1 Introduction

In recent years, infrared laser absorption spectroscopy has been successful for atmospheric measurements of small trace gases, enabling fast online and precise measurements with compact instrumentation. This technique has been implemented in aircraft [1, 2], and is now more and more widespread. Most recently built airborne instruments comprise several variants, i.e., non-resonant cavities using tunable diode lasers (see, e.g., [3,4,5,6] and references therein), quantum cascade lasers (QCL; see, e.g., [7, 8] and references therein), or difference frequency generation lasers [9], and resonant cavities for cavity ring-down spectroscopy (CRDS) [10,11,12], off-axis integrated cavity output spectroscopy (OA-ICOS) [13,14,15], or optical-feedback cavity enhanced absorption spectroscopy (OF-CEAS) [16]. Using lasers allows for narrow linewidths, leading to selectivity of trace gases and very low limits of detection. Using QCLs, which emit in the Mid-Infrared (MIR) domain, leads to stronger absorptions than in the near infrared due to the molecular fundamental transitions taking place and, therefore, again to higher sensitivity. Among other advantages, QCLs possess high optical output power, which is useful for long optical path operation. First used in pulsed mode, QCLs begin to be operated in continuous-wave (CW) mode, resulting in negligible linewidths [17] with respect to the molecular absorption line at the operation pressure. As a result, this leads to measurements of intrinsically higher accuracy and sensitivity.

For more than 20 years, our group, in collaboration with the French space agency CNES (Centre National d’Etudes Spatiales), has developed a series of balloon-borne instruments using optical spectroscopy techniques with remote (stars, moon) or in situ (embarked lasers) light sources for the detection of stratospheric species [18,19,20]. In the present paper, we present a newly built airborne instrument called SPIRIT (SPectromètre Infra-Rouge In situ Toute altitude). SPIRIT performs ultra-high-resolution (<0.001 cm−1) infrared absorption spectroscopy thanks to the use of three MIR CW-QCLs (with FWHM of emission lines lower than 4 × 10−4 cm−1) coupled with a single non-resonant cavity, a Robert cell in the present case [21]. This is a multiple-reflection cell patented by our group, combining the design of the Herriott and White cells, allowing for longer optical paths (for the same size). This arrangement leads to a unique compact and versatile spectrometer for quantifying in situ abundances of at least three atmospheric trace gases, which can be selected using suitable QCLs, depending on the objectives of each project. We first targeted CO, NO2, and CH4 for their potential to answer chemical pollution and air masses transport issues in various projects extending from the lower to the upper troposphere.

After this introduction, we describe in Sect. 2, the opto-mechanical setup, the control–acquisition electronics of the instrument and the data retrieval process. The performances of the measurements obtained in the laboratory and in-flight are given in Sect. 3. The possible applications of SPIRIT measurements are rather for local or regional pollution event studies. Therefore, species concentrations are measured within the range of background to enhancement levels representative of pollution events. Using spectroscopic data from HITRAN database [22] is sufficient to attain the required accuracy (a few % according to this database), since these data contain all the information to model the line shapes and intensities as a function of experimental conditions (temperature, pressure, and gas nature). In its present configuration, SPIRIT is not designed for long-term greenhouse gas monitoring to be reported onto known global calibration scale (e.g., WMO/GAW network). Indeed, this monitoring can be conducted by commercially available instruments (e.g., Picarro Inc., Los Gatos Research Inc., Aerodyne Research Inc.) and needs regular in-flight calibrations [7, 8, 10, 13] at the expense of valuable ambient acquisition such as sudden unexpected increases in abundances of species. Moreover, in-flight calibrations are possible essentially for stable greenhouse gases and CO using embarked cylinders, but more difficult to handle with reactive or sticky species. Besides these commercial spectrometers, developing and implementing home-made instruments such as the one presented here are valuable. This leads to a complete independent mastering of the process for data acquisition and molecular concentration retrievals, and enables immediate diagnosis and swift repair in case of failure during field campaigns.

2 Description of the instrument

2.1 Instrumental design

As a multichannel tunable laser spectrometer, SPIRIT consists of laser sources, an optical cell, where the trace gas is continuously sampled, photodetectors, and an electronic system controlling laser functioning and data acquisition. The instrumental principle for ground measurements with one single channel and the process to retrieve molecular concentrations of the trace gases have been described previously [23]. The present paper only deals with the details related to the three-channel airborne version. Figure 1 shows 3D drawings of the SPIRIT instrument integrated in the aircraft rack, using a computer-aided design (CAD). The instrument weighs 102 kg, its dimensions are 95.8 × 55.9 × 65.4 cm3, and its power consumption is less than 1.2 kW.

Fig. 1
figure 1

3D drawings of the instrument integrated in the Falcon aircraft rack

A schematic representation of the operation principle is given in Fig. 2. SPIRIT operates with three Continuous-Wave Distributed-Feedback Room-Temperature QCLs (CW-DFB-RT-QCLs from Alpes Lasers SA; see characteristics given hereafter, as shown in Table 1), corresponding to the three measurement channels, coupled to one optical cell and two detectors only—one for the measurement channels and the other one for the reference channels. The home-made main electronic module receives commands from a laptop computer via an RS-232 link and drives the current and temperature controllers for each of the three QCLs. Each temperature controller consists of a Peltier thermoelectric module that maintains individually each laser chip at a constant temperature, with a typical precision of 1 mK, allowing for laser single-mode emission at a precise wavenumber. A water–glycol circuit regulated at a constant temperature (13 °C) allows for removing excess heat from each Peltier cooler and the Stirling detector cooler, using a liquid pump (Ecocirc Laing) and a liquid–air thermal exchanger (LA-115 Laird Technologies). The three QCLs are sequentially switched on/off and synchronized by the data acquisition system. For each QCL, the current controller applies a saw-tooth ramp of several mA during 4.1 ms to sweep a spectral micro-window of interest for the species to quantify. This ramp is repeated 110 times for measurement spectra followed by eight times for reference spectra. The whole sequence containing measurements for the three channels are thus recorded within 1.6 s, including averaging and transfer of these data to the laptop computer. All along this paper, volume mixing ratios (vmr) are used as relevant quantities, defined as mole fractions relative to the air (usually expressed in part per billion in volume—ppbv), which is a common practice in the atmospheric community. Roughly calculated preliminary vmr (only based on absorption depths of the line centers corresponding to known vmr measured in the laboratory) are inferred and displayed online to be able to modify live the flight trajectory. In addition, the data are stored on the computer for post-processing, leading to retrievals of the trace-gas vmr, as detailed in Sect. (2.2).

Fig. 2
figure 2

Schematic principle of the instrument

Table 1 Spectral micro-windows and emission conditions of the QCLs

The optical bench is presented in Fig. 3. The main mechanical component is a 5-cm-thick aluminum honeycomb composite panel vertically oriented. It is inserted in the aircraft rack with vibration-shock absorbers suitably dimensioned to the airplane acceleration requirements. In these conditions, the maximum bench amplitude of displacement is less than 3 cm. The optical cell is made of a glass tube (8-cm outside diameter, 3-mm thick, and 62.5-cm length) in low thermal expansion borosilicate. The two tube ends are bonded in two adapted thermal expansion pieces. The optical cell is rigidly fixed on the upper end to the optical bench, and the lower end is free to glide on a V-shaped mount to avoid constraint due to the different thermal expansions between aluminum and glass. The QCL housings include off-axis parabolic mirrors of 5-mm focal length, mounted on the same thermoelectrically cooled copper mechanical support as the laser itself for optical alignment stability. All three laser beams, symbolized by red rays in Fig. 3, go out from these housings through an anti-reflection ZnSe-coated window. An optical multiplexor allows for the spatial superposition of the three beams which then go out onto a BaF2 beam splitter, dividing them into two parts. A minor part is sent through a home-made Fabry–Pérot interferometer etalon used as reference channel for relative wavenumber calibration, since the current ramp does not lead to a precisely linear wavenumber sweep by the QCLs. This beam is then focused onto a photovoltaic HgCdTe detector (Judson J19D10) by a spherical mirror. The detector is linear enough, so that it does not influence the measurements, because it is biased in voltage (typically 100 mV). This has been verified by attenuating the incident intensity of the laser beam using different grids to verify that the molecular line absorptions do not depend on the light power. The major part of the beam is directed and focused on the entrance of the multipass optical cell through a BaF2 window. Before entering this cell, the beam can pass through a small reference cell containing known gaseous species added for retrieving the absolute wavenumber scale and checking the spectral tuning. This is the case for NO2 measurements for which a cell (of 3-cm long) with CH4 (100 hPa at ~120 ppmv) is used, leading to ~1% absorption of a doublet centered at 1630.407 cm−1. Indeed, NO2 atmospheric abundance is highly variable, even down to below the detection limit (0.3 ppbv, as indicated in Sect. 3.2). The measurement output beam goes through a second BaF2 window and is then collected by a flat mirror and focused onto an analogous detector by a spherical mirror similar to the reference beam. Both detectors are cooled by the same Stirling cycle cooler (RM2 Thalès Cryogénie) mounted in a home-made Dewar in which vacuum is maintained by passive pumping using high porosity-sintered materials (CapaciTorr® D 100 Saes Getters). The detected electrical signals are amplified, first by pre-amplifiers with a fixed gain and second by an automatically adjusted gain in the main electronic module, and digitized by a 16-bit analog-to-digital converter.

Fig. 3
figure 3

CAD schematics of the optical bench; the red rays symbolize the laser beams

The optical cell is detailed in a previous paper [21]. In brief, this cell is composed of three mirrors combining the advantages of the Herriott and White cells, which results in very easy and stable operating conditions and allows for very long optical paths, i.e., from 117.45 m (181 reflections) to 167.78 m (259 reflections) in the present experiments, in a compact volume (2.77 L for a cylinder of 7.4-cm internal diameter and 64.5-cm length). Mirror alignments are controlled by two high-performance piezoelectric actuators (Newport PZA12) located at each of its ends, shown in Fig. 3 (in orange color), allowing optical alignment at desired path length and also the change in the pass number by tilting the two mirrors relative to each other [21]. A third actuator (not visible in Fig. 3) is used to move slowly (0.3 Hz) back and forth the stand-alone mirror on a small range. As a result, unavoidable interference fringes between laser beams and scattered light from the optical components are efficiently attenuated. The temperature of the sampled gas in the cell, measured by means of a Pt100 resistance probe (precision of 0.01 K), is accurate and homogeneous to within 1 K (when the air conditioning system is operating properly). Its value is needed to fit the molecular lines using spectroscopic parameters and to derive the gas vmr. The temperature homogeneity is made possible by the use of aluminum, a very good thermal conductor as material constituting the optical bench and the jacket of the optical cell. Homogeneity to within 1 K has been checked by measuring temperature simultaneously at several positions along the optical cell for several hours. The air is sampled through a rear-facing stainless steel inlet of 16-mm inside diameter (i.d.), placed in front of the aircraft on the roof of the cabin to avoid contamination. A ½” (12.7 mm) i.d. PFA Teflon tubing (less porous and adsorbing than PTFE) is inserted inside this inlet and connected via a 6 mm o.d. (Swagelock) stainless steel valve to a pressure regulator (MKS 649) upstream of the cell (Fig. 2). The gas is thermalized along ~3 m of the tube, according to our calculations based on laws of fluid dynamics. This is confirmed under several various experimental conditions that show no correlation between external air temperature and cell air temperature, even for differences greater than 40 °C (as will be illustrated in Fig. 9). The air is continuously drawn through the optical cell using a scroll pump (XDS10 Boc Edwards) fixed with vibration absorbers on the rack floor. The pressure is constantly maintained at a reduced level of 30–35 hPa by the pressure regulator, accurately known (±0.1 hPa) using a temperature regulated 0–100 hPa gauge (DTM Scaime), and the flow is set before the experiments by a manual dosing valve (EVN116 Pfeiffer) located at the cell output, upstream of the pump, as shown in Fig. 2 (labelled flow control). This leads to an effective flow rate of 0.316–0.527 L s−1 in the cell at 35 hPa and 25 °C (corresponding to 0.600–1.000 standard liter per minute—SLM), measured by the mass flowmeter combined with the pressure regulator. The exit line of the vacuum pump is connected to the exhaust tube on the floor or on the roof of the aircraft cabin.

Once switched on, SPIRIT operates fully automated measurements. To compensate for any dimensional drifts, optical realignment (lasting 1 min) can be regularly performed using the two piezoelectric actuators. Its frequency can be chosen, e.g., every 10 min in the case of the absence of a human operator (a compromise between the loss of atmospheric measurements and the degradation of the measurement quality by the decrease of the signal-to-noise ratio). A GPS system (Garmin GPS18) giving universal time, altitude, latitude, and longitude is implemented, on which the laptop computer is synchronized. This has proven to be very useful for the data interpretation, enabling easy comparison with the data of the aircraft and the other scientific instruments on board. Finally, it should be underlined that the materials and functioning of the SPIRIT instrument fulfil the international regulations concerning aircraft equipment and led to certification for installation in three aircrafts: the DLR Falcon-20 (established by Enviscope GmbH) and the CNRS/INSU—Météo France Falcon-20 and ATR-42.

2.2 Data retrieval

The laser tunings over the three spectral micro-windows (channels) indicated in Table 1 allow for matching the wavenumber \(\tilde{\nu }_{0}\) of the ro-vibrational lines of the molecule to quantify, i.e., CO at 2179.772 cm−1, CH4 at 1277.473 cm−1 and N2O at 1277.235 cm−1, and NO2 at 1630.326 cm−1 (with a CH4 doublet centered at 1630.407 cm−1 in the reference cell).

As previously explained in detail [20, 23], the molecular concentrations are obtained by fitting the numerical second derivative with respect to the wavenumber of the measured transmission to that of the theoretical transmission (from the Beer–Lambert law) at each wavenumber \(\tilde{\nu }\) of the spectral micro-window using a linear regression algorithm to minimize the sum of squared residuals. The second derivative method is more advantageous as it does not need the accurate reconstruction of the 100% transmission baseline (the continuous component of the signal), where the molecules absorb, and it reduces the effect of the optical interference fringes. The spectroscopic parameters are taken from the HITRAN 2012 database [22]. Considering the intermediate range of pressure, where neither the pressure (Lorentzian) nor the Doppler (Gaussian) broadenings are predominant, the line shape is assumed to be a Voigt profile. An example of a direct transmission and its second derivative for a CO spectrum is shown in Fig. 4. The fitting errors on the retrieval (the minimum residual) are about 1% at 1σ confidence level in this figure and on average for the measurements of each species.

Fig. 4
figure 4

a Example of experimental (black line) and simulated (red line) transmission spectra for CO at 31.7 °C and 32.0 hPa, with pathlength of 134.22 m, leading to CO vmr of 89.1 ± 0.9 ppbv. b Experimental (black line) and simulated (red line) second derivatives of this transmission signal. c Minimum residuals of the direct transmission (pink line) and of the second derivative (blue line)

In principle, the absorption measurements may be considered absolute and do not require a calibration if accuracy is expected to be about that of the spectroscopic database. The uncertainties on the spectroscopic parameters [22], essentially the lines strengths and air broadening factors (5% maximum each), affect the accuracy of the trace-gas concentrations by 5 and 2.5%, respectively, as derived from a sensitivity analysis.

Vmr are derived knowing the total pressure p and the temperature T of the optical cell. These are measured with accuracies of 0.2 hPa and 1 K, leading to 0.3 and 0.4% accuracies in the inferred vmr, respectively, according to a sensitivity analysis. The dry vmr of the trace gases are then deduced using the water vapor mixing ratios measured by SPIRIT at 1277.270 cm−1 (on the CH4–N2O channel), and the H2O vmr agree with measurements reported by the core instrumentation installed on the aircraft within 10%. Depending on the plane and on the altitude level, this instrumentation is a dew or a frost point temperature hygrometer (CR2 from Buck Research), a lyman-alpha fluorescent spectrometer, or a tunable diode laser absorption spectrometer (WVSS-II from SpectraSensors). This 10% disagreement represents the accuracy in the H2O vmr, and for H2O volume mixing ratios of 1% maximum encountered in the free troposphere, this leads to a maximum contribution of 0.1% to the uncertainty for the dry mixing ratios. By taking into account all the above potential errors, the overall estimated uncertainties calculated as the square root of the quadratic sum are 5.7% for all dry volume mixing ratios.

3 Results: performances of the instrument

Investigations were carried out to assess SPIRIT instrumental performances. As it is designed to be airborne and is mainly used for that purpose, and because environmental conditions drastically differ in a plane from those that may be encountered in a laboratory, emphasis was placed on describing its performances during real flight operations when it was possible.

3.1 Laboratory experiments

Response time has been evaluated by subjecting the instrument to sudden positive and negative variations of carbon monoxide of about 100 ppbv. For six measurements, the mean rise time between 10 and 90% of the step, 8.6 ± 0.3 s, was consistent with the fall time, 8.8 ± 0.3 s, for a standard flow rate set to 0.600 SLM (used for flights in the upper troposphere). These values correspond to the theoretical overall flush time (8.7 s) of the cell, showing the absence of significant dead volumes that could affect the instrument accuracy. This response time is lowered to 5.2 s for a 1.000 SLM flow rate when the flight occurs in the lower troposphere. Larger flows are reachable at ground, but are not sustainable at the highest altitudes, where ambient pressure decreases below 250 hPa, due to the limited size of the pressure regulator orifice and leads to the impossibility of maintaining stable pressure inside the cell. Similar performances are obtained for CH4 and N2O. Concerning stickier molecules such as NO2 or H2O, response time has not been directly evaluated, considering their propensity to adsorb and desorb on surfaces, where they can easily react. One can assume the response time to vary and to somewhat exhibit dependence towards temperature and the chemical passivation state of the cell walls, which itself depends on the concentrations of the gases previously passed through the analyzer and the elapsed time. However, in-flight applications for tracking aircraft contrails in the upper troposphere and oil platform plumes in the boundary layer at 3-year interval have shown that response time for measurement variations of NO2 vmr was not affected by more than a few seconds by this artefact.

A leak rate was found to be about 1.1 × 10−3 hPa L s−1, representing 1.4 × 10−4–8.1 × 10−5 of the sampled gas flow rate. If one considers the worst case, where the ambient trace-gas concentration is greater than that of the sampled air by one order of magnitude (such as might be in the aircraft cabin), the relative error resulting from leaks should not exceed 0.15%.

The linearity of SPIRIT response with respect to the vmr of the gases measured was tested. For CH4 and CO, the test consisted in using five gas cylinders containing calibrated vmr that were alternatively sent to SPIRIT inlet in a pseudo-random sequence. Each gas mixture was sent during 20 min at least three times during the exercise. The reference values for the five cylinders were obtained by CRDS measurement with a G2401 analyzer (PICARRO) at the ICOS ATC Metrology Lab (located at LSCE/IPSL, Gif-sur-Yvette, France) calibrated against the international primary scales (NOAA-2004 for CH4 and WMO-CO-X2004 for CO [24]). As illustrated in Figs. 5a and 6a, we found an excellent linear behavior for both species, with squared correlation coefficients R 2 better than 0.999 and no evidence of a high-order polynomial dependence on the linear fit residuals (Figs. 5b, 6b).

Fig. 5
figure 5

CO measurements of SPIRIT for five calibration cylinders (at 20.0 °C and 32.8 hPa, with pathlength of 134.22 m): a raw measurements with linear fit, b fit residuals calculated as the difference between the measured value and the linear fit, and c difference between the measured vmr and the reference vmr as a function of the reference vmr. Reference calibrated values are 44.73 ± 0.38, 113.94 ± 0.33, 187.31 ± 0.37, 358.10 ± 0.32, and 405.95 ± 0.29 ppbv

Fig. 6
figure 6

Same as Fig. 5, but for CH4 measurements. Reference calibrated values are 1775.59 ± 0.17, 1922.83 ± 0.19, 1955.38 ± 0.23, 2224.91 ± 0.24, and 2811.64 ± 0.35 ppbv

For CO, the main contribution to the difference between measurements and reference gas values for vmr lower than 200 ppbv (Fig. 5c) comes from a constant offset of 3.95 ppbv clearly visible on the linear fit parameters (Fig. 5a). To explain the origin of this bias, we analyzed pure nitrogen (≥999,997%) after it was sent into a CO cold trap (Sofnocat 514). Although the CO residual vmr should have been less than 0.1 ppbv, SPIRIT still detected a clear spectral signature, where the CO line represented 2.6 ± 1 ppbv of CO. A second test was performed, which consisted in interrupting the gas flow in the analyzer cell. After several hours, the CO vmr increased by more than a factor of 5 with respect to the vmr in the ambient air, refuting the possibility that this difference could result from a leak in the optical cell. This rather suggests the presence of a piece of the equipment outgassing CO (or another gas which exhibits an absorption line, where the CO line is measured) in the gas circuitry or inside the measurement cell. Unfortunately, despite the efforts made in this direction, we have not been able to identify this source. If this offset (Fig. 5a) is subtracted, the agreement between reference and measured values is very good (>99.2%). The small difference could result from a constant error on measured pressure, optical path length, or temperature inside the cell, but in all likelihood is due for the most part to the inaccuracy on the spectroscopic parameter given for this line in HITRAN (between 2 and 5%) [22].

For CH4, one can observe (Fig. 6a) a small constant bias of about −19.4 ppbv for an extrapolated reference value of 0 ppbv. However, considering the narrow span of calibrated gases used (from 1776 to 2812 ppbv) and a measurement dispersion of more than 40 ppbv peak-to-peak, this value cannot be considered meaningful, but still could result from a small deviation from linearity as we have no evidence of methane trapping in the cell. Once this bias is removed, the average difference is 2.5%, as shown in Fig. 6a, and is again consistent with the inaccuracy on the spectroscopic parameters given in HITRAN (between 2 and 5% for the line intensity) [22].

Consequently, the fit residuals (Figs. 5b, 6b) show no particular deviation from linear adjustment. At lower concentrations, these residuals cannot be suppressed to attain pure shot noise, since they are due to unavoidable interference fringes between the direct beam of the laser and the light scattered by optical components on the beam path, while at higher concentrations, these residuals mainly originate from inaccuracy in the line fitting, due to limited accuracy of the spectroscopic parameters.

A long-term comparison was made with an Ultraportable Greenhouse Gas Analyzer (UGGA, Los Gatos Research, USA), based on Off-Axis Integrated Cavity Output Spectroscopy (OA-ICOS) [14]. The comparison consisted of sampling outdoor air on the lab roof and regularly (every 2.5 h during 20 min) a target gas from a cylinder (whose CH4 vmr is fixed). As shown in Fig. 7, the two instruments exhibit a remarkable correlation, with relative differences remaining within ±1% for more than 33 h even with ambient CH4 level varying by more than 10%. The vmr sudden decreases to the lower values due to target sampling are repeatable (within 1%). A slight instrumental drift only arose after 13 h, which is difficult to be assigned to one of the instruments. The spikes lasting a few seconds, shown in the lower panel, corresponding to more than 1% difference between both instruments, are due to their different response times during transition regimes between ambient air and target gas samplings.

Fig. 7
figure 7

Comparisons between SPIRIT (grey color) and the UGGA from Los Gatos Research (blue color) for CH4 measurements in laboratory. Upper panel vmr measurements during more than 33 h for both instruments; Lower panel relative difference in % between both instruments

As NO2 propensity to degrade (especially in the presence of humidity) makes it difficult to obtain highly accurate mixing ratios without specially designed equipment even from calibrated cylinders, linearity had to be estimated differently. SPIRIT and an AC-31M (Environnement SA) were installed in parallel at the output of a dilution stage generating volume mixing ratios in the range 0–6 ppbv. The AC-31M is a double-channel chemiluminescence NO x analyzer and was calibrated just before the test. The correlation between both instruments for NO2 is rather good, given the uncertainties on the measurements of AC-31M (±0.5 ppbv) and SPIRIT (estimated by the standard deviation on the NO2 fitting retrieval residuals, largely predominant with respect to all other sources of uncertainties in this vmr range), with a coefficient of determination R 2 of about 0.979, as shown in Fig. 8. The NO2 vmr measured by SPIRIT is on average about 5% above the AC-31M value (indicated from the slope).

Fig. 8
figure 8

Laboratory comparison between SPIRIT (at 29.0 °C and 32.8 hPa, with pathlength of 134.22 m) and a chemiluminescence NO x analyzer. Error bars for SPIRIT are estimated by the standard deviation (1σ) on the NO2 fit residuals, the main contribution to the uncertainty budget in this measurement range. The accuracy displayed for the AC-31M (Environnement SA) is the one reported in the instrument datasheets (0.5 ppbv)

3.2 In-flight performances

An experiment to characterize the in-flight performances of SPIRIT about CO and CH4 was performed during a flight from the ATR-42, operated by the French Airborne Environment Research Service (SAFIRE, INSU—Météo France), on February 5, 2016. It consisted in alternating measurements of sampled air and of a gas mixture cylinder of CO and CH4 calibrated on the international primary scales (from ATC Metrology Lab, LSCE [24]), as illustrated in Fig. 9. The purpose of measuring a reference mixture along the flight was to assess long-term drifts (on the timescale of a flight) under the influences of the cabin temperature and the atmospheric temperature and pressure, in particular during atmospheric changes due to vertical profiling. One can see that there is no significant correlation between the altitude (from ground to 5700 m above sea level), i.e., the external atmospheric pressure and temperature, and the CO and CH4 calibrated values measured by SPIRIT.

Fig. 9
figure 9

ATR-42 flight on February 5, 2016. a Altitude and temperature profiles. b CO measurements. c CH4 measurements. For b and c, the grey crosses represent the moments when the calibrated gas cylinder content was measured and the black crosses the moments when the external air was measured. The transients during the passage from the gas cylinder to the external sampling have been removed from the plot to improve its readability

As expected (Fig. 9b), the CO vmr measured for the calibrated mixture shows a constant bias along the flight with a mean value of +3.8 ppbv and a standard deviation of 0.4 ppbv compared to the cylinder value. Considering the 0.4 ppbv accuracy on the cylinder vmr, this bias is in a very good agreement with the measurements performed on calibration cylinders at ground, which should theoretically lead to an overestimation of the mixing ratio for CO of about 2.9 ± 0.1 ppbv by SPIRIT for this 136.05 ppbv calibrated vmr (given the expression of the linear fit, as shown in Fig. 5a).

For CH4 (Fig. 9c), the average difference during the measurement with the calibration gas is +4.7 ppbv, with a standard deviation of 7.8 ppbv, which is less than the expected value (+29.1 ppbv) deduced from the ground calibration but still consistent; considering the extreme values from measurement of the calibration gas can differ by about 20 ppbv with respect to the reference calibrated gas, as shown in Fig. 6c.

One can also observe in Fig. 9 that the measured vmr for the gas cylinder vary along the flight duration. For CO and CH4, the relative maximum variations represent 0.7 and 1.46%, respectively. This “long-term” drift shows neither direct correlation with the temperature (within these %) nor with the pressure inside the cell, but results from the fringe pattern changing with the mechanical variation of the optical bench with temperature. This pattern can basically be seen as a sum of sinusoids which modulates the spectrum and whose frequencies are integer multiples of c/2L i, with c being the speed of light and L i the optical distance between two parallel optical diopters in the beam path. Unfortunately, the large number of diopters makes this pattern and its evolution eminently non-deterministic, so it is not possible to subtract it from the signal. Nevertheless, as it is inherently limited in amplitude and rate, this drift is hardly detrimental to measurement, since SPIRIT is designed primarily to characterize geophysical phenomena limited in time (a few minutes) and space (a few km), which are clearly distinguishable from background residual vmr.

Another characteristic of the instrument is the 1σ precision or sensitivity, i.e., the lowest vmr variation the instrument is able to reliably detect over a noisy background. Following a current practice in atmospheric trace-gas measurements [7, 24], sensitivities for CO and CH4 are deduced from the Allan deviation (a measure of the precision as a function of averaging time [25]), calculated here for the flight between 9:45 and 10:08 (UT). As illustrated in Fig. 10, these precisions for CO and CH4 are about 0.3 and 5 ppbv at 1.6 s, and 0.2 and 3 ppbv at 9 s averaging (the maximum response time of SPIRIT), respectively. Considering averaging times longer than a few minutes makes no sense in the context in which the instrument is used.

Fig. 10
figure 10

In-flight Allan deviation for CO and CH4. The minima are clearly visible at about 0.06 ppbv for CO (150 s averaging) and 1.5 ppbv for CH4 (for 150 s)

For NO2 in-flight measurement evaluation, a direct comparison has been made with two NO x chemiluminescent analyzers operated by SAFIRE in the ATR-42 aircraft: the Model 42C and 42S from Thermo Environmental Instruments (TEI). Usually, NO2 levels in the free troposphere above the French countryside are relatively low, below the detection limit of SPIRIT (0.3 ppbv; see below). During a flight (on May 27, 2014), the plane executed seven quick loops, which allowed to make measurements inside its own exhaust trails, where NO2 levels are several times higher than the background level. Results presented in Fig. 11 show the agreement between SPIRIT and both instruments on average within about 0.5 ppbv. Moreover, this value is consistent with the accuracy indicated by the TEI Company for these two models and with the results of the laboratory experiments when comparing SPIRIT to another commercial instrument (AC-31M, Fig. 8), giving thus an estimate of the in-flight accuracy for NO2. The instantaneous (1.6 s) 1σ precision of about 0.3 ppbv can also be roughly derived from the SPIRIT data scatter, as shown in Fig. 11c, without taking into account the lower frequency variations, and can be assimilated to the detection limit.

Fig. 11
figure 11

ATR-42 flight on May 27, 2014. Upper panel Latitude, longitude, and altitude profiles showing the six loops completed by the plane between 10 and 11.25 am. Insert zoom of the framed data around 10 UT. Middle panel Comparison of NO2 measurements by SPIRIT and TEI models 42C and 42S. NO2 increases result from the plane flying inside its exhaust wake. Lower panel difference in ppbv between SPIRIT and TEI models

Essential performances concerning the uncertainties of the measurements by SPIRIT are summarized in Table 2. The 1σ precisions at 1.6 s are remembered for each species. The overall uncertainties can be estimated from in-flight measurements. Comparisons with calibrated cylinder gases tied to WMO standard for CO, to NOAA standard for CH4, and with measurements from two different instruments for NO2 (models 42C and 42S from TEI), led to biases and associated standard deviations of 3.8 ± 0.4 = 4.2 ppbv for CO, 4.7 ± 7.8 = 12.5 ppbv for CH4, and 0.5 ppbv for NO2, taking into account the experimental scatters. The long-term drifts of 0.7% for CO and 1.5% for CH4 (Fig. 9) lead to ±0.5 and ±14 ppbv uncertainties at typical atmospheric vmr values of 150 and 1900 ppbv, respectively, whereas the comparison for NO2 with AC-31M instrument (from Environnement SA) in laboratory leads to 5% overestimation of SPIRIT values (Fig. 8). Directly adding these uncertainties with taking into account (or not) calibration leads to estimated overall uncertainties on vmr of 0.9 (or 4.7) ppbv for CO, 22 (or 27) ppbv for CH4, and 0.5 ppbv (+5%) for NO2.

Table 2 Precisions and overall uncertainties of in-flight measurements for CO, CH4, and NO2 vmr

4 Conclusions and perspectives

We presented the newly built instrument SPIRIT, a compact infrared absorption spectrometer with three quantum cascade lasers operating automatically on aircraft to measure at least three trace gases, for instance, CO, NO2, and CH4 in the present study. SPIRIT has already been applied with success in several large field campaigns, such as those linked to FP7 European projects SHIVA [26, 27] and DACCIWA–EUFAR–APSOWA [28, 29] and to national projects such as MISTRAL–CHARMEX [30]. Enhanced volume mixing ratios of CO, NO2, and/or CH4 of a few ppbv to tens of ppbv were observed during flights when crossing, in the upper troposphere convective clouds rapidly transporting near-surface polluted air, in the boundary layer pollutant emission plumes from offshore oil extraction platforms, and in the middle troposphere air masses containing biomass burning products transported over a long distance.

The main advantages of SPIRIT lie in its robustness and versatility, both arising from the use of a broadband multipass cell. It allows the laser to be changed easily to target species according to the scientific objectives. Thereby, other species such as CO2, ozone (O3), carbonyl sulfide (OCS), formaldehyde (HCHO), and hydrogen peroxide (H2O2) may be detected with performances that remain to be determined.

Several improvements can be foreseen. To address the problem of measuring gases particularly sticky or reactive, it appears necessary to improve the gas supply system. One improvement that could be easily achieved consists in reducing the residence time inside the cell and the gas lines. With the present pump capacities, it is possible to achieve gas flows more than five times as high just by replacing the orifice of the pressure regulator valve with a larger one. The residence time of the gas inside the instrument should thus be reduced by the same order of magnitude, without necessarily reducing the quality of the pressure regulation. Another aspect is to improve the compatibility of the wetted materials with gases, some surface treatments such as amorphous silicon coatings which seem to be able to ensure a relative chemical inertness and to drastically reduce adsorption. Heating the gas lines should also be considered as adsorption is highly dependent on the temperature of the walls.

A further improvement also includes regulating the temperature of the optical bench or at least the measurement cell. Although it would not result in an improvement of detection limit, because the short-term precision is mainly limited by relative changes in the optical path length of the order of the wavelength, resulting from vibrations from the pump and the plane, it should drastically enhance long-term accuracy and hence provide the opportunity to study phenomena having low spatial and temporal variability. This temperature regulation should stabilize the fringe pattern, and thus reduce the standard deviation of the average mixing ratios when combined to regular in-flight calibrations using reference cylinders. This should lower the minimum of the Allan deviation such as experienced in ground measurements by our team, reaching precisions below 0.3‰ [31]. The accuracy of the measured values depends on the calibration strategy deployed and the field constraints. When it is not possible to embark calibration gas cylinders, such as in the cases exemplified above, checking the absence of significant drift (i.e., within orders of magnitude of the biases plus the associated standard deviations) of known target gases before and after each flight is enough, since it does not compromise the findings of the pollution event studies. Sometimes also, the aircraft flies over calibrated measurement surface stations from the World Meteorological Organization/Global Atmosphere Watch/ICOS networks, which, taking into account the wind trajectories, allows validations of the measurements.

Finally, the opportunity to use Interband Cascade Lasers (ICLs) in the future should be considered for species having their fundamental vibrational transition lying in the 3–6-µm region. This type of laser offers interesting capabilities compared to QCLs: a higher operating temperature, a larger current tunability, and a better optical efficiency allowing for operation at lower electrical power (at the expense of a lower optical output power that could be limiting for the use of multipass cells). However, for the time being, this technology is still less mature than that of the QCLs, and being mastered by a single manufacturer makes ICLs less affordable than QCLs.