Skip to main content

Part of the book series: IFMBE Proceedings ((IFMBE,volume 31))

  • 186 Accesses

Abstract

Three bi-leaflet prostheses, the St. Jude Medical (SJM) and the On-X valves with straight leaflets and the MIRA valves with curved leaflets were tested in the aortic position under pulsatile-flow condition. Dynamic PIV system, which was employed to analyze the flow field affected by the valve-installed orientations, leaflet shapes, and valve designs, and direct measurement of the coronary flow using ultrasonic flow meter were utilized to correlate the results of aortic flow field to coronary flow. Detailed chronological comparison of the flow fields downstream of the aortic valve and inside the sinus of valsalva associated with the opening, accelerating, peak, and closing phase of the flow revealed the significance of the leaflet and overall valve designs on the flow and possible influence on the coronary flow.

Based on the experimental results, following general conclusions can be made. The SJM and the newer On-X valve generate stronger circulatory flow inside the sinus of valsalva when valve orientation is set at 180°( while the MIRA valve with curved leaflet generates stronger circulatory flow inside the sinus of valsalva when valve orientation is set at 90°). Direct measurement of coronary flow clearly shows the increase of flow rate when the circulatory flow inside the sinus of valsalva increase indicating the positive correlation exist between the circulatory flow inside the sinus of valsalva and the coronary flow.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 259.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 329.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2010 International Federation for Medical and Biological Engineering

About this paper

Cite this paper

Akutsu, T., Matsumoto, A., Takahashi, K. (2010). Correlation between Aortic Flow and Coronary Circulation. In: Lim, C.T., Goh, J.C.H. (eds) 6th World Congress of Biomechanics (WCB 2010). August 1-6, 2010 Singapore. IFMBE Proceedings, vol 31. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-14515-5_113

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-14515-5_113

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-14514-8

  • Online ISBN: 978-3-642-14515-5

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics