Skip to main content

An Avoidable Methodological Failure in Intracranial Pressure Monitoring Using Fiberoptic or Solid State Devices

  • Conference paper
Intracranial Pressure and Neuromonitoring in Brain Injury

Part of the book series: Acta Neurochirurgica Supplements ((NEUROCHIRURGICA,volume 71))

  • 600 Accesses

Summary

Failure of intraventricular pressure (IVP) measurement in case of catheter blockage is believed to be eliminated by using intraventricular microtransducers. We report about an avoidable methodological error, which may affect the reliability of IVP measurement with these devices. Intraventricular fiberoptic or solid state devices were implanted in 43 patients considered to be at risk for catheter occlusion. Two different types were used: devices where the transducer is placed inside the ventriculostomy catheter (Type A), and devices where the transducer is integrated in the external surface of the catheter (Type B).

Of the 15 patients treated with Type A devices, no reliable pressure recording could be obtained in three patients where ventricular puncture was not successful. In four cases of the remaining 12 patients, periods of erroneous pressure readings were revealed. After opening of CSF drainage, all Type A devices failed to reflect real IVP. In patients treated with Type B devices, no erroneous pressure recording could be identified, irrespective if CSF drainage was performed or not.

Transducers, which are simply placed inside the ventriculostomy catheter require fluid coupling. They may fail, either during CSF drainage or when the catheter is blocked or placed within the parenchyma.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 259.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 329.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info
Hardcover Book
USD 329.99
Price excludes VAT (USA)
  • Durable hardcover edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

References

  1. Feldman Z, Narayan RK (1993) Intracranial pressure monitoring: techniques and pitfalls. In: Cooper PR (ed) Head injury. Williams and Wilkins, Baltimore, pp 247–274

    Google Scholar 

  2. Luerssen TG, Shields PF, Vos HR, Marshall LF (1989) Clinical experience with fiberoptic brain parenchymal pressure monitor. In: Hoff JT, Betz AL (eds) Intracranial pressure VII. Springer, Berlin Heidelberg New York Tokyo, pp 35–37

    Chapter  Google Scholar 

  3. Ostrup RC, Luerssen TG, Marshall LF, Zornow MH (1987) Continuous monitoring of intracranial pressure with a miniaturized fiberoptic device. J Neurosurg 67: 206–209

    Article  PubMed  CAS  Google Scholar 

  4. Piek J, Raes P (1996) Pressure-controlled drainage of cerebrospinal-fluid — clinical-experience with a new-type of ventricular catheter (ventcontrol mtc) and an integrated piezoresistive sensor at its tip-technical note. Neurosurgery 38: 216–218

    Article  PubMed  CAS  Google Scholar 

  5. Sundbarg G, Nordstrom CH, Messeter K, Soderstrom S (1987) A comparison of intraparenchymatous and intraventricular pressure recording in clinical practice. J Neurosurg 67: 841–845

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1998 Springer-Verlag Wien

About this paper

Cite this paper

Raabe, A., Stöckel, R., Hohrein, D., Schöche, J. (1998). An Avoidable Methodological Failure in Intracranial Pressure Monitoring Using Fiberoptic or Solid State Devices. In: Marmarou, A., et al. Intracranial Pressure and Neuromonitoring in Brain Injury. Acta Neurochirurgica Supplements, vol 71. Springer, Vienna. https://doi.org/10.1007/978-3-7091-6475-4_19

Download citation

  • DOI: https://doi.org/10.1007/978-3-7091-6475-4_19

  • Publisher Name: Springer, Vienna

  • Print ISBN: 978-3-7091-7331-2

  • Online ISBN: 978-3-7091-6475-4

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics