Abstract
The purpose is to evaluate whether deep learning-based denoising (DLD) algorithm provides sufficient image quality for abdominal computed tomography (CT) with a 30% reduction in radiation dose, compared to standard-dose CT reconstructed with conventional hybrid iterative reconstruction (IR). The subjects consisted of 50 patients who underwent abdominal CT with standard dose and reconstructed with hybrid IR (ASiR-V50%) and another 50 patients who underwent abdominal CT with approximately 30% less dose and reconstructed with ASiR-V50% and DLD at low-, medium- and high-strength (DLD-L, DLD-M and DLD-H, respectively). The standard deviation of attenuation in liver parenchyma was measured as image noise. Contrast-to-noise ratio (CNR) for portal vein on portal venous phase was calculated. Lesion conspicuity in 23 abdominal solid mass on the reduced-dose CT was rated on a 5-point scale: 0 (best) to −4 (markedly inferior). Compared with hybrid IR of standard-dose CT, DLD-H of reduced-dose CT provided significantly lower image noise (portal phase: 9.0 (interquartile range, 8.7–9.4) HU vs 12.0 (11.4–12.7) HU, P < 0.0001) and significantly higher CNR (median, 5.8 (4.4–7.4) vs 4.3 (3.3–5.3), P = 0.0019). As for DLD-M of reduced-dose CT, no significant difference was found in image noise and CNR compared to hybrid IR of standard-dose CT (P > 0.99). Lesion conspicuity scores for DLD-H and DLD-M were significantly better than hybrid IR (P < 0.05). Dynamic contrast-enhanced abdominal CT acquired with approximately 30% lower radiation dose and generated with the DLD algorithm exhibit lower image noise and higher CNR compared to standard-dose CT with hybrid IR.
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10278-023-00808-x/MediaObjects/10278_2023_808_Fig1_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10278-023-00808-x/MediaObjects/10278_2023_808_Fig2_HTML.jpg)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10278-023-00808-x/MediaObjects/10278_2023_808_Fig3_HTML.png)
![](http://media.springernature.com/m312/springer-static/image/art%3A10.1007%2Fs10278-023-00808-x/MediaObjects/10278_2023_808_Fig4_HTML.png)
Similar content being viewed by others
References
Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure. The New England journal of medicine 2007;357:2277-2284.
Han WK, Na JC, Park SY. Low-dose CT angiography using ASiR-V for potential living renal donors: a prospective analysis of image quality and diagnostic accuracy. Eur Radiol 2019;30:798-805.
Iyama Y, Nakaura T, Yokoyama K et al. Impact of Knowledge-Based Iterative Model Reconstruction in Abdominal Dynamic CT With Low Tube Voltage and Low Contrast Dose. AJR American journal of roentgenology 2016;206:687-693.
Khawaja RD, Singh S, Otrakji A et al. Dose reduction in pediatric abdominal CT: use of iterative reconstruction techniques across different CT platforms. Pediatr Radiol 2015;45:1046-1055.
May MS, Wust W, Brand M et al. Dose reduction in abdominal computed tomography: intraindividual comparison of image quality of full-dose standard and half-dose iterative reconstructions with dual-source computed tomography. Invest Radiol 2011;46:465-470.
Willemink MJ, Noël PB. The evolution of image reconstruction for CT-from filtered back projection to artificial intelligence. Eur Radiol 2019;29:2185-2195.
Kwon H, Cho J, Oh J et al. The adaptive statistical iterative reconstruction-V technique for radiation dose reduction in abdominal CT: comparison with the adaptive statistical iterative reconstruction technique. Br J Radiol 2015;88:20150463.
Pickhardt PJ, Lubner MG, Kim DH et al. Abdominal CT with model-based iterative reconstruction (MBIR): initial results of a prospective trial comparing ultralow-dose with standard-dose imaging. AJR American journal of roentgenology 2012;199:1266-1274.
Ehman EC, Yu L, Manduca A et al. Methods for clinical evaluation of noise reduction techniques in abdominopelvic CT. Radiographics 2014;34:849-862.
Jensen CT, Wagner-Bartak NA, Vu LN et al. Detection of Colorectal Hepatic Metastases Is Superior at Standard Radiation Dose CT versus Reduced Dose CT. Radiology 2019;290:400-409.
Ichikawa Y, Kitagawa K, Nagasawa N, Murashima S, Sakuma H. CT of the chest with model-based, fully iterative reconstruction: comparison with adaptive statistical iterative reconstruction. BMC Med Imaging 2013;13:27.
Akagi M, Nakamura Y, Higaki T et al. Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT. Eur Radiol 2019;29:6163-6171.
Jiang Hsieh EL, Brian Nett, Jie Tang, Jean-Baptiste Thibault, Sonia Sahney. A new era of image reconstruction: TrueFidelity™ - Technical white paper on deep learning image reconstruction. https://www.gehealthcare.com/-/jssmedia/040dd213fa89463287155151fdb01922.pdf. 2019
Jensen CT, Liu X, Tamm EP et al. Image Quality Assessment of Abdominal CT by Use of New Deep Learning Image Reconstruction: Initial Experience. AJR American journal of roentgenology 2020;215:1-8.
Ichikawa Y, Kanii Y, Yamazaki A et al. Deep learning image reconstruction for improvement of image quality of abdominal computed tomography: comparison with hybrid iterative reconstruction. Jpn J Radiol 2021;39:598-604.
Noda Y, Kaga T, Kawai N et al (2021) Low-dose whole-body CT using deep learning image reconstruction: image quality and lesion detection. Br J Radiol 20201329
Shin YJ, Chang W, Ye JC et al. Low-Dose Abdominal CT Using a Deep Learning-Based Denoising Algorithm: A Comparison with CT Reconstructed with Filtered Back Projection or Iterative Reconstruction Algorithm. Korean J Radiol 2020;21:356-364.
Kaga T, Noda Y, Fujimoto K et al. Deep-learning-based image reconstruction in dynamic contrast-enhanced abdominal CT: image quality and lesion detection among reconstruction strength levels. Clin Radiol 2021;76:710.e15-710.e24.
Benz DC, Benetos G, Rampidis G et al. Validation of deep-learning image reconstruction for coronary computed tomography angiography: Impact on noise, image quality and diagnostic accuracy. J Cardiovasc Comput Tomogr 2020;14:444-451.
Kim I, Kang H, Yoon HJ, Chung BM, Shin NY. Deep learning-based image reconstruction for brain CT: improved image quality compared with adaptive statistical iterative reconstruction-Veo (ASIR-V). Neuroradiology 2020;63:905-912.
Park C, Choo KS, Jung Y, Jeong HS, Hwang JY, Yun MS. CT iterative vs deep learning reconstruction: comparison of noise and sharpness. Eur Radiol 2021;31:3156-3164.
Greffier J, Hamard A, Pereira F et al. Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study. Eur Radiol 2020;30:3951-3959.
Noda Y, Iritani Y, Kawai N et al. Deep learning image reconstruction for pancreatic low-dose computed tomography: comparison with hybrid iterative reconstruction. Abdom Radiol (NY) 2021;46:4238-4244.
Nam JG, Hong JH, Kim DS, Oh J, Goo JM. Deep learning reconstruction for contrast-enhanced CT of the upper abdomen: similar image quality with lower radiation dose in direct comparison with iterative reconstruction. Eur Radiol 2021;31:5533-5543.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of Interest
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Nagata, M., Ichikawa, Y., Domae, K. et al. Application of Deep Learning-Based Denoising Technique for Radiation Dose Reduction in Dynamic Abdominal CT: Comparison with Standard-Dose CT Using Hybrid Iterative Reconstruction Method. J Digit Imaging 36, 1578–1587 (2023). https://doi.org/10.1007/s10278-023-00808-x
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10278-023-00808-x