Abstract
Purpose
To evaluate the effect on the spinal canal at the treated and adjacent level(s), in patients treated for lumbar spinal stenosis (LSS) with percutaneous interspinous process device (IPD) Aperius™ or open decompressive surgery (ODS), using axial loading of the spine during MRI (alMRI).
Materials
Nineteen LSS patients (mean age 67 years, range 49–78) treated with IPDs in 29 spine levels and 13 LSS patients (mean age 63 years, range 46–76) operated with ODS in 22 spine levels were examined with alMRI pre- and 3 months postoperatively. Radiological effects were evaluated by measuring the dural sac cross-sectional area (DSCSA) and by morphological grading of nerve root affection.
Results
For the IPD group, no DSCSA increase was observed at the operated level (p = 0.42); however, a decrease was observed in adjacent levels (p = 0.05). No effect was seen regarding morphological grading (operated level: p = 0.71/adjacent level: p = 0.94). For the ODS group, beneficial effects were seen for the operated level, both regarding DSCSA (p < 0.001) and for morphological grading (p < 0.0001). No changes were seen for adjacent levels (DSCSA; p = 0.47/morphological grading: p = 0.95). Postoperatively, a significant difference between the groups existed at the operated level regarding both evaluated parameters (p < 0.003).
Conclusions
With the spine imaged in an axial loaded position, no significant radiological effects of an IPD could be detected postoperatively at the treated level, while increased DSCSA was displayed for the ODS group. In addition, reduced DSCSA in adjacent levels was detected for the IPD group. Thus, the beneficial effects of IPD implants on the spinal canal must be questioned.
Graphic abstract
These slides can be retrieved under Electronic Supplementary Material.
![](http://media.springernature.com/lw685/springer-static/image/art%3A10.1007%2Fs00586-019-06159-y/MediaObjects/586_2019_6159_Figa_HTML.png)
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Degenerative lumbar spinal stenosis (LSS) is becoming more frequently diagnosed due to the aging population and, to some extent, better access to advanced imaging facilities such as magnetic resonance imaging (MRI). Spinal stenosis is the most common reason for spinal surgery in patients over 65 years [1] and the most common cause of disability in elderly and middle-aged patients [2]. Common symptoms of LSS include neurogenic claudication, radiating leg pain, decreasing walking distance with or without low back pain and reduced quality of life [3]. However, it is well recognized that clinical findings in LSS not always correlate with radiological findings. Mildly affected patients can have signs of severe stenosis on MRI and vice versa [4, 5].
Conservative treatment is the standard of care in early stages of the disorder and involves analgesics as nonsteroidal anti-inflammatory drugs (NSAIDs), muscle relaxants or paracetamol. Epidural steroid injections and physiotherapy have also been suggested as treatments, but the evidence for the success of non-operative treatments is controversial and lacking [6]. Traditionally, the surgical intervention has been open decompressive surgery (ODS). However, due to risks of morbidity and mortality associated with general anesthesia and open surgery methods, especially in the elderly patients with severe or multiple comorbidities, less invasive surgical treatments such as interspinous process devices (IPDs) have been developed [7]. Such IPDs are inserted between the spinal processes of the stenotic level aiming to increase the interlaminar space, the central spinal canal area and the vertebral foramens at the symptomatic level. Biomechanically, such IPDs aim to position the spinal segment treated in a slightly flexed position to relieve the clinical symptoms [8].
Postoperatively, IPDs have shown good short-time clinical effects, but the long-term results have been less clear and a high reoperation rate compared to ODS has been reported [9,10,11]. In contradiction to the lack of long-term improvement in subjective symptoms and high reoperation rates, some studies have shown beneficial radiological effects [12,13,14].
Not many studies have been published on the radiological effects of IPDs, either by conventional MRI or with the spine in a loaded position, such as upright MRI or axial loading during MRI (alMRI). A few studies have reported a small but statistically significant increase regarding the dural sac and spinal canal cross-sectional area after IPD insertion [12, 15, 16], even up to 2 years postoperatively. However, these studies all lack a control group and do not provide any information regarding the effect on adjacent levels.
The aim of the current study was therefore to evaluate the effect on the spinal canal at the operated level and in adjacent levels, in patients treated for LSS with an IPD in comparison with a group operated with ODS, using alMRI.
Materials and methods
Patients
Patients undergoing surgical treatment for LSS were included: 19 patients (mean age 67 years, range 49–78) operated with Aperius™ PerCLID™ (Medtronic) in a total of 29 spine levels and a group of 13 patients (mean age 63 years, range 46–76) operated with ODS in 22 spine levels.
The patients recruited were a subgroup from a randomized control trial (RCT) (still not published) including 200 patients with neurogenic intermittent claudication (NIC), all with MRI-verified degenerative LSS, defined as a dural sac cross-sectional area (DSCSA) < 100 mm2. These 200 patients were enrolled from referred patients at the outpatient’s clinic at the Department of Orthopedics, The Spine Team Unit at the Sahlgrenska University Hospital. The patients were randomly divided between surgeries with either IPD Aperius™ or conventional ODS, with the aim of 100 patients within each group. During the RCT, 32 patients were included in the present study and the patients enrolled here were all asked whether they agreed to be investigated with alMRI pre- and postoperatively in addition to the conventional MRI that was a part of the RCT study. The selection of patients asked for inclusion was random and based on the available time for adding the alMRI investigation.
The present study was approved by the Regional Ethical Review Board in Gothenburg at The Sahlgrenska Academy, Gothenburg University, Gothenburg, Sweden.
Imaging with axial loading
alMRI was performed preoperatively and 3 months postoperatively on a 1.5 Tesla Philips scanner (Achieva). The alMRI was performed using a non-magnetic loading device with 50% of the patient’s body weight applied, equally distributed to each leg, to simulate the loading conditions during upright position (Dynawell, diagnostics AB, Las Vegas, USA) (Fig. 1) [17]. The load was applied for at least 5 min before the MRI was performed.
alMRI performed with the patient in a supine position with extended hips and knees. To prevent flexion of the spine during compression, and simulate lordosis in upright position, a small cushion was placed beneath the lumbar spine. The feet were positioned against a footplate, attached with side straps to a harness worn by the patient. Applied load is regulated by tightening or loosening adjustment knobs on the compression device
The MRI protocol included sagittal T2-weighted (W) sequences (TR 4500 ms, TE 120 ms, matrix 640 × 640), sagittal T1 W (TR 400 ms, TE 7.4 ms, matrix 704 × 704) and axial T2 W sequences (TR 2700, TE 120, matrix 512 × 512).
Radiological parameters determined
In both cohorts, the DSCSA (mm2), which is a highly rated and well-documented quantitative MRI parameter for LSS evaluation [2, 18, 19], was measured at the operated level and at the adjacent cranial and caudal levels. Due to few observations, no sub-analysis between adjacent cranial and caudal levels was made. The DSCSA was manually outlined with a polygonal measurement tool on mid-axial T2W images at the level of the intervertebral disk (Fig. 2). The smallest DSCSA obtained was used.
As qualitative radiological parameters sometimes are considered superior to quantitative parameters when evaluating LSS [19, 20], the morphological stenos grading, as described by Schizas et al., was included to evaluate affection of the nerve roots (Table 1) [20]. This grading was evaluated at the same axial image as the DSCSA.
All measurements were performed by a senior radiology resident and repeated after 1 month, blinded to previous measurements, in order to evaluate the reproducibility of the measurements. To investigate inter-observer agreement, the measurements were repeated by an experienced senior radiologist.
Statistics
The IBM SPSS Statistics for Windows, version 22.0., Armonk, NY: IBM Corp., was used for all analyses. For comparison between pre- and postoperative radiological measurements, paired t test was employed with significance level set at p < 0.05. Wilcoxon rank test was used for pre- and postoperative comparison regarding morphological stenos grading. For between-group comparison, Mann–Whitney U test was used for the morphological stenos grading and paired t test for the DSCSA. Inter-observer reproducibility and intra-observer reproducibility for the quantitative radiological measurements were calculated using the intraclass correlation coefficient (ICC), two-way random effects, absolute agreement, single rater/measurement [21]. For the qualitative evaluation, weighted kappa was used [22]. ICC values < 0.4 represent poor agreement, values between 0.4 and 0.75 represent fair to good agreement, and values exceeding 0.75 indicate excellent reliability. Weighted kappa values between 0.61 and 0.8 represent good reliability, and values between 0.8 and 1.00 represent excellent reliability.
Results
All the enrolled patients in both the IPD group (n = 19 patients/29 levels) and the ODS group (n = 13 patients/22 levels) were successfully examined with alMRI pre- and 3 months postoperatively.
In the Aperius™ group, there was no significant increase in the DSCSA at the operated level (mean 3 mm2, 95% CI − 4 to 10), while at the adjacent level, a significant decrease was observed pre- versus postoperatively (mean − 10 mm2, 95% CI − 20 to − 0.3). No significant changes were observed regarding the morphological grade, neither at the operated level nor at the adjacent levels (Table 2). A number of levels changing in morphological stenos grade at the postoperative MRI compared to preoperative MRI are displayed in Table 2.
In the ODS group, there was a significant change at the operated level pre- versus postoperatively, with both increased DSCSA (mean 74 mm2, 95% CI 58–90) and reduced affection of the nerve rootlets. Postoperatively, the morphological stenos grade improved at 16 levels (Table 2). At the adjacent level, the morphological stenos grade improved at three levels, but a higher compression of the nerve roots was detected in five levels, hence no significant change (Table 2). Between-group comparison revealed significant differences in postoperative outcome at the operated level regarding both DSCSA (p < 0.0001) and morphological stenos grade (p < 0.003) (Table 2).
ICC for DSCSA was 0.96 for both intra- and inter-observer analyses. The weighted kappa for intra-observer evaluation of morphological stenos grade was 0.83 and was 0.82 for inter-observer grading.
Discussion
No significant DSCSA increase or major change of morphological stenos grade could be seen 3 months after IPD Aperius™ implantation in this alMRI study. In the adjacent level, there was, however, a small but significant decrease in the DSCSA. Contrarily, the ODS group displayed a DSCSA increase in changing from an absolute stenosis preoperatively to close to what is considered normal DSCSA postoperatively (Table 2). The failure of the IPD to show radiological improvement postoperatively in alMRI might, at least to some extent, explain the general lack of long-term functionality and high reported reoperation rates among LSS patients treated with IPD [8,9,10,11, 23].
The reason why postoperative clinical outcome regarding IPD implants is reported poor in spite of improvement in radiological IPD studies [8,9,10,11,12,13,14, 23], could be explained by the fact that previous radiological studies have been performed with the patient in supine relaxed position. As clinical symptoms of LSS are typically induced or exacerbated by walking, standing or hip extension and tend to decrease during forward flexion, squatting or lying supine, the conventional supine MRI may underestimate the severity of the spinal canal stenosis [24]. As such, IPDs are intended to assert its effect during spinal loading. With alMRI, a more realistic image of the spine, as compared with supine MRI without load applied, is achieved by straining the spine during imaging [25]. alMRI is known to induce or increase detection in spinal changes as compared with conventional MRI. Examples of such load-induced changes are as follows: disk bulging, thickening of the ligamentum flavum, altered shape of the dorsal fat pad and deformation of the dural sac [24, 26]. alMRI can also increase the diagnostic accuracy of LSS by revealing LSS “hidden” at conventional MRI and reveal stenosis at more levels compared to the supine non-loaded MRI, changing single-level stenosis to a multi-level stenosis [27]. This positional difference at MRI can obviously impact the LSS diagnosis and affect the choice of treatment. Likely also postoperatively, alMRI provides a more realistic image of the spine morphology during spinal loading, compared to conventional MRI. Hence, the insufficient radiological effects of the IPD in the current study provide a reasonable patho-anatomical explanation to reported clinical treatment failure and high reoperation rates [8,9,10,11].
The few previous studies investigating the radiological effects of IPDs for LSS have been performed with great variability within study design, which makes comparison between studies challenging. For example, there is a large discrepancy in the selected patient groups, differences in radiological methods and parameters investigated, and also great variability at what time point the postoperative effects are evaluated [12, 15, 16, 28,29,30,31]. Nevertheless, there are MRI studies with spinal loading that, opposed to the current study, indicate increase in DSCSA or spinal canal area (SCA) after IPD implantation. Some of these studies have, however, used a patient group with mean preoperative DSCSA above 100 mm2 (150 mm2 for SCA), meaning that those patients can hardly be considered to fulfill the radiological criteria for LSS, and therefore, conclusions based on these studies have to be questioned [30, 31]. Some previous studies evaluating IPD effect with MRI in a loaded position used preoperative DSCSA comparable to the absolute stenosis in the current study, reporting an DSCSA increase from absolute stenosis (DSCSA ~ 75 mm2) up to ranges indicating moderate stenosis (DSCSA ~ 90 mm2) [12, 15, 16, 32]. In the current study, the DSCSA did not change significantly postoperatively in the IPD group while the DSCSA in the ODS group changed from an absolute stenosis preoperatively to close to what is considered normal DSCSA postoperatively (144 mm2). Even though some previous IPD studies report a small increase in spinal canal or DSCSA postoperatively, they never come close to postoperative measures within the normal range and not even to the cutoff value for relative stenosis of 100 mm2. Differences between the current and previous studies could be attributed to methodological issues, such as applied load in an upright position versus supine position.
Comparison between IPD and ODS cohorts has previously not been performed with the spine in a loaded position. Besides from improvements in DSCSA, the ODS group in our study also showed a beneficial change in the morphological stenos grade (Table 2), indicating also less impingement of the neural tissues after the surgery [20], an effect that was not seen for the patients treated with IPD. The current study clearly displays the large postoperative differences between the groups as well as provides data on the IPD effects at adjacent levels, thereby filling a gap in the literature. Although the design of such IPDs varies, resulting in some differences in range of motion and flexibility, they are all based on the same principle, i.e., distracting the spinal processes to limit extension and, to some extent, distracting the neural foramina and unloading the intervertebral disk [8]. It therefore seems reasonable to investigate the spine in a position in which the IPD is intended to affect the spine the most. In spite of the fact that clinical improvement is the most important treatment outcome, a lack of significant improvement in radiological measures indicates insufficient treatment effect and provides a likely patho-anatomical explanation to why IPDs implants lack long-term functionality [9,10,11, 23]. Even though the current data are not directly applicable for all IPDs, this study shows insufficient treatment effect already after 3 months postoperatively, and likely previous studies have failed to show this insufficient radiological effect by using MRI without spinal loading. Another explanation for lack of DSCSA increase postoperatively with the IPD can be due to wear or destruction of the spinal processes during the first month postoperatively. In a biomechanical study, Baranto et al. demonstrated wear of the spinal processes in patients treated with Aperius™ [32]. They concluded that this is due to the fact that the Aperius™ implant is of metal, and that this is the main reason for failure or recurrence of symptoms already after 3–6 months postoperatively in patients treated with IPD.
The small DSCSA decrease in adjacent levels in the IPD group has not previously been reported and might have clinical effects in patients with multi-level disease. This finding strengths the argument for using alMRI to diagnose LSS and also to evaluate the effect of different LSS surgical treatments.
Limitations
The time of follow-up is a potential limitation of this study. It cannot be excluded that the effects of the implanted IPD were better immediately after surgery than at our follow-up time point at 3 months, but if so that effect then diminishes after a short period of time postoperatively. No sub-analysis between adjacent cranial and caudal levels was made because of few observations in this small sample size, which is a limitation.
Conclusion
With the spine imaged in an axial loaded position, no significant radiological effects of the IPD could be detected 3 months postoperatively at the operated level, neither on the DSCSA nor regarding qualitative estimated stenos grade, while beneficial effects clearly were displayed for the ODS group. In addition, reduced DSCSA in adjacent levels was shown for the IPD group. Thus, the beneficial effects of such IPD implants on the spinal canal must be questioned. Studies aiming to investigate the spinal canal in LSS patients, and particularly in studies of new treatment methods, are encouraged to use alMRI for evaluation of the decompressive effect in the treated as well as adjacent levels.
Change history
01 November 2019
Unfortunately, the ninth reference was incorrectly published in the original publication. The complete correct reference is given below.
References
Deyo RA (2010) Treatment of lumbar spinal stenosis: a balancing act. Spine J 10:625–627. https://doi.org/10.1016/j.spinee.2010.05.006
Lim YS, Mun JU, Seo MS et al (2017) Dural sac area is a more sensitive parameter for evaluating lumbar spinal stenosis than spinal canal area: a retrospective study. Med (United States) 96:1–5. https://doi.org/10.1097/MD.0000000000009087
Gevenay S (2011) NIH public access. Best Pr Res Clin Rheumatol 24:253–265. https://doi.org/10.1016/j.berh.2009.11.001.Lumbar
Baker ADL (2014) Abnormal magnetic-resonance scans of the lumbar spine in asymptomatic subjects. A prospective investigation. Class Pap Orthop. https://doi.org/10.1007/978-1-4471-5451-8_60
Sirvanaci M, Bhatia M, Ganiyusufoglu KA et al (2008) Degenerative lumbar spinal stenosis: correlation with Oswestry Disability Index and MR Imaging. Eur Spine J 17:679–685. https://doi.org/10.1007/s00586-008-0646-5
Ammendolia C, Stuber K, de Bruin LK et al (2012) Nonoperative treatment of lumbar spinal stenosis with neurogenic claudication: a systematic review. Spine (Phila Pa 1976) 37:E609–E616. https://doi.org/10.1097/brs.0b013e318240d57d
Ciol MA, Deyo RA, Howell E, Kreif S (1996) An assessment of surgery for spinal stenosis: time trends, geographic variations, complications, and reoperations. J Am Geriatr Soc 44:285–290. https://doi.org/10.1111/j.1532-5415.1996.tb00915.x
Ravindra VM, Ghogawala Z (2017) Is there still a role for interspinous spacers in the management of neurogenic claudication? Neurosurg Clin N Am 28:321–330. https://doi.org/10.1016/j.nec.2017.02.002
LeHuec J-C, Schils F, Zoega B et al (2017) Percutaneous interspinous spacer vs decompression in patients with neurogenic claudication: An alternative in selected patients? Neurosurgery 82:621–629. https://doi.org/10.1093/neuros/nyx326
Pintauro M, Duffy A, Vahedi P et al (2017) Interspinous implants: are the new implants better than the last generation? A review. Curr Rev Musculoskelet Med 10:189–198. https://doi.org/10.1007/s12178-017-9401-z
Ramesh A, Lyons F, Kelleher M (2016) Aperius interspinous device for degenerative lumbar spinal stenosis: a review. Neurosurg Rev 39:197–205. https://doi.org/10.1007/s10143-015-0664-9
Surace MF, Fagetti A, Fozzato S, Cherubino P (2012) Lumbar spinal stenosis treatment with Aperius perclid interspinous system. Eur Spine J 21:69–74. https://doi.org/10.1007/s00586-012-2222-2
Marcia S, Hirsch JA, Chandra RV et al (2015) Midterm clinical and radiologic outcomes after percutaneous interspinous spacer treatment for neurogenic intermittent claudication. J Vasc Interv Radiol 26:1687–1693. https://doi.org/10.1016/j.jvir.2015.05.011
Lee J, Hida K, Seki T et al (2004) An interspinous process distractor (X STOP) for lumbar spinal stenosis in elderly patients: preliminary experiences in 10 consecutive cases. J Spinal Disord Tech 17:72–77. https://doi.org/10.1097/00024720-200402000-00013
Masala S, Fiori R, Bartolucci DA et al (2012) Percutaneous decompression of lumbar spinal stenosis with a new interspinous device. Cardiovasc Intervent Radiol 35:368–374. https://doi.org/10.1007/s00270-011-0167-1
Siddiqui M, Nicol M, Karadimas E et al (2005) The positional magnetic resonance imaging changes in the lumbar spine following insertion of a novel interspinous process distraction device. Spine (Phila Pa 1976) 30:2677–2682. https://doi.org/10.1097/01.brs.0000187878.79676.26
Willén J (1997) Dynamic effect on the lumbar spinal canal axially loaded CT myelography and MRI in patients with sciatica and or neurogenic claudication. Spine (Phila Pa 1976) 22:2968–2976
Hamanishi C, Matukura N, Fujita M et al (1994) Cross-sectional area of the stenotic lumbar dural tube measured from the transverse views of magnetic resonance imaging. J Spinal Disord 7:388–393
Mamisch N, Brumann M, Hodler J et al (2012) Radiologic criteria for the diagnosis of spinal stenosis: results of a delphi survey. Radiology 264:174–179. https://doi.org/10.1148/radiol.12111930/-/DC1
Schizas C (2010) Qualitative grading of severity of lumbar spinal stenosis based on the morphology of the dural sac on magnetic r.pdf. Spine (Phila Pa 1976) 35:1919–1924
Richard Landis J, Koch Gary G (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174
Smith ZA (2014) Interspinous process device versus standard conventional surgical decompression for lumbar spinal stenosis results in increased reoperation rates and costs without improving patient outcomes. Evid Based Med 19:136. https://doi.org/10.1136/eb-2013-101689
Kanno H, Ozawa H, Koizumi Y et al (2012) Dynamic change of dural sac cross-sectional area in axial loaded magnetic resonance imaging correlates with the severity of clinical symptoms in patients with lumbar spinal canal stenosis. Spine (Phila Pa 1976) 37:207–213. https://doi.org/10.1097/brs.0b013e3182134e73
Kimura S, Steinbach GC, Watenpaugh DE, Hargens AR (2001) Lumbar spine disc height and curvature responses to an axial load generated by a compression device compatible with magnetic resonance imaging. Spine (Phila Pa 1976) 26:2596–2600. https://doi.org/10.1097/00007632-200112010-00014
Kanno H, Endo T, Ozawa H et al (2012) Axial loading during magnetic resonance imaging in patients with lumbar spinal canal stenosis: Does it reproduce the positional change of the dural sac detected by upright myelography? Spine (Phila Pa 1976) 37:985–992. https://doi.org/10.1097/brs.0b013e31821038f2
Willén J, Wessberg PJ, Danielsson B (2008) Surgical results in hidden lumbar spinal stenosis detected by axial loaded computed tomography and magnetic resonance imaging. Spine (Phila Pa 1976) 33:E109–E115. https://doi.org/10.1097/brs.0b013e318163f9ab
Sobottke R, Schlüter-Brust K, Kaulhausen T et al (2009) Interspinous implants (X Stop®, Wallis®, Diam®) for the treatment of LSS: Is there a correlation between radiological parameters and clinical outcome? Eur Spine J 18:1494–1503. https://doi.org/10.1007/s00586-009-1081-y
Kim HY, Choi BW (2013) Change of radiological parameters after interspinous implantation (X-stop®) in degenerative spinal stenosis. Eur J Orthop Surg Traumatol 23:281–285. https://doi.org/10.1007/s00590-012-0986-z
Masala S, Marcia S, Taglieri A et al (2016) Degenerative lumbar spinal stenosis treatment with Aperius™ PerCLID™ system and Falena® interspinous spacers: 1-year follow-up of clinical outcome and quality of life. Interv Neuroradiol 22:217–226. https://doi.org/10.1177/1591019915622163
Nandakumar A, Clark NA, Peehal JP et al (2010) The increase in dural sac area is maintained at 2 years after X-stop implantation for the treatment of spinal stenosis with no significant alteration in lumbar spine range of movement. Spine J 10:762–768. https://doi.org/10.1016/j.spinee.2010.06.007
Siddiqui M, Karadimas E, Nicol M et al (2006) Influence of X stop on neural foramina and spinal canal area in spinal stenosis. Spine (Phila Pa 1976) 31:2958–2962. https://doi.org/10.1097/01.brs.0000247797.92847.7d
Halldin K, Ekström L, Hvannberg J et al (2016) Interspinous process implants causes wear of the spinous processes in patients treated for spinal stenosis—an experimental biomechanical study with comparison to clinical cases. Open J Orthop 06:201–210. https://doi.org/10.4236/ojo.2016.67027
Acknowledgements
Open access funding provided by University of Gothenburg. The authors acknowledge the financial support of The Medical Society of Gothenburg, Sweden, financial grants from the Swedish state under the agreement between the Swedish government and the country councils, the ALF-agreement (Id Nos. 238801 and 792231), Hjalmar Svensson Research Foundation, Doktor Felix Neuberghs Foundation, and Christer Johansson, OrigoVerus AB, Gothenburg, Sweden, for assistance with statistics.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Hjaltadottir, H., Hebelka, H., Molinder, C. et al. Axial loading during MRI reveals insufficient effect of percutaneous interspinous implants (Aperius™ PerCLID™) on spinal canal area. Eur Spine J 29, 122–128 (2020). https://doi.org/10.1007/s00586-019-06159-y
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00586-019-06159-y