Kidney diseases are a global health burden with steadily increasing prevalence and incidence [1,2,3,4,5]. Animal studies have provided evidence that acute kidney injury (AKI) of various origins shares a common link in the pathophysiological chain of events, ultimately leading to AKI , as well as to progression from AKI to chronic kidney disease (CKD): imbalance between renal oxygen delivery and oxygen demand [3, 6,7,8,9,10,11,12,13]. Renal tissue hypoperfusion and hypoxia have also been suggested to play a pivotal role in the pathophysiology of other kidney diseases including diabetic kidney disease [14,15,16].

The majority of the preclinical studies that generated this concept utilized a set of invasive probes to measure renal hemodynamics and oxygenation in anesthetized animals [12, 14, 17,18,19,20]. These methods have also been used (1) to study mechanisms of physiological control of renal hemodynamics and oxygenation and (2) the effects of various substances on this control in healthy animals, (3) to test several putative preventive or therapeutic approaches for AKI and CKD in animal models, and (4) for the purpose of calibrating functional magnetic resonance imaging (MRI) including renal blood oxygenation level-dependent MRI (BOLD-MRI) that makes use of the quantitative MR parameter T2*, which is a surrogate for blood oxygenation [18,19,20,21,22,23,24,25,26,27,28].

The invasive probes typically include (1) a perivascular flow probe for measurement of total renal blood flow, (2) laser-Doppler-optodes for assessment of local tissue perfusion, (3) Clark-type electrodes or fluorescence-quenching optodes for measurements of local tissue partial pressure of oxygen (pO2), and (4) devices for invasive measurement of arterial blood pressure. These methods are considered the gold standard for the study of renal hemodynamics and oxygenation, because the methods, with the exception of the laser-Doppler fluxmetry, provide calibrated quantitative data [25, 29, 30]. Step-by-step protocols for application of these techniques in two setups, one for stand-alone experiments in laboratories for integrative physiology, and the other for experiments within small-animal MR scanners by use of a hybrid setup, are provided in the chapter by Cantow K et al. “Monitoring Renal Hemodynamics and Oxygenation by Invasive Probes: Experimental Protocol”.

As with all established modalities available in today’s experimental and translational research, these techniques have shortcomings and methodological limitations. In particular, their invasiveness, for obvious ethical reasons, precludes their use in humans. The invasiveness is, furthermore, a considerable hurdle for implementation of these methods in long-term and longitudinal studies: although invasive probes have been used in conscious, chronically instrumented animals [22, 24, 31,32,33,34,35,36,37], the vast majority of studies has been performed in “terminal experiments” in anesthetized animals. This is a major limitation of all invasive techniques because the measurements are by necessity potentially confounded by the effects of anesthesia (see the chapter by Kaucsar T et al. “Preparation and Monitoring of Small Animals in Renal MRI”).

The measurement of intrarenal tissue perfusion and tissue oxygenation is quite challenging, because of their vast spatial heterogeneity both among the renal layers (cortex, outer medulla, and inner medulla) and within the layers. In this chapter, the causes and the degree for this spatial variability are outlined first, before the basic principles of methodology to quantify renal hemodynamics and oxygenation by means of invasive probes in experimental animals are described together with a discussion of the pros and cons of the various methods.

This introduction chapter is complemented by a separate chapter describing the experimental procedure, which is part of this book.

This chapter is part of the book Pohlmann A, Niendorf T (eds) (2020) Preclinical MRI of the Kidney—Methods and Protocols. Springer, New York.

2 Heterogeneity of Intrarenal Perfusion and Oxygenation—Causes and Degree of Variability

Intrarenal local tissue pO2 is determined by the balance between local O2 supply and local O2 consumption. Additionally, determinants of local O2 diffusion play a role. Local O2 supply depends on local blood perfusion, local vessel density, and local blood pO2 and blood O2 content. Local O2 consumption is foremost determined by local metabolic demand. Blood perfusion within the kidney is much more heterogeneous than within most other organs, due to its particular vascular architecture [8, 38,39,40].

From the branches of the renal artery that enter the kidney at its hilus, interlobar arteries arise, which align with the medullary–cortical axis. From interlobar arteries emerge arcuate arteries that run along the boundary between the cortex and the medulla, that is, perpendicular to the medullary–cortical axis. From the latter, interlobular arteries branch off at regular intervals at angles of about 90 degrees, so that they all enter the renal cortex. For this reason, virtually all the blood flowing into the kidney perfuses the cortex first [8, 25, 38, 41,42,43]. The interlobular arteries are in close proximity to interlobular veins, together forming what might be called “vascular bundles” that run in parallel at regular distances from the corticomedullary boundary toward the capsule. Within the cortex, interlobular arteries branch off into afferent (preglomerular) arterioles, from which glomerular capillaries arise that, in turn, rejoin into efferent (postglomerular) arterioles. The efferent arterioles from the subcapsular and mid-cortical glomeruli branch off into networks of peritubular capillaries. The efferent arterioles of the deep cortical (juxtamedullary) glomeruli, however, branch off at about the corticomedullary boundary to form capillary-like vessels, the descending vasa recta [38,39,40]. These very straight and exceptionally long vessels descend toward the papilla. The descending vasa recta provide virtually all blood that perfuses the medulla. As a consequence, only 10–15% of blood that previously passes through the cortex, will reach the medulla, and the vascular density is much lower in the medulla than the cortex [8, 25, 38, 41,42,43]. The descending vasa recta are in close proximity with the ascending vasa recta, thus forming distinct “vascular bundles” that run in parallel at regular distances through the medulla [38,39,40].

These particularities of the intrarenal vascular architecture have a number of salient consequences with regard to local blood perfusion and thus the local O2 supply. First, the striking difference in the quantity of blood perfusion between the cortex and the medulla, together with the difference in their vascular densities, is one of the major reasons for the low tissue pO2 in the medulla. Second, the Fåhræus–Lindqvist effect that lowers the hematocrit and thereby the O2 content of blood in the vasa recta that supply the medulla, contributes to low medullary O2 delivery [39, 40]. Third, diffusive shunting of O2 between the arcuate and interlobular arteries and their respective veins in the cortex as well as between the descending and ascending vasa recta in the medulla, owing to the close proximities of the respective vessels, are a further major reason behind the “physiological hypoxia” of the medulla. Thus, blood pO2 is higher in the venous vessels than in the capillaries these venous vessels drain. Fourth, due to the particular vascular architecture with distinct “vascular bundles” that comprise arterial vessels with high pO2 and venous vessels with relatively high pO2 (due mainly to the low fractional O2 extraction of the renal tissue), there are huge differences in local pO2 even within a given renal layer under physiological conditions [25, 41, 43,44,45,46,47]. Fifth, plasma skimming at intrarenal vessel branches results in different hematocrit and therefore O2 content of blood perfusing the daughter vessels, which also contributes to the local pO2 differences [39, 48].

The schematic shown in Fig. 1 depicts the range of local pO2 values as measured by invasive Clark-type electrodes in the kidneys of rats and dogs under physiological conditions in vivo [43,44,45]. These measurements reveal that local tissue pO2 levels in the cortex vary dramatically, foremost according to the distance to the “vascular bundles” of interlobular arteries and veins: pO2 values range from 10 mmHg at the greatest distance from two neighboring “bundles,” to up to 70 mmHg around the “bundles.” In addition, local cortical tissue pO2 varies with the distance to the arcuate arteries located at the corticomedullary boundary. Also, the range of tissue pO2 variability in the outer medulla is huge (see Fig. 1): whereas pO2 is less than 10 mmHg in locations at maximum distance from the “vascular bundles” of descending and ascending vasa recta, pO2 is up to 75 mmHg in close proximity to these “bundles.” In the inner medulla, tissue pO2 is rather homogenous: in general, it usually amounts to less than 15 mmHg.

Fig. 1
figure 1

Scheme depicting the range of intrarenal tissue partial pressure of oxygen (pO2). The red curve depicts values at the greatest distance from two neighboring “vascular bundles,” the blue one that directly adjacent to “vascular bundles” (for details see text); OS (outer stripe), IS (inner stripe). Original data were obtained by measurements in in anesthetized dogs and rats [43,44,45]

The huge differences in tissue pO2 among the renal layers not only reflect the differences in O2 supply but also the different O2 consumption of the distinct sections of the tubules located in the respective renal layers. Tubular resorption relies on active transport processes, which account for about 85% of the kidney’s energy expenditure and, therefore, its O2 consumption [8, 49]. The proximal tubule and the distal convoluted tubule, which constitute the majority of tubular segments within the cortex, consume much O2 because a huge amount of osmolytes is resorbed by active transport in these segments. On the other hand, tubular sections such as the thin descending and the thin ascending limbs of Henle’s loop and the medullary collecting duct, which are the most abundant tubular sections within the medulla, feature very low to nonexisting active transport. Yet the thick ascending limbs of Henle’s loop, which are located in the outer medulla, consume much O2, because of their active resorption of osmolytes [8, 25, 41, 43,44,45,46,47, 50]. The high O2 demands of the thick ascending limbs in relation to the rather low O2 supply of the outer medulla, consequent to its particular vascular architecture, exposes this renal layer to the greatest risk for hypoxic tissue damage [6,7,8,9,10,11,12,13,14, 43, 51, 52].

3 Methods: Methodologic Principles, Advantages, and Disadvantages

Techniques for invasive measurements of renal hemodynamics and oxygenation comprise readily (commercially) available methods and devices that are employed in many laboratories of integrative kidney physiology worldwide, but also techniques that are established only in one or a few laboratories. The widely used methods are described first. They include (1) perivascular flow probes for measurement of total renal blood flow, (2) laser-Doppler-optodes for assessment of local tissue perfusion, (3) Clark-type electrodes for measurements of local tissue pO2, (4) fluorescence-quenching optodes for the same purpose, (5) devices for invasive measurement of arterial blood pressure, and (6) pimonidazole adduct immunohistochemistry for assessment of renal tissue hypoxia.

The pimonidazole technique is included here because of its widespread usage, although it is an exceptional case for two reasons. First, in contrast to the other techniques, this approach does not allow investigators to monitor a given parameter over time, but only provides a snapshot: Pimonidazole is injected at a single time point in vivo, and pimonidazole adducts formed in severely hypoxic tissue are visualized in ex vivo samples [53,54,55]. Second, although the technique has been used for two decades, it was subjected to close scrutiny only recently [55]. Upon this careful investigation major pitfalls were identified and detailed—which, at least in part, cast doubts on the reliability of previously published results—and methods to address and overcome these issues have been proposed [55].

In the final part of this chapter, a few selected examples of methods are discussed that are not (yet) generally available but presently established in only one or a few laboratories.

3.1 Transonic Flow Probes

Today’s gold standard for whole kidney blood flow measurements is a flow probe attached to the renal artery or to the renal vein that makes use of ultrasound transit time differences. This method was established in the late 1970s and provides continuous absolute measurements of blood flow (in mL per min) independent of the actual flow velocity profile, the actual vessel diameter, and the actual hematocrit [56]. The commercially available device (Transonic™, Transonic Systems Inc.) comprises an electronic flow detection unit (flowmeter) and perivascular probes [57, 58]. The flow probe consists of two ultrasonic transducers and a fixed acoustic reflector (see Fig. 2a). The flow probe is positioned on the blood vessel so that the vessel lies within the acoustic sensing window defined by the transducers and the metallic reflector. Probes are available in various sizes to fit the dimensions of the blood vessel of interest.

Fig. 2
figure 2

Photographs of typical probes for invasive measurements. (a) Perivascular Transonic™ flow probe for measurement of total renal blood flow. (b) Combined fluorescence-quenching-laser-Doppler probe (OxyLite/OxyFlo™) for simultaneous measurements of local tissue pO2 and assessment of local tissue perfusion in the same circumscribed area. (c) Clark-type electrode for measurement of local tissue pO2

The flowmeter directs the flow probe through measurement cycles [57, 58]. First, a wide beam of ultrasound is emitted by the downstream transducer. It intersects the vessel in the upstream direction, is reflected by the acoustic reflector, then intersects the vessel again before being received by the upstream transducer. The flowmeter records the transit time. Then, the transit–receive sequence is repeated, but with the transmitting and receiving functions of the transducers reversed, that is, the vessel is bisected by the beam in the downstream direction. The transit time during the upstream cycle is higher than during the downstream cycle, because the ultrasound beam travels against the blood flow in the first, and with the blood flow in the latter cycle. The difference in the transit times is directly proportional to blood volume flow. Due to the wide ultrasonic beams, all flowing blood is accounted for; thus, the transit time is sampled at all points across the vessel diameter, and the volume flow measurement is independent of the flow velocity profile [57, 58].

An accurate measure of volumetric flow is obtained by the perivascular probes even when the vessel is smaller than the acoustic window, because the parts of the ultrasonic beam that do not intersect the vessel do not contribute to the signal [57, 58]. However, air effectively blocks ultrasound transmission, and even small air bubbles compromise measurement accuracy. Therefore, all spaces between the vessel and probe must be filled with a suitable acoustic coupling agent. In case of “terminal experiments” in anesthetized animals, where the abdomen is opened, this can be achieved by simply filling the abdominal cavity with isotonic saline (at 37 °C) [27]. Acoustic coupling gels can also be used and are available from the supplier of the probes. Fatty tissue also affects the ultrasonic beam; thus, the vessel of interest must be thoroughly prepared by gently removing fatty tissue [27].

Measurements of total renal blood flow by means of Transonic™ devices have also been performed in chronically instrumented animals [22, 31, 33, 58, 59]. Here, the perivascular probe becomes encapsulated by tissue, which stabilizes its position. En route to calibration of functional MRI for the study of renal hemodynamics and oxygenation, an integrated multimodality approach was developed that enables measurements of invasive gold standard parameters and functional MR parameters simultaneously for the same kidney in an ultrahigh field small animal scanner (MR-PHYSIOL) [25,26,27,28]. For such a setup, a special Transonic™ flow probe is used: its acoustic reflector is made of Macor ceramics instead of stainless steel or brass to meet the safety and compatibility requirements of MRI (see the chapter by Cantow K et al. “Monitoring Renal Hemodynamics and Oxygenation by Invasive Probes: Experimental Protocol”).

With the introduction of the Transonic™ technique, all other methods used for quantitative assessment of total renal blood flow in experimental animals including but not limited to electromagnetic probes and microsphere techniques were rendered obsolete.

3.2 Laser-Doppler Optodes

In the 1970s, laser-Doppler probes were developed that estimate local blood perfusion within circumscribed areas of tissue [60, 61]. For this purpose, pulsed light of a distinct wavelength (e.g., 785 nm) is guided via an optical fiber into the tissue subjacent to the tip of the fiber, where it is scattered within the tissue. A fraction of the light encounters erythrocytes, is reflected, and is fed back via the fiber to the photodetecting device of the apparatus. From the amount of reflected light the concentration of erythrocytes per tissue volume is estimated. Because erythrocytes move with the bloodstream, the reflected light becomes frequency shifted due to the Doppler effect. The photodetected signal comprises a broad spectrum of Doppler frequency shifted signals from which the average velocity of erythrocytes is derived. An estimate of microvascular blood perfusion is than calculated by the apparatus as the product of mean erythrocyte velocity and mean erythrocyte concentration in the volume of tissue under illumination from the probe [61].

The most important limitation of the laser-Doppler technique is that it can only provide arbitrary, nonabsolute, units for blood perfusion rather than an absolute value of blood flow, which is why the method is often termed laser-Doppler-fluxmetry [25]. This is due to the fact that the actual sampling volume could only be determined if it were possible to identify which erythrocytes have interacted with the light reflected from the tissue. The latter depends on the optical scattering and absorption coefficients of the tissue. Since these coefficients vary considerably with regard to the local microvasculature and its perfusion at the time of measurement, it is impossible to determine the actual sampling volume and thus the blood flow in absolute terms [61].

Estimated mean sampling depths for mammal tissues range from 0.5–1.5 mm, corresponding to “captured” volumes of roughly 1 mm3 of tissue. However, these values must be lower in regions with large light absorption due to large amounts of erythrocytes per volume such as the renal cortex, and lower in the inner medulla with its low amount of erythrocytes per volume [38,39,40, 61].

Considering the rather small volume of tissue for which the laser-Doppler data are obtained versus the heterogeneity of intrarenal perfusion, extrapolation of the results of one area to other areas, even of the same renal layer, is unfeasible. Because of the impossibility to provide absolute data, laser-Doppler-fluxmetry allows investigators only to report relative changes occurring while the optode is fixed in place. Therefore, only continuous and quantifiable changes relative to baseline (control) measurements are recorded, typically induced by acute interventions such as changes in renal arterial pressure or the inspiratory O2 fraction, or by the administration of drugs (see Figs. 3 and 4) [18,19,20, 22, 23, 25, 28, 62]. Yet even interpretation of changes in laser-Doppler data recorded during such maneuvers must be done with due caution: they could rely on changes in local hematocrit, for example, due to changes in plasma skimming, rather than changes in blood flow [25, 39, 48].

Fig. 3
figure 3

Original tracings of invasively measured parameters of renal hemodynamic and oxygenation in an anesthetized rat (left panel) and sketch of the left kidney with the locations of the used probes (right panel). The rat was exposed to staircasewise reductions in renal perfusion pressure followed by staircasewise pressure restorations performed by means of a servo-controlled inflatable occluder placed around the suprarenal aorta (pressure servo) [22]. This was done under normoxic (inspiratory fraction of O2, FiO2 = 0.2), hyperoxic (FiO2 = 1.0), and hypoxic (FiO2 = 0.1) conditions

Fig. 4
figure 4

Time courses of invasively measured parameters of renal hemodynamics and oxygenation in two groups of anesthetized rats that both received a 1.5 mL bolus injection of an X-ray contrast medium (iodixanol 320 mg iodine per mL) into the thoracic aorta at time 0. In the blue group, a low dose of nitrite was administered by continuous infusion, while in the red group saline was infused (control). Conductance values (the reciprocal of vascular resistance) were calculated by dividing the respective perfusion values by arterial blood pressure, in order to distinguish flow changes that result from passive circular distension/compression of vessels from those actively exerted by vascular smooth muscles. Values (mean ± SEM) are given as relative changes from baseline. Redrawn from Ref. 20

In many studies on renal hemodynamics, data from two probes are recorded, one placed within the cortex and the other one in the (outer) medulla. This approach has been also employed in chronically instrumented rats [22]. It must be noted, that the potential tissue injury inflicted by the insertion of the optodes (usual tip diameters range from 50 to 300 μm) may confound the results. Laser-Doppler fluxmetry devices are supplied by various manufacturers. Oxford Optronics, Ltd., provides combined fluorescence-quenching and laser-Doppler optodes (OxyLite/OxyFlo™) that enable simultaneous recordings of local tissue pO2 (vide infra) and of relative changes in local tissue perfusion in the same circumscribed area of renal tissue (see Fig. 2b).

3.3 Clark-Type Electrodes

These polarographic electrodes were developed in the 1950s by Clark and colleagues [63]. The electrodes consist of a noble metal (e.g., silver, gold, platinum) which reduces O2 due to a negative polarizing voltage. The difference in voltage between the reference electrode (anode) and the measuring electrode (cathode) is proportional to the amount of O2 molecules being reduced on the cathode. This enables quantification of tissue pO2 (absolute values in mmHg). The critical technical advance was electrical insulation of the anode and cathode with a gas-permeable but liquid-impermeable membrane. This approach provided the foundation for the development of miniaturized versions (see Fig. 2c) with tips as small as 1–5 μm, that enable quantification of pO2 even in specific vascular and tubular elements of the kidney [64, 65]. Using such small electrodes, Lübbers, Baumgärtl, and colleagues were the first to document the vast heterogeneity of renal tissue pO2 among as well as within the renal layers that is depicted by Fig. 1 [43,44,45]. These authors generated “spatial maps” with the frequency distribution of pO2 values (“pO2 histograms”) from the cortex to the inner medulla, by moving the electrodes through the renal tissues using a nanostepper technique (step width 50 μm).

While “pO2 histograms” have also been obtained in a few recent studies [66, 67], the majority of studies uses fixed electrodes, making sure that the tips remain at the same location within the kidney throughout the experiment. Often, one electrode is placed within the cortex and the other one in the (outer) medulla. The tissue volume that is “captured” by the electrodes has probably about the same diameter as the tip of the electrode; thus, standard electrodes of ≤10 μm diameter obtain pO2 values that are representative for a relatively small tissue volume only [67]. In the face of the spatial heterogeneity of intrarenal pO2, this constitutes a major limitation of this technique. On the other hand, the potential tissue injury inflicted by the insertion of these small electrodes is less than with bigger electrodes or fluorescence quenching optodes that enable better “spatial averaging” (vide infra) [24, 35, 67]. Another drawback of Clark-type electrodes is O2 consumption due to their electrochemical reduction, which is particularly unfavorable in areas with a fragile balance between O2 delivery and demand such as the renal medulla [67,68,69,70].

Furthermore, it is not technically feasible to make repeated measurements in the same anesthetized animal over time-frames longer than a few hours. Yet, recently, a telemetric method using carbon paste electrodes has been developed that allows assessment of renal tissue pO2 in conscious rats [35]. This method provides exquisite temporal resolution [24]. The major limitations of the method are (1) that it requires considerable skill and expertise to construct and implant the devices, (2) that tissue pO2 can only be monitored at a single location in each animal, and (3) that only relative changes in tissue pO2, rather than absolute levels are measured.

3.4 Fluorescence-Quenching Optodes

In the 1990s, a fiber-optic oxygen-sensing device was developed that measures pO2 from the lifetime of fluorescence of a (ruthenium- or platinum-based) luminophore at the tip of an optode [70, 71]. The probe is connected to a pO2 meter developed by Oxford Optronics, Ltd., that generates short pulses of blue LED light, which are transmitted via the fiber to the tip of the optode [70]. The resulting emission of fluorescent light is quenched by the presence of O2 molecules. The OxyLite™ instrument measures the lifetime of each pulse of fluorescence, which is approximately inversely proportional to the pO2 according to the Stern-Volmer equation at pO2 values between 0 and 100 mmHg [68, 71].

Direct comparisons between Clark-type electrodes and quenching optodes reveal their respective advantages and disadvantages [67, 68, 70]. Both techniques share the disadvantage of invasiveness, and the advantage of enabling absolute quantification of pO2 (in mmHg). Advantages of the optode include the absence of O2 consumption and a greater accuracy at low pO2, which is of particular importance when it comes to measurements in the renal medulla. In contrast to the electrodes that “capture” a tissue volume that is approximately equal to their tip diameter, optodes do not “capture” a proper volume . The pO2 measured by the optodes is that within the probe-head (at the tip) of the optode itself, which covers a circular area of 230 μm diameter [67, 68, 70]. The larger diameter of the optodes as compared with that of standard pO2 electrodes increases the risk for tissue damage. On the other hand, it facilitates a favorable “spatial averaging” of renal tissue pO2 values in the face of their huge spatial heterogeneity. In fact, “pO2histograms” of renal cortical pO2 obtained in rabbits by a fluorescence optode were more tightly distributed than those provided by a Clark-type electrode, which may aid detection of (patho)physiologically significant changes in intrarenal oxygenation [67]. Finally, the optodes have been proven to be MR compatible and MR safe [25, 26].

As mentioned earlier, combined fluorescence-quenching-laser-Doppler probes (OxyLite/OxyFlo™) enable simultaneous measurements of local tissue pO2 and assessment of local tissue perfusion in the same circumscribed area (see Fig. 2b).

3.5 Monitoring of Arterial Blood Pressure

It is mandatory for in vivo studies in anesthetized experimental animals, be it studies on control of renal hemodynamics and oxygenation or on every other functional parameter, to monitor cardiovascular variables such as arterial blood pressure and heart rate throughout the experiment in order to ascertain that physiological cardiovascular control is maintained. In anesthetized animals this is best achieved by use of invasive techniques: a fluid-filled catheter inserted into a conduit artery (e.g., femoral or carotid arteries) connected to one of the various commercially available pressure transducers and quantitatively calibrated via a recording device (see the chapter by Cantow K et al. “Monitoring Renal Hemodynamics and Oxygenation by Invasive Probes: Experimental Protocol”).

Direct intra-arterial measurements in unanesthetized animals is the most physiologically relevant means of arterial pressure determination. Telemetric pressure monitoring is the gold standard in conscious animals, and various telemetric devices are commercially available [72, 73]. A limitation of this technique is that implantation surgery and maintenance of catheters require considerable skill [72]. The tail-cuff method is a simple noninvasive plethysmographic technique to assess arterial pressure in conscious rats and mice [74, 75]. Since the conditions for tail-cuff measurements necessarily involve heating and restraint of the animal, which almost inevitably alter the measured parameter (arterial pressure), the physiological meaningfulness of the results is questionable [73].

3.6 Pimonidazole Adduct Immunohistochemistry

This method was originally developed for detection of hypoxia in tumors, but subsequently adapted for use in the kidney [53]. For this technique, pimonidazole (usually 60 mg/kg of body mass) is administered to the experimental animal in vivo [54]. Pimonidazole forms adducts within hypoxic (<approximately 10 mmHg) cells. These adducts are immunohistochemically visualized in ex vivo samples of the kidney. The advantage of this technique is that it provides good spatial resolution of renal tissue hypoxia at the cellular level. Its major limitations include (1) that it detects severe hypoxia only, with a cutoff of about 10 mmHg, (2) that hypoxia can only be assessed postmortem, and (3) only at a single time point. Those using this technique must also be aware of two major sources of false-positive detection of hypoxia, as recently revealed [55]. Firstly, pimonidazole adducts can form postmortem in cells that were not hypoxic in the living animal. For this reason, it is recommended that kidney tissue be perfusion fixed, rather than immersion fixed. Secondly, when the standard (mouse ) monoclonal antibody is used, false positive staining has been observed in damaged rat kidney tissue, presumably as a result of the anti-mouse secondary antibody binding to mouse-like antigens in the damaged tissue. This problem can be avoided by use of a newly commercially available polyclonal anti-pimonidazole antibody raised in rabbits [55].

3.7 Examples of Methods That Are Not Generally Available

3.7.1 Bladder pO2

It is not technically feasible to measure renal tissue pO2 in humans. It can also be technically difficult to measure renal tissue pO2 in experimental animals unless they are deeply anesthetized. However, it is technically feasible to measure the pO2 of urine in both experimental animals and man. There has been a long history of interest in this approach for monitoring renal medullary tissue oxygenation [76]. However, with the recent development of fiber-optic probes that can be inserted into a standard Foley bladder catheter, it has been possible to subject this method to close scrutiny [77]. Available evidence now indicates that urinary pO2 can provide a good estimate of renal medullary pO2 [77]. Because this technique can be deployed in humans equipped with a bladder catheter, it provides a translational pathway for development of approaches to improve renal medullary oxygenation in human patients at risk of AKI . It is also noteworthy that low urinary pO2 during cardiac surgery requiring cardiopulmonary bypass predicts development of postoperative AKI . In light of this, urinary pO2 measurements may provide a real-time measure of medullary oxygenation and predict the risk of AKI in critically ill patients. The major limitation of this approach is that the combination of low urine flow and high arterial pO2 can confound the relationship between urinary pO2 and medullary pO2 [77].

3.7.2 Near Infrared Spectroscopy

Near infrared spectroscopy (NIRS), together with models for diffuse optical imaging affords the measurement of (1) tissue concentrations of oxygenated hemoglobin (oxyHb) and deoxygenated hemoglobin (deoxyHb), (2) oxygen saturation of Hb (StO2) within tissues, and (3) tissue blood volume fraction in vivo. In particular, spatially resolved reflectance NIRS (SRR-NIRS), together with Monte-Carlo simulations of diffuse light propagation, has been applied to monitor these parameters within renal cortical tissue [78]. Quantification of the underlying absorption and reduced scattering coefficients of the tissue is based on analyzing the decrease of the intensity of diffusely remitted light with increasing distance from the light injection point. Recording techniques for the renal cortex of anesthetized rats have been established, with a linear fiber array placed on the ventral surface of the exposed kidney. Measurements have been done at selected near-infrared wavelengths between about 670 and 850 nm by multiplexing millisecond pulses from the different laser sources in time. In this way, quantitative measurements of renal cortex parameters have been achieved at a repetition rate of multiple Hz [78].

SRR-NIRS has been successfully combined with “classical” invasive techniques to study renal hemodynamics and oxygenation in rats [78]. Thus, SRR-NIRS-derived parameters such as the amount of Hb per tissue volume and StO2 were obtained simultaneously with parameters including total renal blood flow, local cortical tissue perfusion and tissue pO2 in the same kidney. Control of renal hemodynamics and oxygenation was studied by dedicated reversible test interventions including brief occlusions of the renal artery or the renal vein, short-term hypoxia, hyperoxia, and hypercapnia, and administration of adenosine. The results demonstrate that the combined approach, by providing different but complementary information, enables a more comprehensive characterization of renal hemodynamics and oxygenation [78].

The SRR-NIRS technique could also serve to help calibrate renal T2* MRI. This MR surrogate parameter for blood oxygenation is related to the amount of deoxyHb per tissue volume . Confounders of the T2* to tissue pO2 relationship include the vascular volume fraction, the local hematocrit, and shifts of the oxyHb dissociation curve [25,26,27,28]. Since SRR-NIRS enables quantitative monitoring of the amount of Hb per tissue volume and of StO2 it is an ideal candidate to unravel and quantify the contributions of confounding factors to renal T2* MRI.

By extending the wavelength range beyond 900 nm, the tissue water content can additionally be assessed, which, together with the reduced scattering coefficients at the different wavelengths, can serve as a surrogate for other (patho)physiologically relevant parameters such as the tubular volume fraction.

Application of SRR-NIRS in reflection mode constrains measurements to the renal cortex of rats. Recently it was demonstrated that changes of rat medullary Hb parameters can also be assessed by SRR-NIRS in transmission mode using an additional fiber placed on the dorsal surface of the kidney [79].

3.7.3 Intravital Microscopy

Perhaps the “final frontier” in methods for assessing renal oxygenation are techniques that provide cellular resolution, in real time, in living animals. Hirakawa and colleagues have recently reported use of a technique they call “intravital phosphorescence lifetime imaging microscopy” [47]. This technique allows cellular pO2 in the renal cortex of live animals to be resolved to the level of individual tubular cross-sections. It requires injection of a molecule whose phosphorescence is quenched by O2. Thus, pO2 can be measured by assessing the decay curve of the phosphorescence. Such probes are available that accumulate intracellularly or are retained within the vasculature. Thus, it is possible to use this approach to assess the heterogeneity of renal blood and tissue pO2 in real time with unprecedented cellular resolution. Currently, the technique is limited for use in the superficial cortex.