Abstract
Objectives
To evaluate the application of black-blood CT (BBCT) in carotid artery wall imaging and its accuracy in disclosing stenosis rate and plaque burden of carotid artery.
Methods
A total of 110 patients underwent contrast-enhanced CT scan with two phases, and BBCT images were obtained using contrast-enhancement (CE)-boost technology. Two radiologists independently scored subjective image quality on black-blood computerized tomography (BBCT) images using a 4-point scale and then further analyzed plaque types. The artery stenosis rate on BBCT was measured and compared with CTA. The plaque burden on BBCT was compared with that on high-resolution intracranial vessel wall MR imaging (VW-MR imaging). The kappa value and intraclass correlation coefficient (ICC) were used for consistency analysis. The diagnostic accuracy of BBCT for stenosis rate and plaque burden greater than 50% was evaluated by AUC.
Results
The subjective image quality scores of BBCT had good consistency between the two readers (ICC = 0.836, p < 0.001). BBCT and CTA had a good consistency in the identification of stenosis rate (p < 0.001). There was good consistency between BBCT and VW-MR in diagnosis of plaque burden (p < 0.001). As for plaque burden over 50%, BBCT had good sensitivity (93.10%) and specificity (73.33%), with an AUC of 0.950 (95%CI 0.838–0.993). Compared with CTA, BBCT had higher consistency with VW-MR in disclosing low-density plaques and mixed plaques (ICC = 0.931 vs 0.858, p < 0.001).
Conclusions
BBCT can not only display the carotid artery wall clearly but also accurately diagnose the stenosis rate and plaque burden of carotid artery.
Clinical relevance statement
Black-blood CT, as a novel imaging technology, can assist clinicians and radiologists in better visualizing the structure of the vessel wall and plaques, especially for patients with contraindication to MRI.
Key Points
• Black-blood CT can clearly visualize the carotid artery wall and plaque burden.
• Black-blood CT is superior to conventional CTA with more accurate diagnosis of the carotid stenosis rate and plaque burden features.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Carotid atherosclerotic plaque, especially vulnerable plaque, is one of the important risk factors for ischemic stroke [1]. The vulnerability of carotid plaque and the stenosis extent of carotid artery are the key issues that clinicians concern [2, 3]. Various carotid artery imaging are helpful in identifying atherosclerotic plaques, such as ultrasound, conventional computed tomography angiography (CTA), magnetic resonance imaging (MRI), and digital subtraction angiography (DSA) [4]. Many clinical studies have shown that unstable plaque and stenosis of carotid artery are closely related to ipsilateral stroke [5]. VW-MR imaging is superior to CTA on revealing the arterial wall features and plaque components [6, 7], but VW-MR imaging takes a long scan time, and the claustrophobic patients and the patients with some metal implants are not suitable to MRI [8]. Although qualitative diagnostic methods for plaque have been established using ultrasound and MRI, CT is indispensable for evaluating carotid artery stenosis and plaque for the following reasons: (1) Clinically, CT and CTA are the most commonly used examination in diagnosis of cerebrovascular disease events with the advantages of fast scanning and few contraindications. Doctors can quickly obtain the images and assess the carotid artery stenosis accurately. Besides this, CTA also has the advantages of arbitrary orientation reconstruction and sensitively displaying calcification. (2) Compared with ultrasound, CT is more objective and less affected by the operator. Thus, CT is still kept in the first line of the daily clinical work. However, CTA has limitations in displaying artery wall and vulnerable plaque components.
Contrast-enhancement-boost (CE-boost) is one of the CT technologies using subtraction technique and registration algorithm, which attenuates the motion influence of different phases and provides accurate registration [9]. CE-boost can enhance the brightness of blood vessels by subtracting non-contrast CT images from enhanced-contrast CT images [10]. Similar with CE-boost, subtraction technology and registration algorithm are also used in generating BBCT. A set of iodine mappings are acquired by subtracting arterial-phase CT images from the delayed-phase images, and then, the BBCT images are obtained by superimposing the iodine images on the delayed-phase CT images. Thus, the vascular lumen turns to be black on both BBCT images and the iodine mappings. A previous study has shown that delayed-phase CT images can provide more information of plaques [11]. As the vascular wall has the characteristics of delayed enhancement, the lumen and wall structure of carotid artery can be distinguished on BBCT by superimposing the delayed-phase images and iodine mappings. BBCT improves the exhibition of the arterial wall, which is helpful in the evaluation of carotid stenosis and plaque burden (Fig. 1).
It is well known that unstable carotid plaque and stenosis are closely related to ischemic stroke; in this study, we presented a novel method of carotid artery wall imaging using BBCT to evaluate not only stenosis but also the plaque burden of the carotid arteries. Meanwhile, the image quality and diagnostic value of BBCT in carotid artery imaging were evaluated. The stenosis degree and plaque features on BBCT were analyzed and compared with those on conventional CTA and VW-MR imaging.
Methods
Study population
From January to May 2022, a total of 110 patients received carotid artery enhanced CT scans in our hospital, including 85 males (77.3%) and 25 females (22.7%), with the mean age of 68.7 ± 9.1 years. Inclusion criteria were as follows: (1) age ≥ 18 years old; (2) atherosclerotic plaques in the cervical carotid artery confirmed by ultrasound, CT, or MRI; (3) less than 2 weeks’ interval between VW-MRI and BBCT. Exclusion criteria were as follows: (1) history of iodine allergy; (2) patients with severe heart, liver, lung, renal failure, or hematological diseases; (3) severe artifact. Six patients were excluded due to severe image artifacts. The reason was that the patients were seriously weak and dysphoria during the examination. This study was approved by the local Institutional Research Ethics Committee, and informed consent was obtained from all patients.
CT scan techniques and image postprocessing
All CT angiography examinations were performed on a 320-row CT scanner (Aquilion ONE Genesis, Canon Medical System). Nonionic iodine contrast agent (370 mg/ml) was intravenously injected with the dosage of 0.6 ml/kg, followed by 30 ml saline with a two-channel high-pressure injector. After contrast injection, two phases scanning of CT angiography was performed, including the arterial phase and delayed phase. Arterial-phase scanning parameters: the monitoring layer was located at the middle level of common carotid artery with manually triggered; tube voltage 120 kV, automatic milliampere. The delayed-phase scanning parameters: 150 s after injection of contrast agent, automatic trigger scanning, tube voltage of 120 kV, automatic milliampere. The scanning range of both phases was the same, from the aortic arch to the sellar with slice thickness of 1 mm.
The BBCT images were obtained by subtracting the arterial-phase images from the delayed-phase images using the analysis software installed in the CT workstation (Aquilion ONE, Canon Medical Systems).
VW-MR scan techniques
All scans were performed using a 3.0-T MR scanner (Magnetom Prisma, Siemens Healthineers) with 64-channel coil. The imaging sequences included T1-weighted imaging (TIWI) and contrast-enhanced T1-weighted imaging (CE-T1WI). TIWI using sampling perfection with application-optimized contrasts by using different flip angle evolutions (SPACE) sequence: TR 800 ms, TE 16 ms, FOV 250 mm × 164 mm, voxel size 0.6 mm × 0.6 mm × 0.6 mm. Gadopentetate dimeglumine (Magnevist, Schering) was administered intravenously (0.1 mmol/kg body weight), and T1W imaging was repeated 5 min after contrast administration.
Subjective image quality evaluation
Two radiologists independently evaluated the images of 110 patients. Readers were blinded to the patient’s clinical information. The scoring criteria were as follows: a score of 4 (best image quality) indicated high signal noise ratio (SNR), and the lumen, the wall, and the plaques were most clearly displayed; 3 (good image quality) indicated relatively high SNR, and the lumen, the wall, and the plaque were displayed well; 2 (medium image quality) indicated moderate SNR, and the lumen and wall could be easily distinguished, but the plaque was not clear enough; 1 (poor image quality) indicated low SNR, and the lumen and wall were poorly displayed, and the plaque was blurred. Points ranged from 2 to 4 indicated diagnostic quality, and points of 1 indicated nondiagnostic quality (Fig. 2).
Comparison between BBCT and CTA
Carotid artery plaque recognition
Two radiologists independently analyzed carotid artery plaque on BBCT images using commercial three-dimensional (3D) workstation software (RadiAnt DICOM Viewer, Medixant). Two hundred and twenty vessels from 110 patients were analyzed. The plaque at the narrowest level of the carotid artery was selected for analysis. Calcified plaque, mixed plaque, low-density plaque, and no plaque were recorded as categorical variables 1 to 4, respectively. The calcified plaque was defined with a CT value greater than 130HU, low-density plaque was made of fat and fibers, some with intra-plaque hemorrhage, and mixed plaque contained both calcified and low-density plaque. BBCT results of reader 1 and reader 2 are denoted as BBCTR1 and BBCTR2, respectively. CTA was reviewed by a senior radiologist. The plaque types on BBCT were compared with those on conventional CTA.
Degree of vascular stenosis
Carotid artery stenosis rates were measured according to the European Carotid Surgery Trial (ECST) [12]. Carotid arteries with plaque diameter larger than 5 mm were selected for stenosis rate calculation. All CTA and BBCT original images were loaded in the workstation first to generate curved planar reconstruction (CPR) images. Then, two experienced radiologists checked images and made a manual correction to ensure the centerline was in the center of the lumen. The stenosis rate was measured on CPR images where the plaques were shown best, and the measurement positions were located at the narrowest level of the lumen doublechecked on three consecutive CPR images. The mean value of the measurement results on the three consecutive CPR images was used (Fig. 3).
Comparison of plaque burden between BBCT and VW-MR imaging
Twenty-two of 110 patients underwent VW-MR imaging. The carotid vascular area and lumen area at the most stenotic level were manually delineated on both BBCT images and VW-MR enhanced images; it takes about 3–5 min to sketch each plaque burden. Alignment was performed using software Radiant DICOM viewer with carotid bifurcation or bone structure as the reference. Sagittal or coronal reconstruction will be used if necessary. The vascular area included the vascular wall area and lumen area. The plaque burden was calculated using the following formula [13, 14]: \(\mathrm{plaque}\;\mathrm{burden}\;\left(\%\right)=\left[\left(\mathrm{vascular}\;\mathrm{area}-\mathrm{lumen}\;\mathrm{area}\right)/\mathrm{vascular}\;\mathrm{area}\right]\times100\%\).
All the low-density plaques and mixed plaques were identified in 44 vessels. The plaque burden was analyzed at the level where these plaques were disclosed on BBCT, CTA, and VW-MR images. Then two radiologists measured and compared the plaque volume on BBCT and VW-MRI using software ITK-SNAP 3.6 (Fig. 4).
Statistical analysis
IBM SPSS 26.0 and MedCalc 19.3 software were used for statistical analysis. Numerical variables were expressed as mean ± standard deviation (SD). Two-tailed p < 0.05 indicated a significant difference. The consistency of diagnostic accuracy was evaluated by κ statistics. The κ value indicates the degree of consistency: 0–0.20, slight agreement; 0.21–0.40, fair agreement; 0.41–0.60, moderate agreement; 0.61–0.80, substantial agreement; and 0.81–1.00, perfect agreement. The consistency of the inter-observer image quality score, the stenosis rates of BBCT and CTA, and the plaque burden of BBCT, CTA, and VW-MR imaging were evaluated by intraclass correlation coefficient (ICC). Receiver operating characteristic (ROC) curve and area under curve (AUC) value were used to evaluate the diagnostic accuracy of BBCT for stenosis rate and plaque burden greater than 50%.
Results
Subjective image quality evaluation
Good image quality scores of BBCT were obtained from all the 110 patients; the scores of BBCT obtained by the two readers were 3.32 ± 0.67 and 3.29 ± 0.69, respectively. The consistency of image quality evaluation between the two readers was good (ICC = 0.836, p < 0.001). Regarding the plaque types in 220 carotid arteries, there was a high consistency between the two readers (κ = 0.886, p < 0.001). Meanwhile, the consistency values between BBCT and CTA of the two readers were moderate and high, respectively (κ1 = 0.778, p < 0.001; κ2 = 0.851, p < 0.001) (Table 1).
Degree of vascular stenosis
Among the 220 arteries, 87 carotid arteries were selected for stenosis rate calculation. There was a high consistency between BBCT and conventional CTA (ICC = 0.859; p < 0.001) in displaying the vessel stenosis. Among 87 vessels, stenosis greater than 50% was revealed in 37 vessels on BBCT, and 29 on conventional CTA (Table 2). Sensitivity and specificity of BBCT were 96.55% (28/29) and 84.48% (49/58), respectively. The AUC of BBCT for the detection of stenosis rates greater than 50% was 0.933 (95%CI 0.859–0.976, p < 0.001) (Fig. 5).
Plaque burden
Plaque burden was analyzed in 44 vessels of 22 patients who underwent VW-MR. The plaque burden was highly consistent between BBCT and VW-MR imaging (ICC = 0.921; p < 0.001). Plaque burden was identified greater than 50% in 31 vessels on BBCT, and in 29 vessels on VW-MR imaging (Table 2). Sensitivity and specificity of BBCT were 93.10% (27/29) and 73.33% (11/15), respectively. The AUC of BBCT for the detection of plaque burden greater than 50% was 0.950 (95%CI 0.838–0.993, p < 0.001) (Fig. 5).
A total of 29 low-density plaques and mixed plaques were identified and measured. Compared with CTA, BBCT had higher consistency with VW-MR in disclosing low-density plaques and mixed plaques (ICC = 0.931 vs 0.858, p < 0.001) (Fig. 6). The plaque volume assessed by two radiologists showed a high level of consistency between BBCT and VW-MRI (ICC1 = 0.872, p < 0.001, ICC2 = 0.902, p < 0.001) (Table 1).
Radiation dose
In this study, the radiation dose in the delayed phase was 1.84 ± 0.47 mSv, and the radiation dose of the whole scan was 3.82 ± 0.95 mSv.
Discussion
In this study, we evaluated the subjective image quality of BBCT and its diagnostic accuracy of plaque types. The stenosis rate and plaque burden on BBCT were compared with those on conventional CTA and VW-MR imaging. The BBCT was highly consistent with CTA and VW-MR imaging in displaying carotid artery wall as well as plaque type identification. High specificity and sensitivity of carotid artery stenosis rate and plaque burden greater than 50% were obtained.
Subtraction technology has been widely used in CT [15, 16]. However, highlighting the vascular wall imaging on CT scan is still very challenging to date. In addition to carotid atherosclerosis, there are many other diseases that require imaging of the arterial wall [17,18,19], such as Takayasu arteritis, arterial dissection, and mural thrombus. BBCT depresses the enhanced CT value in the lumen and simultaneously highlights the presentation of the carotid artery wall. To our knowledge, this is the first study to explore the carotid artery wall features based on BBCT images utilizing the subtraction technique. Dr. Steven [20] used Cinematic rendering (CR) to generate black-blood cinematic rendering (BBCR) images for the visualization of cardiac structures. Dr. Rotzinger [21] developed a dual-energy CT algorithm and used two-material decomposition analysis to generate dark-blood CT, aiming to better visualize the aortic wall and aortic intramural hematoma. In this study, we explored BBCT on the carotid arteries. Our results showed that the BBCT can display not only the lumen stenosis but also the wall structure of the carotid artery very well. The BBCT images obtained by subtraction technology can be processed on the workstation with about 3 min on each set of images, which is easy and convenient.
CTA is well known as an effective technique for the diagnosis of carotid artery stenosis with high sensitivity and has a high consistency with DSA which is gold standard [22]. The degree of carotid artery stenosis and plaque type are the main concerns of clinicians. Thus, CTA was used as a criterion for evaluating carotid stenosis in this study, and the results of this study showed a good consistency between BBCT and CTA.
Plaque burden is one of the indicators of the severity of carotid atherosclerosis. Morphology and stability of the plaque are closely related to a previous history of cerebrovascular diseases [23]. Accurate evaluation of plaque burden is important due to its predictive value on ischemic events [24]. Ultrasound and MRI are generally used in revealing the wall and lumen of carotid artery. VW-MR has shown great value in identifying atherosclerotic plaque components compared with pathology, such as intra-plaque hemorrhage, fibrous cap disruption, and lipid-rich necrotic core [25]. We therefore used VW-MR as the gold standard for plaque burden analysis. The results of this study showed that BBCT identified plaque types and stenosis rate with high quality. Compared with CTA, the measurement of plaque burden by BBCT was more consistent with VW-MR. BBCT clearly displayed the carotid artery wall, had better diagnostic value for carotid plaque stenosis rate, and accurately identified the types of plaque. BBCT was superior to VW-MR in short scanning time and less contraindications. Moreover, MPR, curved planar reconstruction (CPR), and other postprocessing techniques can be used in BBCT to display the plaque manifestations more clearly and conveniently. Despite this, MRI still has irreplaceable advantages in the diagnosis of plaque properties.
We added a delayed phase to the examination process, and the radiation dose was 1.84 ± 0.47 mSv for the delayed phase only and the radiation dose of the whole scan was 3.82 ± 0.95 mSv. According to previous studies, dose of routine chest CT scan was about 3–7 mSv [26, 27]. Therefore, BBCT examination does not increase extra radiation risk of patients, and patients with carotid atherosclerosis benefit more from the evaluation.
This study has some limitations. First, as this is a single-center study, the number of patients who underwent VW-MR was small. Second, the available data were not sufficient for in-depth analysis on the details of low-density plaques, and we will continue the work in the future.
In conclusion, BBCT can display the carotid artery wall and lumen more precisely when compared with conventional CTA, improving the detection of low-density plaque. The combination of BBCT and CTA can further identify plaques more comprehensively. Moreover, BBCT is superior to VW-MR with less contraindication and shorter scanning time. It provides a novel option for clinicians to accurately evaluate carotid atherosclerotic plaques.
Abbreviations
- AUC:
-
The area under the receiver operating characteristic curve
- BBCT:
-
Black-blood computerized tomography
- CE-boost:
-
Contrast-enhancement-boost
- CPR:
-
Curved planar reconstruction
- CTA:
-
Computed tomography angiography
- DSA:
-
Digital subtraction angiography
- ECST:
-
European Carotid Surgery Trial
- ICC:
-
Intraclass correlation coefficient
- MPR:
-
Multi-planar reformatting
- MRI:
-
Magnetic resonance imaging
- ROC:
-
Receiver operating characteristic
- SD:
-
Standard deviation
- SNR:
-
Signal noise ratio
- VW-MR:
-
Vessel wall magnetic resonance imaging
References
Saba L, Saam T, Jäger HR et al (2019) Imaging biomarkers of vulnerable carotid plaques for stroke risk prediction and their potential clinical implications. Lancet Neurol 18:559–572
Brinjikji W, Huston J, Rabinstein AA, Kim G, Lerman A, Lanzino G (2016) Contemporary carotid imaging: from degree of stenosis to plaque vulnerability. J Neurosurg 124:27–42
Abbott AL, Paraskevas KI, Kakkos SK et al (2015) Systematic review of guidelines for the management of asymptomatic and symptomatic carotid stenosis. Stroke 46:3288–3301
Motoyama R, Saito K, Tonomura S et al (2019) Utility of complementary magnetic resonance plaque imaging and contrast-enhanced ultrasound to detect carotid vulnerable plaques. J Am Heart Assoc 8:e011302
Mac Grory B, Emmer BJ, Roosendaal SD, Zagzag D, Yaghi S, Nossek E (2020) Carotid web: an occult mechanism of embolic stroke. J Neurol Neurosurg Psychiatry 91:1283–1289
Saba L, Antignani PL, Gupta A et al (2022) International Union of Angiology (IUA) consensus paper on imaging strategies in atherosclerotic carotid artery imaging: from basic strategies to advanced approaches. Atherosclerosis 354:23–40
Saloner D, Acevedo-Bolton G, Wintermark M, Rapp JH (2007) MRI of geometric and compositional features of vulnerable carotid plaque. Stroke 38:637–641
Jungmann PM, Agten CA, Pfirrmann CW, Sutter R (2017) Advances in MRI around metal. J Magn Reson Imaging 46:972–991
Baerends E, Oostveen LJ, Smit CT et al (2018) Comparing dual energy CT and subtraction CT on a phantom: which one provides the best contrast in iodine maps for sub-centimetre details? Eur Radiol 28:5051–5059
Xu J, Wang S, Wang X et al (2022) Effects of contrast enhancement boost postprocessing technique in combination with different reconstruction algorithms on the image quality of abdominal CT angiography. Eur J Radiol 154:110388
Horie N, Morikawa M, Ishizaka S et al (2012) Assessment of carotid plaque stability based on the dynamic enhancement pattern in plaque components with multidetector CT angiography. Stroke 43:393–398
Warlow C (1991) MRC European Carotid Surgery Trial: interim results for symptomatic patients with severe (70–99%) or with mild (0–29%) carotid stenosis. Lancet 337:1235–1243
Brunner G, Virani SS, Sun W et al (2020) Associations between carotid artery plaque burden, plaque characteristics, and cardiovascular events. Jama Cardiol 6:1–8
Lyu J, Ma N, Tian C et al (2019) Perfusion and plaque evaluation to predict recurrent stroke in symptomatic middle cerebral artery stenosis. Stroke Vasc Neurol 4:129–134
Grob DJM, Smit EJ, Prince JL et al (2019) Iodine maps from subtraction CT or dual-energy CT to detect pulmonary emboli with ct angiography: a multiple-observer study. Radiology 292:197–205
Chen GZ, Luo S, Zhou CS, Zhang LJ, Lu GM (2017) Digital subtraction CT angiography for the detection of posterior inferior cerebellar artery aneurysms: comparison with digital subtraction angiography. Eur Radiol 27:3744–3751
Treitl KM, Maurus S, Sommer NN et al (2017) 3D-black-blood 3T-MRI for the diagnosis of thoracic large vessel vasculitis: a feasibility study. Eur Radiol 27:2119–2128
Luo Y, Guo Z, Niu P et al (2016) 3D T1-weighted black blood sequence at 3.0 Tesla for the diagnosis of cervical artery dissection. Stroke Vasc Neurol 1:140–146
Henningsson M, Malik S, Botnar R, Castellanos D, Hussain T, Leiner T (2022) Black-blood contrast in cardiovascular MRI. J Magn Reson Imaging 55:61–80
Rowe SP, Chu LC, Recht HS, Lin CT, Fishman EK (2020) Black-blood cinematic rendering: a new method for cardiac CT intraluminal visualization. J Cardiovasc Comput Tomogr 14:272–274
Rotzinger DC, Si-Mohamed SA, Shapira N, Douek PC, Meuli RA, Boussel L (2020) “Dark-blood” dual-energy computed tomography angiography for thoracic aortic wall imaging. Eur Radiol 30:425–431
Josephson SA, Bryant SO, Mak HK, Johnston SC, Dillon WP, Smith WS (2004) Evaluation of carotid stenosis using CT angiography in the initial evaluation of stroke and TIA. Neurology 63:457–460
Wendorff C, Wendorff H, Pelisek J et al (2015) Carotid plaque morphology is significantly associated with sex, age, and history of neurological symptoms. Stroke 46:3213–3219
Spence JD (2015) Measurement of carotid plaque burden. JAMA Neurol 72:383–384
Cai J, Hatsukami TS, Ferguson MS et al (2005) In vivo quantitative measurement of intact fibrous cap and lipid-rich necrotic core size in atherosclerotic carotid plaque. Circulation 112:3437–3444
Kalra MK, Homayounieh F, Arru C, Holmberg O, Vassileva J (2020) Chest CT practice and protocols for COVID-19 from radiation dose management perspective. Eur Radiol 30:6554–6560
Ohno Y, Koyama H, Seki S, Kishida Y, Yoshikawa T (2019) Radiation dose reduction techniques for chest CT: principles and clinical results. Eur J Radiol 111:93–103
Funding
This study has received funding by 2023 National High Level Hospital Clinical Research Funding (project No. BJ-2023-076), 2020 SKY Imaging Research Fund of the Chinese International Medical Foundation (project No. Z-2014-07-2003-02), 2020 Beijing Hospital “National Natural Science Foundation of China Preliminary Research Project” (project No. BJ -2020-131), and Fundamental Research Funds for the Central Universities (project No. 3332022168).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Juan Chen.
Conflict of interest
Zhuangfei Ma is an employee of Canon Medical Systems (China) Co., Ltd. The remaining authors declare no relationships with any companies whose products or services may be related to the subject matter of the article.
Statistics and biometry
No complex statistical methods were necessary for this paper.
Informed consent
Written informed consent was obtained from all subjects (patients) in this study.
Ethical approval
This study was approved by the Institutional Research Ethics Committee of Beijing Hospital.
Study subjects or cohorts overlap
None
Methodology
• Prospective
• Diagnostic or prognostic study
• Performed at one institution
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lu, Y., Cao, R., Jiao, S. et al. A novel method of carotid artery wall imaging: black-blood CT. Eur Radiol 34, 2407–2415 (2024). https://doi.org/10.1007/s00330-023-10247-5
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-023-10247-5