Introduction

A relatively wide range of incidence rates of developmental dysplasia of the hip (DDH) is reported among the literature. Besides cohort dependent factors different definitions and screening methods for DDH are known issues in this context. Thus, a standardized sonographic hip screening method like the sonographic method according to Graf et al. is expected to increase the reliability and comparability of given incidence rates [1].

Several studies were conducted and proved Graf’s broadly used sonographic screening method for DDH to be reliable and reproducible [24]. However, previous studies focused on the method of measurement on given sonographic images, whereas the process of acquisition of sonographic images was not examined. Considering the clinical importance of Graf’s method in established national screening programs for DDH in many European countries, it is interesting that an essential issue like the link of transducer inclinations to measurement results was not examined before [5, 6].

The screening method according to Graf is based on a defined, standardized transducer position [7, 8]. The transducer position is determined by the appearance of anatomical structures (landmarks) and their defined cross-sectional shape on the sonographic images. Graf described the standardized transducer positon as a position within an ideal plane defined by his criteria. However, in reality the criteria define a set of possible transducer positions including inclinations in the axial and frontal plane. Interestingly there is only little evidence on the tolerance of transducer positions and the resulting effect on measurement results [9]. Despite the lack of evidence on the tolerance of transducer positions and its effect on measurement results, the problem has been reported and technical improvements (“Sono-Guide”) were suggested in order to limit this cause of examiner dependent measurement deviations [10]. However, despite the use of this standardized method a relatively wide range (0.24% to 4.7%) of incidence rates of DDH were reported [1115].

Thus, one question remains: How big is the examiner dependent influence in the process of acquisition of sonographic images?

The objectives of this study were (1) to examine the permissible tolerance of the transducer positioning and (2) to analyse the resulting effect on the hip measurement and grading system.

Materials and methods

Institutional review board approval was obtained for this study at our institution. All patients or legal guardians gave their informed consent prior to the inclusion in this study.

Twelve consecutive newborns (24 hips) delivered at one hospital were included in the study. A feed-and-sleep technique was used to put the newborns to sleep prior to the measurement. The sonographic screening process was video monitored. The examination was interrupted when the newborns started to move.

The newborns were screened using Graf’s ultrasound method within the first neonatal week. Ultrasound screening was performed using a GE logiq 200 pro series system and a 7.5-MHz LH linear transducer (GE Healthcare, Milwaukee, WI, USA). The newborns were examined in lateral position using a positioning device (Sono-Fix, Gebrüder Hirschbeck GmbH, Austria) as recommended by Graf [9]. All newborns underwent a standardized physical examination including the Barlow and Ortolani tests [16, 17], test of hip abduction, assessment of femoral length discrepancies, asymmetries and associated deformities.

Three-dimensional positions in space of the transducer during the whole ultrasound screening process were recorded using an optoelectronic motion capture system Smart-E (BTS Bioengineering, Milan, Italy). The motion capture system is based on optical technology using four infrared cameras sampling at 120 Hz. Passive infrared markers were rigidly attached to the transducer. A three-dimensional reconstruction of trajectories of the markers was preformed using the Smartanalyzer software (BTS Bioengineering, Milan, Italy). Positions in space of the transducer synchronized to points in time of the sonographic image acquisitions were extracted.

In each new-born, sonographic images of the hip were obtained in the following order starting with a standardized position according to Graf which felt most neutral to the examiner. In this neutral position the examiner was encouraged to avoid any specific inclination of the transducer. Subsequently sonographic measurements in a position with a tendency of inclination to anterior, posterior, cranial and caudal were obtained. The amount of inclination in these positions was limited with respect to the criteria described by Graf. Finally, a transducer position using a combination of posterior and cranial inclination was examined (Fig. 1 and Fig. 2).

Fig. 1
figure 1

Photograph of the measurement setup, showing a newborn lying in the positioning device and the passive infrared markers rigidly attached to the transducer. Unit vectors of the transducer position are represented by three white arrows, unit vectors of the coordinate system relative to the positioning device are represented by the blue, red and green arrows. Transducer inclination to anterior and posterior (blue double arrow) and transducer inclination to cranial and caudal (red double arrow) were measured as rotation around the blue and red unit vector relative to the neutral transducer position

Fig. 2
figure 2

Series of sonographic images of the same physiological hip joint (Graf type Ib) acquired in different transducer positions and measurement lines according to Graf (red lines) showing an α-angle range of 54° to 71°. Neutral position (N) α = 64, β = 60; anterior position (A) α = 65, β = 60; dorsal position (D) α = 64, β = 65; caudal position (Ca) α = 54, β = 76; cranial position (Cr) α = 65, β = 62; and dorsocranial position (DCr) α = 71, β = 64

The analysis of the sonographic images was performed blinded to patient and transducer position data. All sonographic images were checked according to Graf’s criteria prior to further analysis and only appropriate sonographic images were processed. Sonographic measurements of alpha- and beta-angels according to Graf and matched transducer position in space were analysed. Based on the defined transducer positions differences of transducer positions in space, and alpha- and beta-angles to the neutral position were calculated.

Statistical analysis

Data were processed using the SPSS 20 software (SPSS Inc., Chicago, IL, USA). Considering the coarsening nature of classification systems α- and β-angels were first analysed directly [18]. Differences between sonographic α- and β-angels on different transducer positions were analysed using two-sided paired t-tests (p < 0.05). The homogeneity of variances was analysed using the Levene-test. Secondly, the ordinal results based on Graf’s grading system in different transducer positions compared to the neutral position were analysed using the Wilcoxon signed-rank test. The level of significance was defined as p < 0.05.

Results

The feed-and-sleep technique was sufficient in ten of the 12 newborns. In seven of the ten newborns both hips were measured according to the study protocol, in three newborns only one hip was measured due to movement of the newborn. Primary sonographic results according to Graf’s technique were type Ia in 3 (17.6%), Ib in 11 (64.7%) and IIa in 3 (17.6%) hips. The results of the physical examination were normal in all newborns.

Three-dimensional (3D) measurement of defined transducer positions was performed in all measured hips: The inclination-angles of transducer positions in space relative to the neutral transducer position for anterior, posterior, cranial and caudal inclination are shown in Fig. 3. The fifth transducer position representing a combination of posterior and cranial inclination showed a mean posterior inclination of 4.2° (range 0.1° to 8.7°) and a mean cranial inclination of 10.5° (range 5.1° to 16.5°). In total we found a standard deviation of 6.6° for the α-angle, and a standard deviation of 6.6° for the β-angle.

Fig. 3
figure 3

Inclination-angles in degrees of transducer position in space relative to the neutral position to anterior, posterior, cranial and caudal

The impact of the defined transducer positions on the sonographic measurement of α- and β-angels relative to the sonographic measurements in neutral transducer position is given in Figs. 4 and 5. Statistical analysis revealed a significant difference of the measured α-angles in posterior-cranial, cranial and caudal transducer positions to α-angles measured in neutral position (t16 = 7.60, p < 0.001; t15 = 3.01, p = 0.009; t16 = −7.00, p < 0.001). Differences of α-angles in anterior and posterior inclined transducer positions to α-angles measured in neutral position were not significant (t16 = 0.11, p = 0.912; t16 = 0.87, p = 0.397).

Fig. 4
figure 4

Impact of inclined transducer positions on sonographic measurement of α-angles

Data on the frequency of hip types according to Graf seen in the defined transducer positions are given in Fig. 5. Compared to the neutral position type IIa was more frequent in anterior and posterior inclined transducer positions, and less frequent in cranial inclined transducer positions. These differences were not statistically significant (p = 0.32, p = 0.66 and p = 0.56). Compared to the neutral position sonographic measurements in caudal inclined transducer positions substantially increased the number of IIa hips, and moreover type IIc and D were seen in this position. This difference was statistically significant (p < 0.001). All hips were classified as Graf type Ia or Ib using a combination of posterior and cranial inclination for the transducer position. However, compared to the neutral position this difference was not significant (p = 0.083).

Fig. 5
figure 5

Frequency (in percent) of the observed sonographic hip types according to Graf (Ia, Ib, IIa, IIc and D) in neutral (highlighted by the red box) and inclined transducer positions

Discussion

It is an interesting fact that incidence rates of DDH based on sonographic screening programs show a relatively wide range of variation (0.24 to 2.5%) [11, 13, 14]. Cohort dependent factors were discussed as a cause of varying incidence rates, but on the other hand examiner dependent factors have also been discussed. The standardization of the sonographic hip screening of DDH according to Graf is an important achievement improving reliability and reproducibility in this field [8]. The achievement of Graf was first to rule out, or minimize examiner dependent factors using a standardized transducer position, but there is only weak evidence on the reliability of the criteria defining the standardized transducer position. Considering the clinical importance of Graf’s method in established national screening programs for DDH in many European countries [10], it is interesting that there is little data on the acceptable amount of transducer inclination and finally on the effect of such transducer movements on sonographic hip screening results. Other authors have shown the method of Graf to give reproducible and reliable results, but focused on the second step of measuring angles on given sonographic images [2, 19]. In contrast, this study focuses on the impact of examiner dependent factors in the process of sonographic image acquisition prior to the analysis of sonographic images.

The objectives of this study were (1) to examine the permissible tolerance of the transducer positioning and (2) to analyse the resulting effect on the hip measurement and grading system.

The results of the three-dimensional analysis of transducer positions show that different inclinations of transducer positions within the criteria according to Graf are possible. In the axial plane we found permissible anterior inclination of 8.8° and a permissible posterior inclination of 8.1°. In the frontal plane we found a permissible caudal inclination 15.4° and a permissible cranial inclination of 7.2°. When testing a combination of posterior and cranial inclination for the transducer position we found a permissible combination of 4.2° of posterior inclination and 10.5° of cranial inclination.

The effect of transducer inclination in the frontal plane is significant for both directions: a caudal inclined transducer position significantly reduces the measured α-angles, whereas a cranial inclined position significantly increases the measured α-angles. A combination of posterior and cranial inclination leads to significantly increased α-angles. Finally, the effect of different transducer positions on the sonographic measurement of α-angles according to Graf showed a non-significant impact considering anterior and posterior inclined transducer positions.

The standard deviation of α- and β-angle found in this study is higher than reported previously. In comparison Roovers et al., who examined the reproducibility of the analysis of sonographic images, found a standard deviation of 3.2° for the α-angle, and a standard deviation of 6.0° for the β-angle [20]. This supports the assumption that measuring inaccuracies of the α-angle, which is the central parameter for Graf’s classification, mainly originate from the image acquisition process. However, we have to consider that measurement deviations as described by Roovers will add up to our results in the second step of analysis of the sonographic images.

The final effect of different transducer inclinations on the results classified according Graf’s system is highly significant for the caudal position. Here, former normal hips were classified as type IIc and D. Therefore, this effect has potential to contribute to higher rates of reported DDH [13, 14].

On the contrary, this effect was reversed using a combination of posterior and cranial inclination of the transducer. In this transducer position all examined hips were classified as type Ia or type Ib, and no hips classified as type IIa anymore. An isolated cranial transducer inclination also shows a limited effect with slightly more hips classified as type Ia and type Ib comprared to the neutral position. Despite significant effects of these transducer positions on the α-angels, the effect on the results according to Graf was not significant. This difference may be explained by the coarsening nature of the classification system. However, these effects may decrease the frequency of DDH especially in the first neonatal weeks in some series [11]. Considering transducer inclinations to anterior or posterior, the effect on the classification system is minimal (see Fig. 5).

In this study, the sonographic measurements showed physiological results in all included subjects. Therefore, the meaning of our results in pathologic conditions cannot be predicted directly. Based on the clinical experience of the investigators, we believe that the definition of the standard plane is less easily applicable in pathologic conditions, like Graf’s type IIc, D. As a consequence, the placement of the standard plane might be less reproducible in these pathologic conditions, likely resulting in an increased variance of the sonographic α- and β-angles. Accordingly, this may lead to uncertainty in the classification of these hips requiring well-skilled examiners to provide a counterbalance. Some authors addressed this issue indirectly by repeating the sonographic examination in cases of mild dysplasia [21]. In this context, we have to reconsider that Graf emphasized the importance of adequate training of sonographic examiners [10].

Limitations

There are several limitations to this study. First, to analyse the relative movement between the examined hip and the transducer, we assumed that there is no movement between the sleeping newborns and the positioning device. Thus, the three-dimensional analysis recorded the transducer position relative to the positioning device. Due to technical aspects movements of the newborns were not traced directly by the optoelectronic motion capture system. The unchanged position of the newborns was monitored by video camera. Newborns, which started to move during the examination where excluded, but still minimal movements in sleeping newborns could influence the measurements.

The second important limitation is that the examiner has to decide on the possible inclination of the transducer position. The method of checking sonographic images according to Graf’s criteria is sufficient to exclude excessive transducer inclinations, but it is possible that the examiner underestimates the range of allowed inclinations in this setting.

Conclusion

Our findings show that the standardized plane defined by Graf’s criteria allows notable inclinations of the transducer positions. Inclinations of the transducer position in the axial plane to anterior or posterior are less critical than inclinations in the frontal plane to caudal or cranial. Most important, a caudal inclination leads to underestimated α-angles, and in contrast a combined posterior and cranial transducer inclination may lead to overestimated α-angles with subsequent effects on the classification system. The use of a mechanical device like the “Sono-Guide” might provide additional technical support. Further research is needed to define criteria and factors to optimize the transducer positioning or to minimize the impact of transducer inclinations on sonographic measurements in screening for DDH.