Abstract
The sensitivity σ of an optically pumped quantum magnetometer is measured directly by comparing the output frequencies of 85Rb and 87Rb magnetometers integrated in a single two-isotope cell. The result σ87=59.6±6 fT Hz−1/2 agrees well with the value σ light87 =53.7±2 fT Hz−1/2 obtained indirectly by measuring 87 the ratio of the resonance steepness to the shot noise of light.
Similar content being viewed by others
References
E. B. Aleksandrov, M. V. Balabas, A. É. Ivanov, et al., Opt. Spektrosk. 78, 325 (1995) [Opt. Spectrosc. 78, 292 (1995)].
E. B. Alexandrov, M. V. Balabas, A. S. Pasgalev, et al., Laser Phys. 6, 244 (1996).
D. Budker, D. F. Kimball, S. M. Rochester, et al., Phys. Rev. A 62, 043403 (2000).
D. Budker, V. Yashchuk, and M. Zolotorev, Phys. Rev. Lett. 81, 5788 (1998).
I. K. Kominis, T. W. Kornack, J. C. Allred, and M. V. Romalis, Nature 422, 596 (2003).
E. Alexandrov, M. Balabas, G. Ban, et al., PSI Sci. Rep. (Paul Sherrer Institute, Villigen, 2002), Vol. 1, pp. 25–26.
C. Cohen-Tannoudji, J. DuPont-Roc, S. Haroche, and F. Laloe, Phys. Rev. Lett. 22, 758 (1969).
A. A. Radtsig and B. M. Smirnov, Reference Data on Atoms, Molecules, and Ions (Énergoatomizdat, Moscow, 1986; Springer-Verlag, Berlin, 1985).
P. J. Mohr and B. N. Taylor, Rev. Mod. Phys. 72, 351 (2000).
Author information
Authors and Affiliations
Additional information
__________
Translated from Zhurnal Tekhnichesko\(\overset{\lower0.5em\hbox{$\smash{\scriptscriptstyle\smile}$}}{l} \) Fiziki, Vol. 74, No. 6, 2004, pp. 118–122.
Original Russian Text Copyright © 2004 by Alexandrov, Balabas, Vershovski, Pazgalev.
Rights and permissions
About this article
Cite this article
Alexandrov, E.B., Balabas, M.V., Vershovski, A.K. et al. Experimental demonstration of the sensitivity of an optically pumped quantum magnetometer. Tech. Phys. 49, 779–783 (2004). https://doi.org/10.1134/1.1767891
Received:
Issue Date:
DOI: https://doi.org/10.1134/1.1767891