Key words

1 Introduction

The water proton longitudinal (or spin-lattice) relaxation time, T1 (s), is a MR parameter commonly used in biomedical imaging. Unlike machine-dependent metrics such as “signal intensity in T1-weighted MRI,” T1 measures a real physical process (i.e., the return to Boltzmann equilibrium of water proton longitudinal magnetisation) measured in absolute units (s) and, for any given magnetic field strength, independent of the machine used. Renal pathology often results in abnormal T1 either globally (e.g., in renal failure) or focally (e.g., in renal cancer). Moreover, disease-induced changes in T1 can often be assigned, at least qualitatively, to pathological changes such as fibrosis, inflammation, and oedema [1]. A second motivation to measure T1 is in so-called dynamic experiments. In this case the object under investigation in the MRI scanner for a short period (e.g., 10 min) while T1 is measured repetitively, during some intervention to perturb T1 (e.g., contrast agent administration). After administration of a contrast agent such as manganese, a gadolinium chelate, or dioxygen, changes in T1 are linearly related to the contrast agent concentration, which could not reliably be determined from signal intensity changes alone.

Here we describe MRI protocols for mapping and monitoring T1 in the kidney of small rodents in a step-by-step experimental protocol. The rationale for choosing different acquisition parameters depending to the aims of the study is discussed, together with specific parameter examples. T1 measurements for dynamic studies are discussed in the chapter by Irrera P et al. “Dynamic Contrast Enhanced (DCE) MRI-Derived Renal Perfusion and Filtration: Experimental Protocol.”

1.1 Motivation for Measuring T1 in Animal Kidneys

Imaging provides many useful metrics in animal models of kidney physiology and pathophysiology. These include “extensive” or size variables, such as renal volume (ml) or thickness (mm) of renal cortex, and “intensive” or tissue characterisation variables, such as the longitudinal relaxation time, T1 (s). Some investigators prefer to report the longitudinal relaxation rate R1, the reciprocal of T1, because from a metrology perspective R1 is a “ratio variable” while T1 is merely an “interval variable” [2]. In this work we use T1. “Intensive” variables can be mapped, the maps themselves being a type of image. Such image metrics used in animal studies are valuable in translational research, as they are directly comparable with the imaging markers [3,4,5] used in studies of human subjects.

1.2 Study Design Concepts

Great care is needed in the choice of method to map T1. No single method is optimal for all in vivo studies. The most accurate methods (e.g., inversion recovery with long TR and short TE readout) are neither fast nor efficient. In vivo studies involve complex trade-offs between accuracy, speed, spatial resolution, field of view, need for fat suppression, sensitivity to inflow, sensitivity to motion artifact, biexponential T1 decay, and other confounding tissue magnetisation behaviours such as transverse relaxation and magnetisation transfer. Moreover, even after a specific method is chosen, errors can be very sensitive to pulse sequence parameters such as delays, flip angles, spoiling and refocussing strategies, mis-set RF pulses and so on. For the renal investigator the first question in designing a study to measure T1 must always be: what is the aim of the study?

  1. 1.

    For studies of global pathology, very high spatial and temporal resolution may be unnecessary provided that the cortex, the zones of the medulla, and the papilla can be resolved. In such studies a 2D image with voxel size of around 0.25 mm (in mice, e.g., [6,7,8]) or around 0.5 mm (in rats, e.g., [9,10,11,12]) may be adequate. It is often desirable to acquire T1 with low uncertainty so that small early pathological changes can be detected, and also to acquire accurate T1 values that are reproducible between centers so that findings are generalizable. A saturation recovery sequence with ten or more inversion times, a 2D RARE readout, moderate spatial resolution, and a two- or three-parameter fit, should provide T1 values which are not biased by B1+ errors. Alternately an inversion recovery sequence with a 2D echo-planar readout [6, 8] may be used although this is less time-efficient. The accuracy of the T1 acquisition and analysis method should be verified using phantoms of known T1 [2].

  2. 2.

    Studies of focal pathology may require high spatial resolution in 3D appropriate to the disease under investigation. Low uncertainty, high accuracy, and high reproducibility are also likely to be important so that scan times may be relatively long. 3D variable flip angle sequences are convenient, although they are quite vulnerable to B1+ errors. In these cases, phantom studies are highly desirable to mitigate B1+ errors, although phantom studies do not mitigate the motion artifact error to which these sequences are also highly vulnerable.

  3. 3.

    Longitudinal studies, during the development and/or treatment of disease have similar demands on spatial and temporal resolution, although if the study endpoint is the change in T1, requirements for accuracy may be relaxed: in a significance test of post vs pre, or treated vs untreated, systematic errors from biased T1 measurement may partly cancel out.

  4. 4.

    Dynamic studies have similar demands on spatial resolution (e.g., high 3D resolution for focal disease) together with very good temporal resolution to allow dynamic changes to be quantitated. In consequence accuracy may be compromised, but since the study endpoint relies on T1 change during the dynamic run, bias may partly cancel out. Propagation of errors in these cases can be exceedingly complex and requires careful simulation.

1.3 Data Quality

It is highly desirable to verify the accuracy and repeatability of T1 using phantoms [2], before experimental animals are exposed to study procedures. The design of MR phantoms can be challenging, and while validated traceable phantoms have been developed by metrologists for clinical MR [13,14,15], these are usually too large for use in rodent MR scanners. Design considerations should include biological, chemical, and physical stability, temperature independence, convenience, and cost. The phantom should cover both the expected T1 range and the spatial range, since it may be advantageous to position the kidneys offset relative to the magnet’s isocenter [16]. Nickel-doped agarose gels [17,18,19] are particularly convenient as simple liquid flood phantoms may introduce artifacts from unfamiliar MR phenomena such as radiation damping, convection, excessive T2 or self-diffusion, standing waves, or abrupt boundaries.

This chapter is part of the book Pohlmann A, Niendorf T (eds) (2020) Preclinical MRI of the Kidney—Methods and Protocols. Springer, New York.

2 General Considerations

2.1 Magnetic Field Strengths

Animal kidney MRI has been performed at a wide range of magnetic field strengths (B0) between at least 0.4 T [20] and 16.4 T [21]. T1 increases with B0 [22]. Hence it is important to consider B0 as an important factor in experiment design, although in practice B0 may not be an adjustable parameter since many facilities only have a single MRI machine operated at a single magnetic field strength available. Higher magnetic field strengths are not always preferable, as the scan times required to properly sample the recovery of longitudinal magnetization will be somewhat longer, and problems from motion artifact, chemical shift artifact or susceptibility variation may increase. For an investigator who wishes to use rodent renal T1 changes to study, say, the development of fibrosis or inflammation, the relevant metric is T1 contrast-to-noise ratio per unit time, which does not necessarily improve as field strength increases: a high-field (≥9.4 T) magnet may be no better than a conventional 3-7 T magnet. On the other hand, for an investigator who measures the uptake of a relaxive substance such as dioxygen, manganese, or gadolinium chelate, a high-field magnet provides a lower baseline R1, making the induced changes ΔR1 more evident.

2.2 Physiologic Motion

Acquisition parameters should ensure that interference from physiologic motion on the signal is kept to a minimum. Indeed, although kidneys are positioned quite low in the abdomen, they are still subject to mild respiratory motion artifacts. It is possible to acquire continuously under shallow breathing, taking advantage of the slow speed of breathing in anesthetized animals [23]. With such datasets, some of this motion can be corrected for during image processing with appropriate coregistration strategies or by intentionally skipping frames presenting a high degree of respiratory motion from the analysis. However, because this strategy potentially removes important datapoints or because it introduces physiological noise into the quantitative estimation, it is still recommended to implement motion compensation strategies at the acquisition level. In particular, respiratory motion compensation takes advantage of a respiratory signal (acquired externally or with MR signals such as navigators) to only allow acquisition during the long periods of breathing motion arrest that are typical of anesthetized animals. An example of such a setup applied to a clinical scanner, is depicted in Fig. 1. Special care is needed to ensure such triggering approaches are compatible with T1 mapping protocols. In such protocols, sequence timing dictates the effective repetition time, upon which the accuracy of the T1 computation is based. It is sometimes preferred to position the trigger on the inversion (or saturation) RF pulse rather than on the slice selection or other imaging loops [16] so that the timing of image acquisition relative to inversion is guaranteed. The only timing affected in this strategy is the delay between saturation or inversion slices. In inversion recovery protocols this delay is already very long, (often more than 10 s), so any additional delay incurred will not significantly affect T1 accuracy, while in saturation recovery the saturation pulse destroys the longitudinal magnetization so that preceding delays are unable to affect T1 accuracy.

Fig. 1
figure 1

Scheme of the experimental setup for a respiratory-gated implementation with monitoring small animal system SA Instruments, Inc., Stony Brook, NY11790 USA

An alternative approach for respiratory motion compensation is retrospective gating or self-gating, which is compatible with gradient echo methods such as Variable Flip Angle (VFA). In this case, data are acquired continuously but an additional “navigator” signal allows data displaced by respiration to be sorted or discarded. This is marketed under the brand-name “IntraGate” (Bruker Biospin, Ettlingen, Germany).

Because kidneys are anatomically close to the bowels, some authors favor imaging sessions done after fasting periods [10]. Fasting should only be done after considering the physiologic consequences, and from the animal welfare perspective should be of the minimum duration consistent with the imaging aims.

3 Imaging Techniques

Conventionally, for T1 measurement, signal intensity is probed at different time points during exponential decay back to Boltzmann equilibrium following inversion or saturation of the longitudinal magnetization: inversion recovery or saturation recovery. Saturation can be achieved using a specific saturation pulse, or simply by varying the repetition time TR of a conventional spin echo or gradient echo acquisition which also incidentally saturates the magnetization. To reduce the total acquisition time, the Look Locker (LL) sequence can be used. With LL sequences, the magnetization is also prepared with an inversion or saturation pulse [24], but instead of a single sample of the recovery curve per TR, multiple samples are taken using a train of low flip angle pulses [25].

The inversion recovery sequence is the reference method to calculate T1 but it suffers from long acquisition times. Therefore, the use of Look-Locker inversion recovery is advantageous, but is limited to single 2D slice acquisition: if multiple slices are needed, they must be acquired sequentially.

The alternative approach involves repeating a gradient echo acquisition at varying flip angles (VFA) while keeping TR and other sequence parameters constant: since the signal depends sinusoidally on flip angle but decays exponentially with T1, the T1 can be calculated from the Ernst equation (chapter by Garteiser P et al. “Analysis Protocols for MRI Mapping of Renal T1”). Three-dimensional acquisition (3D) can be performed with VFA, but since VFA is sensitive to B1+ inhomogeneity, the B1+ field should be mapped to allow correction (chapter by Garteiser P et al. “Analysis Protocols for MRI Mapping of Renal T1”). Although 3D VFA is familiar in clinical imaging, it is not commonly employed in rodent kidney studies, and we have not provided a protocol here.

3.1 RARE-VTR

3.1.1 Protocol Description

Rapid acquisition with relaxation enhancement at variable TR (RARE-VTR) is a type of saturation recovery sequence . Data are acquired as 2D slices, so this sequence is ideal when a single slice provides an adequate sample of the kidney, for example in the case of homogeneous global pathology. If volumetric data are required then a stack of slices must be acquired sequentially, which is time-consuming.

The RARE-VTR protocol is based on the multiecho RARE protocol, with a slice-selective 90° rf pulse followed by a train of slice-selective 180° refocusing pulses separated by an inter-echo time TE (Fig. 2). Each 180° pulse induces a refocused echo comprising spin-echo and stimulated echo contributions, which is digitized, and each echo experiences a different phase encode gradient. Following the last echo there is a delay following which the sequence is repeated with a repetition time TR. The number of 180° pulses is called the echo train length (ETL) or RARE factor: a typical value is 8. TR is then varied to sample the magnetization recovery. A saturation recovery time Tsat is defined as TR  (ETL × TE). Thus if an investigator requires a 1282 matrix, the entire acquisition duration would be ∑TR × 128/ETL, where ∑TR is the sum of all the TR values used to sample the longitudinal magnetization recovery. Recommended acquisition parameters are given in section 4 below.

Fig. 2
figure 2

Simplified timing diagram for 2D saturation recovery sequence with RARE readout. A series of 180° pulses creates f echoes at TE, 2 × TE, 3 × TE, and so on, where f is the RARE factor, typically f = 8. During the phase encoding (PE) step, spatial location is encoded by incrementing the Y gradient NY times so that each echo is encoded with a different phase, thereby accelerating the k-space coverage. During the frequency encoding (FE) readout periods, the spatial location is encoded in the frequency content of the digitized signal from the echo. If the pulses are perfect and if neglecting the longitudinal regrowth occurring during the echo train, all the magnetization is effectively kept in the transverse plane until the last echo (Phase cycling and spoiling not shown)

Because of the presence of refocusing RF pulses, this protocol is relatively insensitive to B0 imperfections and susceptibility artifacts. For 2D single-slice T1 mapping it is quite time-efficient, but acquisition of a multislice stack requires rather long acquisition times.

3.1.2 Repetition Times

The set of repetition times that are sampled needs careful consideration. The TR values should be rationally selected during preliminary studies with the animal model under investigation. In saturation recovery experiments, the signal starts at S0 ~ 0 at short TRs (TR ≪ T1) and increases with approximately linear S/T1 dependence on TR. At longer TRs (TR ≈ T1), the signal enters a transition regime where the increase in signal per unit TR diminishes. Finally at large TRs (TR ≫ T1), the signal asymptotically approaches S. Hence, proper TR sampling strategy should be adopted to cover both regimes and the transition in between.

In determining the set of repetition times to use, the first question may be to select the highest TR that is required. This choice is mainly dictated by the expected T1 in the tissue of interest, the available scan time and the planned postprocessing strategy.

For the first constraint, literature values or preliminary studies may help in the determination of the expected tissue T1 at the corresponding magnetic field strength. The longest TR should be substantially larger than the longest expected tissue T1. Typical T1 values for cortex and medulla are in the range 1–2 s and for urine around 3 s. There is no generally accepted consensus, but authors typically have the maximal TR be 3–5 times the longest expected T1, typically 8–12 s. If the investigator finds this makes the overall study duration too long then spatial resolution can be reduced, particularly when small focal changes are not of interest. Alternatively a maximal TR of two times the longest expected T1 could be considered, but in this case it is recommended to perform careful simulations to ensure the propagated errors are at an acceptable level. A special case arises in contrast agent injection studies. Here, the distribution of the agent into the tissue of interest is quantified by looking at the absolute change in R1 after vs before injection. Hence, care should be taken to consider that the T1 value may be significantly shorter after injection. It may be necessary to adopt a different maximal TR for the pre- and postinjection scans. Contrast agents decrease T1, hence the possible decrease in longest TR that is afforded has the additional consequence of shortening the scan duration. This is potentially advantageous, especially when the dynamics of contrast agent distribution needs to be interrogated and is slow (on the order of minutes). In this case it becomes feasible to monitor the contrast agent distribution dynamically by repetitively acquiring T1 maps. Temporal resolutions of under 2 min can reasonably be achieved by trading off, for example, spatial resolution. This strategy is compatible with some contrast agent dynamics such as in manganese-enhanced MRI [21].

The choice of the longest TR should be done in harmony with the particular postprocessing solution that is planned. Indeed, the dataset obtained with the longest applied TR may be used to provide an estimate for the total equilibrium signal S. In that case, the mathematical problem to be solved is reduced since S can be inserted in the fitting equation, thereby removing a degree of freedom and making the fit more robust. This comes at the cost of the necessity to acquire a scan at very long TR to ensure the signal is close enough to equilibrium values. When such an estimate of S is available, it may no longer be necessary to sample other TRs longer than the tissue T1, concentrating instead on TR values at and below the tissue T1 [21].

Once the largest TR to acquire has been selected, the number of TRs should be determined. The set of TR values should span the range of expected T1. Depending on the fitting equation that is used, a larger number of repetition times may be required to provide sufficient overdetermination for the fitting procedure. In other words, there should be significantly more TRs sampled than there are free parameters in the fitted equation. In an ideal situation with negligible experimental noise, it may be sufficient to select only a number of TR values equal to the number of free parameters in the fit. However, with experimental noise, this rapidly becomes insufficient and introduces bias into the results. Hence, when signal to noise ratio is limited, it is recommended to acquire a larger number of TR values. This comes at the cost of increased scan time, since each TR is generally acquired sequentially. In saturation recovery sequences, the fitting equation only has a limited number of parameters (i.e., S, R1, and maybe S0 and/or an error term); hence, as few as four TR values are acceptable providing sufficient signal to noise is available and sampling is appropriately selected.

In disease models the pathology evolves slowly over days or weeks, the scan duration becomes less important, and it is possible to increase the number of TR values to 6, 8, or even 12, although from an ethical perspective the duration of the procedures should always be as short as possible. The choice of TR values can even be done a priori using Cramer-Rao lower-bound methodology. This technique optimizes the distribution of the sampled variable (in that case TR), for a given number of samples, an expected noise distribution and a signal model. It was used already with some success for cardiac T1 mapping, where it was found that at constant number of acquired averages, increasing the SNR by averaging a carefully selected subset of TR values could prove beneficial relative to a set of uniformly distributed TR values [26]. A similar approach can be proposed to optimize the repetition times to use in kidney T1 mapping.

In typical MRI scanners, the specific TR values can be set manually with a TR table having a user-specified number of TR entries. Manufacturers also often provide utilities that assist the user in prescribing the TR series. These utilities query from the user only a minimal value, a maximal value and a number of steps, or a minimal value, a number of steps and an estimate of the tissue T1. These input values are then used to generate the series of TR values to use during the acquisition.

3.2 Recovery Techniques: Look-Locker and Derivatives

3.2.1 Protocol Description

Look-Locker (LL) is a variant of inversion recovery or saturation recovery sequence . Data are acquired as 2D slices, and if volumetric data are required, a stack of slices must be acquired sequentially. LL involves sampling the magnetization using rapid, small angle imaging readouts while it recovers from the inversion or a saturation RF pulse (Fig. 3). The temporal spacing between imaging readouts provides a rapid way to sample the longitudinal regrowth. Indeed, each imaging shot will occur at a different time during regrowth, hence many different recovery times can be sampled during a single recovery event. Full longitudinal regrowth takes longer after an inversion RF pulse than after a saturation RF pulse, however, adopting an inversion RF pulse increases the dynamic range of the sampled signal. Thus the choice between saturation or inversion for the preparation of magnetization in a recovery sequence may be dictated by the experimental necessity for short acquisition time (for instance in dynamic experiments involving contrast agents with rapid kinetics) or higher measurement precision.

Fig. 3
figure 3

Schematic view of the look-locker inversion recovery pulse sequence (fast low flip angle spoiled gradient echoes variant with spoiler gradient pulses omitted for simplicity). τ repetition time, Ny prescribed number of k-space lines, Nsegments number of k space segments (acceleration factor), TI inversion recovery time. Effective recovery times are considered at the centerline of each imaging segment. τrecovery: delay added to enable full recovery of the longitudinal magnetization before the subsequent inversion pulse is applied. In other variants, the fast low angle shots can be replaced by steady state free precession modules or EPI modules (Accessory gradient lobes are not shown for clarity)

The Modified Look-Locker Inversion Recovery (MOLLI) sequence is an alternative when small animal renal T1 mapping is performed using a clinical MR system. One rationale for choosing a clinical platform for preclinical studies is the similarity between MR protocol played out on small animals and human kidneys. MOLLI was initially designed for myocardial T1 mapping at 1.5 T [27]. A typical MOLLI sequence can be performed by acquiring single shot True FISP images at different inversion times after a single inversion pulse. MOLLI is routinely used clinically for T1 mapping and is usually available on clinical MR systems equipped with a cardiac package. The original MOLLI sequence provided by the manufacturers’ library is a good starting point for MOLLI optimization in the rodent kidney. However, translating the MOLLI technique for scanning rodent models on a clinical MR scanner requires a specific implementation. In human, MOLLI is usually acquired in breath hold. In rodent, a physiological monitoring system to track the respiration and synchronize the sequence is needed (see Fig. 1).

3.2.2 Types of Readouts, and Corresponding Specific Points of Attention

In recovery-type sequences, several options are possible for the imaging shots that are performed to sample the recovery period.

Fast low angle shots, where magnetization is spoiled before subsequent RF pulses, can be used. They have the advantage of a short TR (~3 ms) that is compatible with low flip angles (<10°) [9]. As such, they perturb the longitudinal regrowth of the magnetization only to a small extent.

Balanced readouts such as steady-state free precession (SSFP) can also be used, as they tend to yield higher signal to noise due to the conservation of magnetization from preceding shots [28]. The SSFP condition links the TE and TR. Because FISP involves radiofrequency phase coherence between shots, acquisition can be trueFISP, fid mode or echo mode. It acquires innermost the phase encoding (in-segments), then slices, then phase encoding (segments), then (optionally) averaging. SSFP is especially sensitive to careful shimming of the B0. The segmenting can be performed either sequentially or by interleaving phase encode steps. The interleaved mode is recommended for inversion recovery experiments. The T1 recovery sampling must then be setup by adjusting the time between the inversion pulse and the first imaging shot. This delay is incompressible due to the need for stabilizing the steady state condition of the magnetization. This is either done with a half flip angle pulse or with a starter (dummy) sequence , at the cost of a longer delay. Unfortunately for these reasons, the implementation of respiratory or cardiac triggering is challenging, and it is recommended to proceed by gating on the saturation or inversion pulse.

Finally, echoplanar imaging (EPI) readouts have also been proposed in that context [6, 7, 16, 29, 30]. Although they are quite advantageous due to their speed (in single shot EPI, a full k-space image may be acquired after a single excitation pulse), EPI sequences for kidney T1 mapping are vulnerable to problems associated with EPI in general, particularly if both kidneys are to be assessed. In particular, the EPI train is sensitive to B0 inhomogeneity, which can become problematic at higher magnetic field strengths and in large bore systems. This effect is more important for long echo train EPIs such as single shot EPI. In this situation there is also a long effective echo time to accommodate the train length required by the single shot mode. This can be detrimental because it can increase the point spread function diameter (“smearing”), thus reducing the effective spatial resolution [30]. The long echo time of single shot EPI also comes at a disadvantage in kidneys with short \( {T}_2^{\ast } \) such as kidneys after injection of high concentrations of contrast agents. EPI is also associated with signal dropouts and geometric distortions at regions of strong magnetic susceptibility gradients, which sometimes occur at the kidney edges [30].

These disadvantages can be mitigated by performing segmented EPI acquisitions. In segmented EPI, full k-space acquisition is subdivided over several recovery periods, thereby enabling for decreased echo times. Segmentation will, however, increase the acquisition time, since full k-space coverage will be performed over a greater number of recovery periods. Finally, EPI image quality is also adversely affected by the presence of fat, which may be abundant in some kidney disease models such as obesity-related disease models. This can be remedied by applying fat saturation pulses [7].

3.2.3 Duration of Imaging Readouts

The duration of the imaging readouts should be short relatively to the duration of the recovery. This enables to interrogate precise inversion recovery delays rather than long intervals during which the longitudinal magnetization does vary substantially. Furthermore, short imaging readout durations are also advantageous because more imaging shots can be positioned during longitudinal regrowth. The increased number of data points yields better numerical stability during the subsequent fitting procedure.

3.2.4 Number of Imaging Readouts

The total number of imaging readouts that can be fit in a single regrowth is limited by several factors. First, the imaging sequence (excitation and readout) itself takes an incompressible minimum amount of time. Second, there is a requirement to acquire full images at each of the recovery delays. Hence, if only a single k-space line is acquired at each recovery delay, this also means that to acquire the appropriate number of k-space lines for image reconstruction (e.g., 128), an equal number of saturation or inversion recovery periods will be necessary. Since full relaxation needs to be reached (typically 10 s or more) before the next inversion RF pulse can be applied, this implies very long acquisition times (20 min or more). Hence it is important to consider acceleration schemes such as echoplanar imaging, where partial k-spaces are acquired during the regrowth rather than single k-space lines. Finally, the image acquisition readouts perturb the growing longitudinal magnetization: the application of the imaging RF pulses slightly saturates the signal, which makes the signal regrow with a biased apparent T1 (sometimes denoted T1*) that is different than the true tissue T1. This effect can be mitigated by using smaller flip angles and longer RF pulse repetition intervals. Postprocessing strategies are also available to correct for that effect (cf. the chapter by Garteiser P et al. “Analysis Protocols for MRI Mapping of Renal T1”).

3.2.5 Repetition Time

In saturation or inversion recovery type sequences, it is important to distinguish the time between RF inversion pulses and the time between imaging RF pulses. The time between saturation or inversion must take into account the T1 of the kidney at the considered field strength. A sufficiently long recovery period must be taken into account to prevent to saturate the magnetization. Typically inversion pulses are separated by 10 s [16, 28], 18 s [6, 7] or even 20 s [30].

The repetition time between imaging segments represents the precision with which the inversion recovery will be sampled. If a high accuracy is required for the T1, only short imaging segments with a few to a single k-space line can be adopted. This will ensure that there is only little confounding factor arising from the longitudinal regrowth occurring during each imaging segment. Conversely if a lower T1 recovery sampling is acceptable, then imaging can be noticeably accelerated by k-space segmenting.

3.2.6 Inversion Pulses

Adiabatic passage pulses can be advantageous to ensure saturation (adiabatic half passage) or inversion (adiabatic full passage). These pulses maintain their rotation properties invariant to the \( {B}_1^{+} \) field, provided that the pulse B1 exceeds an adiabaticity threshold. This is of importance in the field of kidney imaging, where the presence of intraabdominal fat can be challenging for B1 homogeneity. Hence adiabatic RF pulses such as the hyperbolic secant RF pulse, are often used in T1-mapping protocols, as half passage for saturation recovery experiments [23, 31] or as full passage for inversion recovery experiments [16, 30]. However, these RF pulses tend to require long duration and RF power levels, which may not be compatible with all RF coils, and unless provided by the manufacturer, their frequency and phase modulation shapes may require careful design.

Generally non selective inversion RF pulses are used. Indeed, slice selective inversion RF pulses are exposed to inflow artifacts. Indeed, when a slice selective RF pulse is applied, the spins entering the imaging slice via perfusion effects, have not been resonantly exposed to the RF pulse. Hence their presence in the imaging slice will contribute to increase on average the magnetization. This effect will be higher if perfusion is elevated, and is in fact the basis for FLAIR sequences. Whenever the true tissue T1 is the parameter of interest, perfusion becomes a confounding factor, and inflow artifacts should be minimized. This can be achieved either by selecting an inversion slice that is significantly larger than the imaging slice to minimize the inflow artifact by creating a large region adjacent to the imaging slice that the noninverted spins need to cross before entering the imaged slice. Alternately, a nonselective saturation pulse may be adopted. In this case, even the spins that enter the imaging slice by perfusion during the imaging train are inverted, and hence they will not modify the apparent tissue T1. Some authors recommend offsetting the position of the kidney within the radiofrequency coil to ensure that the heart is present within the effective region of the coil, to ensure maximally inverted spins throughout the entire body of the animal [16].

Recommended acquisition parameters are given in section 4 below.

3.3 Variable Flip Angle Techniques

Variable flip angle (VFA) sequences vary the flip angle of the rf excitation pulse but keep TR constant. Data can be acquired in 3D volume , so such sequences are ideal when the entire kidney must be sampled, for example in the case of heterogeneous or focal pathology (Fig. 4). However, data acquisition takes longer than for 2D sequences, and they are not robust in the presence of B1+ inhomogeneity.

Fig. 4
figure 4

3D variable flip angle protocol. TR repetition time (maintained constant for the duration of the acquisition), TE echo time, θ flip angle (flip angle is varied during the protocol), nPEY, nPEZ number of prescribed phase encoding lines in the Y and Z dimensions, PE phase encoding, FE frequency encoding. While the TR is maintained and constant during acquisition, the flip angle is varied several times until a complete dataset has been fully acquired for each prescribed flip angle. Accessory gradient lobes are not shown for clarity. Often, investigators need to suppress the signal from fat, in which case the hard pulse would be replaced by a spectrally selective “soft” pulse to excite 1H2O but not C1H2 or C1H3. During the phase-encoding (PE) step, spatial location is encoded by incrementing X and Y gradients iX and GY multiple times nPEX and nPEY, typically 64 or 128. During this period magnetization is prephased using the Z gradient GZ, so that a gradient echo may be formed by inverting the polarity of the gradient, yielding a return to phase at TE at the mid-point of the gradient echo. TE is set as short as possible, typically ~2–3 ms, in order to avoid T2* weighting, and to minimize total acquisition time. Practically, this is limited by the specification and performance of the gradient coils and amplifiers available to the investigator. TR is also set as short as possible, ~5 ms, in order to minimize the overall acquisition time

Variable flip angle sequences are usually based on spoiled gradient echo sequences (fast low angle shot) [12, 32, 33], but the same principles can be applied with UTE sequences [34]. By virtue of their 3D nature, they are well suited for analyzing entire kidney volumes. However, the acquisition scheme where a single angle is acquired at a time also limits the number of different flip angle acquisitions that can be achieved within reasonable scanning durations. A linearization of the VFA signal provides T1 estimates based on as little as two flip angles, although this technique is prone to noise bias. In a study at clinical field strength of 3.0 T using a wrist coil, flip angles of 5° and 26° were proposed for a rat model of acute kidney injury [12]. The use of VFA-based protocols for kidney mapping is not yet widespread at higher magnetic field strengths, but this type of sequence is applicable in other organs [34] or in tumor models [32, 33]. VFA schemes are also prone to B1+ field heterogeneity. Indeed, the effectively achieved flip angle at a specific location in the kidney is function of not only pulse amplitude, but also the shape and electromagnetic properties of the animal, the radiofrequency pulse design and (to a lesser extent when volume RF coils are used for RF transmission) RF coil coverage homogeneity. Thus, the B1 field may require to be mapped first. For instance a reference T1 map using a recovery-based sequence can be obtained and then injected into the signal equation to extract the effective flip angle [35]. Another acquisition strategy consists of measuring the B1+ field directly with dual angle spin echo sequence [36, 37]. Recommended acquisition parameters are given in section 4 below.

4 Methods

4.1 MR Protocol Setup

4.1.1 2D RARE VTR in Bruker ParaVision 6

  1. 1.

    Use the ParaVision RARE T1 saturation-recovery METHOD “T1map_RARE protocol” (Rat/Head/Relaxometry, or Mouse /Head/Relaxometry).

  2. 2.

    Set the geometry to coronal, 58 × 58 mm FOV (rat) or 40 × 30 mm FOC (mouse ). These may need adjustment in studies on large (e.g., obese) animals.

  3. 3.

    Set the matrix size to 128 × 128; if acquisition length is too large, you may decrease the phase encoding resolution.

  4. 4.

    Use a single 90° 1.16 mm slice selection (optionally, apply spectrally selective fat suppression).

  5. 5.

    Select an180° train RARE factor 8, matching pulse bandwidths of 90° and 180°, effective TE 30 ms, echo spacing 7.5 ms.

  6. 6.

    Set five dummy scans; no signal averaging.

  7. 7.

    Set TR = 5500, 2000, 1200, 750, 500, 300, 200, and 100 ms.

  8. 8.

    A manufacturer-provided macro is available for data processing.

4.1.2 Inversion Recovery Sequence in Bruker ParaVision 6

  1. 1.

    Use the ParaVision Fluid-Attenuated Inversion Recovery (FLAIR) method using only the global inversion mode (FairMode set to Nonselective), using Flow-sensitive Alternating Inversion Recovery FAIR-RARE or FAIR-EPI Bruker methods.

  2. 2.

    Set the geometry to axial or coronal, 58 × 58 mm FOV (rat) or 40 × 30 mm FOV (mouse ). These may need adjustment on large (e.g., obese) animals.

  3. 3.

    Use a single 1 mm slice selection.

  4. 4.

    Set the matrix size to 128 × 128 matrix; if acquisition is too long, you may diminish the spatial resolution and adopt partial Fourier acceleration.

  5. 5.

    Set an TE of 36 ms (FAIR-RARE) or longer (FAIR-EPI).

  6. 6.

    Set the inversion pulse properties to calculated, as this provides a full passage adiabatic pulse with appropriately good flip angle homogeneity.

  7. 7.

    Use 21 inversion times starting at 30 ms with increments of 200 ms (to be adapted to the particular study at hand).

  8. 8.

    Select a recovery time of 10 s or greater.

  9. 9.

    Adjust the number of averages based on the available scan time.

  10. 10.

    In case of EPI usage, a careful shimming of B0 is paramount. The manufacturer-recommended method is MapShim using an ellipsoidal shim volume , in which a B0 map is measured (accessible in the adjustments platform) and used to correct the B0 with the shims coils available on the system.

  11. 11.

    A manufacturer-provided macro is provided for data processing.

4.1.3 Modified Look-Locker Inversion Recovery (MOLLI) [27] Sequence for Rats on a Siemens Clinical Scanner

  1. 1.

    Install rat in a wrist coil with conventional pressure pad to monitor breath rate. Figure 1 illustrates a dedicated small animal SA Instruments system connected to a Siemens clinical MR system.

  2. 2.

    Load the MOLLI sequence provided by the MRI system. The original MOLLI 3(3)3(3)5 scheme can be applied for kidney imaging. In this sampling scheme, multiple images are acquired at 11 time points on the recovery curve, with three inversions with slightly shifted inversion times, enabling a pixel-based T1 quantification.

  3. 3.

    Start TI at 117 ms with TI increments (ΔTI) between inversion of 80 ms.

  4. 4.

    The slice thickness and in-plane resolution are particularly important to avoid partial volume effects. Use a slice with the lowest thickness and the highest in-plane resolution allowed by the MR system typically resulting voxel size around 0.7 × 1.1 × 3.5 mm at 3 T.

  5. 5.

    Set TR/TE around 711/1.09 ms.

  6. 6.

    Set flip angle from 28° to 35° (default flip angle of 35°) T1 map relies on kidney being aligned between all TI images. As a consequence it is important to provide an additional motion correction method to mitigate residual respiratory motion. It is recommended to perform at least a rigid registration algorithm between images from various inversion times before the pixelwise analysis of the relaxation curve (cf. the chapter by Garteiser P et al. “Analysis Protocols for MRI Mapping of Renal T1”). Some vendors proposed on line motion correction algorithm, such as MOCCO from Siemens, during the reconstruction process.

4.2 Calibrated Phantom Acquisitions

  1. 1.

    Construct or procure a set of filled nickel/agarose phantoms without air bubbles. Each phantom should use a vial appropriate for the size of the animal abdomen (e.g., 25 mm diameter × 40 mm high for a mouse or 50 mm diameter × 80 mm high for a rat). Use 2% agarose, and a different concentration of nickel chloride in each phantom . Suitable nickel concentrations are 0 mM, 0.5 M, 1 mM, 2 mM, and 4 mM: 8 mM, and 16 mM can be added if very short T1 values are expected (e.g., in DCEMRI). Add a preservative, such as 0.05% sodium azide, and seal.

  2. 2.

    Measure T1 for each phantom using the method of Subheading 4.3 and the chapter by Garteiser P et al. “Analysis Protocols for MRI Mapping of Renal T1.” Repeat the measurements on a different day. Temperature should be controlled carefully, for instance by isolating the T1 phantom in an ice–water mixture, or by inserting the phantom in a larger structure with high thermal inertia and preequilibrated at the target temperature.

  3. 3.

    In consultation with a professional statistician, calculate the repeatability and ensure the sample size is adequate to test the hypothesis of interest. From an ethical perspective, this is mandatory before living animals are entered into the procedures.

  4. 4.

    Repeat steps 2 and 3 in a group of living animals. The group size should be determined in consultation with a professional statistician: group sizes of between 6 and 15 are common.

4.3 Animal Experiments (Dedicated Small Animal MR Systems)

  1. 1.

    Use a volume RF coil of suitable diameter. In dedicated preclinical MRI, volume RF coils with an inner diameter of 38 mm for mice or 72 mm for rats could be used.

  2. 2.

    Anesthetize the animal with isoflurane as described in the chapter by Kaucsar T et al. “Preparation and Monitoring of Small Animals in Renal MRI” and transfer it to the scanner.

  3. 3.

    Set up the temperature monitoring (rectal probe) and respiratory monitoring (balloon on chest) unit.

  4. 4.

    Perform anatomical imaging as described in the chapter by Pohlmann A et al. “Essential Practical Steps for MRI of the Kidney in Experimental Research.”

  5. 5.

    Perform localized shimming on the kidney imaging as described in the chapter by Pohlmann A et al. “Essential Practical Steps for MRI of the Kidney in Experimental Research.” Wherever possible, appropriate tuning and matching of the used RF coil should be done.

  6. 6.

    Make sure the respiratory gating is setup properly, and that the anesthesia is stable.

  7. 7.

    Ensure that B1+ is correct, that is, that the relationship between the RF power and the desired flip angle is known in the region of space occupied by the kidneys. In ParaVision this is performed as part of the initial sequence adjustments when the reference pulse power is calibrated.