Abstract
New MR-techniques for quantitative blood flow registration such as phase-mapping (a two-dimensional space-resolved technique with a time-averaged measurement of blood flow) or RACE (real-time acquisition and evaluation of blood flow in one-dimensional space projection) are available for the diagnosis of valvular heart disease. Initial results of grading aortic valve stenosis by these methods are shown in comparison to continuous wave Ultrasound-Doppler. Two groups of 15 patients were examined by RACE or phase-mapping, 12 respectively 8 of whom suffered from an aortic valve stenosis. The shape of blood flow profiles as well as grading of aortic valve stenosis show high concordance when comparing the results of MR and Doppler technique. Good reliability and practicability of the demonstrated MR-method are shown. With respect to the results of RACE and phase-mapping the development of an alternative and competing MR-method for the evaluation of valvular heart disease and shunt diagnostics seems possible.
Similar content being viewed by others
References
Singer JR (1959) Blood flow rates by NMR measurements. Science 130: 1652
Deimling M, Müller E, Lenz G, Barth K, Fritschy P, Seiderer M (1986) Description of flow phenomena in MRI using a gradient pulse and phase difference technique. Diagn Imaging Clin Med 55: 37–51
Müller E, Deimling M, Reinhardt E (1986) Quantification of pulsatile blood flow in MRI. Magn Reson Med 3: 331–335
Crooks LE, Sheldon PE, Kaufmann L, Rowan W (1982) Quantifications of obstruction in vessels by nuclear magnetic resonance (NMR). IEEE Trans Nucl Sci 29: 1181–1185
Kaufmann C, Crooks LE, Sheldon PE, Rowan W, Miller T (1982) Evaluation of NMR Imaging for quantification of obstructions in vessels. Invest Radiol 17: 554–560
Axel A, Shimakowa A, Macfall IA (1986) A time-of-flight method of measuring flow velocity. Magn Reson Imaging 4: 199–205
Bradley WG, Waluch V (1985) Magnetic resonance imaging. Radiology 154: 443–450
Müller E (1985) Flow measurements. Soc Magn Reson Med, London, V1, Book fo Abstracts
Bryant DJ, Payne JA, Firmin DN, Longmore DB (1984) Measurement of flow with NMR. J Comput Assist Tomogr 8: 588–593
Firmin DN, Nayler GL, Klipstein RH, Underwood SR, Rees RSO, Longmore DB (1987) In vivo validation of MR velocity imaging. J Comput Assist Tomogr 11: 751–756
Underwood SR, Firmin DN, Rees RSO, Longmore DB (1990) Magnetic resonance velocity mapping. Clin Phys Physiol Meas 11: 37–43
Van Rossum A, Sprenger KH, Peels FC (1989) In vivo validation of quantiative flow imaging in arteries and veins using magnetic resonance phase shift techniques. Soc Magn Reson Med, Amsterdam, Book of Abstract: 205
Engels G, Wilke N, Müller E, Bachmann K (1990) Grading of mitral valve stenosis by intracardial real time registration of blood flow. Magn Reson Imaging 8, Suppl. 1: 443
Müller E, Laub G, Graumann R, Loeffler W (1988)_RACE-Real time acquistion and evaluation of pulsatile blood flow on a whole body MR unit. Soc Magn Reson Med. 7th Ann. Meeting, San Francisco, USA, Book of Abstracts: 729
Sechtem U, Sünger B, Kux R, Theissen P, Curtius JM, Hoepp HW, Schicha H (1988) Nichtinvasive Beurteilung von Aortenund Mitralklappeninsuffizienzen mit dynamischer Magnetresonanztomographie. Z Kardiol 77: 145–151
Sechtem U, Pflugfelder PW, Cassidy MM, White RD, Cheitlin MD, Schiller NB, Higgins CB (1988) Mitral or aortic regurgitation: quantification of regurgitant volumes with cine MR imaging. Radiology 167: 425–430
Author information
Authors and Affiliations
Additional information
Correspondence to: G. Engels
Rights and permissions
About this article
Cite this article
Engels, G., Müller, E., Reynen, K. et al. Phase-mapping technique for the evaluation of aortic valve stenosis by MR. Eur. Radiol. 2, 299–304 (1992). https://doi.org/10.1007/BF00175430
Issue Date:
DOI: https://doi.org/10.1007/BF00175430