1 Introduction

Near-infrared spectroscopy (NIRS) provides the capability to non-invasively characterise cerebral haemodynamics and metabolism by using the attenuation of near infrared light to derive concentration changes in oxyhaemoglobin Δ[HbO2 ], deoxyhaemoglobin Δ[HHb ] and the oxidation state of cytochrome c oxidase Δ[oxCCO] [1]. To date NIRS studies of functional brain activation have monitored changes in [HbO2 ] and [HHb] to measure the haemodynamic response resulting from neuronal activation. Detailed [HbO2]/[HHb] maps of localised functional responses have been reported in several studies using diffuse optical imaging [2] and, more recently, with wearable or fibreless systems [3]. Given the role of cytochrome c oxidase as the terminal electron acceptor in the mitochondrial electron transport chain, the NIRS-derived oxCCO measure of the oxidation status of this enzyme is well placed to provide a direct marker of cellular oxygen metabolism . Conventional measures of ΔoxCCO have been achieved in vivo using benchtop broadband near infrared spectroscopy (NIRS) systems [4], but a recently [5] described multichannel fibre-based broadband NIRS system is capable of delivering spatially resolved measures of [oxCCO ]. Here we present, for the first time, a fibreless multiwavelength NIRS system using light emitting diodes (LED ) designed specifically to image localised changes in [oxCCO] and hence neuronal metabolism. We used a well-established hyperoxia protocol to demonstrate the CCO measurement capabilities of the system, and then tested its imaging capabilities using a functional activation paradigm.

2 Methods

The fibreless system consists of four modules each containing two multiwavelength light emitting diodes (LED) sources (OSA_Opt, Germany) and four photodiode detectors (Hamamatsu Photonics, Japan). Each source contains eight LED dies (780, 811, 818, 842, 850, 882, 891 and 901 nm) assembled in an area of 1.5 × 1.5 mm based on the findings of [6]. The optoelectronics for the modular system are described in detail in [7] and the production of the multiwavelength sources in [8]. Figure 1 is a schematic of a single module incorporating two multiwavelength light sources and four detectors.

Fig. 1
figure 1

(a) Schematic of single module containing two multi-wavelength sources and four detectors, (b) photograph of the four-module system

Following ethical approval and volunteer consent, the system was used to collect data during two studies, one involving a hyperoxia challenge in ten healthy volunteers and the other a hemifield stimulation of the visual cortex in a single volunteer.

2.1 Hyperoxia Studies

The array was positioned over Fp1 (10–20 electrode placement system). The experimental protocol involved 2 min of baseline normoxia, 5 min of hyperoxia followed by 5 min of normoxia. NIRS data were collected using all eight wavelengths from one light source on each of two modules and a total of eight detectors, resulting in a source-detector separation of 3.5 cm. Data were averaged across channels and across subjects.

2.2 Hemifield Stimulation Studies

The array was centred over the primary visual cortex (O1 position). A 2 Hz inverting hemifield checkerboard stimulus of 20 s duration on each hemifield (20 s on left and 20 s on right) was repeated 15 times and averaged. Data were collected from all source and detector positions on all four modules but, due to the repetition rate of the switching between wavelengths, only four wavelengths (780, 811, 842, 891) were used at each source as defined by [6]. A multi-spectral image reconstruction approach was used to reconstruct images of Δ[HbO2], Δ[HHb] and Δ[oxCCO ] from the multi-distance, multi-channel optical datasets. TOAST++ software was used to run the forward model on the registered adult volumetric mesh. The LSQR method was employed to solve the inverse problem with first-order Tikhonov regularization. The volumetric images were then projected on the cortical surface mesh [9].

3 Results

Figure 2 shows the data collected during the hyperoxia studies averaged across channels and across subjects; the measurement changes are consistent with previous studies using standard broadband NIRS systems [10]. Figure 3 shows the Δ[HbO2], Δ[HHb] and Δ[oxCCO] images reconstructed from the left hemifield stimulation study.

Fig. 2
figure 2

Mean and standard deviation of Δ[HbO2], Δ[HHb] and Δ[oxCCO ] from ten subjects during hyperoxia challenge

Fig. 3
figure 3

(a) Four-module array centred over the occipital cortex, (b) checkerboard stimulus activating left visual field, (c) images of Δ[HbO2], Δ[HHb] and Δ[oxCCO ] showing activation on right visual cortex (colour scale in μmolar)

4 Discussion

We have demonstrated the development and testing of a novel fibreless multiwavelength NIRS system with specific capability to measure localised changes in oxCCO . The hyperoxia study confirms the eight wavelength capability of the system over two modules without spatial discrimination, and the hemifield stimulation the capability of a four-module system to deliver depth-resolved images of regional changes in oxCCO . These tests demonstrate a flexible approach to multi-wavelength, multi-channel NIRS measures. Although further work is required to increase functionality and implementation in a clinical setting, this approach has potential to deliver an inexpensive, wearable, continuous monitor of cerebral energetics in a range of experimental and clinical scenarios.

Monitoring and management of patients with acute brain injury (ABI ) remain key drivers for the development of NIRS oxCCO systems [11]. Derangements in cerebral metabolism are implicated in the pathophysiology of both secondary brain injury and long-term neurocognitive outcome. Spatially resolved measures of oxCCO may have potential to clarify pathways of injury after ABI, assess the impact of therapies on both injured and non-injured brain regions and, ultimately, guide interventions designed to prevent or minimise secondary brain injury. The advent of fibreless, and therefore wearable, systems capable of making these measures will allow brain monitoring in the post- acute/rehabilitation phase and has potential to elucidate the evolution of long-term neurocognitive damage.

Fibreless systems are also likely to play an increasingly important role in neurodevelopment studies given the additional flexibility they provide for imaging brain function in naturalistic environments, especially for typically and atypically developing young infants and toddlers. Recent studies have demonstrated the first measurement of oxCCO changes during functional activation in the infant brain using a single source detector system [12], and further work is already underway to deliver spatially resolved measures of oxCCO to investigate links between mitochondrial and metabolic dysfunction in infants at high risk of autism [13, 14].

In addition, as the use of NIRS extends to global health projects (see www.globalfnirs.org) the advent of wearable, relatively low cost, flexible, easy to implement systems will further enhance the applicability of NIRS for studies in resource poor settings [15].