Abstract
Background
Radiotherapy is one of the main treatment options for patients with esophageal cancer; however, it has been linked with an increased risk of cardiac toxicities. In the current study, we evaluated the effect of planning the radiation in deep-inspiration breath hold (DIBH) on the dose sparing of cardiac substructures and lung.
Materials and methods
In this study, we analyzed 30 radiation therapy plans from 15 patients diagnosed with esophageal cancer planned for neoadjuvant radiotherapy. Radiation plans were generated for 41.4 Gy and delivered in 1.8 Gy per fraction for free-breathing (FB) and DIBH techniques. We then conducted a comparative dosimetric analysis, evaluating target volume coverage, the impact on cardiac substructures, and lung doses across the two planning techniques for each patient.
Results
There was no significant disparity in target volume dose coverage between DIBH and FB plans. However, the Dmean, D2%, and V30% of the heart experienced substantial reductions in DIBH relative to FB, with values of 6.21 versus 7.02 Gy (p = 0.011), 35.28 versus 35.84 Gy (p = 0.047), and 5% versus 5.8% (p = 0.048), respectively. The Dmean of the left ventricle was notably lower in DIBH compared to FB (4.27 vs. 5.12 Gy, p = 0.0018), accompanied by significant improvements in V10. Additionally, the Dmean and D2% of the left coronary artery, as well as the D2% of the right coronary artery, were significantly lower in DIBH. The dosimetric impact of DIBH on cardiac substructures proved more advantageous for middle esophageal (ME) than distal esophageal (DE) tumors.
Conclusion
Radiotherapy in DIBH could provide a method to reduce the radiation dose to the left ventricle and coronaries, which could reduce the cardiac toxicity of the modality.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Esophageal cancer constitutes a significant clinical challenge, underscored by limited 5‑year survival rates that stand at 47% for localized cases, 26% for regional dissemination, and a stark 6% once metastasized [1]. Consequently, this malignancy ranks as the sixth leading cause of cancer-related deaths globally [2]. Recent decades have witnessed a histological shift in the incidence of esophageal cancer within affluent nations, transitioning predominantly from squamous cell carcinoma to adenocarcinoma [3].
The advent of integrated treatment strategies, in particular neoadjuvant radiochemotherapy, followed by surgical intervention has yielded significant improvements in survival rates for those with advanced esophageal cancer [4, 5]. However, radiation treatment of thoracic malignancies has been associated with an increased risk of cardiac diseases and cardiac mortality [6, 7]. Thoracic radiation therapy, particularly for esophageal cancer, has been correlated with a heightened incidence of cardiac toxicities and cardiac-related mortality. This correlation is evident when comparing patients who underwent radiation therapy to those who did not [8].
Indeed, multiple analyses have revealed an incidence of 16–25% for grade 3 or higher cardiac toxicity after radiation treatment in esophageal cancer, with the most common cardiac events being ischemic heart disease, coronary heart disease, pericardial effusion, and arrhythmia [9,10,11].
Emerging strategies are under exploration to mitigate cardiac toxicities associated with radiotherapy in esophageal cancer treatment. These include investigating proton therapy, which may reduce cardiac exposure [12], and considering radiotherapy exclusion in cases of adenocarcinoma histology [13]. Physiologically, deep inspiration leverages deformation and size reduction of the heart through negative intrathoracic pressure. Radiotherapy in deep-inspiration breath hold (DIBH) has already shown promising results in breast cancer treatment, where it has become a standard approach to protect the heart and the left anterior descending coronary artery (LAD) from radiation exposure in case of radiation treatment of left breast [14]. The same approach has been tested also for intrathoracic tumors, including lung cancer and thymic tumors [15, 16]. While previous studies have examined radiation in DIBH for esophageal cancer to reduce the mean dose to the heart and lung [17], a comprehensive dosimetric analysis of cardiac substructures during DIBH radiotherapy remains outstanding.
In the current study, we investigated the dosimetric advantage of radiotherapy for esophageal cancer in DIBH for heart substructures, namely the heart chambers and coronary arteries.
Materials and methods
This analysis encompassed 30 planning computed tomographies (P-CTs) from 15 patients diagnosed with middle or lower esophageal cancer who were eligible to undergo P‑CT in free breathing (FB) and deep-inspiration breath hold (DIBH) between January 2021 and December 2022. Initially, FB P‑CTs were performed on a 16-slice CT scanner (Brilliance CT Big Bore Oncology, Philips Medical Systems, Inc., Cleveland, Ohio, USA) employing parameters of 120 kV, 146 mAs, a pitch of 0.813, and a slice thickness of 3 mm, following the administration of iodinated contrast media. Subsequently, the DIBH P‑CTs were conducted in the same session using an optical surface imaging system, Catalyst™ (CRAD, Uppsala, Sweden), to monitor sternal motion during DIBH CT acquisition. Planning and diagnostic CTs along with available positron-emission tomography (PET) scans were anonymized and imported into the Pinnacle3 treatment planning system (V.14.0; Philips Healthcare, Amsterdam, Netherlands). The analysis was approved by the local ethics committee (Faculty of Medicine, RWTH Aachen University, approval sign: EK 23-088).
Volume delineations
A radiation oncologist (AM) generated all target volume (TV) delineations to ensure uniformity, and a second radiation oncologist (ME) reviewed them. The gross tumor volume (GTV) included contrast-enhanced thickening and/or FDG-avid lesions on the esophageal wall (GTVp) and enlarged regional lymph nodes greater than 1 cm in short axis diameter and/or FDG avid (GTVn), visualized using a thoracic mediastinal window (window width: 401 HU; window level: 800 HU). Clinical target volumes (CTV) were delineated by extending GTVp by 4 cm craniocaudally and 1.5 cm radially (CTVp) and GTVn by 1.5 cm in all directions (CTVn), excluding the lungs, heart, aorta, and trachea if uninvolved. The planning target volume (PTV) was defined by further expanding the combined CTV (CTVp+n) by 5 mm in all directions.
Subsequently, lung and heart substructures, including the left and right ventricles (LV and RV, respectively), left and right atria (LA and RA, respectively), right coronary (RCA), left coronary (LCA), left anterior descending (LAD), and left circumflex coronary (LCX) arteries were contoured using contouring guidelines as previously described [18] and revised by a radiologist (PB).
Treatment planning
Thirty volumetric modulated arc therapy (VMAT) plans were created. Each patient had one plan based on FB-CT and another on DIBH-CT using the collapsed cone algorithm with a 3 × 3 × 3 mm3 dose grid, all by the same medical physicist (MD). Each plan consisted of two full coplanar arcs at a photon energy of 6 MV, with a normalization mode of 100% and a prescribed PTV dose of 41.4 Gy over 23 fractions. The optimization aimed for uniform PTV coverage with the prescribed dose, ensuring 95% of the dose encompassed 100% of the CTV and over 95% of the PTV.
The following dose–volume parameters were used for plan evaluation:
-
D2%: the near maximum dose.
-
D98%: the near minimum dose.
-
Dmean: the mean dose of TV/OAR.
-
Dx: the dose to defined (x) volume from TV/OAR.
-
VxGy: the volume of TV/OAR that receives a defined (x) dose.
Dose constraints for organ at risk (OAR) were spinal canal D2%< 45 Gy, liver Dmean < 32 Gy, lung Dmean < 17.5 and D20% < 20 Gy, and heart Dmean < 26 Gy.
Dosimetric analysis
A dose–volume histogram (DVH) was generated for each plan. Different dosimetric/volumetric data were extracted from DVHs and used for plan evaluation, including
-
PTV volume, PTV D2% and D98% median dose (D50%), and the volume of PTV enclosed by reference isodose line (V95%).
-
the conformity index RTOG (CI) = VRI/TV, where VRI is the reference isodose volume (95% isodose volume) and TV is the target volume (PTV).
-
the homogeneity index RTOG (HI) = Imax/RI, where Imax is the maximum isodose in the target and RI is the reference isodose.
In addition, dosimetric data from the cardiac substructure and lungs were extracted for comparison between plans in FB and DIBH for each patient, which included the following:
-
the entire heart: volume, Dmean, V30 Gy, and D2%.
-
the heart chambers (ventricles and atria): Dmean, D2%, and V10,20,30, and 40 Gy
-
coronary arteries: Dmean, D2%, and V5,10,20,30, and 40 Gy.
-
lung: volume, Dmean, and V20 Gy.
Statistical analysis
All collected data were transferred to Microsoft Excel (Microsoft Corporation, Redmond, WA, U.S.) for subsequent analysis. We present the data as means with their respective standard deviations (± SD). To assess the distribution of the data, the Shapiro–Wilk test was conducted. Data with a p-value greater than 0.05 were deemed to follow a normal distribution (parametric data), whereas those with a p-value less than 0.05 were classified as non-parametric. For parametric datasets, a paired t-test was used to compare the means of each parameter. Conversely, the Wilcoxon signed-rank test was employed for datasets where at least one group of compared data was non-parametric.
Results
Thirty radiation plans from 15 patients were evaluated in the current analysis: 15 in DIBH and 15 in FB. All participants met the criteria for neoadjuvant radiotherapy (either T3 tumor and/or node positive) including 9 patients with squamous cell histology and 6 patients with adenocarcinoma histology (Table 1). Twelve patients underwent PET scans for radiation planning within 1 week of P‑CT, which were used to assist in target volume delineation.
The average volume of PTV in DIBH was 342.09 cm3 (SD ± 144.12), while for FB it was 363.73 cm3 (SD ± 132.76), with a p-value of 0.088. Dosimetric parameters assessing target volume coverage—including D2%, D98%, D50%, and V95%—demonstrated comparable results between DIBH and FB plans (Table 2). The mean HI was consistent at 1.1 for both cohorts (p-value: 0.125). Additionally, the average CI was 1.02 for plans in DIBH and 1.03 for plans in FB (p-value: 0.33), suggesting similar conformity. The average monitor units per plan showed no significant variance, with 507.7 for DIBH and 535.23 for FB (p-value: 0.107).
Dosimetric outcomes for OAR
The mean cardiac volume was significantly reduced in DIBH compared to FB, with values of 831.5 cm3 and 885.4 cm3, respectively (p-value: 0.012). Similarly, the average mean dose to the heart (heart Dmean) was significantly lower in DIBH-plans, at 6.21 Gy, compared to 7.02 Gy in FB-plans (p-value: 0.011). Moreover, V30 Gy and D2% of the heart in DIBH (5% and 35.28 Gy) were significantly lower than FB (5.8% and 35.84 Gy), with p-values 0.048 and 0.047, respectively (Fig. 1, Table 3).
Regarding the dosimetric impact on cardiac chambers, DIBH plans demonstrated a significantly lower Dmean to the LV compared to FB plans, with a mean dose of 4.27 Gy versus 5.12 Gy, respectively (p-value: 0.0018). Also, V10 and V20 Gy of LV were reduced in DIBH, with 8.5% compared to 11.5% and 2.6% versus 4% in FB, with p-values of 0.011 and 0.079, respectively. However, for the right ventricle and both atria, no significant differences in dose–volume metrics were observed between the DIBH and FB plans (Fig. 2, Table 3).
The Dmean and D2% of the LCA were significantly lower in DIBH at 3 Gy and 3.4 Gy compared to FB at 3.4 Gy and 5.5 Gy, with p-values of 0.019 and 0.021, respectively. Additionally, planning in DIBH yielded lower Dmean and D2% for the RCA at 2.23 Gy and 3.86 Gy, as opposed to 2.99 Gy and 4.96 Gy in FB, with p-values of 0.064 and 0.03, respectively (Table 3). The dosimetric parameters for the other coronary arteries did not demonstrate any significant differences between the FB and DIBH plans (Fig. 2, Table 3).
Further, the mean lung volume was higher for DIBH compared to FB (5138 vs. 3605 cm3, p-value < 0.00001), and this was linked to a significantly reduced Dmean in DIBH vs. FB, 10.3 vs. 10.8 Gy (p-value: 0.047; Fig. 1, Table 3).
Next, we analyzed the dosimetric parameters based on tumor location, where the tumor location was categorized into middle esophageal (ME) or distal esophageal (DE) tumors (proximal edge of the tumor < 32 vs. > 32 cm from the incisor teeth, respectively). When planning in DIBH versus FB for ME tumors, we observed a significant improvement in heart Dmean (6.02 vs. 7.13 Gy, p-value 0.007) and heart V30 Gy (5.3% vs. 6.6%, p-value 0.017), along with a reduction in heart D2% (34.99 Gy vs. 35.8 Gy, p-value 0.064). Moreover, the LV demonstrated significantly lower Dmean, D2%, V10, and V20 Gy in DIBH (3.79 Gy, 14.19 Gy, 6.8%, and 2%, respectively) compared to FB (4.85 Gy, 16.82 Gy, 11%, and 4%, respectively), with respective p-values of 0.0018, 0.041, 0.014, and 0.035. The RA exhibited a reduced Dmean and D2% in DIBH (5.73 Gy and 18.6 Gy) compared to FB (7.11 Gy and 24.02 Gy), with p-values of 0.087 and 0.019. For ME tumors, DIBH planning yielded lower Dmean and D2% for the LCA (3.14 vs. 3.54 Gy and 3.55 vs. 5.88 Gy, p-values 0.016 and 0.067, respectively), lower Dmean and V20 for the LCX (11.4 vs. 13.1 Gy, 21% vs. 28%, p-values 0.09 and 0.025 respectively), and decreased Dmean and D2% for the RCA (1.6 vs. 2.1 Gy, 2.15 vs. 3.03 Gy, p-values 0.057 and 0.046, respectively).
Nonetheless, the Dmean and V20 Gy of the lung in DIBH were 11.6 Gy and 18.2%, and in FB they were 12.5 Gy and 20.2%, with p-values of 0.0013 and 0.067, respectively.
Conversely, planning in DIBH for DE tumors only had a dosimetric advantage with regard to D2% of the LAD (2.52 Gy vs. 2.88, p-value 0.015) and a trend for improved V20 and V30 of LCX in FB compared to DIBH (50% vs. 29.7% and 37% vs. 14.5%, p-value: 0.059 and 0.059, respectively). All other cardiac, subcardiac, or pulmonary parameters evidenced comparable outcomes for DIBH and FB.
Discussion
Cardiac toxicity after radiation therapy in esophageal cancer has emerged as a significant concern [9]. In their analysis, Cai et al. revealed compelling evidence demonstrating that the radiation doses of LV, LAD, and LCX are strong predictors of severe cardiac events of grade 3 or higher [19]. Nonetheless, Burke et al. conducted a prospective study to delve into the structural cardiac alterations using serial cardiac MRI for patients with esophageal cancer treated with neoadjuvant chemoradiotherapy [20]. They found fibrotic and ischemic myocardial changes in almost one third of the patients; these changes were significantly associated with an increase in left ventricular end-systolic volume (LVESV).
The current study aimed to evaluate the possible dosimetric advantage of planning DIBH for cardiac substructures in patients undergoing neoadjuvant radiotherapy for locally advanced esophageal cancer. A previous study assessed the dosimetric advantage of DIBH for OARs in esophageal cancer radiotherapy, concluding a correlation with some dosimetric benefits such as a lower heart V40 compared to planning in FB [17]. However, detailed analysis concerning the subcardiac structures remains scant in the literature.
The current analysis confirmed that planning in DIBH provides superior cardiac sparing compared to FB. The mean heart volume, Dmean, V30 Gy, and D2% of the entire heart were significantly lower in DIBH versus FB. This could be explained by the reduction in intrathoracic pressure and atrial filling associated with increased lung volumes, leading to heart deformation and a consequent decrease in heart volume, thereby diminishing cardiac radiation exposure, as demonstrated in Fig. 3 [21]. These results align with other studies analyzing mediastinal or lung tumors [16, 22]. The clinical relevance of these consistent differences in radiation exposure to the heart between planning in DIBH and FB, even when the magnitude is small, should be considered in light of the results of the analysis from Darby et al., where major coronary events increased linearly by 7.4% per increment with 1 Gy of heart Dmean in female patients with breast cancer who received adjuvant radiation [23].
In examining subcardiac structures, DIBH demonstrated significant reductions in both Dmean and D2% for the LCA, as well as in D2% for the RCA, when compared to FB. Furthermore, Dmean and V10 of LV were significantly lower in DIBH relative to FB. These results gain particular significance in the context of the findings of Cai et al. [19], where they showed LV to be a critical organ for radiation-induced cardiac toxicity. Radiation treatment in DIBH could significantly reduce the dose to LV, and that would reflect on the incidence of cardiac toxicity.
Remarkably, dosimetric analysis revealed that planning in DIBH for ME tumors conferred greater benefits to cardiac substructures than DE tumors. Not only were the Dmean and V30 Gy of the entire heart reduced, but also the Dmean, D2%, V10, and V20 Gy of the LV, along with the D2% of the RA, showed significantly lower doses in DIBH. Furthermore, the Dmean of the LCA, V20 of the LCX, and D2% of the RCA were lower in the DIBH group.
The lung volume increased during DIBH, resulting in a significantly lower Dmean to the lungs across the entire cohort, particularly in patients with middle-third esophageal tumors.
In contrast, while planning in DIBH for DE tumors revealed a dosimetric improvement for D2% of the LAD, there was a trend toward higher doses observed for V20 and V30 of the LCX during DIBH.
Conclusion
This study substantiates that the DIBH technique facilitates significant cardiac sparing during neoadjuvant radiotherapy for patients with locally advanced esophageal cancer. It underscores the necessity of tumor position within the esophagus as a determinant for the efficacy of DIBH in mitigating cardiac and pulmonary doses. Future prospective studies are warranted to corroborate these observations and refine the application of DIBH in the radiotherapeutic management of esophageal cancer.
References
Survival Rates for Esophageal Cancer | Esophageal Cancer Outlook. https://www.cancer.org/cancer/esophagus-cancer/detection-diagnosis-staging/survival-rates.html#references. Accessed 12 Mar 2023
Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A (2018) Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 68(6):394–424
Lin Y, Wang H, Fang K, Zheng Y, Wu J (2022) International trends in esophageal cancer incidence rates by histological subtype (1990–2012) and prediction of the rates to 2030. Esophagus 19(4):560–568
van Hagen P, Hulshof MCCM, van Lanschot JJB, Steyerberg EW, van Berge Henegouwen MI, Wijnhoven BPL, Richel DJ, Nieuwenhuijzen GAP, Hospers GAP, Bonenkamp JJ et al (2012) Preoperative Chemoradiotherapy for Esophageal or Junctional Cancer. N Engl J Med 366(22):2074–2084. https://doi.org/10.1056/NEJMoa1112088
Reynolds JV, Preston SR, O’Neill B, Lowery MA, Baeksgaard L, Crosby T, Cunningham M, Cuffe S, Griffiths GO, Roy R et al (2021) Neo-AEGIS (Neoadjuvant trial in Adenocarcinoma of the Esophagus and Esophago-Gastric Junction International Study): Preliminary results of phase III RCT of CROSS versus perioperative chemotherapy (Modified MAGIC or FLOT protocol). (NCT01726452). J Clin Oncol 39(15_suppl):4004. https://doi.org/10.1200/JCO.2021.39.15_suppl.4004
Piroth MD, Baumann R, Budach W, Dunst J, Feyer P, Fietkau R, Haase W, Harms W, Hehr T, Krug D et al (2019) Heart toxicity from breast cancer radiotherapy. Strahlenther Onkol 195(1):1–12. https://doi.org/10.1007/s00066-018-1378-z
Jaworski C, Mariani JA, Wheeler G et al (2013) Cardiac complications of thoracic irradiation. J Am Coll Cardiol 61(23):2319–2328. https://doi.org/10.1016/j.jacc.2013.01.090
Gharzai L, Verma V, Denniston KA, Bhirud AR, Bennion NR, Lin C (2016) Radiation therapy and cardiac death in long-term survivors of esophageal cancer: an analysis of the surveillance, epidemiology, and end result database. Bruns H, editor. Plos One 11(7):e158916. https://doi.org/10.1371/journal.pone.0158916
Wang X, Palaskas NL, Yusuf SW, Abe J‑I, Lopez-Mattei J, Banchs J, Gladish GW, Lee P, Liao Z, Deswal A et al (2020) Incidence and onset of severe cardiac events after radiotherapy for esophageal cancer. J Thorac Oncol 15(10):1682–1690
Hayashi Y, Iijima H, Isohashi F, Tsujii Y, Fujinaga T, Nagai K, Yoshii S, Sakatani A, Hiyama S, Shinzaki S et al (2019) The heart’s exposure to radiation increases the risk of cardiac toxicity after chemoradiotherapy for superficial esophageal cancer: A retrospective cohort study. BMC Cancer 19(1):1–7. https://doi.org/10.1186/s12885-019-5421-y
Witt JS, Jagodinsky JC, Liu Y, Yadav P, Kuczmarska-Haas A, Yu M, Maloney JD, Ritter MA, Bassetti MF, Baschnagel AM (2019) Cardiac toxicity in operable esophageal cancer patients treated with or without chemoradiation. Am J Clin Oncol 42(8):662
Lin SH, Hobbs BP, Verma V, Tidwell RS, Smith GL, Lei X, Corsini EM, Mok I, Wei X, Yao L et al (2020) Randomized phase IIB trial of proton beam therapy versus intensity-modulated radiation therapy for locally advanced esophageal cancer. J Clin Oncol 38(14):1569
Hoeppner J, Lordick F, Brunner T, Glatz T, Bronsert P, Röthling N, Schmoor C, Lorenz D, Ell C, Hopt UT et al (2016) ESOPEC: Prospective randomized controlled multicenter phase III trial comparing perioperative chemotherapy (FLOT protocol) to neoadjuvant chemoradiation (CROSS protocol) in patients with adenocarcinoma of the esophagus (NCT02509286). BMC Cancer 16(1)
Wolf J, Stoller S, Lübke J, Rothe T, Serpa M, Scholber J, Zamboglou C, Gkika E, Baltas D, Juhasz-Böss I et al (2022) Deep inspiration breath-hold radiation therapy in left-sided breast cancer patients: a single-institution retrospective dosimetric analysis of organs at risk doses. Strahlenther Onkol 199(4):379–388. https://doi.org/10.1007/s00066-022-01998-z
Josipovic M, Aznar MC, Thomsen JB, Scherman J, Damkjaer SMS, Nygård L, Specht L, Pøhl M, Persson GF (2019) Deep inspiration breath hold in locally advanced lung cancer radiotherapy: validation of intrafractional geometric uncertainties in the INHALE trial. Br J Radiol 92(1104)
Yan D, Ning L, Chen Y, Ke S, Huang H, Wang L, Yan S (2022) Analysis of deep inspiration breath-hold technique to improve dosimetric and clinical advantages in postoperative intensity-modulated radiation therapy for thymomas. Quant Imaging Med Surg 12(8):4239–4247
Lorchel F, Dumas JL, Noë A, Wolf D, Bosset JF, Aletti P (2006) Dosimetric consequences of breath-hold respiration in conformal radiotherapy of esophageal cancer. Phys Med 22(4):119–126
Duane F, Aznar MC, Bartlett F, Cutter DJ, Darby SC, Jagsi R, Lorenzen EL, Mcardle O, Mcgale P, Myerson S et al (2017) Cardiac contouring atlas A cardiac contouring atlas for radiotherapy. Radiother Oncol 122:416–422
Cai G, Li C, Li J, Yang J, Li C, Sun L, Li J, Yu J, Meng X (2023) Cardiac substructures dosimetric predictors for cardiac toxicity after definitive radiotherapy in esophageal cancer. Int J Radiat Oncol 115(2):366–381
Burke AM, Yeh C, Kim S, Bergquist P, Krishnan P, Barac A, Srichai MB, Unger K (2020) A prospective study of early radiation associated cardiac toxicity following neoadjuvant chemoradiation for distal esophageal cancer. Front Oncol 10:1169
Carroll RG (2007) Integrated cardiovascular function. In: Elsevier’s Integr Physiol, vol 1, pp 91–98
Persson GF, Scherman Rydhög J, Josipovic M, Maraldo MV, Nygård L, Costa J, Berthelsen AK, Specht L, Aznar MC (2016) Deep inspiration breath-hold volumetric modulated arc radiotherapy decreases dose to mediastinal structures in locally advanced lung cancer. Acta Oncol 55(8):1053–1056. https://doi.org/10.3109/0284186X.2016.1142115
Darby SC, Ewertz M, McGale P, Bennet AM, Blom-Goldman U, Brønnum D, Correa C, Cutter D, Gagliardi G, Gigante B et al (2013) Risk of is che mic heart disease in women after radiotherapy for breast cancer (abstract from the clinical trial service unit mark; the department of medical Epide-miology and biostatistics). N Engl J Med 11:987–998
Funding
Open Access funding enabled and organized by Projekt DEAL.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
A.A. Mohamed, M.N. Douglas, P. Bruners, and M.J. Eble declare that they have no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
66_2024_2197_MOESM1_ESM.docx
Supplementary Table 1: Dosimetric outcomes of planning in deep inspiration breath hold (DIBH) and free breathing (FB) for the patients for the entire cohort. Supplementary Table 2: Dosimetric outcomes of planning in deep inspiration breath hold (DIBH) and free breathing (FB) for the patients with middle esophageal tumors.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Mohamed, A.A., Douglas, M.N., Bruners, P. et al. Dosimetric advantages for cardiac substructures in radiotherapy of esophageal cancer in deep-inspiration breath hold. Strahlenther Onkol 200, 624–632 (2024). https://doi.org/10.1007/s00066-024-02197-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00066-024-02197-8