Acetabular Cup Design Influences Deformational Response in Total Hip Arthroplasty

  • Symposium: Papers Presented at the Annual Meetings of The Hip Society
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Press-fit acetabular components are susceptible to deformation in an underreamed socket, with excessive deformation of metal-on-metal (MOM) components potentially leading to increased torsional friction and micromotion. Specifically, however, it remains unclear how cup diameter, design, and time from implantation affect shell deformation.

Questions/purposes

We asked whether (1) changes in component geometry and material altered maximum shell deformation and (2) time-dependent deformational relaxation processes occurred.

Methods

Diametral deformation was quantified after press-fit implantation of metal shells into a previously validated polyurethane model. Experimental groups (n = 6–8) consisted of 48-, 54-, 60-, and 66-mm MOM cups of 6-mm wall thickness, 58-mm cups of 10-mm wall thickness, and CoCrMo and Ti6Al4V 58-mm modular cups.

Results

Greater cup diameter, thinner wall construction, and Ti6Al4V modular designs generated conditions for maximum shell deformation ranging from 0.047 to 0.267 mm. Relaxation (18%–32%) was observed 120 hours postimplantation in thin-walled and modular designs.

Conclusions

Our findings demonstrate a reduction of shell deformation over time and suggest, under physiologic loading, early component deformation varies with design.

Clinical Relevance

Component deformation should be a design consideration regardless of bearing surface. Designs neglecting to adequately address deformational changes in vivo could be susceptible to diminished cup survival, increased wear, and premature revision.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2A–B
Fig. 3

Similar content being viewed by others

References

  1. Affatato S, Leardini W, Jedenmalm A, Ruggeri O, Toni A. Larger diameter bearings reduce wear in metal-on-metal implants. Clin Orthop Relat Res. 2007;456:153–158.

    Article  PubMed  CAS  Google Scholar 

  2. Cuckler JM, Moore KD, Lombardi AV Jr, McPherson E, Emerson R. Large versus small femoral heads in metal-on-metal total hip arthroplasty. J Arthroplasty. 2004;19:41–44.

    Article  PubMed  Google Scholar 

  3. Dowson D, Hardaker C, Flett M, Isaac GH. A hip joint simulator study of performance of metal-on-metal joints. J Arthroplasty. 2004;19(8 suppl 3):124–130.

    PubMed  Google Scholar 

  4. Everitt H, Evans SL, Holt CA, Bigsby R, Khan I. Acetabular component deformation under rim loading using digital image correlation and finite element methods. Appl Mech Mater. 2010;24–25:275–280.

    Article  Google Scholar 

  5. Jin ZM, Meakins S, Morlock MM, Parsons P, Hardaker C, Flett M, Isaac G. Deformation of press-fitted metallic resurfacing cups. Part 1. Experimental simulation. Proc Inst Mech Eng H. 2006;220:299–309.

    Google Scholar 

  6. Jin ZM, Meakins S, Morlock MM, Parsons P, Hardaker C, Flett M, Isaac G. Deformation of press-fitted metallic resurfacing cups. Part 2. Finite element simulation. Proc Inst Mech Eng H. 2006;220:311–319.

    Article  PubMed  Google Scholar 

  7. Komistek RD, Dennis DA, Ochoa JA, Haas BD, Hammill C. In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty. J Bone Joint Surg Am. 2002;84:1836–1841.

    PubMed  Google Scholar 

  8. Markel D, Day J, Siskey R, Liepins I, Kurtz S, Ong K. Deformation of metal-backed acetabular components and impact of liner thickness in a cadaveric model. Int Orthop. 2001;35:1131–1137.

    Article  Google Scholar 

  9. McKellop H, Park SH, Chiesa R, Doorn P, Lu B, Normand P, Grigoris P, Amstutz H. In vivo wear of three types of metal on metal hip prostheses during two decades of use. Clin Orthop Relat Res. 1996;329(suppl):128–140.

    Article  Google Scholar 

  10. Meding JB, Meding LK, Keating EM, Berend ME. Low incidence of groin pain and early failure with large metal articulation total hip arthroplasty. Clin Orthop Relat Res. 2012;470(suppl):388–394.

    Article  PubMed  Google Scholar 

  11. Ong KL, Rundell S, Liepins I, Laurent R, Markel D, Kurtz SM. Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press-fit implantation. J Orthop Res. 2009;27:1467–1472.

    Article  PubMed  Google Scholar 

  12. Peters CL, McPherson E, Jackson JD, Erickson JA. Reduction in early dislocation rate with large-diameter femoral heads in primary total hip arthroplasty. J Arthroplasty. 2007;22(6 suppl 2):140–144.

    Article  PubMed  Google Scholar 

  13. Rieker CB, Schon R, Konrad R, Liebentritt G, Gnepf P, Shen M, Roberts P, Grigoris P. Influence of the clearance on in-vitro tribology of large diameter metal-on-metal articulations pertaining to resurfacing hip implants. Orthop Clin North Am. 2005;36:135–142.

    Article  PubMed  Google Scholar 

  14. Schmidig G, Patel A, Liepins I, Thakore M, Markel DC. The effects of acetabular shell deformation and liner thickness on frictional torque in ultrahigh-molecular-weight polyethylene acetabular bearings. J Arthroplasty. 2010;25:644–653.

    Article  PubMed  Google Scholar 

  15. Smith TM, Berend KR, Lombardi AV Jr, Emerson RH Jr, Mallory TH. Metal-on-metal total hip arthroplasty with large heads may prevent early dislocation. Clin Orthop Relat Res. 2005;441:137–142.

    Article  PubMed  Google Scholar 

  16. Springer BD, Habet NA, Griffin WL, Nanson CJ, Davies MA. Deformation of 1-piece metal acetabular components. J Arthroplasty. 2012;27:48–54.

    Article  PubMed  Google Scholar 

  17. Squire M, Griffin WL, Mason JB, Peindl RD, Odum S. Acetabular component deformation with press-fit fixation. J Arthroplasty. 2006;21:72–77.

    Article  PubMed  Google Scholar 

  18. Udofia IJ, Jin ZM. Elastohydrodynamic lubrication analysis of metal-on-metal hip-resurfacing prostheses. J Biomech. 2003;36:537–544.

    Article  PubMed  CAS  Google Scholar 

Download references

Acknowledgments

The authors thank Michael Volitich BS, Ron Hofmann BS, and Gary Burgess BS, for their assistance in experimental setup. We additionally thank Kenneth Davis MS, and Matthew Brunsman MS, for their assistance with statistical analysis.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to John B. Meding MD.

Additional information

The institution of one or more of the authors (JBM, SRS, MEB, MAR) has received funding, during the study period, from St Francis Hospital (Mooresville, IN, USA), ERMI, Inc (Atlanta, GA, USA), DePuy Orthopedics, Inc (Warsaw, IN, USA), and Stryker Orthopaedics (Mahwah, NJ, USA). A partnering institution (Rose-Hulman) received funding from the National Science Foundation for the research instrumentation utilized in this study through Major Research Instrumentation Awards 0923135 and 1039716. One of the authors (JBM) certifies that he, or a member of his immediate family, has received or may receive payments or benefits, during the study period, an amount in excess of $100,000, from Biomet, Inc (Warsaw, IN, USA); one of the authors (MEB) certifies that he, or a member of his immediate family, has received or may receive payments or benefits, during the study period, an amount in excess of $100,000, from Biomet, Inc, and OrthAlign, Inc (Aliso Viejo, CA, USA).

All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research editors and board members are on file with the publication and can be viewed on request.

Clinical Orthopaedics and Related Research neither advocates nor endorses the use of any treatment, drug, or device. Readers are encouraged to always seek additional information, including FDA approval status, of any drug or device before clinical use.

Specimen preparation and data collection were conducted at Rose-Hulman Institute of Technology (Terre Haute, IN, USA). Statistical analysis was performed at Joint Replacement Surgeons of Indiana Foundation, Inc (Mooresville, IN, USA).