Abstract
This work describes and characterizes an algorithm for the nonambiguous reduction of multiexponentially decaying luminescence signals to scalar values of corresponding calibration temperatures. Previous evaluation schemes in phosphor thermometry make use of an intermediate step, where data reduction is achieved by fitting a model equation to phosphorescence decays in order to translate one or more fitting parameters into temperature. However, every slight mismatch between model equation and experimental data may lead to substantial errors in connection to noise-related inaccuracies during the retrieval of adequate fitting windows. Additionally, there is a need for fitting windows, capable of automatically adapting to largely varying signal time scales. In this context, the authors propose to set the fitting window length according to the time where the signal falls below a given percentage of the initial intensity. In comparison to fitting windows, defined by multiple decay times, modeling results suggest substantial precision benefits for as long as signal-to-noise ratios stay above 4. Nevertheless, by comparing signal shapes of measured curves directly with a library of temperature-calibrated decay signals, all necessary assumptions on the mathematical description of measured signals become redundant and evaluation errors connected to uncertain fitting windows are largely circumvented. Resulting capabilities of the proposed signal shape recognition method (SSR) in terms of temperature precision and accuracy were compared to a conventional least-squares fitting approach, using a set of temperature-calibrated phosphorescence decay signals from CdWO4. Accordingly, the SSR algorithm was found to reduce statistical temperature errors by at least 9 %.
Similar content being viewed by others
Notes
Some phosphor coatings may exhibit thermal history effects in the initial part of the decay, which become annealed after long thermal exposures. Hence, there are cases where selecting a fitting window that begins further from the start of the decay might be a prudent choice.
References
Allison SW, Gillies GT. Remote thermometry with thermographic phosphors: instrumentation and applications. Rev Sci Instrum. 1997;68:2615–50.
Khalid AH, Kontis K. Thermographic phosphors for high temperature measurements: principles, current state of the art and recent applications. Sensors. 2008;8:5673–744.
Aldén M, Omrane A, Richter M, Särner G. Thermographic phosphors for thermometry: a survey of combustion applications. Prog Energy Combust. 2011;37:422–61.
Brübach J, Pflitsch C, Dreizler A, Atakan B. On surface temperature measurements with thermographic phosphors: a review. Prog Energ Combust. 2013;39:37–60.
Krauss RH, Hellier RG, McDaniel JC. Surface temperature imaging below 300 K using La2O2S:eu. Appl Opt. 1994;33:3901–4.
Cates MR, Allison SW, Jaiswal SL, Beshears DL. YAG:dy and YAG:Tm fluorescence to 1700 °C. Proc Int Instrum Symp. 2003;49:389–400.
Istratov A, Vyvenko O. Exponential analysis in physical phenomena. Rev Sci Instrum. 1999;70:1233–57.
Brühbach J, Janicka J, Dreizler A. An algorithm for the characterisation of multi-exponential decay curves. Opt Las Eng. 2009;47:75–9.
Dowell LD, Gillies GT. Errors caused by baseline offset and noise in the estimation of exponential lifetimes. Rev Sci Instrum. 1991;62:242–3.
Ballew RM, Demas JN. An error analysis of the rapid lifetime determination method for the evaluation of single exponential decays. Anal Chem. 1989;61:30–3.
Ballew RM, Demas JN. Error analysis of the rapid lifetime determination method for single exponential decays with a non-zero baseline. Anal Chim Acta. 1991;245:121–7.
Rosso L, Fernicola V. Time- and frequency-domain analyses of fluorescence lifetime for temperature sensing. Rev Sci Instrum. 2006;77:034901.
Levenberg K. A method for the solution of certain non-linear problems in least squares. Quart J Appl Math. 1944;2:164–8.
Marquardt DW. An algorithm for least-squares estimation of non linear parameters. J Soc Ind Appl Math. 1963;11:431–41.
Knappe C, Lindén J, Abou Nada F, Richter M, Aldén M. Investigation and compensation of the nonlinear response in photomultiplier tubes for quantitative single-shot measurements. Rev Sci Instrum. 2012;83:034901.
Knappe C, Lindén J, Abou Nada F, Richter M, Aldén M. Comparison of photo detectors and operating conditions for decay time determination in phosphor thermometry. Rev Sci Instrum. 2012;83:094901.
Knappe C, Algotsson M, Andersson P, Richter M, Tunér M, Johansson B, Aldén M. Thickness dependent variations in surface phosphor thermometry during transient combustion in an HCCI engine. Combust Flame. 2013;160:1466–75.
Abou Nada F, Knappe C, Xu X, Richter M, Aldén M. Development of an automatic routine for thermographic phosphors calibration. submitted to Meas. Sci. Instrum. 2013.
Acknowledgements
For financial support the authors would like to acknowledge the European Union Seventh Framework Programme (FP7/2007–2011) under grant agreement no. 265861 (Helios).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Knappe, C., Pfeiffer, K., Richter, M. et al. A library-based algorithm for evaluation of luminescent decay curves by shape recognition in time domain phosphor thermometry. J Therm Anal Calorim 115, 545–554 (2014). https://doi.org/10.1007/s10973-013-3337-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10973-013-3337-3