Abstract
Objective.Evaluation in animals of a non-invasive and continuous cardiac output monitoring system based on partial carbon-dioxide (CO2) rebreathing indirect Fick technique. Methods.We have developed a non-invasive cardiac output (NICO) monitoring system, based on the partial rebreathing method. The partial rebreathing technique employs a differential form of the Fick equation for calculating cardiac output (QT) using non-invasive measurements. Changes in CO2 elimination (ΔVCO2) and partial pressure of end-tidal CO2 (Δ PETCO2) in response to a brief period of partial rebreathing are used to measure pulmonary capillary blood flow (QPCBF). A non-invasive estimate of anatomic and intrapulmonary shunt fraction (QS/QT), based on oxygen saturation from pulse oximetry (SpO2) and inspired oxygen concentration (FIO2), is added to compute total cardiac output [QT=QPCBF/(1−QS/QT)]. The performance of the NICO was compared with iced 5% dextrose bolus thermodilution cardiac output (TDco) measurements in 6 dogs. Cardiac output was varied using dobutamine, and halothane, and by clamping of the inferior vena cava. Two hundred and forty-six (n = 246) paired measurements of TDco and NICO over a range of cardiac outputs (TDco range = 0.60–8.87 l/min) were compared using Bland-Altman analysis and weighted correlation coefficient. Results.The Bland–Altman technique yielded a NICO precision of ± 0.70 l/min (13.8%) with a mean bias of −0.07 l/min (−1.4%) compared to TDco. The weighted correlation coefficient between TDco and NICO values was: r= 0.93 (n= 246). Conclusion.The partial CO2 rebreathing technique for measurement of cardiac output is non-invasive, automated, and based on the well accepted Fick principle. The limits of agreement between NICO and TDco is within the recommended value for NICO to be a clinically acceptable method for cardiac output measurement. The results of this canine study show that NICO performed as well, and in some cases better, than other currently available non-invasive cardiac output techniques over a wide range of cardiac outputs.
Similar content being viewed by others
REFERENCES
Ganz W, Swan HJ. Measurement of blood £ow by thermo-dilution. AmJ Cardiol 1972; 29 (2): 241–246
Levett JM, Replogle RL. Thermodilution cardiac out-put: A critical analysis and review of the literature. J Surg Res 1979; 27 (6): 392–404
Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 1993; 40 (2): 142–153
Sola JE, Bender JS. Use of the pulmonary artery catheter to reduce operative complications. Surgical Clinic of North America 1993; 73 (2): 253–264
Connors AF Jr, Speroff T, Dawson NV et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. JAMA 1996; 276 (11): 889–97
Dalen JE, Bone RC. Is it time to pull the pulmonary artery catheter? JAMA 1996 Sep 18; 276 (11): 916–918
Fick A. The Measurement of CardiacVentricular Output (in German). Phys Med Ges 1870; 16: XVI (Translated in Geddes LA, Reference 8)
Geddes LA. Cardiovascular devices and their applications. 1st Edition. New York: John Wiley & Sons, 1984: 182–183
Collier CR. Determination of mixed venous CO2 tensions by rebreathing. J Appl Physiol 1956; 9: 25–29
Cerretelli P, Cruz JC, Farhi LE, Rahn H. Determination of mixed venous O2 and CO2 tensions and cardiac output by rebreathing method. Respir Physiol 1966; 1: 258–264
Franciosa JA. Evaluation of the CO2 rebreathing cardiac output method in seriously ill patients. Circulation 1977; 55: 449–455
Orr JA, Westenskow D, Kofoed S, Turner R. Evaluation of a system to measure CO2 production. J Clin Monit 1996; 12 (6): 471–472
Capek JM, Roy RJ. Fick Techniques. 1302–1314. In: Webster JG (ed): Encyclopedia of medical devices and instrumentation, vol. 2. New York: John Wiley and Sons, 1988
Gedeon A, Forslund L, Hedenstierna G, Romano E. A new method for non-invasive bedside determination of pulmonary blood flow. Med Biol Eng Comput 1980; 18: 411–418
Gedeon A. Non-invasive pulmonary blood flow for optimal PEEP. Clin Physiol 1985; 5 (3): 49–58
Roy RJ, Capek J. Non-invasive cardiac output monitoring using partial CO2 rebreathing. Anesthesiol 1985; 63 (3A): A172
Capek JM, Roy RJ. Non-invasive measurement of cardiac output using partial CO2 rebreathing. IEEE Trans BME 1988; 35 (9): 653–661
Nunn JF. Nunn's applied respiratory physiology, 4th edition. Oxford: Butterworth-Heinemann, 1993
Severinghaus JW. Simple, accurate equations for human blood O2 dissociation computations. J Appl Physiol 1979; 46: 599–602
Lobdell DD. An invertible simple equation for computation of blood O2 dissociation relations. J Appl Physiol 1981; 50 (5): 971–3
Haryadi DG, Orr JA, Kuck K, Jaffe MB. Methods of non-invasively estimating intrapulmonary shunt fraction and measuring cardiac output. US Patent no. 6,042,550, 2000
Farhi LE, Herman R. Dynamics of changes in carbon dioxide stores. Anesthesiol 1960; 21 (6): 604–14
Mathews CM, Laszlo G, Campbell EJ, Read DJ. A model for the distribution and transport of CO2 in the body and the ventilatory response to CO2. Respir Physiol 1968; 6: 45–87
Jaffe MB. Partial CO2 rebreathing cardiac output — operating principles of the NICOTM system. J Clin Monit 1999; 15: 387–401
Williams JE, Pfau SE, Deckelbaum LI. Effect of injectate temperature and thermistor position on reproducibility of thermodilution cardiac output determinations. Chest 1994; 106: 895–898
Haryadi DG, Orr JA, McJames S, Westenskow DR. Measurement of injectate temperature at right atrial port improves accuracy of thermodilution cardiac output. J Clin Monit 1998; 14: 518–519
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical assessment. Lancet 1986; 2: 307–310
Bland JM, Altman DG. Calculating correlation coe¤cients with repeated observations: Part 1 — correlation within subjects. BMJ 1995; 310: 446
Bland JM, Altman DG. Calculating correlation coe¤cients with repeated observations: Part 2 — correlation between subjects. BMJ 1995; 310: 633
Critchley LAH, Critchley JAJH. A meta-analysis of bias and precision statistics to compare cardiac output measurement techniques. J ClinMonit 1999; 15: 85–91
Heigenhauser GJF, Jones NL. Measurement of cardiac output by carbon dioxide rebreathing methods. Clin ChestMed 1989; 10 (2): 255–264
Blomqvist H, Frostell C, Hedenstrierna G. A Non-invasive technique for measurement of lung perfusion. Intens CareMed 1986; 12: 172
Capek JM. Noninvasive measurement of cardiac output using partial carbon dioxide rebreathing. 1988 PhD Dissertation. Rensselaer Polytechnic Institute, Troy, New York, U.S.A.
Bosman RJ, Stoutenbeek ChP, Zandstra DF. Non-invasive pulmonary blood £ow measurement by means of CO2 analysis of expiratory gases. Intens Care Med 1991; 17 (2): 98–102
Gedeon A, Krill P, Kristensen J, Gottlieb I. Non-invasive cardiac output determined with a new method based on gas exchange measurements and carbon dioxide rebreathing: A study in animals/pigs. J Clin Monit 1992; 8 (4): 267–278
Osterlund B, Gedeon A, Krill P et al. A new method of using gas exchange measurements for the non-invasive determination of cardiac output: Clinical experiences in adults following cardiac surgery. Acta Anaesthesiol Scand 1995; 39: 727–32
Gama De Abreu M. Prediction of ARDS and MOF in critically ill patients and Reliability of the partial CO2 rebreathing technique for measurement of cardiac output. 1995 Dissertation, Ruprecht-Karls-Universitat, Heidelberg
Orr JA, Kofoed S, Westenskow D, Turner R. A non-invasive cardiac output system using the partial rebreathing Fick method. J ClinMonit 1996; 12 (6): 464–465
Gama de Abreu M, Quintel M, Ragaller M, Albrecht DM. Partial carbon dioxide rebreathing: A reliable technique for non-invasive measurement of nonshunted pulmonary capillary blood flow. Crit Care Med 1997; 25 (4): 675–683
Orr JA, Westenskow DR, Haryadi DG, Kofoed S. Evaluation of partial re-breathing cardiac output measurement in animals. J Clin Monit 1998; 14: 527–528
Bailey PL, Haryadi DG, Orr JA, Westenskow DR. Partial CO2 rebreathing Fick technique for non-invasive measurement of cardiac output. Anesth Analg 1998; 86, SCA53
Haryadi DG, Orr JA, Kuck K, McJames S, Westenskow DR. Evaluation of partial CO2 rebreathing Fick technique for measurement of cardiac output. Anesthesiol 1998; 89 (3A): A534
Johnson KB, Haryadi DG, Orr JA, McJames S, Kuck K, Westenskow DR. Influence of pulmonary edema on non-invasive measurements of cardiac output using partial CO2 rebreathing in a canine model. Anesthesiol 1998; 89 (3A): A535
Watt RC, Loeb RG, Orr JA. Comparison of a new non-invasive cardiac output technique with invasive bolus and continuous thermodilution. Anesthesiol 1998; 89 (3A): A536
Kuck K, Haryadi DG, Orr JA, Bailey PL. Evaluation of partial rebreathing cardiac output measurement during surgery. Anesthesiol 1998; 89 (3A): A542
Guzzi L, Jaffe MB, Orr JA. Clinical evaluation of a new noninvasive method of cardiac output measurement — preliminary results in CABG patient. Anesthesiol 1998; 89: A543
Jopling MW. Noninvasive cardiac output determination utilizing the method of partial CO2 rebreathing. A comparison with continuous and bolus thermodilution cardiac output. Anesthesiol 1998; 89: A544
Orr JA, Kuck K, Haryadi DG, Jaffe MB. Clinical validation of a new non-invasive breath to breath Fick method for cardiac output estimation. J Clin Monit 1999; 15 (3–4): 241–242
Haryadi DG, Bailey PL, Kuck K, Orr JA, Westenskow DR. Clinical evaluation of partial CO2 rebreathing Fick technique for non-invasive measurement of cardiac out-put. J Clin Monit 1999; 15 (3–4): 255–256
Loeb RG, Watt RC, DiNardo JA, Orr J. Comparison of non-invasive, partial rebreathing cardiac output with invasive bolus and continuous thermodilution. J Clin Monit 1999; 15 (3–4): 259–260
Haryadi DG, Kuck K, Orr JA, Westenskow DR. Reliability of invasive versus non-invasive indices in reflecting intrapulmonary shunt. Crit Care Med 1999; 27 (1, Suppl): A116
Haryadi DG, Orr JA, Johnson KB, Kuck K, Westenskow DR. Influence of error in estimation of intrapulmonary shunt on the measurement of cardiac output using re-breathing techniques. J Clin Monit 1999; 15 (3–4): 256–257
Kuck K, Haryadi DG, Orr JA, Westenskow DR, Jaffe MB. Haryadi et al: Partial CO2 Rebreathing Fick CardiacOutput 373 Arterial blood gas measurements improve non-invasive cardiac output estimates from partial CO2 rebreathing Fick technique. Crit Care Med 1999; 27 (1, Suppl): A109
Haryadi DG, Orr JA, Kuck K, Westenskow DR. Limited reproducibility of thermodilution hampers evaluation of new cardiac output monitoring devices. J Clin Monit 1999; 15 (3–4): 255
Orr JA, Westenskow DR. Anesthesiol 1997; 87 (3A): A427
Haller M, Zollner C, Briegel J, Forst H. Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: A prospective criterion standard study. Crit Care Med 1995; 23 (5): 860–866
Comroe JH. The lung, 2nd edition. Chicago: Year Book Medical Publishers, 1962.
West JB. Ventilation-perfusion inequality and overall gas exchange in computer models of the lung. Respir Physiol 1969; 7: 88–110
Fowler WS. Lung Function Studies II, The Respiratory Dead Space. AmJ Physiol 1948; 154: 405
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Haryadi, D.G., Orr, J.A., Kuck, K. et al. Partial CO2 Rebreathing Indirect Fick Technique for Non-Invasive Measurement of Cardiac Output. J Clin Monit Comput 16, 361–374 (2000). https://doi.org/10.1023/A:1011403717822
Issue Date:
DOI: https://doi.org/10.1023/A:1011403717822