Abstract
Background
The objective was to evaluate the use of fruit juice with an interactive inversion recovery (IR) MR pulse sequence to visualise the gastrointestinal tract.
Methods
We investigated the relaxation properties of 12 different natural fruit juices in vitro, to identify which could be used as oral contrast. We then describe our initial experience using an interactive MR pulse sequence to allow optimal visualisation after administering pineapple juice orally, and suppressing pre-existing bowel fluid contents, with variable TI in three adult and one child volunteer.
Results
Pineapple juice (PJ) had both the shortest T1 (243 ms) and shortest T2 (48 ms) of the fruit juices tested. Optimal signal differentiation between pre-existing bowel contents and oral PJ administration was obtained with TIs of between 900 and 1100 ms.
Conclusion
The use of an inversion recovery preparation allowed long T1 pre-existing bowel contents to be suppressed whilst the short T1 of fruit juice acts as a positive contrast medium. Pineapple juice could be used as oral contrast agent for neonatal gastrointestinal magnetic resonance imaging.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Background
Congenital abnormalities of gut rotation in neonates and infants are usually investigated by the upper gastrointestinal contrast study (UGI), using traditional X-ray fluoroscopy techniques. Malrotation is typically excluded on the basis of confidently identifying a normally positioned duodeno-jejunal (DJ) flexure, usually on the first pass of radio-opaque oral contrast medium through the duodenum at UGI [1]. However, these procedures require skilled interpretation, carry a significant radiation burden to the child, and may not always be diagnostic [2]. Optimal imaging of the neonatal GI tract is currently under review, with ongoing debate in the literature as to whether ultrasound or cross sectional imaging (such as CT) could replace the UGI series [3, 4], although both have limitations, including ionising radiation doses in CT higher than that of UGI, and a significant false negative rate just as with UGI.
Recent advances in magnetic resonance imaging (MRI), particularly the development of interactive multi-contrast pulse sequences, may now provide an alternative imaging technique, thus avoiding ionising radiation. Interactive MRI allows real-time control over image plane location as well as a number of image acquisition parameters, including the development of an IR based Single Shot Fast Spin Echo (IR-SSFSE) sequence [5, 6]. This could be useful in the setting of possible gut malrotation, just as interactive MRI has been used to visualise vesicoureteric reflux in the neonatal urinary tract [7]. However, in order to closely replicate the UGI series using MRI, an easily visualised contrast medium is needed.
The ideal medium should be easily tolerated, provide high positive signal enhancement, and help to distinguish the DJ flexure from pre-existing bowel contents in adjacent small bowel (likely to be a combination of milk or water-based material in neonates). There are several practical limitations in neonatal gut MRI, particularly regarding the choice of oral contrast agent. This is largely determined by their signal properties, the imaging modality used, the degree of gut distension required, and therefore the minimum volume of contrast agent required. Traditionally, barium sulphate is used in conventional radiographic gastrointestinal studies, but where there is a risk of spillage into the respiratory tract or perforation (particularly in paediatric studies) it has been replaced by non-ionic, low osmolality compounds (e.g. Iopamidol; Gastromiro, Bracco, High Wycombe, UK). High density barium has been used as a negative oral contrast agent for MR imaging of the gut [8], but typically requires high volumes of contrast medium to be ingested, which is not practical for neonates and small children.
Administering water (high contrast on T2w images) may not allow confident identification of the DJ flexure separately from the pre-existing fluid in adjacent jejunal loops, a key diagnostic factor for the UGI series. Currently, there are no gadolinium-based contrast agents licensed for oral use in children, but several commercially available (i.e. commodity) fruit juices have been used as both positive and negative oral contrast agents. For example, the paramagnetic properties of fruit juices such as concentrated pineapple or blueberry juice have been used to suppress the signal from bowel fluid in paediatric magnetic resonance cholangio-pancreatography (MRCP) studies which employ heavily T2 -weighted imaging [9, 10]. As well as their T2 effects, the T1 shortening properties of these juices may also make them suitable positive contrast agents. However, they cannot be administered in high concentrations to neonates, due to the inability of the immature gut to deal with complex carbohydrate loads (of greater than 10%; [11]).
Administering an oral contrast medium with unknown positive signal properties, to an infant with unknown signal properties of small bowel content provides a wide range of variables. In this study, we first investigated the relaxation properties of several diluted fruit juices in vitro, to identify which could be used as oral contrast. We then describe our initial experience using an interactive MR pulse sequence, whereby real-time control over imaging parameters such as IR pulse timing could allow optimal contrast medium visualisation by suppressing pre-existing bowel fluid contents.
Methods
We optimised the interactive IR-SSFSE sequence as part of an ethically approved study into optimising gut imaging (Cambridgeshire 3 Research Ethics Committee formal approval 08/H0306/7). Three adult volunteers and one child participated, who were nil-by-mouth for 6 hours and 3 hours respectively prior to the MR examination. Full informed written consent was obtained from each participant and the child’s parent. No sedation or anaesthesia was used. Ethical approval was not required for in-vitro studies.
Contrast agent testing
12 different contrast media were prepared, as follows: Tap water, standard infant formula milk (SMA Gold, SMA nutrition, Berkshire, UK), Iopamidol (Gastromiro, Bracco, High Wycombe, UK), and the following 9 commercially available (standard supermarket stock) fruit juices: orange juice, two examples of pineapple juice (one from juice; PJ, and one reconstituted from concentrate, PJ2), apple & beetroot juice, prune juice, blackcurrant juice, blueberry juice, raspberry juice, and blackberry juice. Three of these products (PJ2, prune, and blueberry juice) required further dilution to achieve a carbohydrate concentration of <10%, all other media were imaged undiluted (see Table 1). Estimated nutritional and paramagnetic agent contents of each juice are given (Table 2).Each proposed contrast agent was placed within a multi-compartment phantom positioned inside an 8 channel brain array coil, and images acquired on a standard 1.5 T whole-body MR system (Signa HDx; GE Healthcare, Waukesha, WI; Figure 1).
The T1 values were measured using a 2D inversion recovery (IR) prepared fast radiofrequency (RF) spoiled gradient echo (FSPGR) sequence with variable inversion times (TI) at nine different TIs (40, 80, 160, 320, 640 ms, and 1.28, 2.56, 5.12, 10.24 sec). Other imaging parameters were: field of view (FOV) 25×25 cm, slice thickness 5 mm, matrix 256 × 256, TR 7.2, TE 3.3 ms.
The T2 values were measured using an 8-echo spin echo sequence. Eight different echo times (TE) were used, 30, 60, 90, 120, 150, 180, 210, 240 ms. Other imaging parameters were: FOV 25×25 cm, slice thickness 5 mm, matrix 256 × 256, TR 2000 ms.
We then measured the signal intensity of each contrast medium using the IR-SSFSE sequences at variable TI times (0–4000 ms), to simulate the in-vivo environment, and subtracted the signal intensity of milk to predict the possible optimal TI times in vivo.
Interactive IR-SSFSE
Interactive MRI in-vivo was performed using an IR-SSFSE pulse sequence. This modified pulse sequence has been custom developed to operate within a proprietary interactive imaging interface (i/Drive Pro Plus; GE Healthcare) which allows interactive control over many imaging parameters such as image plane location, phase encode ordering, and the use of a 180° inversion pulse that can be interactively toggled on and off prior to the SSFSE readout. The inversion time (TI) between the 180° inversion pulse and the SSFSE readout could also be interactively controlled during imaging, by inputting different IR times in ms. Initial imaging parameters for the SSFSE sequence were: FOV 25×25 cm, matrix 256×256, TR 2500 ms, TE 50.1 ms, single slice thickness of 10 mm. No motion suppression or breathing suppression technique was used.In volunteer imaging, initial localiser images and non-interactive axial and coronal FIESTA and SSFSE sequences were obtained in a supine position to outline upper gastrointestinal anatomy and localise pre-existing small bowel contents. Coronal interactive IR-SSFSE was then used during free breathing to vary the TI at 100 ms intervals (between 0 and 2000 ms), and signal intensities were measured in 3 regions of interest (ROIs): stomach, proximal small bowel (duodenum) and mid-small bowel (jejunum; Figure 2), in order to characterise the signal properties of pre-existing bowel contents.
We then repeated the study with increasing TIs within 10 minutes of the oral administration of 200 ml pineapple juice to three adult volunteers, and after 20 ml pineapple juice administration to a single neonatal infant (2 week old 3 kg male infant). We did not use any form of respiratory or cardiac gating.
Results
In vitro testing
Figure 1 demonstrates the phantom used for in-vitro testing. The T1 and T2 relaxation times for the contrast media tested are reported in Table 1: pineapple juice from juice (PJ) was found to have both the shortest T1 (243 ms) and T2 (48 ms) relaxation times.Assuming that milk was likely to be the main constituent of pre-existing bowel contents in pre-weaned infants, a representative graph of the relative difference in signal intensity between PJ and formula milk is shown in Figure 3, demonstrating that maximal differences are seen in-vitro with TIs between 500 – 900 ms.
In vivo strategy
We subsequently tested pineapple juice as an in vivo contrast agent using the IR-SSFSE sequence: we measured signal intensity in three different ROIs (Figures 2), with increasing T1 at 100 ms intervals (0 – 2000 ms; Figure 4) before and after oral PJ administration. By subtracting the signal intensity of distal (jejunal) small bowel contents, we plotted differences in signal intensity (Figure 5) and showed that maximum differences between stomach and jejunum contents were observed at TIs of 1000 ms, 1100 ms, and 900 ms in the three adult volunteers respectively (Figure 6). We were unable to obtain good images of every TI in the neonate following 20 ml PJ administration due to patient motion, but maximal signal differences were seen at 1000 ms (Figure 7).
Discussion
This study demonstrates that pineapple juice could be used as a positive contrast agent for the proximal neonatal gut. By tailoring an IR-SSFSE sequence to null the signal from existing bowel fluid and thereby optimise the signal from PJ, we have developed a strategy for bowel visualisation using MRI which could be used in a similar fashion to the traditional UGI series in children. The true diagnostic performance of an interactive MRI strategy under these conditions remains to be evaluated.
IR-SSFSE sequences have previously been used for MR imaging of the abdomen [12], but not with interactive control. Our strategy used the IR-SSFSE sequence to effectively suppress bowel fluid signal where its properties were unknown, and to optimally tune the IR timing to provide optimal signal from a contrast agent with known signal properties. Several vendors may offer an IR-prepared sequence, such as T1-scout series or modified look-locker inversion recovery (MOLLI) sequences, where changes in inversion recovery time can be used to null myocardium during delayed enhanced imaging, for instance. However they do not allow for on-the-fly adjustment of the TI, which is necessary for rapid optimisation in a paediatric setting. Our strategy helps to tailor the examination to the individual, and has the potential to reduce overall examination time, although this remains to be evaluated in clinical practice.
One of the advantages of using the T1 shortening properties of fruit juices are that smaller volume can be used as gut distension is not a requirement. Here, we were able to administer 200 ml to an adult, less than some studies using blueberry juice in adults (400–600 ml) [9, 13, 14]. Using only 20 ml in a neonate is comparable to that used in an UGI and is less than normal meal intakes. The majority of other substances used for MR gut imaging are typically combined with gut-distending agents such as mannitol, known to give gastrointestinal side effects such as flatulence and bloating, but still widely used for small bowel MR studies in older children and adults.
Despite differences in commercial products and dilutions, our measurements closely mirror those found in other studies, particularly regarding the short T1 and T2 of PJ [15]. We included formula milk in our evaluation and obtained similar values to the range previously measured (640–1750 ms) [16]. Several other fruit juices have been proposed as good gut contrast agents, but have only been evaluated in adults, including concentrated blueberry juice and blackberry juice [13, 17, 18].
The main difference between our phantom study and some previous experiments is likely to relate to the dilutions used; several previous studies measured the signal characteristics of concentrated fruit juices, or indeed fruit pulp itself [17]. Concentrated fruit juices can stimulate gut transit in non-weaned infants, causing diarrhoea and nappy irritation, which can be both uncomfortable and distressing for the child and care-giver, which is why we used diluted products. It is widely reported that the signal suppression effect is due to T2 relaxation caused by their inherently high iron (blackberry) or manganese (pineapple, beet juice, green tea) content, although the exact contents of different commercial fruit juices vary widely, and the precise paramagnetic component (iron, copper or manganese) that gives the relaxation shortening property is debatable. We estimated the paramagnetic agent contents of each of our fruit juices (Table 2) and hypothesise that it may be the manganese content of pineapple juice which gave favourable properties in this study.
Several other oral agents have been developed and have been comprehensively reviewed elsewhere [19]. These can be grouped into predominantly positive gastrointestinal MR imaging contrast media, typically gadolinium-based compounds e.g. Magnevist Enteral® (Schering AG, Berlin, Germany) or Gadolinium zeolite (Gadolite®; Pharmacyclics, Sunnyvale, CA, now GlaxoSmithKline Ltd, Brentford, UK), or negative gastrointestinal MR imaging contrast agents, particularly iron-containing solutions such as Perflubron® (Imagent GI®, Alliance Pharma), or Abdoscan® (Nycomed-Amersham, Oslo, Norway). Most of these are not commercially available, and are not licensed for children. Carbon dioxide gas has also been used as a negative contrast agent for proximal gut MR imaging and early studies demonstrated acceptable visualisation of the duodenum but were limited in the jejunum and small bowel [20], but this remains to be investigated in neonates.
Limitations of this study
One limitation of this feasibility study could be the variable content of commercially available fruit juice. We did not determine the exact constituents of the different fruit juices used in this study, but relied on the manufacturers details provided on the packaging and estimated provided by the USDA National Nutrient Database (http://ndb.nal.usda.gov). Since the concentrations of fruit pulp and therefore paramagnetic properties of diluted commercial fruit juices may vary, the ability to interactively optimise the IR timing becomes even more important as a way to compensate for these variations.
We acknowledge that we did not include any patients in this volunteer feasibility study, and have tested only a limited number of volunteers. The full diagnostic potential of this technique now needs to be evaluated in a larger patient population.
Conclusion
In conclusion, this work indicates that pineapple juice is the most promising natural commercially available short T1 contrast medium suitable for neonatal GI tract imaging using an interactively adjusted inversion-recovery SSFSE strategy. This allows for pre-existing longer T1, bowel contents to be suppressed, whilst the short T1 of the PJ acts as a positive contrast medium to delineate the gut. This approach allows individual tailoring of the examination, to give optimal discrimination of an orally administered contrast agent from the background signal of gut contents. Further work is required using this strategy in a larger cohort of neonates and infants to evaluate both technical and diagnostic performance.
References
Long FR, Kramer SS, Markowitz RI, Taylor GE: Radiographic patterns of intestinal malrotation in children. Radiographics. 1996, 16: 547-556. 10.1148/radiographics.16.3.8897623.
Sizemore AW, Rabbani KZ, Ladd A, Applegate KE: Diagnostic performance of the upper gastrointestinal series in the evaluation of children with clinically suspected malrotation. Pediatr Radiol. 2008, 38: 518-528.
Daneman A: Malrotation: the balance of evidence. Pediatr Radiol. 2009, 39: S164-S166.
Taylor GA: CT appearance of the duodenum and mesenteric vessels in children with normal and abnormal gut rotation. Pediatr Radiol. 2011, 41: 1378-1383. 10.1007/s00247-011-2118-z.
Graves MJ, Wong P, Black RT, Lomas DJ: MR fluoroscopy employing interactive pulse sequence switching. Proc ISMRM. 2007, 15: 1654-
Makki M, Graves MJ, Lomas DJ: Interactive body MR fluoroscopy using SSFSE RARE with multiparameter control. J MRI. 2002, 16: 85-93.
Arthurs OJ, Edwards AD, Joubert I, Graves MJ, Set PAK, Lomas DJ: Interactive magnetic resonance voiding cystourethrography (iMRVC) for vesicoureteric reflux (VUR) in unsedated infants: a feasibility study. Eur Radiol. 2011, 21: 1874-1881. 10.1007/s00330-011-2124-4.
Li KCP, Tart RP, Fitzsimmons JR, Storm BL, Mao J, Rolfes RJ: Barium sulfate suspension as a negative oral MRI contrast agent: In vitro and human optimization studies. MRI. 1991, 9: 141-150. 10.1016/0730-725X(91)90002-4.
Riordan RD, Khonsari M, Jeffries J, Maskell GF, Cook PG: Pineapple juice as a negative oral contrast agent in magnetic resonance cholangiopancreatography: a preliminary evaluation. Br J Radiol. 2004, 77: 991-999. 10.1259/bjr/36674326.
Darge K, Anupindi S: Pancreatitis and the role of US, MRCP and ERCP. Pediatr Radiol. 2009, 39: S153-7. 10.1007/s00247-009-1145-5.
Baker SS on behalf of the Committee on Nutrition: The use and misuse of fruit juice in paediatrics. Paediatrics. 2001, 107: 1210-1213.
Lauenstein TC, Sharma P, Hughes T, Heberlein K, Tudorascu D, Martin DR: Evaluation of optimized inversion-recovery fat-suppression techniques for T2-weighted abdominal MR imaging. J Magn Reson Imaging. 2008, 27: 1448-54. 10.1002/jmri.21350.
Hiraishi K, Narabayashi I, Fujita O, Yamamoto K, Sagami A, Hisada Y, Saika Y, Adachi I, Hasegawa H: Blueberry juice: preliminary evaluation as an oral contrast agent in gastrointestinal MR imaging. Radiology. 1995, 194: 119-23. 10.1148/radiology.194.1.7997537.
Papanikolaou N, Karantanas A, Maris T, Gourtsoyannis N: MR cholangiopancreatography before and after oral blueberry juice administration. J Comput Assist Tomogr. 2000, 24: 229-234. 10.1097/00004728-200003000-00008.
Asbach P, Breitwieser C, Diederichs G, Eisele S, Kivelitz D, Taupitz M, Zeitz M, Hamm B, Klessen C: Cine magnetic resonance imaging of the small bowel: comparison of different oral contrast media. Acta Radiol. 2006, 47: 899-906. 10.1080/02841850600965054.
Balzarini L, Aimeb S, Barberob L, Ceglia E, Petrillo R: Magnetic resonance imaging of the gastrointestinal tract: investigation of baby milk as a low cost contrast medium. Eur Radiol. 1992, 15: 17l-174.
Karantanas AH, Papanikolaou N, Kalef-Ezra J, Challa A, Gourtsoyiannis N: Blueberry juice used per os in upper abdominal MR imaging: composition and initial clinical data. Eur Radiol. 2000, 10: 909-13. 10.1007/s003300051035.
Espinosa MG, Sosa M, De León-Rodríguez LM, Córdova T, Bernal-Alvarado J, Avila-Rodríguez M, Reyes-Aguilera JA, Ortíz JJ, Barrios FA: Blackberry (Rubus spp.): a pH-dependent oral contrast medium for gastrointestinal tract images by magnetic resonance imaging. Magn Reson Imaging. 2006, 24: 195-200. 10.1016/j.mri.2005.10.026.
Giovagnoni A, Fabbri A, Maccioni F: Oral contrast agents in MRI of the gastrointestinal tract. Abdom Imaging. 2002, 27: 367-375. 10.1007/s00261-001-0117-5.
Lomas DJ, Habib SH, Joubert IJ, Sala E, Graves MJ: CO2 as a distending medium for gastric and small bowel MRI: a feasibility study. Eur Radiol. 2005, 15: 672-6. 10.1007/s00330-004-2580-1.
Pre-publication history
The pre-publication history for this paper can be accessed here:http://www.biomedcentral.com/1471-2342/14/33/prepub
Acknowledgements
OJA was funded by a Medical Research Council/Royal College of Radiologists Clinical Research Training Fellowship. The views expressed in this publication are those of the author(s). The project was supported by the Addenbrookes Charitable Trust, and the NIHR comprehensive Biomedical Research Centre award to Cambridge University Hospitals NHS Foundation Trust in partnership with the University of Cambridge. None of the funding bodies had any role in analysis of data, results or conclusions of the study.
Author information
Authors and Affiliations
Corresponding author
Additional information
Competing interests
The authors declare that they have no competing interests, financial or non-financial
Authors’ contributions
OA, MJG, IJ and ADE carried out the in-vitro studies, and OA, MJG, IJ and PS carried out the in-vivo interactive MRI studies. All authors participated in the design of the study and helped to draft the manuscript. All authors read and approved the final manuscript.
Authors’ original submitted files for images
Below are the links to the authors’ original submitted files for images.
Rights and permissions
This article is published under an open access license. Please check the 'Copyright Information' section either on this page or in the PDF for details of this license and what re-use is permitted. If your intended use exceeds what is permitted by the license or if you are unable to locate the licence and re-use information, please contact the Rights and Permissions team.
About this article
Cite this article
Arthurs, O.J., Graves, M.J., Edwards, A.D. et al. Interactive neonatal gastrointestinal magnetic resonance imaging using fruit juice as an oral contrast media. BMC Med Imaging 14, 33 (2014). https://doi.org/10.1186/1471-2342-14-33
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/1471-2342-14-33