and the spiral-in/out cine (3.8 ± 0.6) than the spiral-out cine (2.9 ± 0.5). Edges (mm−1) were sharper in the standard cine (0.69 ± 0.14) than the spiral-out (0.58 ± 0.13) and spiral-in/out cine (0.59 ± 0.13), as shown in Fig. 5c. The difference in edge sharpness between the standard cine and either the spiral-out or spiral-in/out cine was significant (p < 0.05), while that was not significant between the spiral-out and spiral-in/out cine.
Ejection fraction
The Bland–Altman plots show the comparison of LV EF between Cartesian cine and the spiral-out cine (Fig. 6a), and between Cartesian cine and the spiral-in/out cine (Fig. 6b). The mean differences in LV EF were − 1.5% and − 1.6% for the spiral-out cine and spiral-in/out cine, respectively, compared to the breath-hold reference. There was no statistical significance among these three methods (p > 0.1), establishing the accuracy of estimating LV EF using these two dynamic spiral cine techniques. There was no significant difference in mean LV EDV (p > 0.1). However, the mean LV ESV was significantly overestimated (p < 0.05) using both of the spiral cine techniques compared to those using the conventional breath-hold cine (see Supporting Table 1).
Discussion
In this work, we developed and validated both spiral-out and spiral-in/out bSSFP sequences combined with a L + S reconstruction method to achieve highly accelerated, ungated, free-breathing dynamic cine with 1.5 mm in-plane isotropic spatial resolution and 36 ms temporal resolution. The total scan time for whole LV coverage was greatly reduced from approximately 5–6 min with multiple breath-holds to 30–45 s with continuous scanning during free-breathing. The performance of these two sequences for cardiac function was compared to the standard Cartesian cine using protocols with similar spatiotemporal resolution, demonstrating there was good agreement for LV EF calculation between dynamic spiral techniques and the conventional breath-hold Cartesian cine.
Spiral k-space scanning has been previously investigated for accelerating ungated, free-breathing cardiac imaging, either with spoiled gradient echo (GRE) sequences [13, 39] or with bSSFP sequences [21]. A GRE strategy may have certain advantages over bSSFP acquisition for its insensitivity to off-resonance effects, allowing a longer spiral readout for a more efficient sampling; however, image contrast of GRE images between the blood pool and the myocardium is inferior compared to that of bSSFP images, which may lead to difficulty and inaccuracy in demarcation of the endocardial surface when calculating LV EF. Steeden et al. [21] used a short uniform density spiral-out (UDS) trajectory with a zeroth order moment rewinder to reduce the off-resonance effects in bSSFP sequences, yet this approach limits its spatial resolution due to a relatively low sampling efficiency within each TR. The unbalanced first moment of the readout gradient may also induce undesirable phase-related in-plane motion and inflow artifacts [20]. In this work, a VDS design instead of the UDS design was used for both spiral-out and spiral-in/out trajectories, which has been demonstrated to suppress under-sampling aliasing artifacts [40, 41] by sampling more at the center but less at the outer portion of k-space, thus increasing the level of incoherence that is important for CS-related approaches. Furthermore, the spiral-out trajectory was designed with time-optimized first-order moment nulling gradients, and the spiral-in/out trajectory has intrinsic first-order compensation via symmetry, with each resulting in better flow artifact suppression than a spiral trajectory with zeroth order compensation only.
Compared to the standard cine images, both spiral cine images show some image degradation, primarily due to the slight loss of spatiotemporal resolution and residual aliasing artifacts, as shown in Fig. 5a, c. The reduced spatiotemporal resolution mainly came from (1) a high under-sampling ratio at the outer of k-space; and (2) the trade-off between the residual aliasing artifacts and the temporal blurring during image reconstruction process. Comparing these two spiral cine sequences, the refocusing mechanism of the spiral-in/out readout, similar to that in the Cartesian readout, may account for a better image contrast compared to the spiral-out readout, because the signal at the middle of TR has more T2-weighted refocused spins (e.g., a darker myocardium). Furthermore, as demonstrated both in Figs. 3 and 4, this refocusing mechanism also produced clear right ventricular endocardial borders in the spiral-in/out cine and standard cine but not in the spiral-out cine, mainly because of a pi phase difference between the blood pool (0 Hz) and fat (− 210 Hz) signal at 1.5 T which results in a signal cancelation. This property could be utilized for improving the border contour of the right ventricle (RV) when calculating RV EF, although this work did not explore it. The magnitude and phase modulation as a function of off-resonance frequency for both the spiral-out and spiral-in/out readout based on current sequence protocols were simulated and can be found in Supporting Fig. 4.
k-space misregistration in spiral imaging due to gradient infidelity affects the image quality of spiral images, especially for cardiac scanning with oblique slice orientations. In our implementation, the measured system parameters only need to be calibrated once and can then be used for future scans [19, 30]. B0 inhomogeneities can degrade spiral images. In this work, short spiral readouts with acquisition windows of around 2 ms resulted in acceptable blurring artifacts at 1.5 T, which were further reduced by a simple linear off-resonance correction, in case the shimming was not good enough.
The choice of lambda parameters is the key for CS-related image reconstruction methods. Although we have not yet conducted a rigorous search for the optimal values because of the difficulty in quantitatively evaluating the cardiac images in this work, we observed that, for a single dataset, the performance remained unchanged within a small range of the chosen parameters. Thus, among different datasets, we fixed the FOV and spatial/temporal resolution for all seven volunteers, in order to have stable performance of the reconstruction method. In addition, the datasets were normalized to [0 1] before running the reconstruction kernel. We tested the fixed lambdas (\({\lambda }_{L}\;{and\; \lambda }_{S})\) on all datasets in this work, finding that the lambdas we chose yielded almost the best overall qualitative image quality, in terms of residual aliasing artifacts and visual temporal blurring. However, these lambda values may still be suboptimal for some in vivo scans because of variations in FOV, SNR, and banding artifacts. More general methods for determining lambdas warrant future investigation, but this is outside the scope of this manuscript.
This work has several limitations. The main limitation is that no patients were studied, so a patient study is needed for clinical validation of the proposed spiral cine methods. For example, patients with arrhythmia, which may produce inferior image quality associated with irregular ECG signals used in conventional Cartesian cine, can be recruited for clinical evaluation. Second, the comparison of LV EF values among methods validated that these two dynamic spiral cine sequences achieved good agreement in the accuracy of LV EF quantification, but a larger study could provide a more reliable analysis of functional parameter quantification. Third, end-diastolic and end-systolic frames from spiral cine scans without ECG-gating may not be matched to the same time points from the standard ECG-gated cine scans. The frame mismatch problem becomes more severe when imaging subjects with variable breathing patterns, although in general, a consistent breathing pattern induced through-plane motion, assuming that all short-axis slices were sampled during a similar respiratory cycle, seems to have an acceptable effect on the LV EF quantitative measurements when summing up the total volume across the whole left ventricle. One solution to minimize the discrepancy in respiratory motion is to perform dynamic cine imaging with 3–4 s short breath-holds per short-axis slice, or to combine this method with a simultaneous multi-slice technique for whole left ventricular coverage within a single breath-hold. Another potential direction is to use a cardiac motion- and respiratory motion-resolved multi-dimensional imaging technique, but this would result in the sacrifice of much-increased total scan time [13, 42, 43]. Fourth, the computational cost for offline image reconstruction in MATLAB is currently around 2–3 min. Optimizing the code and implementing the pipeline in C + + may enable real-time image reconstruction on commercial scanners.
Conclusions
Two spiral-based (spiral-out, spiral-in/out) bSSFP sequences with low-rank plus sparse image reconstruction were presented for highly accelerated dynamic cine to evaluate the cardiac function. These methods demonstrated cardiac cine imaging without ECG-gating during free breathing and with diagnostic image quality in normal volunteers, suggesting the potential for clinical dynamic cardiac MR imaging.
Data availability
Data presented in this study are available upon reasonable request.
References
Carr JC, Simonetti O, Bundy J, Li D, Pereles S, Finn JP (2001) Cine MR angiography of the heart with segmented true fast imaging with steady-state precession. Radiology 219:828–834
Setser RM, Fischer SE, Lorenz CH (2000) Quantification of left ventricular function with magnetic resonance images acquired in real time. J Magn Reson Imaging 12:430–438
Barkhausen J, Goyen M, Ruhm SG, Eggebrecht H, Debatin JF, Ladd ME (2002) Assessment of ventricular function with single breath-hold real-time steady-state free precession cine MR imaging. AJR Am J Roent-genol 178:731–735
Nita N, Kersten J, Pott A, Weber F, Tesfay T, Benea M-T, Metze P, Li H, Rottbauer W, Rasche V, Buckert D (2022) Real-time spiral CMR Is superior to conventional segmented cine-imaging for left-ventricular functional assessment in patients with arrhythmia. J Clin Med 11(8):2088
Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195
Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) SENSE: Sensitivity encoding for fast MRI. Magn Reson Med 42:952–962
Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A (2002) Generalized auto-calibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47:1202–1210
Otazo R, Kim D, Axel L, Sodickson DK (2010) Combination of compressed sensing and parallel imaging for highly accelerated first-pass cardiac perfusion MRI. Magn Reson Med 64:767–776
Feng L, Srichai MB, Lim RP, Harrison A, King W, Adluru G, Dibella EV, Sodickson DK, Otazo R, Kim D (2013) Highly accelerated real-time cardiac cine MRI using k-t SPARSE-SENSE. Magn Reson Med 70:64–74
Feng L, Grimm R, Block KT, Chandarana H, Kim S, Xu J, Axel L, Sodickson DK, Otazo R (2014) Golden-angle radial sparse parallel MRI: combination of compressed sensing, parallel imaging, and golden angle radial sampling for fast and flexible dynamic volumetric MRI. Magn Reson Med 72:707–717
Lingala S, Hu Y, Dibella E, Jacob M (2011) Accelerated dynamic MRI exploiting sparsity and low-rank structure: k-t SLR. IEEE Trans Med Imag 30:1042–1054
Otazo R, Candes E, Sodickson DK (2014) Low-rank plus sparse matrix decomposition for accelerated dynamic MRI with separation of background and dynamic components. Magn Reson Med 73(3):1125–1136
Zhou R, Yang Y, Mathew RC et al (2019) Free-breathing cine imaging with motion-corrected reconstruction at 3T using SPiral Acquisition with respiratory correction and cardiac self-gating (SPARCS). Magn Reson Med 82:706–720
Feng X, Wang Z, Meyer CH (2021) Real-time dynamic vocal tract imaging using an accelerated spiral GRE sequence and low rank plus sparse reconstruction. Magn Reson Imaging 80:106–112
Cao Z, Gore JC, Grissom WA (2019) Low-rank plus sparse compressed sensing for accelerated proton resonance frequency shift MR temperature imaging. Magn Reson Med 81(6):3555–3566
Bauer RW, Radtke I, Block KT, Larson MC, Kerl JM, Hammerstingl R, Graf TG, Vogl TJ, Zhang S (2013) True real-time cardiac MRI in free breathing without ECG synchronization using a novel sequence with radial k-space sampling and balanced SSFP contrast mode. Int J Cardiovasc Imaging 29:1059–1067
Muthurangu V, Lurz P, Critchely J, Deanfield J, Taylor A, Hansen MS (2008) Real-time assessment of right and left ventricular volumes and function in patients with congenital heart disease by using high spatiotemporal resolution radial k-t SENSE. Radiology 248:782–791
Haji-Valizadeh H, Rahsepar AA, Collins JD et al (2018) Validation of highly accelerated real-time cardiac cine MRI with radial k-space sampling and compressed sensing in patients at 1.5T and 3T. Magn Reson Med 79:2745–2751
Feng X, Salerno M, Kramer CM, Meyer CH (2016) Non-Cartesian balanced steady-state free precession pulse sequences for real-time cardiac MRI. Magn Reson Med 75:1546–1555
Nayak KS, Hargreaves BA, Hu BS, Nishimura DG, Pauly JM, Meyer CH (2005) Spiral balanced steady-state free precession cardiac imaging. Magn Reson Med 53:1468–1473
Steeden JA, Kowalik GT, Tann O et al (2018) Real-time assessment of right and left ventricular volumes and function in children using high spatiotemporal resolution spiral bSSFP with compressed sensing. J Cardiovasc Magn Reson 20:1–11
Wang Z, Feng X, Dou Q, Meyer CH (2018) Accelerated spiral imaging for real-time cardiac MRI. Proc Intl Soc Magn Reson Med 26:0935
Wang Z, Feng X, Yang Y, Zhou R, Salerno M, Meyer CH (2019) High resolution real-time spiral cine with whole heart coverage in under 30 seconds. Proc Intl Soc Magn Reson Med 27:2124
Meyer CH, Pauly JM, Macovski A, Nishimura DG (1990) Simultaneous spatial and spectral selective excitation. Magn Reson Med 15:287–304
Uecker M, Lai P, Murphy MJ, Virtue P, Elad M, Pauly JM, Vasanawala SS, Lustig M (2014) ESPIRiT–an eigenvalue approach to auto-calibrating parallel MRI: where SENSE meets GRAPPA. Magn Reson Med 71:990–1001
Hargreaves BA, Vasanawala SS, Pauly JM, Nishimura DG (2001) Characterization and reduction of the transient response in steady-state MR imaging. Magn Reson Med 46:149–158
Meyer CH, Hu BS, Nishimura DG, Macovski A (1992) Fast spiral coronary artery imaging. Magn Reson Med 28:202–213
Wundrak S, Paul J, Ulrici J, Hell E, Geibel MA, Bernhardt P, Rottbauer W, Rasche V (2016) Golden ratio sparse MRI using tiny golden angles. Magn Reson Med 75:2372–2378
Chan RW, Ramsay EA, Cheung EY, Plewes DB (2012) The influence of radial undersampling schemes on compressed sensing reconstruction in breast MRI. Magn Reson Med 67:363–377
Aldefeld B, Bornert P (1998) Effects of gradient anisotropy in MRI. Magn Reson Med 39:606–614
Tan H, Meyer CH (2009) Estimation of k-space trajectories in spiral MRI. Magn Reson Med 61:1396–1404
Irarrazabal P, Meyer CH, Nishimura DG, Macovski A (1996) Inhomogeneity correction using an estimated linear field map. Magn Reson Med 35:278–282
Fessler JA (2007) On NUFFT-based gridding for non-Cartesian MRI. J Magn Reson 188:191–195
Fessler JA. Michigan image reconstruction toolbox (MIRT). https://web.eecs.umich.edu/~fessler/code/index.html. Accessed September 15, 2018.
Foo TK, Berstein MA, Aisen AM et al (1995) Improved ejection fraction and flow velocity estimates with use of view sharing and uniform repetition time excitation with fast cardiac techniques. Radiology 195:471–478
Lin CY, Fessler JA (2019) Efficient dynamic parallel MRI reconstruction for the low-rank plus sparse model. IEEE Trans Comput Imaging 5:17–26
Wissmann L, Santelli C, Segars WP, Kozerke S (2014) MRXCAT: realistic numerical phantoms for cardiovascular magnetic resonance. J Cardiovasc Magn Reson 16:2–11
Wang Z, Bovik AC, Sheikh HR, Simoncelli EP (2004) Image quality assessment: from error visibility to structural similarity. IEEE Trans Image Proc 13:600–612
Eirich P, Wech T, Heidenreich JF et al (2021) Cardiac real-time MRI using a pre-emphasized spiral acquisition based on the gradient system transfer function. Magn Reson Med 85:2747–2760
Liao J-R, Pauly JM, Brosnan TJ, Pelc NJ (1997) Reduction of motion artifacts in cine MRI using variable-density spiral trajectories. Magn Reson Med 37:569–575
Tsai CM, Nishimura DG (2000) Reduced aliasing artifacts using variable density k-space sampling trajectories. Magn Reson Med 33:452–458
Feng L, Coppo S, Piccini D et al (2018) 5D whole-heart sparse MRI. Magn Reson Med 79(2):826–838
Christodoulou AG, Shaw JL, Nguyen C et al (2018) Magnetic resonance multitasking for motion-resolved quantitative cardiovascular imaging. Nat Biomed Eng 2(4):215–226
Acknowledgements
This work was supported in part by Siemens Medical Solutions (projects UVA-2018-MR-27-01-Mugler_C00227955 and UVA-2021-MR-29-01-Mugler_C00239406), by the National Institute of Biomedical Imaging and Bioengineering (R01 EB028773), and by the National Heart, Lung, and Blood Institute (R01 HL131919 and R01 HL155962).
Author information
Authors and Affiliations
Contributions
ZW: study conception and design, acquisition of data, analysis and interpretation of data, drafting of manuscript, critical revision; XF: study conception and design, acquisition of data; MS: acquisition of data, analysis and interpretation of data; CMK: acquisition of data, analysis and interpretation of data; CHM: study conception and design, acquisition of data, analysis and interpretation of data, critical revision.
Corresponding author
Ethics declarations
Conflict of interest
All authors declare that they have no conflicts of interest.
Ethical standard
Imaging was approved by the local Institutional Review Board (# 17557) and written informed consent was obtained from all participants.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Supplementary file2 (MP4 924 KB)
Supplementary file3 (MP4 1832 KB)
Supplementary file4 (MP4 1394 KB)
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Wang, Z., Feng, X., Salerno, M. et al. Dynamic cardiac MRI with high spatiotemporal resolution using accelerated spiral-out and spiral-in/out bSSFP pulse sequences at 1.5 T. Magn Reson Mater Phy 36, 857–867 (2023). https://doi.org/10.1007/s10334-023-01116-9
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10334-023-01116-9